Apex Knee System

K201611 · Omnilife Science · JWH · Apr 2, 2021 · Orthopedic

Device Facts

Record IDK201611
Device NameApex Knee System
ApplicantOmnilife Science
Product CodeJWH · Orthopedic
Decision DateApr 2, 2021
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3560
Device ClassClass 2
AttributesTherapeutic

Intended Use

The Apex Knee™ System is intended for use as a primary or revision total knee replacement. This prosthesis may be used for the following conditions, as appropriate: - Non-inflammatory degenerative joint disease, including osteoarthritis and avascular necrosis; - Rheumatoid arthritis; - Correction of functional deformity: - Revision procedures where other treatments or devices have failed; The porous coated femoral component may be used cemented (biological fixation). The porous coated tibial baseplate is to be used uncemented (biological fixation). All other femoral, tibial baseplate, and patellar components are indicated for cemented use only. The Apex Knee™ Modular Tibial Augments are intended to the Tibia baseplate and cemented to the prepared tibia.

Device Story

Apex Knee™ System is a semi-constrained total knee prosthesis for primary or revision arthroplasty. Components include pegged tibial baseplates (Cobalt Chrome, ASTM F75) for enhanced torsional stability and ECiMa tibial inserts/patellae made of Vitamin E-stabilized highly crosslinked polyethylene (VE-XLPE). System components are implanted by orthopedic surgeons in a clinical/OR setting. Pegged baseplates are secured via pegs; inserts attach to baseplates via dovetails and a retaining mechanism. Device provides mechanical joint replacement to restore function and alleviate pain in patients with degenerative or inflammatory joint disease. System is compatible with existing Apex Knee components.

Clinical Evidence

Bench testing only. Non-clinical evidence includes FEA simulation, peg location templating, material characterization, PS post strength testing, insert disassembly strength testing, and wear testing. Biocompatibility testing included intramuscular and subcutaneous implantation in animal models.

Technological Characteristics

Materials: Cobalt Chrome (ASTM F75), Vitamin E-stabilized highly crosslinked polyethylene (VE-XLPE). Design: Semi-constrained patellofemorotibial prosthesis. Features: Pegged tibial baseplates, modular tibial augments, dovetail insert attachment. Sterilization: Ethylene oxide (EO) to SAL 10^-6. Connectivity: None (mechanical implant).

Indications for Use

Indicated for patients requiring primary or revision total knee replacement due to non-inflammatory degenerative joint disease (osteoarthritis, avascular necrosis), rheumatoid arthritis, functional deformity, or failed previous treatments.

Regulatory Classification

Identification

A knee joint patellofemorotibial polymer/metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace a knee joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that have a femoral component made of alloys, such as cobalt-chromium-molybdenum, and a tibial component or components and a retropatellar resurfacing component made of ultra-high molecular weight polyethylene. This generic type of device is limited to those prostheses intended for use with bone cement (§ 888.3027).

Predicate Devices

Reference Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ April 2, 2021 Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services seal on the left and the FDA acronym with the full name of the agency on the right. The seal features a stylized depiction of human figures, while the FDA part includes a blue square with "FDA" in white and the words "U.S. FOOD & DRUG ADMINISTRATION" in blue. OMNIlife science Dipti Dharia Chief Quality, Regulatory, & Clinical Officer 480 Paramount Drive Raynham, Massachusetts 02767 ## Re: K201611/S001 Trade/Device Name: Apex Knee™ System Regulation Number: 21 CFR 888.3560 Regulation Name: Knee Joint Patellofemorotibial Polymer Semi-Constrained Cemented Prosthesis Regulatory Class: Class II Product Code: JWH, MBH Dated: February 19, 2021 Received: March 4, 2021 Dear Dipti Dharia: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's {1}------------------------------------------------ requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Ting Song, Ph.D., R.A.C. Assistant Director DHT6A: Division of Joint Arthroplasty Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) #### K201611 Device Name Apex Knee™ System Indications for Use (Describe) The Apex Knee™ System is intended for use as a primary or revision total knee replacement. This prosthesis may be used for the following conditions, as appropriate: - · Non-inflammatory degenerative joint disease, including osteoarthritis and avascular necrosis; - · Rheumatoid arthritis; - · Correction of functional deformity: - · Revision procedures where other treatments or devices have failed; The porous coated femoral component may be used cemented (biological fixation). The porous coated tibial baseplate is to be used uncemented (biological fixation). All other femoral, tibial baseplate, and patellar components are indicated for cemented use only. The Apex Knee™ Modular Tibial Augments are intended to the Tibia baseplate and cemented to the prepared tibia. | Type of Use (Select one or both, as applicable) | | |------------------------------------------------------------------------------------------------|----------------------------------------------------------| | <div><span style="font-size:100%;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D)</div> | <div>□ Over-The-Counter Use (21 CFR 801 Subpart C)</div> | ### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ Image /page/3/Picture/1 description: The image shows the logo for OMNI. The logo consists of a circular graphic on the left and the word "OMNI" on the right. The circular graphic is blue and green and appears to be a stylized representation of a person. Below the word "OMNI" is the phrase "Engineering Orthopaedic Innovation". # 510(k) Summary | TABLE 1: 510(k) SUMMARY | | |------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Date Summary Prepared | 06/12/2020 | | Manufacturer/Distributor/Sponsor | OMNIlife Science<br>480 Paramount Drive<br>Raynham, MA 02767 | | 510(k) Contact | Dipti Dharia<br>Chief Quality, Regulatory & Clinical Officer<br>Corin Ltd.,<br>480 Paramount Drive<br>Raynham, MA 02767<br>Telephone: +1-732-754-5466<br>Email: Dipti.Dharia@coringroup.com<br><br>Secondary Contact:<br>Jack Liang<br>Senior Regulatory Affairs Specialist<br>Corin Ltd / Corin Australia<br>+61 2 9497 7400<br>Jack.Liang@coringroup.com | | Trade Name | Apex Knee™ System | | Common Name | prosthesis, knee, patellofemorotibial, semi-constrained,<br>cemented, polymer/metal/polymer<br><br>prosthesis, knee, patello/femorotibial, semi-constrained,<br>uncemented, porous, coated | | Classification | 21 CFR 888.3560. Class II. Knee joint<br>patellofemorotibial polymer/metal/polymer semi-<br>constrained cemented prosthesis. JWH<br><br>21 CFR 888.3565. Class II. Knee joint<br>patellofemorotibial metal/polymer porous-coated<br>uncemented prosthesis. MBH | | Predicate Device | OMNI TiN Coated Apex Knee™ System- K191765 | | Purpose of Submission | This traditional 510(k) premarket notification is being<br>submitted to obtain clearance for the modifications to<br>the Apex Knee™ System to expand OMNI's product<br>offering for total knee arthroplasty. | | Indications for Use | The Apex Knee™ System is intended for use as a<br>primary or revision total knee replacement. This<br>prosthesis may be used for the following conditions, as<br>appropriate:<br>Non-inflammatory degenerative joint disease,<br>including osteoarthritis and avascular necrosis; Rheumatoid arthritis; Correction of functional deformity; Revision procedures where other treatments or<br>devices have failed; The porous coated femoral component may be used<br>cemented or uncemented (biological fixation).<br>The Porous coated tibial baseplate is to be used<br>uncemented (biological fixation).<br>All other femoral, tibial baseplate, and patellar<br>components are indicated for cemented use only.<br>The Apex Knee™ Modular Tibia System Tibial<br>Augments are intended to be bolted to the Tibia<br>baseplate and cemented to the prepared tibia. | | Device Description | The proposed devices are intended to be used for<br>primary and revision total knee replacement.<br><br>The Pegged Tibial Baseplates offer additional torsional<br>stability by adding pegs to the posterior end of the<br>device. The material of the proposed Pegged Tibial<br>Baseplates are Cobalt Chrome, CoCr (ASTM F75).<br><br>The ECiMa tibial inserts are machined from<br>compression molded highly crosslinked polyethylene<br>with Vitamin E (VE-XLPE) and attaches to the cobalt<br>chrome tibial baseplate via dovetails and a retaining | | Engineering Orthopaedic Innovation | | | Sterility and Biocompatibility | The proposed devices undergo the same validated<br>sterilization process, using ethylene oxide (EO), as the<br>predicate devices (K060192, K102578, K111184,<br>K112891) under the sterility assurance level (SAL) of<br>10-6. All ethylene oxide residuals are monitored and<br>well below standard limits.<br>In addition, OMNI has developed a plan to test<br>endotoxins on all OMNIlife science device groups<br>through Limulus amebocyte lysate (LAL) testing.<br>Products have been segregated into product families<br>based on manufacturing process and material type and<br>have generated 8 product groups that will be tested for<br>endotoxins on a quarterly basis on a yearly rotation, and<br>the subject devices are included in one of the product<br>groups. Product will not be released if the 4 EU/device<br>limit is exceeded. | | Substantial Equivalence Summary | The proposed devices with modifications are<br>substantially equivalent to the existing Apex Knee™<br>System cleared in K191765 as the basic design,<br>interface, fundamental technology and intended use are<br>the same.<br>The use of the new Apex Knee™ System components<br>with the existing knee components do not introduce any<br>new risks of safety or efficacy.<br>Tibial Baseplate:<br>In addition to the testing performed for the Apex Knee<br>System, non-clinical testing specifically for the Pegged<br>Porous Coated Cementless Tibial Baseplates include<br>FEA simulation and peg location templating study.<br>The results of the testing show that the Apex Knee<br>Pegged Porous Coated Cementless Tibial Baseplates are | | | safe and effective for the proposed indications and is<br>substantially equivalent to the predicate devices.<br>The justifications are described in Section 10, Device<br>Description. | | | <b>ECiMa Tibial Inserts &amp; Patellae</b> | | | In addition to the testing already performed for the Apex<br>Knee System, non-clinical testing was conducted<br>specifically for the ECiMa Tibial Inserts & Patellae.<br>These tests include:<br>• material characterization,<br>• PS post strength testing,<br>• insert disassembly strength testing,<br>• wear testing. | | | Biological testing of the ECiMa material was conducted<br>for biocompatibility. These tests include:<br>• intramuscular implantation in animal models,<br>• implant toxicity and subcutaneous implantation in<br>animal models. | | | The results of the testing show that the Apex Knee<br>ECiMa Tibial Inserts & Patellae are safe and effective<br>for the proposed indications and is substantially<br>equivalent to the predicate devices. | | | Based on the design, fundamental technology, material,<br>intended use and technological characteristics,<br>OMNIlife science believes the proposed Apex KneeTM<br>System devices to be substantially equivalent to legally<br>marketed predicates. | | Conclusion Statement | The conclusions drawn from the nonclinical tests<br>demonstrate that the devices are safe, as effective, and<br>perform as well as or better than the legally marketed<br>device | {4}------------------------------------------------ Image /page/4/Picture/1 description: The image shows the logo for OMNI. The logo consists of a blue and green circular graphic to the left of the word "OMNI" in large, bold, blue letters. Below the word "OMNI" is the text "Engineering Orthopaedic Innovation" in a smaller font size. The logo is clean and professional, and the colors are eye-catching. {5}------------------------------------------------ Image /page/5/Picture/1 description: The image shows the logo for OMNI Engineering Orthopaedic Innovation. The logo features a stylized human figure inside of a circle on the left. The word "OMNI" is in large, bold, blue letters to the right of the figure. Below the word "OMNI" is the phrase "Engineering Orthopaedic Innovation" in a smaller font. {6}------------------------------------------------ Image /page/6/Picture/1 description: The image shows the logo for OMNI. The logo consists of a blue and green circular design on the left, followed by the word "OMNI" in large, bold, blue letters. Below the word "OMNI" is the text "Engineering Orthopaedic Innovation" in a smaller font.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...