K102578 · Omni Life Science, Inc. · JWH · Mar 25, 2011 · Orthopedic
Device Facts
Record ID
K102578
Device Name
APEX PS KNEE SYSTEM
Applicant
Omni Life Science, Inc.
Product Code
JWH · Orthopedic
Decision Date
Mar 25, 2011
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3560
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Apex Knee™ System is intended for use as a primary or revision total knee replacement. This prosthesis may be used for the following conditions, as appropriate: • Non-inflammatory degenerative joint disease, including osteoarthritis and avascular necrosis; • Rheumatoid arthritis; • Correction of functional deformity; • Revision procedures where other treatments or devices have failed; The porous coated femoral component may be used cemented or uncemented (biological fixation). All other femoral, tibial baseplate and patellar components are indicated for cemented use only. The Apex Knee™ Modular Tibia System Tibial Augment are intended to be bolted to the Tibia baseplate and cemented to the prepared tibia.
Device Story
Apex PS Knee System is a semi-constrained total knee replacement prosthesis; includes posterior stabilized (PS) femur component with box, PS tibial insert, and tibial baseplate. Designed for primary or revision total knee arthroplasty. Femur components available in cemented or uncemented versions; other components cemented only. PS insert features cam-and-post mechanism to control kinematics and provide stability. Used by orthopedic surgeons in clinical/hospital settings. Provides range of motion up to 140 degrees. Benefits include restoration of joint function, correction of deformity, and relief from degenerative joint disease symptoms. Output is physical joint replacement; surgeon uses intraoperative assessment to select appropriate UHMWPE insert thickness for soft tissue balance.
Clinical Evidence
No clinical studies were performed. Substantial equivalence is based on non-clinical bench testing, including range of motion, contact area, constraint (ASTM F1223-03), post strength, patello-femoral contact, wear testing (ISO 14243-2/3), and surface finish analysis.
Technological Characteristics
Semi-constrained knee prosthesis; materials include metal (CoCr) and polymer (UHMWPE). Features posterior stabilized cam-and-post design. Standards: ASTM F2083-08 (Total Knee Prosthesis), ASTM F1223-03 (Constraint), ISO 14243 (Wear), ISO 7207-2, ISO 21534, ISO 21536. Femoral components available with porous coating for biological fixation. Modular tibial augments. No software or electronic components.
Indications for Use
Indicated for skeletally mature patients requiring primary or revision total knee replacement due to non-inflammatory degenerative joint disease (osteoarthritis, avascular necrosis), rheumatoid arthritis, functional deformity, or failed previous procedures.
Regulatory Classification
Identification
A knee joint patellofemorotibial polymer/metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace a knee joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that have a femoral component made of alloys, such as cobalt-chromium-molybdenum, and a tibial component or components and a retropatellar resurfacing component made of ultra-high molecular weight polyethylene. This generic type of device is limited to those prostheses intended for use with bone cement (§ 888.3027).
{0}------------------------------------------------
# '14
MAR 2 5 2011
# 510(k) Summary
# K102578 - Apex PS Knee ™ System
# 24 March, 2010
| Submitter | OMNIlife science, Inc.<br>50 O'Connell Way<br>E. Taunton MA 02718 | Contact | Radhika Pondicherry<br>Regulatory Affairs<br>774-226-1852<br>(508) 822-6030 (fax) |
|------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------|-----------------------------------------------------------------------------------|
| Preparation Date | 24 March, 2010 | | |
| Device Name<br>Trade Name<br>Common/Classification<br>Name | Apex PS Knee™ System<br>Knee joint patellofemorotibial polymer/metal/polymer semi-constrained cemented<br>prosthesis<br>Knee joint patellofemorotibial metal/polymer porous-coated uncemented<br>prosthesis. | | |
| Regulatory Class<br>Product Code | Class II per 21 CFR §888.3560, §888.3565<br>JWH, MBH | | |
| Legally Marketed<br>Predicate Device(s) | • K060192- Apex Knee™ System- cleared 15Jul2006<br>• K073602- Apex Knee™ System Porous Coated Femoral<br>Components(cementless), cleared 14Feb2008<br>• K950010- Darwin Knee System, cleared 15May1995<br>• K936159- The Natural-Knee® II knee system, cleared 22May2005 | | |
| Device Description | The APEX PS Knee System includes a posterior stabilized Femur Component<br>incorporating a proportionally sized box. The Femur Component has the same<br>bone cuts as the Apex Knee™ System (K060192) with the addition of a cut for the<br>Femur Component box. Femur Components will be available in both cemented<br>and uncemented versions. Size ranges, high flex, and all other design features of<br>the Apex Knee System are retained. The PS Insert has a medio-lateral constraint<br>and utilizes the Apex Knee System Tibial Baseplate. For each PS Insert, a range of<br>UHMWPE thicknesses are available to aid in obtaining the proper soft tissue<br>balance across the knee joint. | | |
| Indications for Use | The Apex Knee™ System is intended for use as a primary or revision total knee<br>replacement. This prosthesis may be used for the following conditions, as<br>appropriate:<br>• Non-inflammatory degenerative joint disease, including osteoarthritis and<br>avascular necrosis;<br>• Rheumatoid arthritis;<br>• Correction of functional deformity;<br>• Revision procedures where other treatments or devices have failed;<br><br>The porous coated femoral component may be used cemented or uncemented<br>(biological fixation). All other femoral, tibial baseplate and patellar components<br>are indicated for cemented use only.<br>The Apex Knee™ Modular Tibia System Tibial Augment are intended to be bolted<br>to the Tibia baseplate and cemented to the prepared tibia. | | |
{1}------------------------------------------------
K102578 #2/4
## Predicate Device Comparison
| | | Device Comparison | | | |
|--------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|
| | APEX PS TKA<br>(Subject<br>Device) | APEX CR TKA<br>(K060192,<br>K073602) | Darwin Knee<br>System<br>(K950010) | Natural Knee II<br>(K936159) | |
| Body Site | Knee | Knee | Knee | Knee | |
| Intended use | Primary and<br>revision 3<br>compartment<br>TKA.<br>Is intended for<br>use as a<br>primary or<br>revision total<br>knee<br>replacement. | Primary and<br>revision 3<br>compartment<br>TKA.<br>Is intended for<br>use as a primary<br>or revision total<br>knee<br>replacement. | Primary and<br>revision 3<br>compartment<br>TKA.<br>Provides<br>seamless<br>transitions<br>between<br>primary and<br>the most<br>difficult<br>revision cases. | Primary and<br>revision 3<br>compartment<br>TKA. | |
| Patient<br>Population | Skeletally<br>mature<br>patients. | Skeletally<br>mature patients. | Skeletally<br>mature<br>patients. | Skeletally<br>mature<br>patients. | |
| Similar Design and Specifications | | | | | |
| Femur<br>Component | Femur<br>Component:<br>Identical to<br>K060192-<br>noncoated and<br>K073602-<br>coated | Femur<br>Component:<br>Coated and<br>noncoated | Surface<br>preparation<br>for use with<br>cement. | Femoral<br>Component:<br>CSTi™ titanium<br>on CoCr porous<br>coating | |
| Tibial<br>Baseplate/<br>Component | Tibial<br>Component:<br>Identical to<br>K060192 -non<br>coated | Tibial<br>Component:<br>Non-coated | | Tibial<br>Component:<br>Titanium alloy | |
| Asymmetric<br>femur,<br>anatomic<br>patellar<br>groove | Yes<br>Designed with a<br>deep, wide<br>patellar groove | Identical<br>Designed with a<br>deep, wide<br>patellar groove | Similar<br>Deep single<br>radius<br>Trochlear<br>groove,<br>enables the<br>patella to sit<br>deeply in the<br>groove even at<br>high flexion<br>angles. | Similar<br>Deepened<br>trochlear<br>groove<br>prevents<br>excessive load<br>on<br>the patellar<br>component<br>while<br>providing<br>excellent range<br>of motion. | |
| Anatomic<br>asymmetric<br>tibial baseplate | Monobloc: Yes<br>Modular: No | Monobloc: Yes<br>Modular: No | No | Yes | |
| | | | | | |
| Condylar Box<br>(cam) and<br>Tibial Post<br>(spine) | Rounded, open,<br>and sloped box<br>- Minimizes<br>tibial spine<br>edge loading<br>and potential<br>for<br>polyethylene<br>wear or tibial<br>spine fracture. | Not Applicable,<br>CR device | Squared box<br>(cam) | Squared box<br>(cam) | |
| PS Insert | PS style insert<br>with cam and<br>post to control<br>kinematics | N/A | PS style insert<br>with cam and<br>post to control<br>kinematics | PS style insert<br>with cam and<br>post to control<br>kinematics | |
| PS Insert Post | Yes | No | Yes | Yes | |
| High Flexion<br>Design Option | Full flexion to<br>140° | Full flexion to<br>140° | Similar | Similar | |
{2}------------------------------------------------
K102578 #3/4
Non-Clinical Test Summary
#### Apex PS Knee Flexion Range of Motion
- FDA -Class II Special Controls Guidance Document: Knee Joint Patellofemorotibial . and Femorotibial Metal/Polymer Porous-Coated Uncemented Prostheses: Guidance for Industry and FDA (January 16, 2003)
- . ASTM F2083-08- Standard Specification for Total Knee Prosthesis
- Contact Area of the Apex PS Knee
- FDA -Class II Special Controls Guidance Document: Knee Joint Patellofemorotibial . and Femorotibial Metal/Polymer Porous-Coated Uncemented Prostheses: Guidance for Industry and FDA (January 16, 2003)
- ASTM F2083-08- Standard Specification for Total Knee Prosthesis .
#### Tibio-Femoral Constraint of the Apex PS Knee, PS Insert
- ASTM F2083-08- Standard Specification for Total Knee Prosthesis .
- ASTM F1223-03- Standard Test Method for Determination of Total Knee . Replacement Constraint
### Apex Knee PS Tibial Insert Post Strength Testing
- FDA Class II Special Controls Guidance Document: Knee Joint Patellofemorotibial . and Femorotibial Metal/Polymer Porous-Coated Uncemented Prostheses; Guidance for Industry and FDA, issued on January 16, 2003
#### Apex PS Knee Patello-Femoral Contact Area and Stability
- ASTM F2083-08 Standard Specification for Total Knee Prosthesis .
- . FDA Class II Special Controls Guidance Document: Knee Joint Patellofemorotibial and Femorotibial Metal/Polymer Porous-Coated Uncemented Prostheses; Guidance for Industry and FDA, issued on January 16, 2003
# Apex PS Knee Lift-off Comparison to DePuy P.F.C. Sigma
### Apex PS Knee Wear Review
- ISO 14243-3- Implants for surgery -- Wear of total knee-joint prostheses -- Part 3: . Loading and displacement parameters for wear-testing machines with
{3}------------------------------------------------
K102578 * 4/4
- displacement control and corresponding environmental conditions for test.
- ISO 14243-2- Implants for surgery -- Wear of total knee-joint prostheses -- Part 2: Methods of measurement
- ASTM F1877- Standard Practice for Characterization of Particles
- Apex PS Insert Minimum PS Thickness
- FDA Class II Special Controls Guidance Document (section 5)
- ISO 21536 (2007) - Non-active surgical implants -- Joint replacement implants -- Specific requirements for knee-joint replacement implants
Apex PS Knee Component Surface finish Review
- ISO 7207-2 (1998) - Components for partial and total knee joint prostheses -- Part 2: Articulating surfaces made of metal, ceramic and plastics materials
- ISO 21534 (2007)- Non-active surgical implants -- Joint replacement implants -- Particular requirements
- ISO 21536 (1998)- Non-active surgical implants -- Joint replacement implants -- Specific requirements for knee-joint replacement implants
Apex PS Knee Instrument Review
Apex PS Knee Tibio-Femoral Conformity Ratios
All samples tested met the acceptance criteria.
| Clinical Test Summary | No clinical studies were performed. |
|-----------------------|-------------------------------------------------------------------------------|
| Conclusions | The Apex PS Knee System is substantially equivalent to the predicate devices. |
{4}------------------------------------------------
# DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/4/Picture/1 description: The image shows the logo for the Department of Health & Human Services USA. The logo features a stylized eagle with three lines representing its body and wings. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" are arranged in a circular pattern around the eagle.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Omni Life Science, Inc. % Ms. Radhika Pondicherry 50 O'Connell Way East Taunton, Massachusetts 02718
MAR 2 5 2" 11
Re: K102578
Trade/Device Name: Apex PS Knee™ System Regulation Number: 21 CFR 888.3560 Regulation Name: Knee joint patellofemorotibial polymer/metal/polymer semi-constrained cemented prosthesis Regulatory Class: Class II Product Code: JWH, MBH Dated: March 18, 2011 Received: March 21, 2011
Dear Ms. Pondicherry:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act
{5}------------------------------------------------
Page 2 – Ms. Radhika Pondicherry
or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Mark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
Indications for Use
**510(k) Number:** K102578
Device Name: Apex PS Knee System
The Apex Knee™ System is intended for use as a primary or revision total knee replacement. This prosthesis may be used for the following conditions, as appropriate:
- . Non-inflammatory degenerative joint disease, including osteoarthritis and avascular necrosis;
- Rheumatoid arthritis;
- . Correction of functional deformity;
- Revision procedures where other treatments or devices have failed; .
The porous coated femoral component may be used cemented (biological fixation). All other femoral, tibial baseplate and patellar components are indicated for cemented use only.
The Apex Knee™ Modular Tibial Augment are intended to be bolted to the Tibia baseplate and cemented to the prepared tibia.
Prescription Use _ X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Page 1 of 1
A. B. n. h. In no 11
Division Off
Division of Surgical, Orthopedic, and Restorative Devices
510(k) Number K102578
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.