The Apex Knee System is intended for use as a primary or revision total knee replacement. This knee replacement system is intended for cemented fixation, with the option of using a porous coated femoral component for uncemented or cemented fixation. This prosthesis may be used for the following conditions, as appropriate: . Non-inflammatory degenerative joint disease, including osteoarthritis and avascular necrosis; . . Rheumatoid arthritis; . Correction of functional deformity: . Revision procedures where other treatments or devices have failed. .
Device Story
Apex Knee System is a total knee replacement prosthesis for primary or revision surgery. Components include cobalt chrome femoral components (with or without sintered porous coating), dome-shaped UHMWPE patella resurfacing components, UHMWPE tibial inserts, cobalt chrome tibial trays, and titanium alloy locking bolts. Surgeon selects component sizes and tibial insert geometry (cruciate retaining or ultra-congruent) to match patient anatomy and soft tissue balance. Fixation achieved via PMMA bone cement or biological tissue ingrowth into porous coating. Device used in clinical/OR settings by orthopedic surgeons to restore joint function, correct deformity, and alleviate pain associated with degenerative or inflammatory joint disease.
Clinical Evidence
Bench testing only. No clinical data provided.
Technological Characteristics
Materials: Cobalt chrome femoral components, UHMWPE patella/tibial inserts, titanium alloy locking bolts. Porous coating: Sintered cobalt chrome (ASTM F1377). Fixation: PMMA bone cement or biological tissue ingrowth. Design: Modular, asymmetric femoral/tibial components, CR or Ultra-congruent tibial inserts. Sterilization: Not specified.
Indications for Use
Indicated for primary or revision total knee replacement in patients with non-inflammatory degenerative joint disease (osteoarthritis, avascular necrosis), rheumatoid arthritis, functional deformity, or congenital dislocation, and for revision of failed prior treatments.
Regulatory Classification
Identification
A knee joint patellofemorotibial polymer/metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace a knee joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that have a femoral component made of alloys, such as cobalt-chromium-molybdenum, and a tibial component or components and a retropatellar resurfacing component made of ultra-high molecular weight polyethylene. This generic type of device is limited to those prostheses intended for use with bone cement (§ 888.3027).
{0}------------------------------------------------
# Apex Knee System
2. Device Name
#### December 20, 2007
1. Submitter: OMNI life science™, Inc. 175 Paramount Drive
Contact: Mr. William S. McCallum
Raynham, MA 02767
- Director of Requlatory and Quality Systems (508) 824-2444 (voice) (508) 822-6030 (fax)
FEB 14 23
| Proprietary Name: | Apex Knee System Porous Coated Femoral Components |
|-----------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Common Name: | Total Knee Replacement Prosthesis |
| Classification Names: | Prosthesis, knee, patellofemorotibial, semi-constrained,<br>cemented, polymer/metal/polymer<br>Prosthesis, knee, patellofemorotibial, semi-constrained,<br>uncemented, polymer/metal/polymer |
| Regulatory Classes: | Class II per 21 CFR §888.3560<br>Class II (special controls) per 21 CFR §888.3565 |
| Product Codes: | JWH, MBH |
#### 3. Intended Use
The Apex Knee System is intended for use as a primary or revision total knee replacement. This knee replacement system is intended for cemented fixation, with the option of using a porous coated femoral component for uncemented or cemented fixation. This prosthesis may be used for the following conditions, as appropriate:
- . Non-inflammatory degenerative joint disease, including osteoarthritis and avascular necrosis;
- . Rheumatoid arthritis;
- . Correction of functional deformity:
- Revision procedures where other treatments or devices have failed. .
## 4. Device Description
The Apex Knee System Porous Coated Femoral Components are part of the Apex Knee System for primary or revision total knee replacement. These cobalt chrome femoral components have a sintered porous coating (ASTM F1377). Fixation of the femoral component is achieved using either PMMA bone cement or by biological fixation via tissue ingrowth into the porous coating. The Apex Knee System consists of a range of sizes of femoral components with a deep patellar groove, dome shaped UHMWPE patella resurfacing components, UHMWPE tibial inserts, cobalt chrome tibial trays, and a titanium alloy bolt for locking the tibial insert to the tibial tray. This modular configuration allows the surgeon user to choose a combination of femoral and tibial tray component sizes to appropriately fit the anatomy of the patient, and to use a tibial insert with a sizefor-size match to the femoral component. There are two different articular geometries for the tibial insert, a cruciate retaining ("CR") design, which allows retention of the posterior cruciate ligament, and an ultra-congruent posterior cruciate substituting ("Ultra") design. For each tibial insert, a range of UHMWPE thicknesses are available to aid in obtaining the proper soft tissue balance across the knee joint.
{1}------------------------------------------------
## 5. Predicate Device Comparison
Substantial equivalence is claimed to the Apex Knee System (K060192), distributed by OMNI life science, Inc., and the Sigma Cruciate-Retaining Porocoat® Femoral Components distributed by DePuy Orthopedics (K062654). The following table summarizes the similarities and differences between the subject uncemented femoral components of the Apex Knee System and these predicate devices:
| | Apex Knee<br>Porous Coated<br>Femoral<br>Components | Apex Knee<br>Cemented<br>Femoral<br>Components<br>(K060192) | Sigma*<br>Porocoat®<br>Femoral<br>Components<br>(K062654) |
|-----------------------------|-----------------------------------------------------|-------------------------------------------------------------|-----------------------------------------------------------|
| INTENDED USE | | | |
| Primary and revision, 3 | Yes, cemented or | Yes, cemented | Yes, cemented or |
| compartment | uncemented | | uncemented |
| DESIGN | | | |
| Porous coated | Yes | No | Yes |
| Asymmetric femur, | Yes | Yes | Yes |
| anatomic patella groove | | | |
| Anatomic (asymmetric) | Yes | Yes | No |
| tibial tray | | | |
| Metal-backed UHMWPE | Yes | Yes | Yes |
| tibial component | | | |
| Tibial insert designs | CR and Ultra | CR and Ultra | CR ("Curved") |
| Tibial tray distal features | Central post and | Central post and | Central post and 2 |
| | 2 keels | 2 short keels | short keels |
| Patella design | Round, single | Round, single | Round, single |
| | radius dome, 3 | radius dome, 3 | radius dome, 3 |
| | pegs | pegs | pegs |
| MATERIALS | | | |
| Femoral component | Cobalt chrome | Cobalt chrome | Cobalt chrome |
| Femoral porous coating | Yes, sintered | No | Yes, sintered |
| | cobalt chrome | | cobalt chrome |
The only change to the Sponsor's predicate Apex Knee System (K060192) is the addition of a sintered cobalt chrome porous coating to the femoral components. This porous coating is similar to the sintered cobalt chrome porous coating on the predicate DePuy femoral components, and has the same bead configuration (sizes and shapes) as the Sponsor's Apex Modular Acetabular Cup (K031110). All other materials, designs, and manufacturing, packaging, and sterilization methods are identical to the predicate Apex Knee System.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three stripes forming its body and wings. The eagle faces right. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - U.S.A." is arranged in a circular fashion around the eagle.
Public Health Service
FEB 14 2008
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
OMNI life science, Inc. % Mr. William McCallum Director, Regulatory and Quality 175 Paramount Drive, Suite 302 Raynham, Massachusetts 02767
Re: K073602
Trade/Device Name: Apex Knee System Regulation Number: 21 CFR 888.3560 Regulation Name: Knee joint patellofemorotibial polymer semi-constrained cemented prosthesis Regulatory Class: II Product Code: JWH, MBH Dated: January 25, 2008 Received: January 28, 2008
Dear Mr. McCallum:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set
{3}------------------------------------------------
Page 2 - Mr. William McCallum
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (240) 276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at (240) 276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Mark M. Mullens
Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# Indications for Use
#### 510(k) Number (if known): K073602
Device Name: OMNI life science Apex Knee System
Indications For Use:
The Apex Knee System is intended for use as a primary or revision total knee replacement. This knee replacement system is intended fixation, with the option of using a porous coated femoral component for uncemented or cemented fixation. This prosthesis may be used for the following conditions, as appropriate:
- Non-inflammatory degenerative joint disease, including osteoarthritis and avascular necrosis; .
- . Rheumatoid arthritis;
- Correction of functional deformity; .
- Congenital dislocation; ●
- Revision procedures where other treatments or devices have failed. .
Prescription Use X (Per 21 CFR 801 Subpart D)
AND/OR Over-The-Counter Use
(21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Chubare BouellD
Page 1 of 1
Restorative, Division of Ge and Neurological Devices
510(k) Number K073602
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.