ORTHOLOC(TM) 3DI LOCKING SCREWS ORTHOLOC(TM) BONE SCREWS
Applicant
Wrightmedicaltechnologyinc
Product Code
HRS · Orthopedic
Decision Date
Apr 2, 2012
Decision
SESE
Submission Type
Special
Regulation
21 CFR 888.3030
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Wright's ORTHOLOC™ 3Di Ankle Plating System is intended for fixation of fractures, osteotomies, and non-unions of the distal tibia and fibula such as: - Lateral Malleolar Fractures . - Syndesmosis injuries - Medial Malleolar Fractures - Bi-Malleolar Fractures - Tri-Malleolar Fractures - Posterior Malleolar Fractures . - Distal Anterior Tibia Fractures - Vertical Shear Fractures of the Medial Malleolous - Pilon Fractures - Distal Tibia Shaft Fractures - Distal Fibula Shaft Fractures - Distal Tibia Periarticular Fractures - Medial Malleolar Avulsion Fractures - Lateral Malleolar Avusion Fractures The ORTHOLOC™ 3Di locking screws are intended for use with Wright's ORTHOLOCTM 3Di Plating Systems of the same base material. ORTHOLOCTM Bone Screws are indicated for use in bone reconstruction, osteotomy, arthrodesis, joint fusion, fracture repair, and fracture fixation, appropriate for the size of the device. Washer Wright's washers are intended to prevent a screw head from breaking through the cortex of the bone by distributing the forces/load over a large area when used for fracture fixation of bone fragments.
Device Story
ORTHOLOC™ 3Di Locking Screws and ORTHOLOC™ Bone Screws are orthopedic fixation implants. Subject devices include 2.7mm diameter locking screws (26-30mm lengths) and 2.7mm low-profile non-locking bone screws. Implants are manufactured from titanium alloy with solid cores; intended for single use. Used by orthopedic surgeons in clinical/surgical settings for fracture repair, osteotomy, and joint fusion. Screws provide mechanical stabilization of bone fragments; washers distribute load to prevent cortical breakthrough. Performance verified via bench testing (torque, worst-case analysis).
Clinical Evidence
No clinical data; bench testing only. Evidence includes design comparison, materials information, worst-case analyses, and mechanical torque testing (ultimate, insertion, and removal).
Technological Characteristics
Materials: Titanium alloy. Form factor: 2.7mm diameter locking and non-locking bone screws. Solid core construction. Single-use. Mechanical fixation principle.
Indications for Use
Indicated for patients requiring fixation of fractures, osteotomies, non-unions, bone reconstruction, arthrodesis, or joint fusion in the distal tibia, fibula, or other bones appropriate for device size.
Regulatory Classification
Identification
Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.
Predicate Devices
ORTHOLOC™ 3Di Ankle Plating and ORTHOLOCTM System Bone Screws (K102429)
K121713 — PMT BONE SCREWS · Providence Medical Technology, Inc. · Sep 27, 2012
K090692 — ORTHOLOC 2.0/2.4 PLATE SYSTEM · Wrightmedicaltechnologyinc · May 8, 2009
K132248 — FLOWER BONE SCREW SET · Flower Orthopedics Corporation · Oct 30, 2013
K241359 — Medline UNITE® MIS Foot Recon Screw System · Medline Industries, LP · Aug 2, 2024
Submission Summary (Full Text)
{0}------------------------------------------------
K 120802 (113) APR - 2 2012
Image /page/0/Picture/1 description: The image shows the word "WRIGHT." in bold, sans-serif font. Below the word is a graphic of two horizontal lines that are angled slightly downward. There is a small design in the middle of the two lines.
# 510(K) SUMMARY OF SAFETY AND EFFECTIVENESS
In accordance with the Food and Drug Administration Rule to implement provisions of the Safe Medical Devices Act of 1990 and in conformance with 21 CFR 807.92, this information serves as a Summary of Safety and Effectiveness for the URTHOLOC™ 3Di Locking Screws and ORTHOLOCTM Bone Screws.
A.1. Submitted By:
Date:
Contact Person:
Wright Medical Technology, Inc. 5677 Airline Rd Arlington, TN 38002
March 15, 2012
Sarah Holtgrewe Regulatory Affairs Project Specialist (901) 867-4476
A.2. Proprietary Name:
Common Name:
Device Classification Regulation:
Device Product Code & Panel:
A.3. Predicate Devices:
ORTHOLOC™ 3Di Locking Screws & ORTHOLOCTM Bone Screws
Locking Plate Screw & Bone Screw
21 CFR 888.3040-Class II
HWC: Screw, Fixation Bone HRS: Plate, Fixation, Bone 87 Orthopedics
K102429—ORTHOLOC™ 3Di Ankle Plating and ORTHOLOCTM System Bone Screws K112772—ORTHOLOC™ Bone Screws --Modification
## A.4. Device Description
The ORTHOLOC™ 3Di Locking Screws subject to this premarket notification include the 2.7mm diameter-26, 28, and 30mm length locking screws. The ORTHOLOC™ Bone Screws subject to this premarket notification include the non-locking 2.7mm low profile bone screws. All screws are manufactured from titanium alloy and have a solid core. The implants are single use only devices.
{1}------------------------------------------------
K120802(2/3)
Image /page/1/Picture/1 description: The image shows the word "WRIGHT." in a bold, sans-serif font. Below the word is a stylized graphic that appears to be a stylized letter "W". The graphic is composed of two thick, horizontal lines that converge towards the center, creating a sense of depth and movement. The overall design is clean and modern, with a focus on simplicity and visual impact.
### A.5. Intended Use
#### Ankle Plates:
Wright's ORTHOLOC™ Ankle Plating System is intended for fixation of fractures, osteotomies, and non-unions of the distal tibia and fibula such as:
- Lateral Malleolar Fractures .
- Syndesmosis injuries
- Medial Malleolar Fractures
- Bi-Malleolar Fractures
- Tri-Malleolar Fractures
- Posterior Malleolar Fractures .
- Distal Anterior Tibia Fractures
- Vertical Shear Fractures of the Medial Malleolous
- Pilon Fractures
- Distal Tibia Shaft Fractures
- Distal Fibula Shaft Fractures
- Distal Tibia Periarticular Fractures
- Medial Malleolar Avulsion Fractures
- Lateral Malleolar Avusion Fractures
### ORTHOLOC™ 3D iLocking Screws:
The ORTHOLOC™ 3Di locking screws are intended for use with Wright's ORTHOLOCTM 3Di Plating Systems of the same base material.
#### ORTHOLOCTM Bone Screws:
ORTHOLOC™ Screws are indicated for use in bone reconstruction, osteotomy, arthrodesis, joint fusion, fracture repair, and fracture fixation, appropriate for the size of the device.
#### Washer
Wright's washers are intended to prevent a screw head from breaking through the cortex of the bone by distributing the forces/load over a large area when used for fracture fixation of bone fragments.
# A.6. Technological Characteristics Comparison
The subject 2.7mm locking screws are identical to the predicate 2.7mm locking screws with the exception of screw length. The subject 2.7mm bone/non-locking screws have an identical low-profile head and neck as the predicate 3.5 mm low-profile non-locking bone screws and identical thread form as the predicate 2.7mm locking screw.
# B.1. Substantial Equivalence - Non-Clinical Evidence
Substantial equivalence is shown through comparison of design characteristics, materials information, worst-case analyses, and ultimate, insertion, and removal torque testing. The
headquarters Wright Medical Technology, Inc
Arlington, TN 38002
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the word "WRIGHT." in a bold, sans-serif font. Below the word is a stylized design that appears to be two horizontal lines that are angled slightly upwards. The design is black and white, and the word "WRIGHT." is also black and white.
results show that the subject screws can be expected to perform at least as well as the legally marketed predicate screws.
### B.2. Substantial Equivalence - Clinical Evidence
N/A-clinical evidence not required
#### B.3. Substantial Equivalence - Conclusions
Substantial equivalence is shown through design characteristics, materials information, worst-case analysis, and ultimate, insertion, and removal torque testing, The subject locking screws are identical in indication for use, and all dimensional characteristics except screw length. The subject bone screws have an identical low-profile head and neck as the predicate 3.5 mm low-profile non-locking bone screws and identical thread form as the predicate 2.7mm locking screws. No new types of safety and effectiveness questions can be expected. From the evidence given in the Premarket Notification, the subject devices can be expected to perform at least as well as the predicate devices.
www.wmt.com
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image is a black and white logo for the Department of Health & Human Services - USA. The logo features a stylized eagle with three lines forming its body and wings. The eagle is facing to the right. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the eagle.
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Wright Medical Technology, Incorporated % Ms. Sarah Holtgrewe Regulatory Affairs Project Specialist 5677 Airline Road Arlington, Tennessee 38002
APR - 2 2012
#### Re: K120802
Trade/Device Name: ORTHOLOC™ 3Di Locking Screws & ORTHOLOC™ Bone Screws Regulation Number: 21 CFR 888.3040
Regulation Name: Smooth or threaded metallic bone fixation fastener Regulatory Class: Class II Product Code: HRS, HWC
Dated: March 15, 2012
Received: March 16, 2012
### Dear Ms. Holtgrewe:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{4}------------------------------------------------
# Page 2 – Ms. Sarah Holtgrewe
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Mark N. Melkerson
for Peter
Dan Cur
Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# Indications for Use
510(k) Number (if known): K120802
Device Name: ORTHOLOC™ 3Di Locking Screws & ORTHOLOC™ Bone Screws
Indications For Use:
Wright's ORTHOLOC™ 3Di Ankle Plating System is intended for fixation of fractures, osteotomies, and non-unions of the distal tibia and fibula such as:
- Lateral Malleolar Fractures .
- Syndesmosis injuries
- Medial Malleolar Fractures
- Bi-Malleolar Fractures
- Tri-Malleolar Fractures
- Posterior Malleolar Fractures
- Distal Anterior Tibia Fractures
- Vertical Shear Fractures of the Medial Malleolous
- Pilon Fractures
- Distal Tibia Shaft Fractures
- Distal Fibula Shaft Fractures
- Distal Tibia Periarticular Fractures
- Medial Malleolar Avulsion Fractures
- Lateral Malleolar Avusion Fractures
The ORTHOLOC™ 3Di locking screws are intended for use with Wright's ORTHOLOC™ 3Di Plating Systems of the same base material.
ORTHOLOC™ Bone Screws are indicated for use in bone reconstruction, osteotomy, arthrodesis, joint fusion, fracture repair, and fracture fixation, appropriate for the size of the device.
Wright's washers are intended to prevent a screw head from breaking through the cortex of the bone by distributing the forces/load over a large area when used for fracture fixation of bone fragments.
Prescription Use × (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
{6}------------------------------------------------
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
D. Ovens
(Division Sign-Off) Division of Surgical, Orthopedic, and Restorative Devices
510(k) Number K120802
1 of 1
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.