The Medline UNITE® MIS Foot Recon Screw System is indicated for the stabilization and fixation of bone reconstructions, osteotomies, joint fusions (arthrodeses), and fractures of short (tarsals) and long (metatarsals, phalanges, distal tibia and fibula) bones comprising the foot and ankle. Screws are intended for single use only.
Device Story
System consists of titanium alloy bone screws and associated instrumentation (drill bits, countersinks) for orthopedic fixation. Screws feature Torx drive mechanism, self-tapping, and self-drilling design. Used by surgeons in clinical settings for stabilization of foot/ankle bone reconstructions, osteotomies, joint fusions, and fractures. Implants are single-use, provided non-sterile, and require steam sterilization prior to use. Output is mechanical fixation of bone segments; aids in bone healing and structural stability.
Clinical Evidence
No clinical data. Bench testing only. Torsional yield strength and drive torque evaluated per ASTM F543-23. Axial pullout strength evaluated via engineering analysis.
Technological Characteristics
Materials: Titanium Alloy (Ti-6Al-4V ELI) per ASTM F136 or ISO 5832-3. Design: Torx drive, self-tapping, self-drilling. Dimensions: 2.5mm-4.0mm diameter; 14mm-70mm length. Sterilization: Steam (non-sterile provided).
Indications for Use
Indicated for stabilization and fixation of bone reconstructions, osteotomies, joint fusions (arthrodeses), and fractures of short (tarsals) and long (metatarsals, phalanges, distal tibia and fibula) bones of the foot and ankle.
Regulatory Classification
Identification
A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
August 2, 2024
Medline Industries, LP Jennifer Mason Regulatory Affairs Principal Three Lakes Drive Northfield, Illinois 60030
Re: K241359
Trade/Device Name: Medline UNITE® MIS Foot Recon Screw System Regulation Number: 21 CFR 888.3040 Regulation Name: Smooth Or Threaded Metallic Bone Fixation Fastener Regulatory Class: Class II Product Code: HWC Dated: May 13, 2024 Received: May 14, 2024
Dear Jennifer Mason:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
{1}------------------------------------------------
### K241359 - Jennifer Mason
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100. Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Shumaya Ali -S
Shumaya Ali, M.P.H. Assistant Director DHT6C: Division of Restorative, Repair and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
{2}------------------------------------------------
## Indications for Use
Submission Number (if known)
K241359
Device Name
Medline UNITE® MIS Foot Recon Screw System
Indications for Use (Describe)
The Medline UNITE® MIS Foot Recon Screw System is indicated for the stabilization and fixation of bone reconstructions, osteotomies, joint fusions (arthrodeses), and fractures of short (tarsals) and long (metatarsals, phalanges, distal tibia and fibula) bones comprising the foot and ankle. Screws are intended for single use only.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the Medline logo. The logo is set against a blue background. The word "MEDLINE" is in white, bold, sans-serif font. Above and below the word is a white, stylized star shape, with the bottom point extending further down than the top point.
ledline Industries, LP Three Lakes Drive Northfield. IL 60093
# K241359 510(k) SUMMARY [AS REQUIRED BY 21CFR807.92(c)]
### Submitter / 510(k) Sponsor
Medline Industries, LP Three Lakes Drive Northfield, IL 60093
Registration Number: 1417592
## Contact Person
Contact Person: Jennifer Mason, Regulatory Affairs Principal Phone: 847-643-3652 Email: jamason@medline.com
Summary Preparation Date July 18, 2024
Type of 510(k) Submission Traditional
### Device Name / Classification
Trade Name: Medline UNITE® MIS Foot Recon Screw System Common Name: Screw, Fixation, Bone Classification Name: Smooth or threaded metallic bone fixation fastener Product Code: HWC Classification Panel: Orthopedics Regulatory Class: Class II Regulation Number: 21 CFR 888.3040
### Predicate Device
Medline Cannulated Screw K130319
### Device Description
The Medline UNITE® MIS Foot Recon Screw System includes screws manufactured from Titanium Alloy (Ti-6Al-4V ELI) per ASTM F136 or IOS 5832-3. The screws are offered in various diameters from 2.5mm up to 4.0mm and overall lengths ranging from 14mm up to 70mm. The screws are single use and provided
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the Medline logo. The logo consists of the word "MEDLINE" in white, set against a blue square background. Above the word "MEDLINE" is a white, four-pointed star-like shape. The logo is simple and clean, with a professional appearance.
edline Industries. L Three Lakes Drive Northfield II 60093
non-sterile intended to be cleaned and steam sterilized prior to use. The system also includes non-sterile, single use instrumentation necessary to implant the screws, e.g. drill bits and countersinks. Validated cleaning and sterilization instructions are included in the instructions for use.
#### Indications for Use
The Medline UNITE® MIS Foot Recon Screw System is indicated for the stabilization and fixation of bone reconstructions, osteotomies, joint fusions (arthrodeses), and fractures of short (tarsals) and long (metatarsals, phalanges, distal tibia and fibula) bones comprising the foot and ankle. Screws are intended for single use only.
### Summary of Technological Characteristics
The proposed Medline UNITE® MIS Foot Recon Screw System and the predicate device have the same intended use.
The design features of the Medline UNITE® MIS Foot Recon Screw System are compared to the predicate, Medline Cannulated Screws (K130319) in the table below.
| Device Characteristic | Proposed Device | Predicate Device | Comparison<br>Analysis |
|-----------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------|
| Product Name | Medline UNITE® MIS Foot<br>Recon Screw System | Medline Cannulated Screw | Different |
| 510(k) Reference | N/A | K130319 | Different |
| Product Owner | Medline Industries, LP | Medline Industries, LP | Same |
| Product Code | HWC | HWC | Same |
| Intended Use | The Medline UNITE® MIS<br>Foot Recon Screw System is<br>indicated for the stabilization<br>and fixation of bone<br>reconstructions, osteotomies,<br>joint fusions (arthrodeses),<br>and fractures of short<br>(tarsals) and long<br>(metatarsals, phalanges,<br>distal tibia and fibula) bones<br>comprising the foot and<br>ankle. Screws are intended<br>for single use only. | The Medline Cannulated<br>Screws are indicated for use in<br>bone reconstruction,<br>osteotomies, arthrodesis, joint<br>fusion, fracture repair, and<br>fracture fixation of bones<br>appropriate for the size of the<br>device. Screws are intended for<br>single use only. | Same |
| Regulation Number | 21 CFR 888.3040 | 21 CFR 888.3040 | Same |
## TABLE 1: COMPARISON OF PROPOSED AND PREDICATE DEVICES
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the Medline logo. The logo consists of the word "MEDLINE" in white, set against a blue square background. A white, stylized cross shape is positioned behind the text, with its arms extending beyond the letters. A small gray rectangle is visible in the upper right corner of the image.
Medline Industries, LP Three Lakes Drive Northfield. IL 60093
| Screw Diameters | 2.5mm, 3.0mm, 3.5mm,<br>4.0mm | 2.0mm, 2.5mm, 3.0mm, 3.5mm,<br>4.0mm, 6.5mm, 7.5mm | Similar |
|-----------------------------------|-------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------|---------|
| Screw Lengths | 2.5mm: 14-40mm<br>3.0mm: 14-50mm<br>3.5mm: 16-70mm<br>4.0mm: 20-70mm | 2.0mm: 10-24mm<br>2.5mm: 10-40mm<br>3.0mm: 10-40mm<br>3.5mm: 12-50mm<br>4.0mm: 14-50mm<br>6.5mm: 40-130mm<br>7.5mm: 40-130mm | Similar |
| Materials | Titanium Alloy | Titanium Alloy | Same |
| Design Features | Torx drive mechanism,<br>self-tapping and self-drilling | Torx drive mechanism,<br>self-tapping and self-drilling | Same |
| Sterile vs. Non-Sterile | Non-Sterile | Non-Sterile | Same |
| Disposable vs. Non-<br>Disposable | Non-Disposable - Screw<br>Implants<br>Disposable - Drills and<br>countersinks | Non-Disposable - Screw<br>Implants | Similar |
| Single Use vs. Reusable | Single use | Single use | Same |
### Summary of Non-Clinical Testing
The subject Medline UNITE® MIS Foot Recon Screw System does not represent a new worst-case when compared to the previously cleared Medline Cannulated Screws.
- Torsional Yield Strength O
- I The worst-case analysis predicted the subject screws, Medline UNITE® MIS Foot Recon Screw System (K241359), did not create a new worst-case compared to the predicate (K130319) with regard to torsional yield strength. However, testing per ASTM F543-23 was conducted to confirm the analysis. This is highlighted in the provided test reports (TR3270-001 Medline ASTM F543 Torsional Properties and Driving Torque Testing for 2.0mm Headed Screws Final Report.pdf, TR3270-002 Medline ASTM F543 Torsional Properties and Driving Torque Testing for 2.5mm PT Screws Final Report.pdf). The worst-case subject screws, Ø2.5mm PT screws (MSCP2526), were evaluated for torsional yield strength in accordance with ASTM F543-23 A1, exhibiting an average torsional vield strength of 0.66 Nm for six samples compared to the predicate screws, Ø2.0mm Headed Screws (MSD02016), which exhibited an average torsional vield strength of 0.50 N-m . The results from the Ø2.5mm PT Screws exceeds the FDA's Bone Screw SPBP Guidance (https://www.fda.gov/regulatory-information/search-fda-guidance-documents/orthopedic-
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the Medline logo. The logo is set against a blue background. The word "MEDLINE" is in white, sans-serif font, and to the right of the word is a white starburst symbol. There is a gray bar on the right side of the image.
Medline Industries, LP Three Lakes Drive Northfield. IL 60093
non-spinal-metallic-bone-screws-and-washers-performance-criteria-safety-and-performance ), which is 0.6 N-m for a 2.5mm nominal major diameter screw
#### o Drive Torque
- I The subject screws, the driving and removal torque was only 6.36% of the torsional yield strength. which falls within the the FDA's Bone Screw SPBP Guidance (https://www.fda.gov/regulatory-information/search-fda-guidance-documents/orthopedicnon-spinal-metallic-bone-screws-and-washers-performance-criteria-safety-and-performance ) acceptance criteria of driving or removal torque being less than 50% of the torsional vield.
- 트 The worst-case subject screws were evaluated for drive torque (insertion, removal) in accordance with ASTM F543-23 Annex A2 and met the FDA'S Bone Screw Guidance acceptance criteria for driving torque being less than 50% of the torsional yield strength of the subject screw.
- 0 Axial Pullout Strength
- 트 An engineering analysis was performed to evaluate the axial pullout strength of the worstcase subject screws. The results demonstrated substantially equivalent performance compared to the predicate, Medline Cannulated Screws (K130319) and met the FDA'S Bone Screw Guidance acceptance criteria for predicted shear failure force.
#### Performance Testing (Animal)
This section does not apply. No animal testing was performed.
# Performance Testing (Clinical)
This section does not apply. No clinical testing was performed.
### Summary of Clinical Testing
Not applicable.
### Conclusion
In accordance with 21 CFR Part 807, and based on the information provided in this premarket notification, Medline Industries, LP concludes that the Medline UNITE® MIS Foot Recon Screw System is as safe and as effective for their intended use as the predicate device, Medline Cannulated Screws (K130319).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.