DEN200035 · Miach Orthopaedics, Inc. · QNI · Dec 16, 2020 · Orthopedic
Device Facts
Record ID
DEN200035
Device Name
BEAR (Bridge-Enhanced ACL Repair) Implant
Applicant
Miach Orthopaedics, Inc.
Product Code
QNI · Orthopedic
Decision Date
Dec 16, 2020
Decision
DENG
Submission Type
Direct
Regulation
21 CFR 888.3044
Device Class
Class 2
Attributes
Therapeutic, Real-World Evidence
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
DEN200035 · Dec 16, 2020
BEAR (Bridge-Enhanced ACL Repair) Implant
Miach Orthopaedics, Inc.
Historical control subject-level data; Structured literature review
The sponsor used historical control data and a structured literature review to compare the safety profile (specifically ACL re-tear rates) of the BEAR implant against standard ACL reconstruction (ACLR) outcomes.
Historical control; Literature review; Safety comparison; ACL reconstruction
Clinical Evidence
Study Design
Population
Comparator
Key Endpoints
Historical Control Comparison; Historical control comparison and structured literature review
Patients with ACL rupture undergoing ACL reconstruction (ACLR)
Standard ACL reconstruction (ACLR)
ACL re-tear rate
Indications for Use
The BEAR® (Bridge Enhanced ACL Repair) Implant is a bovine extracellular matrix collagen-based implant for treatment of anterior cruciate ligament (ACL) injuries. The BEAR® Implant is indicated for skeletally-mature patients at least 14 years of age with a complete rupture of the ACL, as confirmed by MRI. Patients must have an ACL stump attached to the tibia to construct the repair.
Device Story
BEAR® Implant is a bovine extracellular matrix collagen-based scaffold; used during surgical ACL repair. Device bridges/surrounds torn ACL ends; protects biological healing process from intraarticular environment; does not replace biomechanical fixation via sutures. Surgeon places implant during procedure; device resorbs over time while supporting ligament healing. Benefits include potential for natural ACL healing rather than traditional reconstruction. Used in clinical/surgical settings.
Clinical Evidence
Clinical performance testing required to demonstrate device effectiveness. Metrics include post-operative knee pain and function, and durability assessed by re-tear or re-operation rates.
Technological Characteristics
Cylindrical scaffold (22mm diameter, 45mm length) composed of bovine collagen and extracellular matrix. Sterilized via electron-beam irradiation (ISO 11137-1). Biocompatible per ISO 10993-1. Non-ferromagnetic (MR Safe). Resorbable material; no reusable components.
Indications for Use
Indicated for skeletally-mature patients ≥14 years old with complete ACL rupture confirmed by MRI, requiring an existing tibial ACL stump for repair construction.
Regulatory Classification
Identification
A resorbable implant for anterior cruciate ligament (ACL) repair is a degradable material that allows for healing of a torn ACL that is biomechanically stabilized by traditional suturing procedures. The device is intended to protect the biological healing process from the surrounding intraarticular environment and not intended to replace biomechanical fixation via suturing. This classification includes devices that bridge or surround the torn ends of a ruptured ACL.
Special Controls
In combination with the general controls of the FD&C Act, the resorbable implant for ACL repair is subject to the following special controls:
*Classification.* Class II (special controls). The special controls for this device are:(1) Clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and include the following:
(i) Post-operative evaluation of knee pain and function; and
(ii) Durability as assessed by re-tear or re-operation rate.
(2) Animal performance testing must demonstrate that the device performs as intended under anticipated conditions of use and include the following:
(i) Device performance characteristics, including resorption and ligament healing at repair site; and
(ii) Adverse effects as assessed by gross necropsy and histopathology.
(3) Non-clinical testing must demonstrate that the device performs as intended under anticipated conditions of use and include the following:
(i) Characterization of materials, including chemical composition, resorption profile, and mechanical properties; and
(ii) Simulated use testing, including device preparation, device handling, compatibility with other ACL repair instrumentation, and user interface.
(4) The device must be demonstrated to be biocompatible.
(5) Performance data must demonstrate the device to be sterile and non-pyrogenic.
(6) Performance data must support the shelf life of the device by demonstrating continued sterility, package integrity, and device functionality over the identified shelf life.
(7) Labeling must include the following:
(i) Identification of device materials and specifications;
(ii) A summary of the clinical performance testing conducted with the device;
(iii) Instructions for use, including compatibility with other ACL repair instrumentation or devices;
(iv) Warnings regarding post-operative rehabilitation requirements; and
(v) A shelf life.
Submission Summary (Full Text)
{0}------------------------------------------------
## DE NOVO CLASSIFICATION REQUEST FOR BEAR® (BRIDGE-ENHANCED ACL REPAIR) IMPLANT
### REGULATORY INFORMATION
FDA identifies this generic type of device as:
Resorbable implant for anterior cruciate ligament (ACL) repair. A resorbable implant for anterior cruciate ligament (ACL) repair is a degradable material that allows for healing of a torn ACL that is biomechanically stabilized by traditional suturing procedures. The device is intended to protect the biological healing process from the surrounding intraarticular environment and not to replace biomechanical fixation via suturing. This can include devices that bridge or surround the torn ends of a ruptured ACL.
NEW REGULATION NUMBER: 21 CFR 888.3044
CLASSIFICATION: Class II
PRODUCT CODE: QNI
### BACKGROUND
DEVICE NAME: BEAR® (Bridge-Enhanced ACL Repair) Implant
SUBMISSION NUMBER: DEN200035
DATE DE NOVO RECEIVED: June 4, 2020
#### SPONSOR INFORMATION:
Miach Orthopaedics, Inc. 69 Milk Street, Suite 100 Westborough, Massachusetts 01581
### INDICATIONS FOR USE
The BEAR® (Bridge-Enhanced ACL Repair) Implant is indicated as follows:
The BEAR® (Bridge Enhanced ACL Repair) Implant is a bovine extracellular matrix collagen-based implant for treatment of anterior cruciate ligament (ACL) injuries. The BEAR® Implant is indicated for skeletally-mature patients at least 14 years of age with a complete rupture of the ACL, as confirmed by MRI. Patients must have an ACL stump attached to the tibia to construct the repair.
{1}------------------------------------------------
# LIMITATIONS
The sale, distribution, and use of the BEAR® Implant are restricted to prescription use in accordance with 21 CFR 801.109.
## PLEASE REFER TO THE LABELING FOR A COMPLETE LIST OF WARNINGS, PRECAUTIONS AND CONTRAINDICATIONS.
### DEVICE DESCRIPTION
The BEAR® Implant (22 mm in diameter and 45mm in length) is cylindrical in shape and comprised of collagen and extracellular matrix derived from bovine connective tissue, which has been cleaned, disinfected and processed by a proprietary manufacturing method. The implant has been terminally sterilized by electron-beam irradiation and is intended to be used with up to 10 ml of autologous blood drawn during the surgical implantation procedure. The BEAR® Implant stabilizes the blood in the gap between the torn ligament ends. The BEAR® Implant is resorbed within 8 weeks and replaced with a fibrovascular repair tissue.
Image /page/1/Picture/5 description: The image shows a series of five diagrams illustrating a medical procedure on a knee joint. The diagrams depict the progression of the procedure, starting with the initial state of the knee and ending with the final placement of a prosthetic component. The diagrams show the knee joint from a side view, with the femur and tibia bones clearly visible. The procedure involves the use of various medical instruments and materials, such as sutures, screws, and a prosthetic component, to repair or replace damaged tissue in the knee joint.
### SUMMARY OF NONCLINICAL/BENCH STUDIES
### BIOCOMPATIBILITY/MATERIALS
The BEAR® Implant is manufactured from the following materials:
| Description | Material | Direct Patient<br>Contact | Contact Duration |
|-------------|---------------------------------------------|---------------------------|-------------------|
| Implant | Bovine collagen and<br>extracellular matrix | Yes | Permanent (>30 d) |
Biocompatibility evaluation has been completed according to FDA Guidance, Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process"
### SHELF LIFE/STERILITY
{2}------------------------------------------------
## E-beam Sterilization:
The subject implant is provided sterile to the end user. The sterilization method is e-beam radiation at a dose of " kGy. Sterilization was validated using the VDmax method as per ISO 11137-1:2006(R/2018) Sterilization of health care products – Radiation Requirements for development, validation, and routine control of a sterilization process for medical devices to ensure that a Sterility Assurance Level (SAL) of 10th is achieved.
Representative sterilized samples real-time aged to " years were used to determine the shelf life of the device. Seal width, seal strength, and package integrity (bubble test) were used on accelerated aged samples to determine the sterile barrier packaging shelf life. Non-clinical performance testing of the representative devices was used to assess the performance shelf life.
## Viral Inactivation and Titer Testing:
The Viral Inactivation properties of the BEAR® Implant manufacturing process have model viruses and at least (b) (4) been validated to at least (b) (4) reduction for (b) (4) reduction for the (b) (4) model virus, following the guidance in ISO 22442-3:2007 Medical devices utilizing animal tissues and their derivatives- Part 3: Validation of the elimination and/or inactivation of viruses and transmissible spongiform encephalopathy (TSE), as well as the FDA guidance document "Medical Devices Containing Materials Derived from Animal Sources (Except for In Vitro Diagnostic Devices, March 2019)". This addresses the ability of the process to inactivate/eliminate viruses that might enter the process via the tissue used as a starting material. In addition, viral safety was confirmed via titer testing of representative product lots and incorporation of viral titer testing into the lot release criteria.
### Reprocessing:
There are no reusable or reprocessed components in this device.
### MAGNETIC RESONANCE (MR) COMPATIBILITY
The BEAR® Implant is a non-ferromagnetic, collagen-based material. The subject device is considered MR Safe.
### PERFORMANCE TESTING - BENCH
The sponsor provided both biochemical characterization and bench performance testing to demonstrate the device's ability to absorb blood and be sutured.
| Test | Purpose | Method | Acceptance Criteria | Results |
|----------------------|----------------------|---------------------------------|---------------------|---------------------------|
| Collagen content | Lot release criteria | Biochemical<br>characterization | >(b) (4) | Mean (b) (4) |
| DNA content | Lot release criteria | Biochemical<br>characterization | < (b) (4) | Mean: (b) (4) |
| Phospholipid content | Lot release criteria | Biochemical<br>characterization | <(b) (4) | Mean: <loq< td=""></loq<> |
{3}------------------------------------------------
| Pepsin activity | Lot release criteria | Biochemical<br>characterization | <(b) (4) | Mean: <LOQ |
|-------------------|----------------------|-------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------|------------------------------------------------------------------------------------|
| GAG content | Lot release criteria | Biochemical<br>characterization | >(b) (4) | Mean: (b) (4) µg/g |
| SDS-PAGE | Lot release criteria | Biochemical<br>characterization | Presence of (b) (4)<br>(b) (4)<br>typical of Type I<br>Collagen | All samples show α,<br>β, and γ protein<br>banding typical of<br>Type I Collagen |
| DSC | Lot release criteria | Biochemical<br>characterization | (b) (4) °C average<br>peak<br>temperature | Mean: (b) (4) °C |
| Endotoxin content | Lot release criteria | LAL test per<br>ANSI/AAMI ST72 | <(b)(4) EU/device | Mean: <(b) (4)<br>EU/device |
| Density | Lot release criteria | Mass and<br>dimensional<br>measurement | (b) (4) | Mean:<br>(b) (4)<br>g/cm3 |
| Blood absorption | Structural integrity | 10mm thick disc<br>sample placed in<br>blood | (b) (4) | Mean absorption<br>time: (b) (4)<br>sec;<br>Mean height<br>reduction:<br>(b) (4) % |
| Keith needle test | Structural integrity | Functional testing of<br>the device's ability<br>to retain mass while<br>having 4 Keith<br>needles with sutures<br>passed through its<br>length | (b) (4) | Mean: (b) (4) % |
## PERFORMANCE TESTING - ANIMAL
(b) (4) ACL transection models were utilized for pivotal animal studies on an investigational version of the subject device. These tests were utilized for design validation of the sterilization process, histologic evidence of device resorption within 8 weeks, and biomechanical evaluation of healed ligaments.
| | VIV-003 | VIV-004 |
|-------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Title | Retention Time of E-Beam Sterilized Miach<br>[Implant] in the Porcine Knee | Aseptic vs. Ebeam Process In Vivo Trial |
| Date of Study | July 2013 | November 2013 |
| Objective | To determine how long residual particles of e-beam<br>sterilized BEAR® Implant are found in synovium,<br>ligament and popliteal lymph nodes after<br>implantation in ACL wound site. | To determine if terminal sterilization with e-beam<br>has any significant effect on the mechanical<br>properties of primary ACL repairs performed with<br>the BEAR® Implant eight weeks in vivo. |
| Animal Model | (b) (4) | (b) (4) |
| Number of Animals | (b) | (b) (4) per group) |
| Study Design | ACL transection created surgically in one knee of<br>each animal; ACL repaired with BEAR Implant;<br>treated ACL and contralateral ACL harvested and<br>examined histopathologically; synovium and<br>popliteal lymph nodes also harvested and examined<br>histopathologically | ACL transection created surgically in one knee of<br>each animal; ACL repaired with BEAR Implant;<br>treated ACL and contralateral ACL harvested,<br>tested for biomechanical function and examined<br>histopathologically; synovium and popliteal lymph<br>nodes also harvested and examined<br>histopathologically |
{4}------------------------------------------------
| | VIV-003 | VIV-004 |
|--------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Survival | (b) animals each at 4 weeks and 6 weeks | 8 weeks for all animals in both groups |
| Test Article | BEAR® Implant (30 mm x 22 mm) sterilized by e-<br>beam at (b)(4)kGy | Group 1: BEAR® Implant (30 mm x 22 mm)<br>sterilized by e-beam at (d)(4) kGy<br>Group 2: BEAR® Implant (30 mm x 22 mm)<br>aseptically prepared, no terminal sterilization |
| Results | No serious adverse effects to ACL, synovium, or<br>popliteal lymph nodes. Implant material resorbed<br>rapidly with near complete resorption by 6 weeks<br>post-surgery. Implant material was associated with<br>an expected mild mononuclear inflammatory<br>reaction that was not considered excessive, likely<br>contributing to its resorption and possibly helping<br>establish a framework for local healing. BEAR®<br>synovium mildly inflamed and hyperplastic<br>compared to control joints. This was considered a<br>normal reaction to surgery and expected to resolve<br>over time. BEAR®-related popliteal lymph nodes<br>had mild follicular and paracortical hyperplasia and<br>sinus histiocytosis. This is consistent with<br>inflammation associated with surgery. No implant<br>was visualized in lymph nodes. | Biomechanical properties (linear stiffness, yield<br>load, maximum load ratio, anteroposterior [AP]<br>laxity) numerically lower in e-beam group than<br>aseptic group but differences not statistically<br>significant.<br>Devices in both groups completely resorbed by 8<br>weeks post-surgery.<br>Overall, no significant differences in the histologic<br>appearance of the ligament, synovium or popliteal<br>lymph nodes between the two treatment groups. |
| Conclusions | Near-complete resorption of the implant by 6 weeks<br>post-surgery; no evidence of implant material in<br>synovium or popliteal lymph nodes. | Minor alterations in the implant itself caused by<br>irradiation most likely do not have a detrimental<br>effect on the outcome after repair. Electron beam<br>irradiation at (b)(4) kGy effectively sterilizes the<br>implant without significantly harming the in vivo<br>function of the implant as indicated by<br>histopathological and biomechanical testing. |
## SUMMARY OF CLINICAL INFORMATION
### Study Design
There were two completed clinical studies using the BEAR® Implant, including an early feasibility study (BEAR I; G140151) with "" patients, "" of whom received the BEAR® Implant, and a larger pivotal study (BEAR II; G150268). The sponsor relied on the pivotal BEAR II study to support the clinical performance of the BEAR® Implant.
In the BEAR II Study, the BEAR® Implant was studied in a randomized (2:1 ratio) controlled trial of (4) subjects with complete ACL rupture, performed at one U.S. site by three surgeons. subjects were randomized to the BEAR® Implant and "" to the control treatment. (b) (4) ACL reconstruction (ACLR) with autograft (b)(0) subjects received a hamstring graft and received a bone-patellar-tendon-bone [BPTB] graft). Following surgery, subjects underwent a prescribed physical therapy regimen and were followed up at 1-2 and 6 weeks, and 3, 6, 12 and 24 months. Various outcomes were measured at the follow-up visits, including patient-reported outcomes, strength and functional measurements and imaging (X-ray, magnetic resonance imaging [MRI]). The primary endpoints, International Knee Documentation Committee (IKDC) Subjective Score, KT-instrumented AP knee laxity, and various safety parameters, were evaluated at 24 months (two years) post-surgery.
The primary analysis population was the modified Intent-to-Treat (mITT) population, which consisted of all ITT patients who had the BEAR procedure attempted. This included 10/4 subjects, in the BEAR group and in the control group to (4) subjects were consented and
{5}------------------------------------------------
randomized, but did not undergo surgery for various reasons. Thus, the ITT population was 109 subjects. The As-Treated (AT) population, analyzed for safety, was the same as the mITT population.
## Subject Demographics
Subjects participating in the BEAR II study were young, with an overall mean age of 19.6±5.2 years and a median age of 17.5 years; overall, 64.2% of subjects were 18 years and younger, and 35.8% were 19 years and older. To be eligible for the study, all patients had to have closed femoral and tibial physes and were therefore skeletally mature. More females than males were enrolled in the study (55.0% female, 45.0% male). Time from injury to surgery averaged 35.5±7.9 days, with a range of 12.0 to 46.0 days. There were no significant differences between the treatment groups at baseline.
| | BEAR<br>N=73 | Control<br>N=36 | Total<br>N=109 | p-value [1] |
|------------------------------------|-------------------|-------------------|-------------------|-------------|
| Age (years) [2] | | | | |
| Mean ± SD (N) | 19.5 ± 5.2 (73) | 19.8 ± 5.3 (36) | 19.6 ± 5.2 (109) | 0.784 |
| Median (Min, Max) | 17.4 (13.8, 35.6) | 17.7 (14.1, 35.6) | 17.5 (13.8, 35.6) | |
| Age Group % (n/N) | | | | 0.674 |
| 18 Years Old and Under | 65.8% (48/73) | 61.1% (22/36) | 64.2% (70/109) | |
| 19 Years Old and Over | 34.2% (25/73) | 38.9% (14/36) | 35.8% (39/109) | |
| Gender % (n/N) | | | | 0.838 |
| Female | 56.2% (41/73) | 52.8% (19/36) | 55.0% (60/109) | |
| Male | 43.8% (32/73) | 47.2% (17/36) | 45.0% (49/109) | |
| BMI (kg/m^2) | | | | |
| Mean ± SD (N) | 24.7 ± 3.8 (72) | 23.5 ± 4.6 (36) | 24.3 ± 4.1 (108) | 0.147 |
| Median (Min, Max) | 24.5 (18.1, 36.9) | 22.2 (17.2, 38.3) | 24.0 (17.2, 38.3) | |
| Time from Injury to Surgery (days) | | | | |
| Mean ± SD (N) | 34.7 ± 8.1 (65) | 36.9 ± 7.6 (35) | 35.5 ± 7.9 (100) | 0.189 |
| Median (Min, Max) | 36.0 (12.0, 46.0) | 39.0 (15.0, 46.0) | 37.5 (12.0, 46.0) | |
## Safety Endpoints
There were no cases of deep joint infection or incision and drainage of deep surgical site infection and no evidence of graft/implant rejection in either group. Graft or repair failure occurred in nine BEAR subjects (13.8%) and two control subjects (5.7%), p=0.320. Additional surgical procedures (other than ACL surgery) were required on the study knee in eight BEAR subjects (12.3%) and four control subjects (11.4%), p=1.000. Neither comparison reached statistical significance. Boyine IgE antibody levels were positive at the 6-month follow-up in two BEAR subjects (3.1%) and no control subjects; both results were low positive (0.39 kUL, just
{6}------------------------------------------------
slightly above the threshold of 0.35 kUL) and resolved at 15 months and two years post-surgery. Neither subject had any adverse events related to the transient antibody elevation.
| | BEAR<br>N=65 | Control<br>N=35 | p-value<br>[1] |
|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------|-----------------|----------------|
| Deep Joint Infection/Incision and Drainage of Deep<br>Surgical Site Infection | 0% (0/65) | 0% (0/35) | 1.000 |
| Evidence of Graft or Implant Rejection | 0% (0/65) | 0% (0/35) | 1.000 |
| Graft or Repair Failure | 13.8% (9/65) | 5.7% (2/35) | 0.320 |
| Additional Surgical Procedures Required on Study<br>Knee [2] | 12.3% (8/65) | 11.4% (4/35) | 1.000 |
| Bovine IgE Ant body Levels >=0.35kU/L [3] | 3.1% (2/64) | 0% (0/33) | 0.546 |
| Bovine Antibody Level (kU/L) | | | |
| Mean ± SD (N) | 0.39 $\pm$ 0.00 (2) | | |
| Median (Min, Max) | 0.39 (0.39,<br>0.39) | | |
| [1] p-value from a two-sided Fisher's Exact Test, testing the null hypothesis that the true proportions<br>are equal for the two treatments versus the alternative hypothesis that they are not equal.<br>[2] Not including subjects requiring a second ACL surgery.<br>[3] Subjects who tested positive resolved after 15 months and 2 years post procedure date. | | | |
Graft or repair failure was determined by positive pivot shift exam, Lachman exam with >6 mm side to side difference, absence of tissue in expected ACL location on MRI, evidence of graft or repair loss of continuity on MRI or symptomatic instability requiring revision ACL surgery. Of the nine BEAR subjects who experienced repair failure, five were non-compliant with postoperative requirements (physical therapy and/or brace use), returned to sports prior to surgeon clearance, had an accident or had a very high body mass index (BMI), and three returned to sports prior to 9 months post-surgery. All subjects who re-tore the ACL, in both groups, were age 18 years or younger. Results of the BEAR II study were compared to data from a historical control for which the manufacturer was able to access subject-level data and to data from a structured literature review. The analyses demonstrated that the rate of ACL re-tear with the BEAR® Implant was similar to the historical control and was consistent with the published literature. In conclusion, the BEAR® Implant had a similar safety profile to ACLR, and repair failure was more likely to occur in younger subjects, which is consistent with the experience of ACLR as documented in the literature.
## Primary Effectiveness Endpoints
The BEAR II Study had two co-primary effectiveness endpoints. IKDC score and instrumented AP knee laxity, both at 24 months (two years) post-surgery (Table 3). In the primary analysis of the mITT population using multiple imputation for missing data, IKDC score for the BEAR group at 24 months was found to be non-inferior to control based on the null hypothesis that the true difference in the means between treatment groups was less than or equal to -11.5, which is considered a clinically significant difference and was the pre-specified non-inferiority delta.
{7}------------------------------------------------
Mean IKDC score in the BEAR group was 88.6±13.4 and in the control group 84.6±13.3. The 95% confidence interval for the difference in the means was 4.03 (-1.55, 9.61) (p<0.001).
Instrumented AP knee laxity using the KT device at 24 months was found to be non-inferior to control based on the null hypothesis that the true difference in the means between treatment groups was greater than or equal to 2.0 mm, which is considered a clinically significant difference and was the pre-specified non-inferiority delta. Mean instrumented AP knee laxity in the BEAR group was 1.7±3.2 mm and in the control group 1.8±2.8 mm. The 95% confidence interval for the difference in the means was -0.10 (-1.45, 1.25) (p<0.001).
Both primary endpoints were confirmed by multiple sensitivity analyses. including a tipping point analysis.
| | BEAR<br>N=65 | Control<br>N=35 | Difference in Means<br>BEAR - Control<br>(95% CI) [2] | p-value |
|-------------------------------------------------------------------------------------------|--------------------------|--------------------------|-------------------------------------------------------|---------|
| IKDC Patient Reported Score at 24 Months [3] | | | | |
| Mean ± SD | 88.6 ± 13.4 | 84.6 ± 13.3 | 4.03 (-1.55, 9.61) | <0.001 |
| Median (Min, Max) [4] | 91.95 (35.63,<br>100.00) | 89.08 (47.13,<br>100.00) | | |
| KT Instrumented AP Knee Laxity (mm) at 24 Months<br>(Injured Knee - Non-Injured Knee) [5] | | | | |
| Mean ± SD | 1.7 ± 3.2 | 1.8 ± 2.8 | -0.10 (-1.45, 1.25) | 0.001 |
| Median (Min, Max) [4] | 1.88 (-8.50, 7.00) | 1.38 (-6.00, 6.00) | | |
[1] Analysis done on mITT population with multiple imputation used for missing data. In the BEAR group, 3 (4,6%) patients are missing IKOC and 7 (10.8%) patients are missing AP knee laxity at 24 months. In the control group. 1 (2.9%) patient is missing IKDC and 3 (8.6%) are missing AP knee laxity at 24 months.
[2] Confidence interval based on the t-distribution.
[3] p-value from a one-sided, two-sample t-test of the null hypothesis that the true difference in means is less than or equal to -11,5 versus the alternative hypothesis that it is greater than -11.5.
[4] Median, minimum and maximum values are shown for the observed data only, and do not include imputed values.
[5] p-value from a one-sided, two-sample t-test of the true difference in means is greater than or equal to 2.0 versus the alternative hypothesis that it is less than 2.0.
## Secondary Effectiveness Endpoints
Twelve secondary effectiveness endpoints were statistically tested using multiple imputation for missing data and were tested hierarchically in the order specified below to control the Type I error rate and adjust for multiple testing, whereby further testing would stop if a result was not significant. These endpoints were:
- . Hamstring strength, reported as percentage of the contralateral side, and as determined by hand-held dynamometer at 6 months post-surgery (superiority)
- . Hamstring strength. reported as percentage of the contralateral side, as determined by handheld dynamometer at 12 months post-surgery (superiority)
- . Hamstring to quadriceps ratio for the operated knee at 6 months post-surgery (superiority)
- . Hamstring to quadriceps ratio for the operated knee at 12 months post-surgery (superiority)
- . ACL Return-to-Sport Index (RSI) score at 6 months post-surgery (superiority),
{8}------------------------------------------------
- 9 Knee Injury and Osteoarthritis Outcome Score (KOOS) at 12 months post-surgery - Pain (non-inferiority)
- KOOS at 12 months post-surgery Symptoms (non-inferiority) 9
- KOOS at 12 months post-surgery Sports and Recreation (non-inferiority) .
- KOOS at 12 months post-surgery Ouality of Life (OOL: non-inferiority) .
- . KOOS at 12 months post-surgery - Activities of Daily Living (ADL; non-inferiority)
- . KOOS at 12 months post-surgery - Pain (superiority)
- . KOOS at 12 months post-surgery - Symptoms (superiority)
All 12 endpoints were statistically significant, either for non-inferiority or for superiority, as defined in the statistical analysis plan (SAP). Prone hamstring strength and hamstring to quadriceps ratio, both tested for superiority at both 6- and 12-months post-surgery, were significantly better in the BEAR group than the control group. Mean prone hamstring strength. which is measured as the proportion of the strength of the injured knee to the non-injured knee, was more than (absolute) 30% higher in the BEAR group than control at 6 months (on average. 93.3% vs. 59.1%, respectively [p<0.001]), and this finding was sustained at 12 months (on average, 96.6% vs. 65.2%. respectively [p<0.001]). Similarly, mean hamstring to quadriceps ratio at 6 months was 0.5 ± 0.2 in the BEAR group vs. 0.3 ± 0.1 in the control group (p<0.001); at 12 months, the difference between treatment groups was slightly smaller but still statistically significant in favor of BEAR (0.4+0.1 vs. 0.3+0.1, p<0.001).
The mean ACL RSI in the BEAR group was superior to control by 12 points at 6 months postsurgery (71.5±19.5 compared to 58.9±24.1, p=0.005), the timepoint that was tested for this analysis.
All five KOOS domains, including pain, symptoms, sports and recreation, QOL and ADL, were tested for non-inferiority at 12 months; all were statistically significant for non-inferiority, and in all cases the BEAR value was numerically higher than the control value. KOOS-pain and KOOS-symptoms were also tested for superiority at 12 months and found to be significantly better in the BEAR group than control.
| | BEAR<br>N=65 | Control<br>N=35 | Difference in Means<br>BEAR - Control<br>(95% CI) [2] | p-value [3] |
|-------------------------------------------------------------------------------------------------------------|---------------------|---------------------|-------------------------------------------------------|-------------|
| Prone Hamstring Strength at 6 Months (%)<br>(100*(Injured Knee/Non-injured Knee)) (superiority)<br>[4] | | | | |
| Mean ± SD | 93.3 ± 23.6 | 59.1 ± 21.3 | 34.21 (24.70, 43.72) | <0.001 [S] |
| Median (Min, Max) [5] | 91.7 (29.6, 188.5) | 56.4 (27.0, 124.0) | | |
| Prone Hamstring Strength at 12 Months (%)<br>(100*(Injured Knee/Non-injured Knee)) (superiority)<br>[6] | | | | |
| Mean ± SD | 96.6 ± 16.7 | 65.2 ± 18.5 | 31.37 (24.08, 38.66) | <0.001 [S] |
| | BEAR<br>N=65 | Control<br>N=35 | Difference in Means<br>BEAR - Control<br>(95% CI) [2] | p-value [3] |
| Median (Min, Max) [5] | 96.8 (40.0, 164.0) | 61.9 (36.0, 114.5) | | |
| Hamstring to Quadriceps Ratio at 6 Months<br>(Hamstring Strength/Quadriceps Strength)<br>(superiority) [4] | | | | |
| Mean ± SD | 0.5 ± 0.2 | 0.3 ± 0.1 | 0.16 (0.10, 0.22) | <0.001 [S] |
| Median (Min, Max) [5] | 0.4 (0.2, 1.2) | 0.3 (0.1, 0.7) | | |
| Hamstring to Quadriceps Ratio at 12 Months<br>(Hamstring Strength/Quadriceps Strength)<br>(superiority) [6] | | | | |
| Mean ± SD | 0.4 ± 0.1 | 0.3 ± 0.1 | 0.13 (0.09, 0.17) | <0.001 [S] |
| Median (Min, Max) [5] | 0.4 (0.2, 0.7) | 0.3 (0.2, 0.5) | | |
| ACL RSI Score at 6 Months (superiority) [7] | | | | |
| Mean ± SD | 71.5 ± 19.5 | 58.9 ± 24.1 | 12.59 (3.74, 21.44) | 0.005 [S] |
| Median (Min, Max) [5] | 75.0 (0.8, 100.0) | 64.2 (11.7, 95.0) | | |
| KOOS at 12 months (Pain) (non-inferiority) [8] | | | | |
| Mean ± SD | 94.4 ± 6.6 | 91.2 ± 7.1 | 3.19 (0.37, 6.02) | <0.001 [N] |
| Median (Min, Max) [5] | 97.2 (66.7, 100.0) | 91.7 (77.8, 100.0) | | |
| KOOS at 12 months (Symptoms) (non-inferiority) [8] | | | | |
| Mean ± SD | 88.3 ± 9.3 | 82.4 ± 12.0 | 5.87 (1.54, 10.19) | <0.001 [N] |
| Median (Min, Max) [5] | 89.3 (57.1, 100.0) | 85.7 (57.1, 100.0) | | |
| KOOS at 12 months (Sports and Recreation) (non-<br>inferiority) [8] | | | | |
| Mean ± SD | 86.0 ± 15.7 | 83.0 ± 18.9 | 2.96 (-4.05, 9.98) | <0.001 [N] |
| Median (Min, Max) [5] | 87.5 (15.0, 100.0) | 85.0 (15.0, 100.0) | | |
| KOOS at 12 months (Quality of Life) (non-inferiority)<br>[8] | | | | |
| Mean ± SD | 69.4 ± 19.7 | 64.6 ± 17.5 | 4.76 (-3.19, 12.72) | <0.001 [N] |
| Median (Min, Max) [5] | 68.8 (25.0, 100.0) | 62.5 (37.5, 100.0) | | |
| KOOS at 12 months (Activities of Daily Living) (non-<br>inferiority) [8] | | | | |
| Mean ± SD | 98.8 ± 2.4 | 98.0 ± 4.2 | 0.74 (-0.59, 2.07) | <0.001 [N] |
| Median (Min, Max) [5] | 100.0 (88.2, 100.0) | 100.0 (77.9, 100.0) | | |
| KOOS at 12 months (Pain) (superiority) [8] | | | | |
| Mean ± SD | 94.4 ±6.6 | 91.2 ± 7.1 | 3.19 (0.37, 6.02) | 0.027 [S] |
| Median (Min, Max) [5] | 97.2 (66.7, 100.0) | 91.7 (77.8, 100.0) | | |
| | BEAR<br>N=65 | Control<br>N=35 | Difference in Means<br>BEAR - Control<br>(95% CI) [2] | p-value [3] |
| KOOS at 12 months (Symptoms) (superiority) [8] | | | | |
| Mean ± SD | 88.3 ± 9.3 | 82.4 ± 12.0 | 5.87 (1.54, 10.19) | 0.008 [S] |
| Median (Min, Max) [5] | 89.3 (57.1, 100.0) | 85.7 (57.1, 100.0) | | |
The secondary endpoints were confirmed with sensitivity analysis.
{9}------------------------------------------------
{10}------------------------------------------------
N=non-inferiority test; S=superiority test
[1] Analysis done on mITT population with multiple imputation used for missing data.
[2] Confidence interval based on the t-distribution.
[3] These are to be Tested in a hierarchical manner so that if a significant result is reached the next variable will be testlit is not significant (p>0.05) then testing will not continue.
For tests of superiority, the p-value is from a two-sided, two-sample t-test, testing the true means are equal versus the alternative hypothesis that they are not equal.
For tests of non-inferiority, the p-value is from a one-sided, two-sample t-test of the true difference in means is less than or equal to -10 versus the alternative hypothesis that it is greater than -10.
[4] Data for prone hamsting strength and hamsting to quadriceps ratio at 6 months was imputed for 1 (1.5%) patient in the BEAR group, and 1 (2.9%) patient in the control group.
[5] Median, minimum and maximum values are shown for the observed data only, and do not include imputed values.
[6] Data for prone hamstring strength and hamsting to quadriceps ratio at 12 morths was imputed for 3 (4.6%) patients in the BEAR group, and 3 (8.6%) patients in the control group.
[7] Data for ACL RSI Score at 6 months was imputed for 1 (1.5%) patient in the BEAR group, and 1 (2.9%) patient in the control group.
[8] Data for KOOS (all parts) at 12 months was imputed for 1 (1.5%) patient in the BEAR group, and 2 (5.7%) of patients in the control group.
## LABELING
The labeling consists of the following: device description, indications for use, instructions for use including surgical steps, compatibility of device with other soft tissue repair devices, principles of device operation, identification of device materials, contraindications, warnings, precautions, MR compatibility, a list of potential adverse effects, importance of patient compliance with post-operative activity restrictions, and a summary of the clinical data. Furthermore, the sterile packaging includes a shelf life for the device. The labeling meets the requirements of 21 CFR 801.109 for prescription devices.
### RISKS TO HEALTH
The table below identifies the risks to health that may be associated with use of the resorbable implant for ACL repair and the measures necessary to mitigate these risks.
| Identified Risks to Health | Mitigation Measures |
|--------------------------------------------------------------------------------|------------------------------------------------------------|
| Repaired ACL has inadequate durability,<br>leading to re-tear | Animal testing<br>Clinical performance testing<br>Labeling |
| Repaired ACL is loose or functionally limited,<br>leading to joint instability | Clinical performance testing |
| ACL does not heal due to inadequate<br>resorption or migration of implant | Non-clinical performance testing<br>Animal testing |
| Adverse tissue reaction | Biocompatibility evaluation<br>Labeling |
{11}------------------------------------------------
| Infection | Sterilization validation |
|---------------------------------------------------------------------------------------------------------------------------|----------------------------------|
| | Shelf life testing |
| | Labeling |
| Febrile response due to endotoxins | Pyrogenicity testing |
| Implant is incompatible with other ACL repair<br>instrumentation and sutures, leading to<br>inability to complete surgery | Non-clinical performance testing |
| | Labeling |
## SPECIAL CONTROLS
In combination with the general controls of the FD&C Act, the resorbable implant for ACL repair is subject to the following special controls:
- (1) Clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and include the following:
- (i) Post-operative evaluation of knee pain and function; and
- (ii) Durability as assessed by re-tear or re-operation rate.
- (2) Animal performance testing must demonstrate that the device performs as intended under anticipated conditions of use and include the following:
- (i) Device performance characteristics, including resorption and ligament healing at repair site; and
- (ii) Adverse effects as assessed by gross necropsy and histopathology.
- (3) Non-clinical testing must demonstrate that the device performs as intended under anticipated conditions of use and include the following:
- Characterization of materials, including chemical composition, resorption profile, (i) and mechanical properties; and
- (ii) Simulated use testing, including device preparation, device handling, compatibility with other ACL repair instrumentation, and user interface.
- (4) The device must be demonstrated to be biocompatible.
- (5) Performance data must demonstrate the device to be sterile and non-pyrogenic.
- (6) Performance data must support the shelf life of the device by demonstrating continued sterility, package integrity, and device functionality over the identified shelf life.
- (7) Labeling must include the following:
- (i) Identification of device materials and specifications;
- (ii) A summary of the clinical performance testing conducted with the device;
- (iii) Instructions for use, including compatibility with other ACL repair instrumentation or devices;
- (iv) Warnings regarding post-operative rehabilitation requirements; and
- (v) A shelf life.
## BENEFIT-RISK DETERMINATION
The sponsor has collected adequate data to assess the safety profile of the subject device and has identified that there are benefits. Compared to the standard-of-care ACL reconstruction procedures, treatment with the subject device results in no donor site morbidity, which is confirmed via superiority in hamstring strength secondary endpoints at 6 and 12 months post
{12}------------------------------------------------
operative. The KOOS pain and function subscales and RSI scores also demonstrated superiority at the 6- and 12-month post-operative time points. There is also a presumed benefit from a repair procedure preserving more native anatomy than a reconstruction, which requires wider bone tunnels. Device-related serious adverse events such as infection or rejection/immunogenic response were not observed in the clinical data and are mitigated by design controls and processing controls. Serious adverse events that necessitated reoperation (i.e., re-tear) were observed with similar frequency between ACL repairs with the subject device and ACL reconstructions. In conclusion, the benefits of using the subject device for its intended use/indications for use outweigh the risks to health.
## PATIENT PERSPECTIVES
This submission did not include specific information on patient perspectives for this device.
### BENEFIT/RISK CONCLUSION
In conclusion, given the available information above, for the following indication statement:
The BEAR® (Bridge Enhanced ACL Repair) Implant is a bovine extracellular matrix collagen-based implant for treatment of anterior cruciate ligament (ACL) injuries. The BEAR® Implant is indicated for skeletally-mature patients at least 14 years of age with a complete rupture of the ACL, as confirmed by MRI. Patients must have an ACL stump attached to the tibia to construct the repair.
The probable benefits outweigh the probable risks for the BEAR® Implant. The device provides benefits and the risks can be mitigated by the use of general and the identified special controls.
### CONCLUSION
The De Novo request for the BEAR® (Bridge-Enhanced ACL Repair) Implant is granted and the device is classified as follows:
Product Code: QNI Device Type: Resorbable implant for anterior cruciate ligament (ACL) repair Regulation Number: 21 CFR 888.3044 Class: II
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.