XIA 3 SPINAL SYSTEM, UNIPLANNER/REDUCTION SCREWS & VITALLIUM ROD

K091291 · Stryker Spine · NKB · Jun 24, 2009 · Orthopedic

Device Facts

Record IDK091291
Device NameXIA 3 SPINAL SYSTEM, UNIPLANNER/REDUCTION SCREWS & VITALLIUM ROD
ApplicantStryker Spine
Product CodeNKB · Orthopedic
Decision DateJun 24, 2009
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 888.3070
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Xia 3 Spinal System is intended for use in the noncervical spine. When used as an anterior/anterolateral and posterior, noncervical pedicle and non-pedicle fixation, the Xia® 3 Spinal System is intended to provide additional support during fusion using autograft or allograft in skeletally mature patients in the treatment of the following acute and chronic instabilities or deformities: degenerative disc disease (DDD) (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); spondylolisthesis; trauma (i.e., fracture or dislocation); spinal stenosis; curvatures (i.e., scoliosis, kyphosis, and/or lordosis); tumor; pseudoarthrosis; and failed previous fusion. The Ø5.5mm titanium and Vitallium® rods from the Stryker Spine Radius® Spinal System and Ø6.0mm Vitallium® rods from Xia® Spinal System are intended to be used with the other components of Xia 3 Spinal System.

Device Story

Xia 3 Spinal System line extension; provides spinal fixation/support during fusion. Components include titanium Uniplanar Screws, Uniplanar Reduction Screws, and Vitallium rods (Ø5.5mm and Ø6.0mm). Used in noncervical spine; implanted by surgeons during spinal fusion procedures. System stabilizes spine segments using autograft or allograft; facilitates bone healing; addresses instabilities/deformities. No software or algorithmic components.

Clinical Evidence

Bench testing only; no clinical data provided.

Technological Characteristics

Spinal fixation system; materials include titanium and Vitallium®; components include Uniplanar Screws, Uniplanar Reduction Screws, and rods (Ø5.5mm, Ø6.0mm). Mechanical assembly; non-powered; non-software.

Indications for Use

Indicated for skeletally mature patients requiring noncervical spinal fixation (anterior/anterolateral/posterior) for DDD, spondylolisthesis, trauma, spinal stenosis, curvatures (scoliosis, kyphosis, lordosis), tumor, pseudoarthrosis, or failed fusion.

Regulatory Classification

Identification

(1) Rigid pedicle screw systems are comprised of multiple components, made from a variety of materials that allow the surgeon to build an implant system to fit the patient's anatomical and physiological requirements. Such a spinal implant assembly consists of a combination of screws, longitudinal members (e.g., plates, rods including dual diameter rods, plate/rod combinations), transverse or cross connectors, and interconnection mechanisms (e.g., rod-to-rod connectors, offset connectors).(2) Semi-rigid systems are defined as systems that contain one or more of the following features (including but not limited to): Non-uniform longitudinal elements, or features that allow more motion or flexibility compared to rigid systems.

Special Controls

*Classification.* (1) Class II (special controls), when intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra; degenerative spondylolisthesis with objective evidence of neurologic impairment; fracture; dislocation; scoliosis; kyphosis; spinal tumor; and failed previous fusion (pseudarthrosis). These pedicle screw spinal systems must comply with the following special controls:(i) Compliance with material standards; (ii) Compliance with mechanical testing standards; (iii) Compliance with biocompatibility standards; and (iv) Labeling that contains these two statements in addition to other appropriate labeling information: “Warning: The safety and effectiveness of pedicle screw spinal systems have been established only for spinal conditions with significant mechanical instability or deformity requiring fusion with instrumentation. These conditions are significant mechanical instability or deformity of the thoracic, lumbar, and sacral spine secondary to severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra, degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudarthrosis). The safety and effectiveness of these devices for any other conditions are unknown.” “Precaution: The implantation of pedicle screw spinal systems should be performed only by experienced spinal surgeons with specific training in the use of this pedicle screw spinal system because this is a technically demanding procedure presenting a risk of serious injury to the patient.” (2) Class II (special controls), when a rigid pedicle screw system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion in the treatment of degenerative disc disease and spondylolisthesis other than either severe spondylolisthesis (grades 3 and 4) at L5-S1 or degenerative spondylolisthesis with objective evidence of neurologic impairment. These pedicle screw systems must comply with the following special controls: (i) The design characteristics of the device, including engineering schematics, must ensure that the geometry and material composition are consistent with the intended use. (ii) Non-clinical performance testing must demonstrate the mechanical function and durability of the implant. (iii) Device components must be demonstrated to be biocompatible. (iv) Validation testing must demonstrate the cleanliness and sterility of, or the ability to clean and sterilize, the device components and device-specific instruments. (v) Labeling must include the following: (A) A clear description of the technological features of the device including identification of device materials and the principles of device operation; (B) Intended use and indications for use, including levels of fixation; (C) Identification of magnetic resonance (MR) compatibility status; (D) Cleaning and sterilization instructions for devices and instruments that are provided non-sterile to the end user; and (E) Detailed instructions of each surgical step, including device removal. (3) Class II (special controls), when a semi-rigid system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion for any indication. In addition to complying with the special controls in paragraphs (b)(2)(i) through (v) of this section, these pedicle screw systems must comply with the following special controls: (i) Demonstration that clinical performance characteristics of the device support the intended use of the product, including assessment of fusion compared to a clinically acceptable fusion rate. (ii) Semi-rigid systems marketed prior to the effective date of this reclassification must submit an amendment to their previously cleared premarket notification (510(k)) demonstrating compliance with the special controls in paragraphs (b)(2)(i) through (v) and paragraph (b)(3)(i) of this section.

Predicate Devices

Submission Summary (Full Text)

{1}------------------------------------------------ K091291 Special 510(k) Premarket Notification # Special 510(k) Summary of Safety and Effectiveness: JUN 2 4 2009 | Proprietary Name: | Xia® 3 Spinal System – Line Extension | |------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Common Name: | Spinal Fixation Appliances | | Classification Name and Reference: | 1) Spinal Interlaminal Fixation Orthosis, 21 CFR §888.3050<br>2) Spinal Intervertebral Body Fixation Orthosis, 21 CFR §888.3060<br>3) Pedicle Screw Spinal System, 21 CFR §888.3070 (b) (1) & (b) (2) | | Device Product Code: | NKB, KWP, KWQ, MNH, MNI | | Proposed Regulatory Class: | Class III | | For Information contact: | Curtis Truesdale<br>Regulatory Affairs Project Manager<br>2 Pearl Court<br>Allendale, NJ 07401<br>Telephone: (201) 760-8296<br>Fax: (201) 760-8496<br>Email: Curtis.Truesdale@Stryker.com | | Date Summary Prepared: | April 30, 2009 | | Predicate Devices | Stryker Spine Xia® 3 Spinal System, K071373; Stryker Spine Xia® Spinal System, K080928; Stryker Spine Xia® Spinal System, K063428; Stryker Spine Radius® Spinal System, K062270, K082608; Stryker Spine Xia II Spinal System, K013823; and DePuy Spine Moss Miami Spinal System, K950697 | {2}------------------------------------------------ K091291 #### Xia 3 Spinal System - Line Extension Special 510(k) Premarket Notification #### Description of Device Modification Summary of the Technological Characteristics Intended Use This 510(k) is intended to introduce an extension to the existing Xia 3 Spinal System. The proposed line extension includes the addition of titanium Uniplanar Screws, titanium Uniplanar Reduction Screws, and use of the Stryker Spine Radius Ø5.5 mm Vitallium® rod with the Xia® 3 Spinal System. The Xia 3 Spinal System is intended for use in the noncervical spine. When used as an anterior/anterolateral and posterior, noncervical pedicle and non-pedicle fixation, the Xia® 3 Spinal System is intended to provide additional support during fusion using autograft or allograft in skeletally mature patients in the treatment of the following acute and chronic instabilities or deformities: degenerative disc disease (DDD) (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); spondylolisthesis; trauma (i.e., fracture or dislocation); spinal stenosis; curvatures (i.e., scoliosis, kyphosis, and/or lordosis); tumor; pseudoarthrosis; and failed previous fusion. The Ø5.5mm titanium and Vitallium® rods from the Stryker Spine Radius® Spinal System and Ø6.0mm Vitallium® rods from Xia® Spinal System are intended to be used with the other components of Xia 3 Spinal System. The Stryker Spine Xia® 3 Spinal System, with the incorporation of the subject components, is substantially equivalent to the predicate devices in terms of material, design, and indications for use. Engineering analysis and testing to demonstrate compliance with FDA's Guidance for Spinal System 510(k)'s May 3, 2004 were completed for the Stryker Spine Xia 3 Spinal System. Pg 2 of 2 {3}------------------------------------------------ DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three stripes forming its body and wing. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle. Public Health Service JUN 24 2009 Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 Stryker Spine % Mr. Curtis D. Truesdale Regulatory Affairs Project Manager 2 Pearl Court Allendale, New Jersey 07401 Re: K091291 Trade/Device Name: Stryker Spine XIA® 3 Spinal System-Line Extension Regulation Number: 21 CFR 888.3070 Regulation Name: Pedicle Screw Spinal System Regulatory Class: III Product Code: NKB, MNH, MNI, KWP, KWQ Dated: June 2, 3009 Received: June 3, 2009 Dear Mr. Truesdale: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions a jainst misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing {4}------------------------------------------------ Page 2-Mr. Curtis D. Truesdale practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/cdrh/mdr/ for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours, Saubare Bucher M. N. Muller Mark N. Melkerso Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ XIA 3 Spinal System -- Line Extension ## Indications for Use ### 510(k) Number (if known): Krog 129 Device Name: Stryker Spine XIA 9 3 Spinal System - Line Extension Indications for Use: The Stryker Spine XIA 3 Spinal System is intended for use in the noncervical spine. When used as an anterior/anterolateral and posterior, noncervical pedicle and non-pedicle fixation system, the XIA 3 Spinal System is intended to provide additional support during fusion using autograft or allograft in skeletally mature patients in the treatment of the following acute and chronic instabilities or deformities: - Degenerative disc disease (DDD) (defined as back pain of discogenic origin with . degeneration of the disc confirmed by history and radiographic studies); - Spondylolisthesis; - Trauma (i.e., fracture or dislocation); - Spinal Stenosis: - Curvatures (i.e., scoliosis, kyphosis, and/or lordosis); - Tumor; - Pseudoarthrosis; and - Failed previous fusion. The Ø5.5 mm titanium and Vitallium rods from the Stryker Spine Radius® Spinal System and Ø6.0 mm Vitallium rods from XIA Spinal System are intended to be used with the other components of XIA® 3 Spinal System. | Prescription Use _X_ | AND/OR | Over-The-Counter Use _ | |------------------------|--------|------------------------| | (21 CFR 801 Subpart D) | | (21 CFR 807 Subpart C) | (PLEASE DO NOT WRITE BELOW THIS LINE -- CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) 2-2 (EXT for UyxM) (Division Sign-Off) Division of Surgics Orthopedic, and Restorative Devices Number K 091291 Pg 1 of 1
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...