Sterile Bioabsorbable bone screw (Bioabsorbable ACL screw)
K210122 · Osteonic Co., Ltd. · MAI · Aug 27, 2021 · Orthopedic
Device Facts
Record ID
K210122
Device Name
Sterile Bioabsorbable bone screw (Bioabsorbable ACL screw)
Applicant
Osteonic Co., Ltd.
Product Code
MAI · Orthopedic
Decision Date
Aug 27, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3030
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Sterile bioabsorbable bone screw (Biocomposite ACL screw) is a fixation screw that fixes soft tissue such as ligaments, tendons, and the articular capsules to bone, and is used in orthopedic surgery(Cruciate ligament reconstruction procedures).
Device Story
Sterile bioabsorbable bone screw (Biocomposite ACL screw) is an implantable device used in orthopedic surgery to fix soft tissues (ligaments, tendons, articular capsules) to bone during cruciate ligament reconstruction. The screw is composed of poly(L-lactide-co-glycolide) (PLGA) and tri-calcium phosphate (β-TCP). It is designed to be gradually decomposed in vivo via hydrolysis, with PLGA excreted as carbon dioxide and water, and β-TCP re-used as a calcium source. The device is supplied sterile via ethylene oxide gas in an aluminum pouch for single-use application by orthopedic surgeons. It provides mechanical fixation of grafts to bone, facilitating healing and reconstruction.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing including axial pullout strength, torsional yield strength, driving torque, degradation testing per ISO 13781, and fatigue testing (1000 cycles at 1 Hz). Biocompatibility evaluated per ISO 10993.
Technological Characteristics
Material: Poly(L-lactide-co-glycolide) (PLGA) and tri-calcium phosphate (β-TCP). Form factor: Bone screw. Sterilization: Ethylene oxide (EO) gas. Standards: ASTM F2502 (mechanical), ISO 13781 (degradation), ISO 10993 (biocompatibility), ISO 11135/11138/11140/11737/11607 (sterilization/packaging).
Indications for Use
Indicated for fixation of soft tissue (ligaments, tendons, articular capsules) to bone in orthopedic surgery, specifically cruciate ligament reconstruction procedures.
Regulatory Classification
Identification
Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.
{0}------------------------------------------------
August 27, 2021
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the seal of the Department of Health & Human Services USA. To the right of the seal is the FDA logo, with the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, and "ADMINISTRATION" in a smaller font below.
OSTEONIC Co., Ltd. % Sanglok Lee Manager WISE COMPANY Inc. #507, #508, 166 Gasan digital 2-ro Geumcheon-gu, Seoul 08507 South Korea
#### Re: K210122
Trade/Device Name: Sterile Bioabsorbable Bone Screw (Biocomposite ACL Screw) Regulation Number: 21 CFR 888.3030 Regulation Name: Single/Multiple Component Metallic Bone Fixation Appliances And Accessories Regulatory Class: Class II Product Code: MAI Dated: July 20, 2021 Received: July 22, 2021
#### Dear Sanglok Lee:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
{1}------------------------------------------------
devices or postmarketing safety reporting (21 CFR 4. Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE(@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Laura C. Rose -S
Laura C. Rose, Ph.D. Assistant Director DHT6C: Division of Restorative, Repair and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K210122
Device Name
Sterile bioabsorbable bone screw (Biocomposite ACL screw)
Indications for Use (Describe)
Sterile bioabsorbable bone screw (Biocomposite ACL screw) is a fixation screw that fixes soft tissue such as ligaments, tendons, and the articular capsules to bone, and is used in orthopedic surgery(Cruciate ligament reconstruction procedures).
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the Osteonic logo. The logo consists of a blue geometric shape on the left and the word "OSTEONIC" in dark gray on the right. The geometric shape is a diamond with a smaller diamond inside it.
# 1206Ho, 38, Digital-ro 29 Tel: +82(2)-6902-8400 Fax: +82(2)-6902-8401 Home page: http://www.osteonic.com
K210122 - Page 1 of 5
# The assigned 510(k) Number: K210122
# 1. 510(k) Summary Date: 15.01.2021
# 2. Applicant
DaKyung Ham OSTEONIC Co., Ltd. 303Ho, 405Ho, 505-3Ho, 505-3Ho, 902Ho, 1004Ho, 1201Ho, 1206Ho, 38 Digital-ro 29-gil, Gurogu, Seoul, Korea, 102Ho, 103Ho, 24, Digital-ro 27-gil, Guro-gu, Seoul, Korea TEL: +82-2-6902-8426 FAX: +82-2-6902-8401 Email: dakham@osteonic.com
# 3. Submission Correspondent
Sang-lok, Lee Wise COMPANY Inc. #507, #508, 166 Gasan digital 2-ro, Geumcheon-gu, Seoul, Korea TEL: +82 70 8812 3619 / +82 2 831 3615 FAX: +82 50 4031 3619 Email: info@wisecompany.org
# 4. Subject Device Identification
- 1) Trade Name: Sterile bioabsorbable bone screw (Biocomposite ACL screw)
- 2) Common Name: Sterile bioabsorbable bone screw (Biocomposite ACL screw)
- 3) Classification Name: fastener, fixation, biodegradable, soft tissue
- 4) Product Code: MAI
- 5) Panel: Orthopedic
- 6) Regulation Number: 21 CFR 888.3030
- 7) Device Class: II
# 5. Indication for use
Sterile bioabsorbable bone screw (Biocomposite ACL screw) is a fixation screw that fixes soft tissue such as ligaments, tendons, and the articular capsules to bone, and is used in orthopedic surgery(Cruciate ligament reconstruction procedures).
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image contains the logo for Osteonic. The logo consists of a blue geometric shape resembling a stylized diamond or square with an opening in the center, positioned to the left of the company name. The text "OSTEONIC" is written in a bold, dark gray sans-serif font to the right of the geometric shape.
1206Ho. 38. Digital-ro 29 Tel: +82(2)-6902-8400 Fax: +82(2)-6902-8401 Home page: http://www.osteonic.com
## K210122 - Page 2 of 5
## 6. Predicate devices
Predicate device
- 1) 510(k) Number: K143660
- 2) Device Name: Milagro/Milagro Advance Interference Screw
- 3) Manufacturer: DePuy Mitek, Inc.
Reference device
- 1) 510(k) Number: K103831
- 2) Device Name: Milagro Interference screw
- 3) Manufacturer: DePuy Mitek, Inc.
Reference device
- 1) 510(k) Number: K202806
- 2) Device Name: Fix2Lock (Biocomposite medial, Biocomposite lateral, Biocombi SelfPunching)
3) Manufacturer: Osteonic Co., Ltd.
# 7. Device Description
The Sterile bioabsorbable bone screw (Biocomposite ACL screw) is an absorbable bone fixation screw that fixes soft tissues such as ligament, tendon, and the articular capsules to bone, and is used in orthopedic surgery.
The implanted screw is made of poly(L-lactide-co-glycolide), PLGA and tri-calciumphosphate(S-TCP). Poly(L-lactide-co-glycolide), PLGA is gradually decomposed in vivo by hydrolysis and eventually excreted in carbon dioxide and water. Tri-calciumphosphate(ß-TCP) is gradually degraded in vivo by hydrolysis and eventually re-used as a calcium source.
Sterile bioabsorbable bone screw (Biocomposite ACL screw) is supplied EO gas sterile state and it is packed in Aluminum Pouch.
# 8. Non-Clinical Test Conclusion
- 1) General Information
Bench tests were conducted to verify that the subject device met all design specifications as was Substantially Equivalent (SE) to the predicate device. The following standards were used to support substantial equivalence:
#### ▪ Mechanical Performance
- ASTM F2502:17 Standard Specification and test methods for absorbable plates and screws for internal fixation implants
- ISO 13781:2017 Implants for surgery Homopolymers, copolymers and blends on poly(lactide) In vitro degradation testing
- = Sterilization, Shelf-life and Packaging for Sterile Product
- ISO 11135:2014, Sterilization of health care products Ethylene oxide Requirements for the development validation and routine control of a sterilization and routine control of a sterilization process for medical devices.
- ISO 11138-1:2006, Sterilization of health care products Biological indicators Part 1: General requirements
- ISO 11138-2:2009, Sterilization of health care products Biological indicators Part 2: Biological
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the Osteonic logo. The logo consists of a blue geometric shape on the left and the word "OSTEONIC" in dark gray on the right. The geometric shape is a diamond with a stylized "O" inside, created by overlapping lines. The word "OSTEONIC" is in a bold, sans-serif font.
1206Ho. 38. Digital-ro 29-gil. Tel: +82(2)-6902-8400 Fax: +82(2)-6902-8401 Home page: http://www.osteonic.com
## K210122 - Page 3 of 5
indicators for ethylene oxide sterilization processes
- ISO 11140-1:2014, Sterilization of health care products Chemical indicators Part 1: General requirements
- ISO 11737-1:2018 Sterilization of medical devices Microbiological methods- Part 1: Estimation of population of microorganisms on products
- ISO 11737-2:2009 Sterilization of medical devices Microbiological methods- Part 2: Tests of sterility performed in the validation of a sterilization process
- ISO 11607-1:2006/AMD1:2014 Packaging for terminally sterilized medical devices part 1: requirements for materials, sterile barrier systems and packaging system
- ISO 11607-2:2006/AMD1:2014 Packaging for terminally sterilized medical devices Part 2: Validation requirements for forming, sealing and assembly processes
- ASTM F1980:2016 Standard guide for accelerated aging of sterile barrier systems for medical devices
- ASTM F88/F88M:2015 Standard test method for seal strength of flexible barrier materials.
- ASTM F1929:2015 Standard test method for detecting seal leaks in porous medical packaging by dye penetration
#### ▪ Bacterial Endotoxin
- USP <85> Bacterial Endotoxin Test
- USP <161> Medical Devices-Bacterial Endotoxin and Pyrogen Tests
#### 2) Summary of Biocompatibility
Sterile bioabsorbable bone screw (Biocomposite ACL screw) has been evaluated for biocompatibility according to ISO 10993.
#### 3) Summary of Performance
All specimens selected (worst-case) to verify the performance of the test criteria established based on literatures and comparison with the predicate device. Therefore, the entire model was proven.
| No. | Test Article | Test Method | Result |
|-----|--------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------|
| 1 | Axial Pullout Strength | Test referring ASTM F2502-A3.Test Method for Axial<br>Pullout of Absorbable Bone Screws.<br>After fully inserting the screw into the artificial bone (20PCF),<br>Apply tensile load to the specimen at a rate 5mm/min until the<br>screw fails or releases from the block. Measure the maximum<br>load. | PASS |
| 2 | Torsional Yield Strength | Test referring ASTM F2502-A1.Test Method for Determining<br>the Torsion Properties of Absorbable Bone Screws.<br>After placing the specimen in the holding device, Rotate the<br>screw in inserting direction at a rate of 1 revolution/min.<br>Torsional yield strength is determined by an off-set method (2<br>degree off-set) on the torque versus rotation angle curve<br>measured in the test. | PASS |
| 3 | Driving Torque | Test referring ASTM F2502-A2.Test Method for Driving<br>Torque of Absorbable Bone Screws.<br>Make a hole in the fixed artificial bone (20PCF) using an<br>instrument compatible with the product. Measure the<br>maximum strength when the screw is inserted 4 revolutions at<br>a rate of 5 revolutions per minute.<br>* Inserting load (axial load) = 1.14kgf | PASS |
| 4 | Degradation testing | Mechanical performance assessed after degradation according<br>to ISO 13781 | PASS |
| 5 | Fatigue | After fully inserting the screw with the tendon into the fixed<br>artificial bone (20PCF), a fatigue tensile load of 70 to 220N | PASS |
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for Osteonic. The logo consists of a blue diamond shape with a stylized "O" inside, followed by the word "OSTEONIC" in bold, dark blue letters. The logo is simple and modern, with a focus on the company name.
OSTEONIC Tel: +82(2)-6902-8400 Fax: +82(2)-6902-8401 Home page: http://www.osteonic.com
K210122 - Page 4 of 5
on the tendon is repeated 1000 cycles at 1 Hz.
# 9. Substantially Equivalent Conclusion
#### Table 1: Substantial Equivalence Comparison
| Product<br>Name | SUBJECT Device | PREDICATE Device<br>Milagro/Milagro Advance<br>Interference Screw (K143660) | REFERENCE Device 1<br>Milagro Interference screw<br>(K103831) | Equivalence<br>Discussion |
|-------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------|
| Product<br>code | MAI | MAI | MAI, HWC | Same |
| Regulatory<br>class | Class II | Class II | Class II | Same |
| Regulation<br>Number | 21 CFR 888.3030 | 21 CFR 888.3030 | 21 CFR 888.3040 | Same |
| Intended<br>Use | Sterile bioabsorbable bone<br>screw (Biocomposite ACL<br>screw) is a fixation screw that<br>fixes soft tissue such as<br>ligaments, tendons, and the<br>articular capsules to bone, and<br>is used in orthopedic<br>surgery(Cruciate ligament<br>reconstruction procedures). | Milagro Interference<br>screw(5x12,6x12,7x15,8x15)<br>The MILAGRO BR<br>Interference screws are<br>designed to attach soft tissues<br>to bone in Ortohpedic surgical<br>procedures for the following<br>indications:<br>Shoulder : Proximal Biceps<br>Tenodesis, Acromio-<br>Clavicular Repair<br>Elbow: Distal Biceps<br>Tenodesis, Ulnar Collateral<br>Ligament Repair<br>Knee: Collateral Ligament<br>Repair, Medial Patellofemoral<br>Ligament Reconstruction<br>(patella fixation)<br>Milagro Advance Interference<br>screw(7x23,8x23,9x23)<br>The MILAGRO ADVANCE<br>Interference screw is intended<br>for attachment of soft tissue<br>grafts or bone-tendon-bone<br>grafts to the tibia and/or femer<br>during cruciate ligament<br>reconstruction procedures.<br>Additionally, the 7, 8 and<br>9mm x 23mm screws are<br>indicated for : medial and<br>lateral collateral ligament<br>repair, medial patellofemoral<br>ligament reconstruction(femer<br>fixation) of the knee, proximal<br>bicep tenodesis in the shoulder<br>and distal bicep tenodesis in<br>the elbow. | Knee :<br>- Attachment of a bone-tendon-<br>bone(BTB)graft to the tibia<br>and/or femur during cruciate<br>ligament<br>reconstruction<br>procedure.<br>- Attachment of a soft tissue<br>(ST)graft to the tibia and/or<br>femur during cruciate ligament<br>reconstruction procedures.<br>- Medial and lateral collateral<br>ligament repair<br>Shoulder :<br>Proximal bicep tenodesis<br>Elbow :<br>Distal bicep tenodesis | Same |
| Operating<br>Principles | Sterile bioabsorbable bone<br>screw (Biocomposite ACL<br>screw) is a fixation screw that<br>fixes soft tissue such as<br>ligaments, tendons, and the<br>articular capsules to bone, and | Milagro/Milagro Advance<br>Interference Screw is a<br>fixation screw that fixes soft<br>tissue such as ligaments,<br>tendons, and the articular<br>capsules to bone, and is used | Milagro Interference screw<br>is a fixation screw that fixes<br>soft tissue such as ligaments,<br>tendons, and the articular<br>capsules to bone, and is used<br>in orthopedic surgery(Cruciate | Same |
| | | | | |
| | is used in orthopedic<br>surgery(Cruciate ligament<br>reconstruction procedures). | in orthopedic surgery(Cruciate<br>ligament reconstruction<br>procedures). | ligament reconstruction<br>procedures). | |
| Material | PLGA + β-TCP | Composite of β-TCP and<br>PLGA copolymer | Absorbable Poly(lactide -co-<br>glycolide) polymer and<br>Tricalcium Phosphate(TCP) | Same |
| Product<br>Size | Ø7.0 x (25, 30) mm<br>Ø8.0 x (23, 25, 28, 30) mm<br>Ø9.0 x (23, 25, 28, 30) mm<br>Ø10.0 x (23, 25, 28, 30) mm | Ø5.0 x 12 mm<br>Ø6.0 x 12 mm<br>Ø7.0 x (15, 23) mm<br>Ø8.0 x (15, 23) mm<br>Ø9.0 x 23 mm | Ø5.0 x (23, 30) mm<br>Ø6.0 x (23, 30) mm<br>Ø7.0 x (23, 30) mm<br>Ø8.0 x (23, 30) mm<br>Ø9.0 x (23, 30, 35) mm<br>Ø10.0 x (23, 30, 35) mm<br>Ø11.0 x (30, 35) mm<br>Ø12.0 x (30, 35) mm | Similar |
| Sterilization | Sterile (EO gas sterilization) | Sterile (EO gas sterilization) | Sterile (EO gas sterilization) | Same |
| Single Use/<br>Reuse | Single use | Single use | Single use | Same |
| Packaging | 1 EA / BOX | 1 EA / BOX | 1 EA / BOX | Same |
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo for Osteonic. The logo consists of a blue geometric shape resembling a stylized diamond or square with internal lines, positioned to the left of the company name. The text "OSTEONIC" is written in a bold, dark gray sans-serif font.
1206Ho, 38, Digital-ro 29-gil, Guro-gu, Seoul, Korea Tel: +82(2)-6902-8400 Fax: +82(2)-6902-8401 Home page: http://www.osteonic.com
#### K210122 - Page 5 of 5
Based on above, the subject device, Sterile absorbable bone screw (Biocomposite ACL screw), is determined to be Substantially Equivalent (SE) to the predicate devices, Milagro Advance Interference Screw (K143660) in respect of safety and effectiveness.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.