Components of the Signature Orthopaedics hip replacement range are intended to replace a hip joint where bone stock is sufficient to support the implant. When a surgeon has selected prosthetic replacement as the preferred treatment, the devices are indicated for: - . Non-inflammatory degenerative joint disease including osteoarthritis or avascular necrosis - Inflammatory joint disease including rheumatoid arthritis - . Correction of functional deformity including congenital hip dysplasia - . Traumatic injury involving the hip joint including traumatic arthritis or femoral head or neck fracture - . Failed previous hip surgery including internal fixation or joint fusion, reconstruction, hemiarthroplasty, surface replacement, or total replacement Signature Orthopaedics' Origin, Aria, Remedy, TSI, Pegasus and Spartan femoral stems and Logical acetabular cups are intended for cementless fixation only. Signature Orthopaedics' Evolve, Cemented TSI (both CoCr and HNSS variants) and Cemented Origin femoral stems are intended for cemented fixation only. Signature Orthopaedics' constrained liner components are indicated particularly for patients at high risk of hip dislocation due to a history of prior dislocation, bone loss, joint or soft tissue laxity, neuromuscular disease or intraoperative instability. Signature Orthopaedics' Evolve UniPolar Head and BiPolar Head are intended for hemihip arthroplasty only, where the natural acetabulum does not require replacement. The Evolve UniPolar Head and BiPolar Head are indicated for bone fractures or pathologies involving only the femoral head/neck and/or proximal femur, such as: - . Acute femoral head or neck fracture - Fracture dislocation of the hip ● - Avascular necrosis of the femoral head - Non-union of femoral neck fractures ● - Certain high subcapital and femoral neck fractures in the elderly - Degenerative arthritis involving only the femoral head
Device Story
Spartan Hip Stem is a cementless femoral component for total hip arthroplasty; used by orthopedic surgeons in clinical settings. Device features triple-tapered geometry; manufactured from forged titanium alloy (Ti6Al4V) with proximal titanium and hydroxyapatite plasma spray coating for osseointegration. Implanted into the femur to replace the hip joint; provides structural support for prosthetic hip components. Benefits include restoration of joint function and stability in patients with degenerative or traumatic hip conditions. Surgeon selects device based on patient bone stock; device functions as a mechanical prosthesis.
Clinical Evidence
Bench testing only. No clinical or animal data required. Testing included ISO 21535 range of motion, stem/neck fatigue (FEA and physical testing), CoCr head disassembly, and ceramic head static/fatigue burst, pull-off, and torsion tests. Device validated as non-pyrogenic.
Technological Characteristics
Material: Forged titanium alloy (Ti6Al4V) per ISO 5832-3 and ASTM F136. Design: Triple-tapered geometry. Fixation: Cementless, with proximal titanium and hydroxyapatite plasma spray coating. Connectivity: None (mechanical implant). Sterilization: Not specified.
Indications for Use
Indicated for patients requiring hip joint replacement due to non-inflammatory degenerative joint disease (osteoarthritis, avascular necrosis), inflammatory joint disease (rheumatoid arthritis), functional deformity (congenital hip dysplasia), traumatic injury (traumatic arthritis, femoral head/neck fracture), or failed previous hip surgery. Constrained liners indicated for high dislocation risk (prior dislocation, bone loss, laxity, neuromuscular disease). Hemi-arthroplasty heads indicated for femoral head/neck fractures, fracture dislocation, avascular necrosis, non-union, or degenerative arthritis of the femoral head.
Regulatory Classification
Identification
A hip joint metal/ceramic/polymer semi-constrained cemented or nonporous uncemented prosthesis is a device intended to be implanted to replace a hip joint. This device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. The two-part femoral component consists of a femoral stem made of alloys to be fixed in the intramedullary canal of the femur by impaction with or without use of bone cement. The proximal end of the femoral stem is tapered with a surface that ensures positive locking with the spherical ceramic (aluminium oxide, A12 03 ) head of the femoral component. The acetabular component is made of ultra-high molecular weight polyethylene or ultra-high molecular weight polyethylene reinforced with nonporous metal alloys, and used with or without bone cement.
Signature Orthopaedics' Cemented TSI Stem (K181340)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square. To the right of that is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
August 5, 2020
Signature Orthopaedics Pty Ltd. Declan Brazil Managing Director 7 Sirius Road Lane Cove, 2066 Au
Re: K192883
Trade/Device Name: Spartan Hip Stem Regulation Number: 21 CFR 888.3353 Regulation Name: Hip Joint Metal/Ceramic/Polymer Semi-Constrained Cemented Or Nonporous Uncemented Prosthesis Regulatory Class: Class II Product Code: LZO Dated: July 1, 2020 Received: July 8, 2020
Dear Declan Brazil:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Vesa Vuniqi Assistant Director DHT6A: Division of Joint Arthroplasty Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### 1 INDICATIONS FOR USE STATEMENT
510(k) Number (if Known): K192883
Device Name: Spartan Hip Stem
Indications For Use:
Components of the Signature Orthopaedics hip replacement range are intended to replace a hip joint where bone stock is sufficient to support the implant. When a surgeon has selected prosthetic replacement as the preferred treatment, the devices are indicated for:
- . Non-inflammatory degenerative joint disease including osteoarthritis or avascular necrosis
- Inflammatory joint disease including rheumatoid arthritis
- . Correction of functional deformity including congenital hip dysplasia
- . Traumatic injury involving the hip joint including traumatic arthritis or femoral head or neck fracture
- Failed previous hip surgery including internal fixation or joint fusion, reconstruction, hemiarthroplasty, surface replacement, or total replacement
Signature Orthopaedics' Origin, Aria, Remedy, TSI, Pegasus and Spartan femoral stems and Logical acetabular cups are intended for cementless fixation only. Signature Orthopaedics' Evolve, Cemented TSI (both CoCr and HNSS variants) and Cemented Origin femoral stems are intended for cemented fixation only.
Signature Orthopaedics' constrained liner components are indicated particularly for patients at high risk of hip dislocation due to a history of prior dislocation, bone loss, joint or soft tissue laxity, neuromuscular disease or intraoperative instability.
Signature Orthopaedics' Evolve UniPolar Head and BiPolar Head are intended for hemihip arthroplasty only, where the natural acetabulum does not require replacement. The Evolve UniPolar Head and BiPolar Head are indicated for bone fractures or pathologies involving only the femoral head/neck and/or proximal femur, such as:
- . Acute femoral head or neck fracture
- Fracture dislocation of the hip ●
- Avascular necrosis of the femoral head
- Non-union of femoral neck fractures ●
- Certain high subcapital and femoral neck fractures in the elderly
- Degenerative arthritis involving only the femoral head
Prescription Use: Yes Over-The-Counter Use: No AND/OR (Part 29 CFR 801 Subpart D) (Part 29 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
{3}------------------------------------------------
#### 2 510(K) SUMMARY
| Manufacturer: | Signature Orthopaedics Pty Ltd<br>7 Sirius Road<br>Lane Cove, NSW 2066<br>Australia |
|-----------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Trade<br>Name: | Spartan Hip Stem |
| Common Name: | Cementless Hip Replacement Prosthesis |
| Contact: | Dr. Declan Brazil<br>Managing Director of Signature Orthopaedics |
| Prepared By: | Signature Orthopaedics Pty Ltd<br>7 Sirius Road<br>Lane Cove, NSW 2066<br>Australia<br>Phone: +61 (2) 9428 5181<br>Fax: +61 (2) 8456 6065 |
| Date Prepared: | 1st of July, 2020 |
| Classification: | Class II as per 21 CFR 888.3353 Hip joint<br>metal/ceramic/polymer semi-constrained cemented or<br>nonporous uncemented prosthesis (MEH)<br><br>Class II as per 21 CFR 888.3353 Hip joint<br>metal/ceramic/polymer semi-constrained cemented or<br>nonporous uncemented prosthesis. (LZO) |
| Predicate Devices: | Primary Predicate<br>• DePuy's Actis Stem (K150862, K160907)<br><br>Additional Predicate<br>• Signature Orthopaedics' Origin Stem (K121297, K161155)<br>• Signature Orthopaedics' Logical C-Series Shell (K153131)<br>• Signature Orthopaedics' Cemented TSI Stem (K181340) |
# Device Description:
Signature Orthopaedics' Spartan Hip Stem is manufactured from forged titanium alloy (Ti6Al4V) as per ISO 5832-3 and ASTM F136. The Spartan Stem has a triple tapered
{4}------------------------------------------------
geometry and features a titanium and hydroxyapatite plasma spray proximal coating below its resection line for cementless fixation in total hip arthroplasty.
## Indications for Use:
Components of the Signature Orthopaedics hip replacement range are intended to replace a hip joint where bone stock is sufficient to support the implant. When a surgeon has selected prosthetic replacement as the preferred treatment, the devices are indicated for:
- Non-inflammatory degenerative joint disease including osteoarthritis or . avascular necrosis
- . Inflammatory joint disease including rheumatoid arthritis
- Correction of functional deformity including congenital hip dysplasia .
- Traumatic injury involving the hip joint including traumatic arthritis or . femoral head or neck fracture
- . Failed previous hip surgery including internal fixation or joint fusion. reconstruction, hemiarthroplasty, surface replacement, or total replacement
Signature Orthopaedics' Origin, Aria, Remedy, TSI, Pegasus and Spartan femoral stems and Logical acetabular cups are intended for cementless fixation only. Signature Orthopaedics' Evolve, Cemented TSI (both CoCr and HNSS variants) and Cemented Origin femoral stems are intended for cemented fixation only.
Signature Orthopaedics' constrained liner components are indicated particularly for patients at high risk of hip dislocation due to a history of prior dislocation, bone loss, ioint or soft tissue laxity, neuromuscular disease or intraoperative instability.
Signature Orthopaedics' Evolve UniPolar Head and BiPolar Head are intended for hemi-hip arthroplasty only, where the natural acetabulum does not require replacement. The Evolve UniPolar Head and BiPolar Head are indicated for bone fractures or pathologies involving only the femoral head/neck and/or proximal femur, such as:
- Acute femoral head or neck fracture .
- . Fracture dislocation of the hip
- Avascular necrosis of the femoral head
- Non-union of femoral neck fractures
- o Certain high subcapital and femoral neck fractures in the elderly
- . Degenerative arthritis involving only the femoral head
## Summary of Technological Characteristics:
The Spartan Hip Stem shares the following technological characteristics as its predicate devices:
- . The Spartan Stem is intended for cementless fixation which is the same intended use as one or more of the predicate devices.
- The Spartan Stem has the same indications for use as its predicate devices.
- . The Spartan Stem is manufactured from the same materials as its predicate devices.
- The Spartan Stem has the same proximal coating as Signature Orthopaedics' Logical C-Series Shell.
- . The Spartan Stem's taper connection is identical to Signature Orthopaedics' Origin Stem and Cemented TSI Stem.
{5}------------------------------------------------
- The Spartan Stem has the same femoral head options as Signature Orthopaedics' Origin Stem and Cemented TSI Stem.
- . The Spartan Stem's geometry is similar to DePuy's Actis Stem.
- . The Spartan Stem has the same size range as DePuy's Actis Stem.
The following are the technological differences between the Spartan Stem and its predicate devices:
- . Some of the Spartan Stem's design features differ from the predicate devices.
- The Spartan Stem's coating differs from Signature Orthopaedics' Origin Hip Stem.
- The Spartan Stem's trunnion surface finish differs from Signature Orthopaedics' Cemented TSI Stem.
# Performance Testing:
Non-clinical testing and engineering evaluations were conducted to verify that the performance of the Spartan Hip Stem is adequate for anticipated in-vivo use. No animal or clinical testing was required to support substantial equivalence. Non-clinical testing carried out included:
- Range of Motion testing as per ISO 21535
- . Stem and Neck Fatigue FEA
- . Stem and Neck Fatigue Testing
- . CoCr Head Disassembly Test
- Ceramic Head Static Burst Test, Fatigue and Post-Fatigue Burst Test
- . Ceramic Head Pull-Off and Torsion Test
The subject device has been validated as non-pyrogenic.
## Substantial Equivalence Conclusion:
Signature Orthopaedics' Spartan Stem has the same intended use, indications for use, material and similar design features as DePuy's Actis Stem (K150862 and K160907), and Signature Orthopaedics' Origin Stem (K121297), Logical C-Series Shell (K153131) and Cemented TSI Stem (K181340) and non-clinical testing support the substantial equivalence claim.