Components of the Signature Orthopaedics hip replacement range are intended to replace a hip joint where bone stock is sufficient to support the implant. When a surgeon has selected prosthetic replacement as the preferred treatment, the devices are indicated for: - Non-inflammatory degenerative joint disease including osteoarthritis or avascular necrosis - . Inflammatory joint disease including rheumatoid arthritis - . Correction of functional deformity including congenital hip dysplasia - . Traumatic injury involving the hip joint including traumatic arthritis or femoral head or neck fracture - . Failed previous hip surgery including internal fixation or joint fusion, reconstruction, hemiarthroplasty, surface replacement, or total replacement Signature Orthopaedics' Origin, Aria, Remedy, TSI, Pegasus, Spartan, World and Encore Hip femoral stems, SignaSure Cementless Cups, Logical and World Acetabular Cups are intended for cementless fixation only. Signature Orthopaedics' Evolve, Cemented TSI (both CoCr and HNSS variants), and Cemented Origin femoral stems and SignaSure Cemented Cups are intended for cemented fixation only. Signature Orthopaedics' SignaSure Insert is indicated for use with a cementless Signature Orthopaedics' Logical Acetabular Cup to provide dual mobility articulation. Signature Orthopaedics' constrained liner components are indicated particularly for patients at high risk of hip dislocation due to a history of prior dislocation, bone loss, joint or soft tissue laxity, neuromuscular disease or intraoperative instability. Signature Orthopaedics' Evolve UniPolar Head and BiPolar Head are intended for hemi-hip arthroplasty only, where the natural acetabulum does not require replacement. The Evolve UniPolar Head and BiPolar Head are indicated for bone fractures or pathologies involving only the femoral head/neck and/or proximal femur, such as: - . Acute femoral head or neck fracture - Fracture dislocation of the hip - Avascular necrosis of the femoral head - Non-union of femoral neck fractures - . Certain high subcapital and femoral neck fractures in the elderly - . Degenerative arthritis involving only the femoral head
Device Story
Encore Hip Stem is a straight, tapered, rectangular-cross-section femoral stem for cementless hip arthroplasty. Implanted by orthopedic surgeons in clinical settings to replace diseased or damaged hip joints. Device features polished distal tip to minimize stress shielding and titanium/hydroxyapatite plasma-sprayed proximal coating to promote osseointegration. Surgeon selects appropriate stem size based on patient anatomy; device provides stable femoral fixation. Benefits include restoration of hip function and mobility for patients with degenerative or traumatic hip conditions.
Clinical Evidence
Bench testing only. No clinical or animal data required. Non-clinical testing included range of motion testing, stem and neck fatigue FEA, stem and neck fatigue testing, and rim impingement analysis.
Technological Characteristics
Forged titanium alloy (Ti6Al4V) per ISO 5832-3 and ASTM F136. Straight, tapered, rectangular cross-section geometry. Proximal titanium and hydroxyapatite plasma spray coating. Cementless fixation. Polished distal tip. Mechanical device; no software or electronics.
Indications for Use
Indicated for patients requiring hip joint replacement due to non-inflammatory degenerative joint disease (osteoarthritis, avascular necrosis), inflammatory joint disease (rheumatoid arthritis), functional deformity (congenital hip dysplasia), traumatic injury (traumatic arthritis, femoral head/neck fracture), or failed previous hip surgery. Specific components indicated for hemi-hip arthroplasty in patients with femoral head/neck fractures or pathologies. Constrained liners indicated for high dislocation risk patients.
Regulatory Classification
Identification
A hip joint metal/ceramic/polymer semi-constrained cemented or nonporous uncemented prosthesis is a device intended to be implanted to replace a hip joint. This device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. The two-part femoral component consists of a femoral stem made of alloys to be fixed in the intramedullary canal of the femur by impaction with or without use of bone cement. The proximal end of the femoral stem is tapered with a surface that ensures positive locking with the spherical ceramic (aluminium oxide, A12 03 ) head of the femoral component. The acetabular component is made of ultra-high molecular weight polyethylene or ultra-high molecular weight polyethylene reinforced with nonporous metal alloys, and used with or without bone cement.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The FDA logo is composed of two parts: the Department of Health and Human Services logo on the left and the FDA acronym and full name on the right. The Department of Health and Human Services logo is a stylized depiction of a human figure, while the FDA acronym and full name are written in blue.
September 24, 2021
Signature Orthopaedics Pty Ltd. Declan Brazil Managing Director 7 Sirius Road Lane Cove, NSW 2066 Australia
Re: K211505
Trade/Device Name: Encore Hip Stem Regulation Number: 21 CFR 888.3353 Regulation Name: Hip Joint Metal/Ceramic/Polymer Semi-Constrained Cemented Or Nonporous Uncemented Prosthesis Regulatory Class: Class II Product Code: MEH, LZO Dated: August 18, 2021 Received: August 24, 2021
Dear Declan Brazil:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Limin Sun, Ph.D. Acting Assistant Director DHT6A: Division of Joint Arthroplasty Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### INDICATIONS FOR USE STATEMENT 1
510(k) Number (if Known): K211505
Device Name: Encore Hip Stem
Indications For Use:
Components of the Signature Orthopaedics hip replacement range are intended to replace a hip joint where bone stock is sufficient to support the implant. When a surgeon has selected prosthetic replacement as the preferred treatment, the devices are indicated for:
- Non-inflammatory degenerative joint disease including osteoarthritis or avascular necrosis
- . Inflammatory joint disease including rheumatoid arthritis
- . Correction of functional deformity including congenital hip dysplasia
- . Traumatic injury involving the hip joint including traumatic arthritis or femoral head or neck fracture
- . Failed previous hip surgery including internal fixation or joint fusion, reconstruction, hemiarthroplasty, surface replacement, or total replacement
Signature Orthopaedics' Origin, Aria, Remedy, TSI, Pegasus, Spartan, World and Encore Hip femoral stems, SignaSure Cementless Cups, Logical and World Acetabular Cups are intended for cementless fixation only.
Signature Orthopaedics' Evolve, Cemented TSI (both CoCr and HNSS variants), and Cemented Origin femoral stems and SignaSure Cemented Cups are intended for cemented fixation only.
Signature Orthopaedics' SignaSure Insert is indicated for use with a cementless Signature Orthopaedics' Logical Acetabular Cup to provide dual mobility articulation.
Signature Orthopaedics' constrained liner components are indicated particularly for patients at high risk of hip dislocation due to a history of prior dislocation, bone loss, joint or soft tissue laxity, neuromuscular disease or intraoperative instability.
Signature Orthopaedics' Evolve UniPolar Head and BiPolar Head are intended for hemi-hip arthroplasty only, where the natural acetabulum does not require replacement. The Evolve UniPolar Head and BiPolar Head are indicated for bone fractures or pathologies involving only the femoral head/neck and/or proximal femur, such as:
- . Acute femoral head or neck fracture
- Fracture dislocation of the hip
- Avascular necrosis of the femoral head
- Non-union of femoral neck fractures
- . Certain high subcapital and femoral neck fractures in the elderly
- . Degenerative arthritis involving only the femoral head
Prescription Use: Yes Over-The-Counter Use: No AND/OR (Part 29 CFR 801 Subpart D) (Part 29 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
{3}------------------------------------------------
### 510(K) SUMMARY 2
| Manufacturer: | Signature Orthopaedics Pty Ltd<br>7 Sirius Road<br>Lane Cove, NSW 2066<br>Australia |
|-----------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Trade<br>Name: | Encore Hip Stem |
| Common Name: | Cementless Hip Replacement Prosthesis |
| Contact: | Dr. Declan Brazil<br>Managing Director of Signature Orthopaedics |
| Prepared By: | Signature Orthopaedics Pty Ltd<br>7 Sirius Road<br>Lane Cove, NSW 2066<br>Australia<br>Phone: +61 (2) 9428 5181<br>Fax: +61 (2) 8456 6065 |
| Date Prepared: | 18th of August, 2021 |
| Classification: | Class II as per 21 CFR 888.3353 Hip joint<br>metal/ceramic/polymer semi-constrained cemented or<br>nonporous uncemented prosthesis (MEH)<br>Class II as per 21 CFR 888.3353 Hip joint<br>metal/ceramic/polymer semi-constrained cemented or<br>nonporous uncemented prosthesis (LZO) |
| Predicate Devices: | Primary Predicate<br>• Signature Orthopaedics' Spartan Stem (K192883)<br>Reference Predicate<br>• Signature Orthopaedics' Origin Stem (K121297)<br>• Zimmer Biomet's Fitmore Stem (K071723) |
# Device Description:
Signature Orthopaedics' Encore Hip Stem is manufactured from forged titanium alloy (Ti6Al4V) as per ISO 5832-3 and ASTM F136. The Encore Stem is straight and tapered with a rectangular cross-section. The stem has a polished distal tip to reduce fixation leading to proximal stress shielding. It features a titanium and hydroxyapatite plasma spray proximal coating below its resection line.
### Indications for Use:
{4}------------------------------------------------
Components of the Signature Orthopaedics hip replacement range are intended to replace a hip joint where bone stock is sufficient to support the implant. When a surgeon has selected prosthetic replacement as the preferred treatment, the devices are indicated for:
- . Non-inflammatory degenerative joint disease including osteoarthritis or avascular necrosis
- . Inflammatory joint disease including rheumatoid arthritis
- . Correction of functional deformity including congenital hip dysplasia
- Traumatic injury involving the hip joint including traumatic arthritis or femoral head or neck fracture
- . Failed previous hip surgery including internal fixation or joint fusion. reconstruction, hemiarthroplasty, surface replacement, or total replacement
Signature Orthopaedics' Origin, Aria, Remedy, TSI, Pegasus, Spartan, World and Encore Hip femoral stems, SignaSure Cementless Cups, Logical and World Acetabular Cups are intended for cementless fixation only.
Signature Orthopaedics' Evolve, Cemented TSI (both CoCr and HNSS variants), and Cemented Origin femoral stems and SignaSure Cemented Cups are intended for cemented fixation only.
Signature Orthopaedics' SignaSure Insert is indicated for use with a cementless Signature Orthopaedics' Logical Acetabular Cup to provide dual mobility articulation. Signature Orthopaedics' constrained liner components are indicated particularly for patients at high risk of hip dislocation due to a history of prior dislocation, bone loss, joint or soft tissue laxity, neuromuscular disease or intraoperative instability. Signature Orthopaedics' Evolve UniPolar Head and BiPolar Head are intended for hemi-hip arthroplasty only, where the natural acetabulum does not require replacement. The Evolve UniPolar Head and BiPolar Head are indicated for bone fractures or pathologies involving only the femoral head/neck and/or proximal femur, such as:
- . Acute femoral head or neck fracture
- . Fracture dislocation of the hip
- Avascular necrosis of the femoral head .
- . Non-union of femoral neck fractures
- Certain high subcapital and femoral neck fractures in the elderly ●
- . Degenerative arthritis involving only the femoral head
# Summary of Technological Characteristics:
The Encore Hip Stem shares the following technological characteristics as its predicate devices:
- . The Encore Stem is intended for cementless fixation which is the same intended use as its predicate device.
- . The Encore Stem has the same indications for use as its predicate device.
- . The Encore Stem is manufactured from the same materials as its predicate device.
- . The Encore Stem has the same proximal coating as its predicate device.
- . The Encore Stem has the same taper connection as its predicate device.
- . The Encore Stem has the same surface finish as its predicate device
- . The Encore Stem has the same femoral head options as its predicate device.
{5}------------------------------------------------
The following are the technological differences between the Encore Stem and its predicate device:
- . The Encore Stem is sterilised via a different method compared to its predicate device
# Performance Testing:
Non-clinical testing and engineering evaluations were conducted to verify that the performance of the Encore Hip Stem is adequate for anticipated in-vivo use. No animal or clinical testing was required to support substantial equivalence. Non-clinical testing carried out included:
- Range of Motion Testing
- . Stem and Neck Fatigue FEA
- . Stem and Neck Fatigue Testing
- . Rim Impingement Analysis
# Substantial Equivalence Conclusion:
Signature Orthopaedics' Encore Stem has the same intended use, indications for use, material and similar design features as Signature Orthopaedics' Spartan Stem (K192883) and non-clinical testing support the substantial equivalence claim.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.