K190535 · Waldemar Link GmbH & Co. KG · LZO · Aug 6, 2019 · Orthopedic
Device Facts
Record ID
K190535
Device Name
BiMobile Dual Mobility System - E-Dur Inserts
Applicant
Waldemar Link GmbH & Co. KG
Product Code
LZO · Orthopedic
Decision Date
Aug 6, 2019
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3353
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The BiMobile Dual Mobility System is indicated for patients with mobility-limiting diseases, fractures or defects of the hip joint or proximal femur, which cannot be treated by conservative or osteosynthetic procedures. The BiMobile Dual Mobility System is indicated for the following conditions: - · Primary and secondary osteoarthritis - · Rheumatoid arthritis - · Correction of functional deformities - Avascular necrosis - Femoral neck fractures - · Revision after implant loosening dependent on bone mass and quality - · Dislocation risks The device is intended for cemented and cementless use.
Device Story
BiMobile Dual Mobility System is a hip acetabular implant system; provides dual mobility articulation. Subject device adds Vitamin E highly crosslinked UHMWPE (E-Dur) liners to existing system. System components include CoCrMo metal shells (cemented or cementless with PlasmaLink/TiCaP coatings) and UHMWPE liners; compatible with previously cleared femoral heads, stems, and neck segments. Used by orthopedic surgeons in hip arthroplasty procedures to restore joint function, reduce dislocation risk, and address bone defects/fractures. Output is a reconstructed hip joint interface; clinical benefit includes improved mobility and stability for patients with degenerative or traumatic hip conditions.
Clinical Evidence
No clinical data. Substantial equivalence demonstrated via non-clinical bench testing, including material characterization, biocompatibility, axial disassembly, lever-out testing, standard/adverse wear analyses (including particle analysis), and range of motion analysis.
Technological Characteristics
Acetabular cup system. Shells: Wrought CoCrMo (ISO 5832-12). Coatings: PlasmaLink (Ti plasma spray, ISO 5832-2/ASTM F1580) or TiCaP (Ti plasma spray/CaP, ISO 5832-2/ASTM F1580/F1609). Liners: Vitamin E highly crosslinked UHMWPE (E-Dur, ASTM F2695, ASTM F2565) or conventional UHMWPE (ISO 5834-2/ASTM F648). Modular design for cemented/cementless use.
Indications for Use
Indicated for patients with mobility-limiting hip/proximal femur disease, fractures, or defects untreatable by conservative/osteosynthetic methods. Conditions include primary/secondary osteoarthritis, rheumatoid arthritis, functional deformities, avascular necrosis, femoral neck fractures, revision for implant loosening, and dislocation risks.
Regulatory Classification
Identification
A hip joint metal/ceramic/polymer semi-constrained cemented or nonporous uncemented prosthesis is a device intended to be implanted to replace a hip joint. This device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. The two-part femoral component consists of a femoral stem made of alloys to be fixed in the intramedullary canal of the femur by impaction with or without use of bone cement. The proximal end of the femoral stem is tapered with a surface that ensures positive locking with the spherical ceramic (aluminium oxide, A12 03 ) head of the femoral component. The acetabular component is made of ultra-high molecular weight polyethylene or ultra-high molecular weight polyethylene reinforced with nonporous metal alloys, and used with or without bone cement.
{0}------------------------------------------------
August 6, 2019
Image /page/0/Picture/1 description: The image shows the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
Waldemar Link GmbH & Co. KG Stefanie Fuchs Regulatory Affairs Manager Oststraße 4-10 Norderstedt, Germany 22844
### Re: K190535
Trade/Device Name: BiMobile Dual Mobility System - E-Dur Inserts Regulation Number: 21 CFR 888.3353 Regulation Name: Hip Joint Metal/Ceramic/Polymer Semi-Constrained Cemented Or Nonporous Uncemented Prosthesis Regulatory Class: Class II Product Code: LZO, MEH Dated: July 8, 2019 Received: July 11, 2019
Dear Stefanie Fuchs:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
{1}------------------------------------------------
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Vesa Vuniqi Acting Assistant Director DHT6A: Division of Joint Arthroplasty Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K190535
Device Name
BiMobile Dual Mobility System - E-Dur Inserts
Indications for Use (Describe)
The BiMobile Dual Mobility System is indicated for patients with mobility-limiting diseases, fractures or defects of the hip joint or proximal femur, which cannot be treated by conservative or osteosynthetic procedures. The BiMobile Dual Mobility System is indicated for the following conditions:
- · Primary and secondary osteoarthritis
- · Rheumatoid arthritis
- · Correction of functional deformities
- Avascular necrosis
- Femoral neck fractures
- · Revision after implant loosening dependent on bone mass and quality
- · Dislocation risks
The device is intended for cemented and cementless use.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
| 510(k) Submitter: | Waldemar Link GmbH & Co. KG<br>Barkhausenweg 10<br>22339 Hamburg, Germany<br>Phone: +49-40-539950<br>Facility Registration #:3004371426 (Oststraße 4-10)<br>Facility Registration #:3007118403 (Harckesheyde 95) |
|---------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Waldemar Link GmbH & Co. KG<br>Stefanie Fuchs (Regulatory Affairs)<br>Oststraße 4-10<br>Norderstedt, GERMANY 22844<br>Phone: +49-40 53995-530<br>Fax: +49-40 53995-174<br>E-Mail: st.fuchs@linkhh.de |
| Date Prepared: | July 07, 2019 |
| Trade Name: | BiMobile Dual Mobility System – E-Dur Inserts |
| Common Name: | Acetabular Cup |
| Classification Name: | Prosthesis, hip, semi-constrained, metal/ceramic/polymer,<br>cemented or non-porous; 21 CFR §888.3353, product code<br>LZO<br>Prosthesis, hip, semi-constrained, uncemented,<br>metal/polymer, non-porous, calcium-phosphate; 21 CFR<br>§888.3353, product code MEH |
| Classification and Panel: | Class II, Orthopedic / 87 |
| Predicate Devices: | LINK BiMobile Dual Mobility System, K171273,<br>cleared January 19, 2018 |
| | Trinity Dual Mobility System with ECiMa Liners by Corin,<br>K170359, cleared October 31, 2017 |
| | EXp Acetabular Shell Liner by Stelkast, Inc., K094035,<br>cleared March 24, 2011 |
| | MobileLink Acetabular Cup System, K182321, cleared May<br>24, 2019 |
| Device Description: | The BiMobile cup system is a dual mobility acetabular<br>implant system. The system is consisting of wrought cobalt<br>chromium molybdenum alloy (CoCrMo acc. to ISO 5832-12) |
{4}------------------------------------------------
metal shells with a highly polished inner surface, which are available in three different outer surface modifications. There is the cemented version with a glass blasted surface of concentric grooves, and two cementless versions where a pressfit generating macrostructure is covered by either a PlasmaLink (titanium plasma spray acc. to ISO 5832-2/ASTM F1580) or TiCaP (titanium plasma spray and calcium phosphate acc. to ISO 5832-2/ASTM F1580 and F1609) coating. For each size, from 42 to 70, a dedicated liner made of Ultra-high-molecular-weight polyethylene (conventional UHMWPE acc. to ISO 5834-2/ASTM F648) is available. These can be combined with ceramic or CoCr alloy femoral heads of sizes 22 mm (for Size 42-46) and 28 mm (for size 48-70). The purpose of this submission is to add acetabular liners in Vitamin E highly crosslinked UHMWPE (E-Dur) acc. to ASTM F2695, ASTM F2565 to the BiMobile Dual Mobility System. This premarket notification does not include the femoral components. The BiMobile Dual Mobility System is compatible with previously cleared femoral heads (K171273, K161840, K953653, K931571, and K920756), femoral stems and neck segments (K183141, K161840, K151008, K142187, and K953653,) of Waldemar Link GmbH & Co. KG. Indications for Use: The BiMobile Dual Mobility System is indicated for patients with mobility-limiting diseases, fractures or defects of the hip joint or proximal femur, which cannot be treated by conservative or osteosynthetic procedures. The BiMobile Dual Mobility System is indicated for the following conditions: 1) Primary and secondary osteoarthritis 2) Rheumatoid arthritis 3) Correction of functional deformities 4) Avascular necrosis 5) Femoral neck fractures 6) Revision after implant loosening dependent on bone mass and quality 7) Dislocation risks The device is intended for cemented and cementless use. The additional components of the BiMobile Dual Mobility Comparison to Predicate Device:
System are similar to the predicate devices in terms of intended use and indications, materials, sizes, designs and performance.
{5}------------------------------------------------
Based on these similarities, the additional components of the BiMobile Dual Mobility System are believed to be substantially equivalent to the predicate device.
| Performance Testing: | Non-Clinical Performance and Conclusions: |
|----------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | The following material characterization, biocompatibility<br>testing/analysis, wear analyses, mechanical testing were<br>performed to demonstrate that the BiMobile E-Dur® Inserts |
| | perform as intended and are substantially equivalent to the<br>identified devices: |
| | - Material Characterization<br>- Biocompatibility Testing and Analysis<br>- Axial Disassembly Testing<br>- Lever-out Testing<br>- Standard and Adverse Wear Anaylses including Particles<br>Analyses<br>- Range of Motion Analysis |
| | The results of non-clinical performance testing demonstrate<br>that the device is as safe, as effective, and substantially<br>equivalent to the predicate devices. |
| Clinical Testing: | Clinical performance Testing was not required to demonstrate<br>the substantial equivalence of this device. |
| Conclusion: | The subject device BiMobile E-Dur inserts is substantially<br>equivalent to the predicate devices identified in this premarket<br>notification. |
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.