LINK® SP-CL® Hip System PoroLink® (microporous) and HX® (CaP) coated & LINK® LCU® Hip System PoroLink® (microporous) and HX® (CaP) coated
K161840 · Waldemar Link GmbH & Co. KG · LZO · Mar 16, 2017 · Orthopedic
Device Facts
Record ID
K161840
Device Name
LINK® SP-CL® Hip System PoroLink® (microporous) and HX® (CaP) coated & LINK® LCU® Hip System PoroLink® (microporous) and HX® (CaP) coated
Applicant
Waldemar Link GmbH & Co. KG
Product Code
LZO · Orthopedic
Decision Date
Mar 16, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3353
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The LINK® SP-CL® and LCU® Hip System PoroLink® (microporous) and HX® (CaP) coated is indicated for patients with mobility-limiting hip diseases, fractures or defects which cannot be treated by conservative or osteosynthetic procedures. The hip system is indicated for the following conditions: - Primary and secondary coxarthrosis - Osteoarthritis - Necrosis of the femoral head - Femoral neck fractures The stems are indicated for cementless use only.
Device Story
Hip prosthesis system comprising straight (LCU) and anatomically curved (SP-CL) femoral stems; titanium alloy (Ti6Al4V) construction; available with PoroLink (microporous) or HX (calcium phosphate) coatings. Stems connect to CoCrMo or ceramic (BIOLOX forte/delta) femoral heads via 12/14 Morse taper. Used in total hip arthroplasty; femoral heads articulate against crosslinked acetabular liners. Intended for cementless fixation. Surgeon-operated in clinical/OR settings. Replaces diseased/damaged hip joint to restore mobility and function. Benefits include pain relief and improved joint mechanics for patients with degenerative hip conditions.
Clinical Evidence
No clinical performance testing was required or performed. Substantial equivalence is supported by non-clinical bench testing, including femoral stem and neck fatigue testing (ISO 7206-4, 7206-8, 7206-6), fretting corrosion testing (ASTM F1875-98), and range of motion analysis.
Technological Characteristics
Materials: Titanium alloy (Ti6Al4V, ISO 5832-3, ASTM F136), CoCrMo, ceramic (BIOLOX forte/delta). Coatings: PoroLink (microporous) or HX (calcium phosphate, ASTM F1609). Design: Straight (LCU) and curved (SP-CL) stems with 12/14 Morse taper. Cementless fixation. Sterilization per ISO 17665-1.
Indications for Use
Indicated for patients with mobility-limiting hip diseases, fractures, or defects (primary/secondary coxarthrosis, osteoarthritis, femoral head necrosis, femoral neck fractures) untreatable by conservative/osteosynthetic procedures. Stems for cementless use only.
Regulatory Classification
Identification
A hip joint metal/ceramic/polymer semi-constrained cemented or nonporous uncemented prosthesis is a device intended to be implanted to replace a hip joint. This device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. The two-part femoral component consists of a femoral stem made of alloys to be fixed in the intramedullary canal of the femur by impaction with or without use of bone cement. The proximal end of the femoral stem is tapered with a surface that ensures positive locking with the spherical ceramic (aluminium oxide, A12 03 ) head of the femoral component. The acetabular component is made of ultra-high molecular weight polyethylene or ultra-high molecular weight polyethylene reinforced with nonporous metal alloys, and used with or without bone cement.
{1}------------------------------------------------
Image /page/0/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized image of an eagle with three human profiles incorporated into its design. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
March 17, 2017
Waldemar Link GmbH & Co. KG Mr. André Von Malotki Regulatory Affairs Manager Oststraße 4-10 Norderstedt, Germany 22844
Re: K161840
Trade/Device Name: LINK® SP-CL® Hip System PoroLink® (microporous) and HX® (CaP) coated & LINK® LCU® Hip System PoroLink® (microporous) and HX® (CaP) coated Regulation Number: 21 CFR 888.3353 Regulation Name: Hip Joint Metal/Ceramic/Polymer Semi-Constrained Cemented Or Nonporous Uncemented Prosthesis Regulatory Class: Class II Product Code: LZO, MEH Dated: February 16, 2017 Received: February 16, 2017
Dear Mr. André Von Malotki:
This letter corrects our substantially equivalent letter of March 16, 2017.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply
{2}------------------------------------------------
with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
# Lori A. Wiggins -S
for Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K161840
#### Device Name
LINK® SP-CL® Hip System PoroLink® (microporous) and HX® (CaP) coated & LINK® LCU® Hip System PoroLink® (microporous) and HX® (CaP) coated
#### Indications for Use (Describe)
The LINK® SP-CL® and LCU® Hip System PoroLink® (microporous) and HX® (CaP) coated is indicated for patients with mobility-limiting hip diseases, fractures or defects which cannot be treated by conservative or osteosynthetic procedures.
The hip system is indicated for the following conditions:
- Primary and secondary coxarthrosis
- Osteoarthritis
- Necrosis of the femoral head
- Femoral neck fractures
The stems are indicated for cementless use only.
Type of Use (Select one or both, as applicable)
2 Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
# PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
# FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(k) Summary
| 510(k) Submitter: | Waldemar Link GmbH & CO. KG |
|---------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Barkhausenweg 10 |
| | 22339 Hamburg, Germany |
| | Phone: +49-40-539950 |
| | Facility Registration #:3004371426 (Oststraße 4-10) |
| Contact Person: | Waldemar Link GmbH & Co. KG |
| | André von Malotki (Regulatory Affairs) |
| | Oststraße 4-10 |
| | Norderstedt, GERMANY 22844 |
| | Phone: +49-40 53995-448 |
| | Fax: +49-40 53995-174 |
| | E-Mail: a.vonmalotki@linkhh.de |
| Date Prepared: | March 10th, 2017 |
| Trade Name: | LINK® SP-CL® Hip System PoroLink® (microporous) and<br>HX® (CaP) coated |
| | & |
| | LINK® LCU® Hip System PoroLink® (microporous) and<br>HX® (CaP) coated |
| Common Name: | Hip Prosthesis |
| Classification Name: | Hip joint metal/ceramic/polymer semi-constrained cemented<br>or nonporous uncemented prosthesis; 21 CFR §888.3353,<br>product code LZO, MEH |
| Classification and Panel: | Class II, Orthopedic / 87 |
| Predicate Devices: | Corail AMTTM Hip Prosthesis by DePuy Orthopaedics,<br>#K123991 cleared September 16, 2013; #K070554 cleared<br>September 11, 2007; #K042992 cleared February 11, 2005 |
| | LINK® Contour Ribbed Hip Prosthesis,<br>#K854465 cleared February 2, 1986 |
{5}------------------------------------------------
# Device Description:
The LINK® SP-CL® and LCU® Systems include a straight stem (LCU® Hip Prosthesis) and an anatomically curved stem (SP-CL® Hip Prosthesis). Both stems are comprised of titanium alloy (Ti6A14V, ISO 5832-3, ASTM F136) available with corundum blasted PoroLink® (microporous) surface or HX® (calcium phosphate) (ASTM F1609) coated. The stems are available in a range of sizes, lengths, and offsets. The stems connect to femoral heads that are manufactured from Cobalt Chromium Molybdenum casting alloy (CoCrMo) or ceramic (BIOLOX® forte and BIOLOX® delta) via 12/14 Morse taper. In total joint use, the femoral head articulates against Signature Orthopedics' Origin crosslinked acetabular liners (#K121297) fixed in a cementless shell. Femoral heads part of this submission (BIOLOX® forte / delta head and LINK® Prosthesis Head B) are not intended for hemiarthroplasty use. Both the LCU® and SP-CL® stems are intended for cementless use to replace the hip joint.
# Indications for Use:
The LINK® SP-CL® and LCU® Hip System PoroLink® (microporous) and HX® (CaP) coated is indicated for patients with mobility-limiting hip diseases, fractures or defects which cannot be treated by conservative or osteosynthetic procedures. The hip system is indicated for the following conditions:
- -Primary and secondary coxarthrosis
- -Osteoarthritis
- Necrosis of the femoral head -
- -Femoral neck fractures
The stems are indicated for cementless use only.
# Comparison to Predicate Device:
Consistent with FDA's guidance document, "The 510(k) Program: Evaluating Substantial Equivalence in Premarket Notifications [510(k)]" (July 28, 2014) (510(k) Guidance Document), the LINK® SP-CL® and LCU® Hip Systems are substantially equivalent to DePuy Orthopaedics' Corail AMT Hip Prosthesis (#K123991, #K070554, #K042992) and the LINK® Contour Ribbed Hip Prosthesis (#K854465). The LINK® SP-CL® and LCU® Hip Systems have the same intended use and the same or similar indications, technological characteristics, and principles of operation as the predicate devices.
{6}------------------------------------------------
# Performance Data:
# Non-Clinical Performance and Conclusions:
Non-Clinical performance testing was conducted with consideration to "Guidance for Industry and FDA Staff, Non-Clinical Information for Femoral Stem Prosthesis, September 17, 2007".
Non-clinical performance testing included: femoral stem fatigue tests per ISO 7206-4 and ISO 7206-8; femoral neck segment fatigue tests per ISO 7206-6; ASTM F2068; Modular Connections, Fretting Corrosion Testing per ASTM F1875-98; Range of Motion analysis of the LINK® SP-CL® and LCU® Hip Systems and Component Testing of ceramic heads.
The results of non-clinical performance testing demonstrate that the device is as safe, as effective, and substantially equivalent to the predicate device.
## Clinical Performance:
There was no clinical performance testing required for this device.
## Cleaning, Sterilization and Routine Endotoxin Testing:
Sterilization and cleaning studies validate the procedures used, per the applicable standards of AAMI ST 77:2006, AAMI ST79, ISO 17665-1, ISO 11138-3, AAMI TIR 12:2010, and AAMI TIR 30.
Routine Endotoxin Testing is performed according to AAMI ST 72.
# Conclusion:
The LINK® LCU® and SP-CL® Systems have the same intended use and the same or similar indications, technological characteristics, and principles of operation as its predicate device. The minor technological differences between the LINK® LCU® and SP-CL® Systems and its predicate device do not raise any new issues of safety or effectiveness.
The subject devices LINK® SP-CL® and LCU® Hip System PoroLink® (microporous) and HX® (CaP) coated are substantially equivalent to the predicate device identified in this premarket notification.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.