K142187 · Waldemar Link GmbH & Co. KG · LZO · Nov 6, 2014 · Orthopedic
Device Facts
Record ID
K142187
Device Name
LINK MP RECONSTRUCTION PROSTHESIS
Applicant
Waldemar Link GmbH & Co. KG
Product Code
LZO · Orthopedic
Decision Date
Nov 6, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3353
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The LINK® MP® Reconstruction Prosthesis is intended for revision hip arthroplasty in patients whose bone stock is of poor quality or inadequate for other reconstruction techniques as indicated by deficiencies of the femoral head, neck, or portions of the proximal femur. The LINK® MP® Reconstruction Hip Prosthesis is indicated for the following conditions: 1) Revision arthroplasty due to juxta-articular bone defects. 2) Revision of loosened femoral prosthesis components involving extensive bone resorption of the proximal femur and widening of the medullary cavity or marked thinning of proximal femoral cortical bone. 3) Revision of loosened femoral prosthesis components by peri- / subprosthetic fracture. 4) Deformed proximal femur due to fractures or osteotomies. 5) Correction of bone deficiencies, e.g. due to tumors. 6) Large post-revision and post-trauma segmental bone defects. The LINK® MP® Reconstruction Prosthesis is for cementless use only.
Device Story
Modular hip revision prosthesis system; components include femoral stems, neck segments, proximal spacers, and expansion bolts. Stems feature tapered design with microporous surface and longitudinal fluting for cementless fixation. Modular design allows independent positioning and adjustment of leg length. Used by orthopedic surgeons in clinical settings for complex revision hip arthroplasty. Device replaces damaged or deficient proximal femoral bone stock; restores joint function and stability. Surgeon selects appropriate stem diameter, length, and neck segment angle to match patient anatomy. Expansion bolts secure modular components. Benefits include improved structural support in patients with significant bone loss.
Clinical Evidence
No clinical data. Substantial equivalence demonstrated via non-clinical bench testing, including femoral stem fatigue (ISO 7206-4/8), femoral neck segment fatigue (ISO 7206-6), and modular connection fretting/corrosion testing (ASTM F1875-98).
Technological Characteristics
Modular hip prosthesis. Stems/neck segments: Ti-6Al-4V. Spacers/bolts: CoCrMo. Cementless fixation. Features: tapered stem, microporous surface, longitudinal fluting. Testing standards: ISO 7206-4, ISO 7206-6, ISO 7206-8, ASTM F2068, ASTM F1875-98.
Indications for Use
Indicated for revision hip arthroplasty in patients with poor bone stock or deficiencies of the femoral head, neck, or proximal femur. Conditions include juxta-articular bone defects, revision of loosened femoral components with extensive bone resorption or cortical thinning, peri-/subprosthetic fractures, deformed proximal femur (fractures/osteotomies), bone deficiencies (e.g., tumors), and large post-revision/post-trauma segmental bone defects. Cementless use only.
Regulatory Classification
Identification
A hip joint metal/ceramic/polymer semi-constrained cemented or nonporous uncemented prosthesis is a device intended to be implanted to replace a hip joint. This device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. The two-part femoral component consists of a femoral stem made of alloys to be fixed in the intramedullary canal of the femur by impaction with or without use of bone cement. The proximal end of the femoral stem is tapered with a surface that ensures positive locking with the spherical ceramic (aluminium oxide, A12 03 ) head of the femoral component. The acetabular component is made of ultra-high molecular weight polyethylene or ultra-high molecular weight polyethylene reinforced with nonporous metal alloys, and used with or without bone cement.
{1}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the circle is an abstract symbol that resembles three human profiles facing to the right.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
November 6, 2014
Waldemar Link GmbH and Company, KG Thomas Mehler, Ph.D. Director Quality Management Barkhausenweg 10 22339 Hamburg GERMANY
Re: K142187
Trade/Device Name: LINK® MP® Reconstruction Prosthesis Regulation Number: 21 CFR 888.3353 Regulation Name: Hip joint metal/ceramic/polymer semi-constrained cemented or nonporous uncemented prosthesis Regulatory Class: Class II Product Code: LZO, LWJ, JDI Dated: August 4, 2014 Received: August 8, 2014
Dear Dr. Thomas Mehler:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing
{2}------------------------------------------------
(21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Lori A. Wiggins -S
for Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
#### Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
510(k) Number (if known)
K142187 (page 1/1)
Device Name
LINK® MP® Reconstruction Prosthesis
Indications for Use (Describe)
The LINK® MP® Reconstruction Prosthesis is intended for revision hip arthroplasty in patients whose bone stock is of poor quality or inadequate for other reconstruction techniques as indicated by deficiencies of the femoral head, neck, or portions of the proximal femur. The LINK® MP® Reconstruction Hip Prosthesis is indicated for the following conditions:
1) Revision arthroplasty due to juxta-articular bone defects.
2) Revision of loosened femoral prosthesis components involving extensive bone resorption of the proximal femur and widening of the medullary cavity or marked thinning of proximal femoral cortical bone.
3) Revision of loosened femoral prosthesis components by peri- / subprosthetic fracture.
- 4) Deformed proximal femur due to fractures or osteotomies.
- 5) Correction of bone deficiencies, e.g. due to tumors.
- 6) Large post-revision and post-trauma segmental bone defects.
The LINK® MP® Reconstruction Prosthesis is for cementless use only.
Type of Use (Select one or both, as applicable)
2 Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
## FOR FDA USE ONLY Concurrence of Center for Devices and Radiological Health (CDRH) (Signature) N/A This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW." The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration
Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (1/14)
Page 1 of 1
PSC Publishing Services (301) 443-6740 EF
{4}------------------------------------------------
# 510(k) Summary
| 510(k) Submitter: | Waldemar Link GmbH & CO. KG<br>Barkhausenweg 10<br>22339 Hamburg, Germany<br>Phone: +49-40-539950<br>Facility Registration #:3004371426 (Oststraße 4-10) |
|---------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Waldemar Link GmbH & Co. KG<br>Stefanie Reimers (Regulatory Affairs)<br>Oststraße 4-10<br>Norderstedt, GERMANY 22844<br>Phone: +49-40 53995-530<br>Fax: +49-40 53995-174<br>E-Mail: S.Reimers@linkhh.de |
| Date Prepared: | August 4th, 2014 |
| Trade Name: | LINK® MP® Reconstruction Prosthesis |
| Common Name: | Hip Revision Prosthesis |
| Classification Name: | Prosthesis Hip, Semi-Constrained, Metal/Polymer,<br>Uncemented; 21 CFR §888.3360, product code LWJ<br>Prosthesis, Hip, Semi-Constrained, Metal/Polymer,<br>Cemented; 21 CFR §888.3350, product code JDI<br>Prosthesis, Hip, Semi-Constrained, Metal / Ceramic /<br>Polymer, Cemented or Non-Porous, Uncemented; 21 CFR<br>§888.3353, product code LZO |
| Classification and Panel: | Class II, Orthopedic / 87 |
| Predicate Devices: | LINK® MP® Reconstruction Hip (K955296) |
| Device Description: | The LINK® MP® Reconstruction Prosthesis modular system<br>consists of Prosthesis Heads, Stems, Neck Segments,<br>Proximal Spacers, and Expansion Bolts. The modular<br>components are interchangeable allowing for independent<br>positioning. The Prosthesis Stems are available in a variety of<br>diameters and lengths. Neck Segments are available in a<br>variety of CCD angles and sizes. Proximal Spacers are<br>available in a variety of heights and can be used<br>independently or combined to add leg length. Expansion<br>Bolts are used to secure the Neck Segments and Proximal<br>Spacers, when used, to the modular Prosthesis Stems.<br>Additional features include a tapered stem with a. |
{5}------------------------------------------------
microporous surface. The Prosthesis Stems also include longitudinal fluting.
The MP® Reconstruction Prosthesis Stems (cementless) and Neck Segments are produced of Titanium Aluminum Vanadium alloy (Ti-6Al-4V). Proximal Spacers and Expansion Bolts are made of Cobalt Chromium Molybdenum casting alloy (CoCrMo) materials.
The LINK® MP® Reconstruction Prosthesis is intended for Indications for Use: revision hip arthroplasty in patients whose bone stock is of poor quality or inadequate for other reconstruction techniques as indicated by deficiencies of the femoral head, neck, or portions of the proximal femur. The LINK® MP® Reconstruction Hip Prosthesis is indicated for the following conditions:
- 1) Revision arthroplasty due to juxta-articular bone defects.
- 2) Revision of loosened femoral prosthesis components involving extensive bone resorption of the proximal femur and widening of the medullary cavity or marked thinning of proximal femoral cortical bone.
- Revision of loosened femoral prosthesis components by 3) peri-/subprosthetic fracture.
- 4) Deformed proximal femur due to to fractures Or osteotomies.
- 5) Correction of bone deficiencies, e.g. due to tumors.
- 6) Large post-revision and post-trauma segmental bone defects.
The LINK® MP® Reconstruction Prosthesis is for cementless use only.
Comparison to Predicate Device: The subject device has the same intended use as the predicate and uses materials identical to those used in the predicate. The design of the subject and predicate devices has the same technological characteristics except: the subject device Neck Segments are made of Ti-6Al-4V instead of CoCrMo; additional Prosthesis Stem and Neck Segment sizes and configurations have been introduced; and a hex head Expansion Bolt is used in place of the predicates Slot Head Fixation Screw.
> The performance testing is sufficient to demonstrate that the subject and predicate devices are substantially equivalent with regard to design. Any difference between the subject and predicate device does not change the intended use or fundamental scientific technology.
{6}------------------------------------------------
### Performance Data:
### Non-Clinical Performance and Conclusions:
Non-Clinical performance testing was conducted with consideration to Guidance for Industry and FDA Staff, Non-Clinical Information for Femoral Stem Prosthesis, September 17, 2007 and Guidance Document For Testing Non-Articulating,"Mechanically Locked", Components, May 1, 1995.
Non-clinical performance testing included: femoral stem fatigue tests per ISO 7206-4 and ISO 7206-8; femoral neck segment fatigue tests per ISO 7206-6; ASTM F2068; Modular Connections, Fretting, and Corrosion Testing per ASTM F1875-98.
The results of non-clinical performance testing demonstrated that the device is as safe, as effective, and substantially equivalent to the predicate devices.
### Clinical Performance and Conclusions:
There was no clinical performance testing required for this device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.