K183141 · Waldemar Link GmbH & Co. KG · LZO · Feb 8, 2019 · Orthopedic
Device Facts
Record ID
K183141
Device Name
LINK MP Monoblock Hip Prosthesis
Applicant
Waldemar Link GmbH & Co. KG
Product Code
LZO · Orthopedic
Decision Date
Feb 8, 2019
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3353
Device Class
Class 2
Attributes
Therapeutic
Intended Use
The LINK ® MP® Monoblock is indicated for patients with Mobility-limiting diseases, fractures or defects of the hip joint or proximal femur which cannot be treated by conservative or osteosynthetic procedures The LINK® MP® Monoblock is indicated for the following conditions: - Revision of loosened femoral prosthesis components involving extensive bone resorption of the proximal femur and widening of the medullary cavity or marked thinning of proximal femoral cortical bone the fixation of a new standard prosthesis is not possible - Revision of loosened femoral prosthesis components by peri-/subprosthetic fracture The device is intended for cementless use.
Device Story
Straight hip prosthesis stem; manufactured from Ti-6Al-4V titanium alloy; features tapered design with microporous glass-blasted surface. Used in hip arthroplasty; implanted by orthopedic surgeons in clinical settings. Functions as femoral component in total hip replacement; articulates with femoral heads (CoCrMo or ceramic) via 12/14 Morse taper and LINK BiMobile Cup System. Provides cementless fixation to bone; restores joint function in revision cases involving significant bone loss or periprosthetic fracture. Benefits patients by providing stable reconstruction when standard prostheses are unsuitable.
Clinical Evidence
No clinical performance testing was required or performed. Substantial equivalence is supported by non-clinical bench testing, including femoral stem fatigue, femoral neck segment fatigue, fretting corrosion testing, and range of motion analysis.
Technological Characteristics
Material: Ti-6Al-4V titanium alloy. Design: Straight, tapered stem with microporous glass-blasted surface. Fixation: Cementless. Interface: 12/14 Morse taper for femoral head connection. Available in various sizes, lengths, CCD angles, and offsets.
Indications for Use
Indicated for patients with mobility-limiting diseases, fractures, or defects of the hip joint or proximal femur untreatable by conservative/osteosynthetic procedures. Specific conditions: revision of loosened femoral components with extensive proximal bone resorption, widened medullary cavity, or marked cortical thinning where standard prosthesis fixation is impossible; revision of loosened femoral components due to peri-/subprosthetic fracture. Intended for cementless use.
Regulatory Classification
Identification
A hip joint metal/ceramic/polymer semi-constrained cemented or nonporous uncemented prosthesis is a device intended to be implanted to replace a hip joint. This device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. The two-part femoral component consists of a femoral stem made of alloys to be fixed in the intramedullary canal of the femur by impaction with or without use of bone cement. The proximal end of the femoral stem is tapered with a surface that ensures positive locking with the spherical ceramic (aluminium oxide, A12 03 ) head of the femoral component. The acetabular component is made of ultra-high molecular weight polyethylene or ultra-high molecular weight polyethylene reinforced with nonporous metal alloys, and used with or without bone cement.
K161840 — LINK® SP-CL® Hip System PoroLink® (microporous) and HX® (CaP) coated & LINK® LCU® Hip System PoroLink® (microporous) and HX® (CaP) coated · Waldemar Link GmbH & Co. KG · Mar 16, 2017
K061786 — VERSYS FIBER METAL MIDCOAT LOW HEAD CENTER HIP PROSTHESIS · Zimmer, Inc. · Jul 24, 2006
K974294 — ENCORE LINEAR POROUS COATED HIP · Encore Orthopedics, Inc. · Jan 12, 1998
K994070 — REVELATION POROUS HIP STEM · Encore Orthopedics, Inc. · Jan 20, 2000
K080625 — SMITH & NEPHEW MIS HIP STEM WITH STIKTITE · Smith & Nephew, Inc. · May 8, 2008
Submission Summary (Full Text)
{0}------------------------------------------------
February 8, 2019
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Waldemar Link GmbH & Co. KG Andre Malotki Official Correspondent Oststraße 4-10 Norderstedt, 22844 De
## Re: K183141
Trade/Device Name: LINK MP Monoblock Hip Prosthesis Regulation Number: 21 CFR 888.3353 Regulation Name: Hip Joint Metal/Ceramic/Polymer Semi-Constrained Cemented Or Nonporous Uncemented Prosthesis Regulatory Class: Class II Product Code: LZO Dated: November 13. 2018 Received: November 13, 2018
### Dear Andre Malotki:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
{1}------------------------------------------------
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Daniel S. Ramsey -S 2019.02.08 16:39:00 -05'00'
FOR Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Form Approved: OMB No. 0910-0120
Expiration Date: January 31, 2017
See PRA Statement below.
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
## Indications for Use
510(k) Number (if known)
#### K183141
Device Name
LINK® MP® Monoblock Hip Prosthesis
Indications for Use (Describe)
The LINK ® MP® Monoblock is indicated for patients with Mobility-limiting diseases, fractures or defects of the hip joint or proximal femur which cannot be treated by conservative or osteosynthetic procedures
The LINK® MP® Monoblock is indicated for the following conditions:
- Revision of loosened femoral prosthesis components involving extensive bone resorption of the proximal femur and widening of the medullary cavity or marked thinning of proximal femoral cortical bone the fixation of a new standard prosthesis is not possible
- Revision of loosened femoral prosthesis components by peri-/subprosthetic fracture
The device is intended for cementless use.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
#### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW!*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## SECTION 5: 510(k) Summary
| 510(k)<br>Submitter: | Waldemar Link GmbH & Co. KG<br>Barkhausenweg 10<br>22339 Hamburg, Germany<br>Phone: +49-40-539950<br>Facility Registration: 3003386935 (Oststraße 4-10)<br>Facility Registration: 3007118403 (Harckesheyde 95) |
|------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact<br>Person: | Waldemar Link GmbH & Co. KG<br>André von Malotki (Regulatory Affairs)<br>Oststraße 4-10<br>22844 Norderstedt, Germany<br>Phone: +49-40 53995-530<br>Fax: +49-40 53995-174<br>E-Mail: a.vonmalotki@linkhh.de |
| Date<br>Prepared: | November 8th, 2018 |
| Trade Name: | LINK® MP® Monoblock Hip Prosthesis |
| Common<br>Name: | Hip Revision Prosthesis |
| Classification<br>Name: | 21 CFR §888.3353, Hip joint metal/ceramic/polymer semi-<br>constrained cemented or nonporous uncemented prosthesis. |
| Product Code | LZO |
| Classification<br>and Panel: | Class II, Orthopedic / 87 |
| Predicate<br>Devices: | Wagner SL Revision Stem by Zimmer Biomet Holding, Inc.,<br>K973837, cleared January 6, 1998 |
| Reference<br>Device: | LINK® MP® Reconstruction Prostheses by Waldemar Link GmbH &<br>Co. KG, K142187, cleared November 6, 2014 |
| | LINK® SP-CL® Hip System PoroLink® (Microporous) And HX® (CaP)<br>Coated & LINK® LCU® Hip System PoroLink® (Microporous) And<br>HX® (CaP) Coated by Waldemar Link GmbH & Co. KG, K161840,<br>cleared March 16, 2017 |
| Device<br>Description: | The LINK® MP® Monoblock Hip Prosthesis is a straight hip<br>prosthesis stem. The hip prosthesis is made of Titanium Aluminum<br>Vanadium alloy (Ti-6Al-4V). The prosthesis stems are available in a<br>range of sizes, lengths, CCD angles and offsets. Additional features<br>include a tapered stem with a microporous glass-blasted surface.<br><br>The LINK® MP® Monoblock Hip Prosthesis is designed to be used in<br>conjunction with femoral heads that are manufactured from Cobalt<br>Chromium Molybdenum casting alloy (CoCrMo) or ceramic |
{4}------------------------------------------------
(BIOLOX® forte and BIOLOX® delta) via 12/14 morse taper (K161840). In total joint use, the femoral head articulates aqainst LINK® BiMobile Cup System (K171273). The LINK® MP® Monoblock Hip Prosthesis provides cementless fixation to the bone.
The LINK® MP® Monoblock Hip Prosthesis is indicated for patients Indications for Use: with Mobility-limiting diseases, fractures or defects of the hip ioint or proximal femur which cannot be treated by conservative or osteosynthetic procedures
> The LINK® MP® Monoblock Hip Prosthesis is indicated for the following conditions:
- Revision of loosened femoral prosthesis components involving . extensive bone resorption of the proximal femur and widening of the medullary cavity or marked thinning of proximal femoral cortical bone the fixation of a new standard prosthesis is not possible
- . Revision of loosened femoral prosthesis components by peri-/subprosthetic fracture
The device is intended for cementless use.
#### Non-Clinical Performance and Conclusions: Performance Data:
Non-clinical performance testing and analysis were provided, including bench testing. Specifically, the following tests were performed:
- Femoral stem fatigue
- Femoral neck segment fatigue
- . Fretting corrosion testing
- Range of motion analysis
The results of non-clinical performance testing demonstrate that the device is as safe, as effective and substantially equivalent to the predicate device.
### Clinical Performance and Conclusions:
There was no clinical performance testing required for this device.
- The subject device LINK® MP® Monoblock Hip Prosthesis is Conclusion: substantially equivalent to the predicate devices identified in this premarket notification.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.