K973231 · Corin USA · KTT · Nov 20, 1997 · Orthopedic
Device Facts
Record ID
K973231
Device Name
THE CORIFIX DYNAMIC HIP SCREW SYSTEM
Applicant
Corin USA
Product Code
KTT · Orthopedic
Decision Date
Nov 20, 1997
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3030
Device Class
Class 2
Attributes
Therapeutic
Intended Use
The Dynamic Hip Screw System devices are used in the treatment of femoral neck fractures in the pertrochanteric, intertrochanteric and high subtrochanteric regions of the femur, or in the treatment of combinations of these fractures. Pertrochanteric fractures occur from the greater to lesser trochanter. Intertrochanteric fractures occur within the trochanteric area. High subtrochanteric fractures occur immediately below the trochanter. The devices provide support to the femur and femoral head during the natural process of bone regeneration to repair the femoral fracture. The devices are designed to withstand the in vivo loads experienced during normal activity throughout the life of the implant.
Device Story
System comprises stainless steel compression plates and compression lag screws; optional hexagonal screw included. Used for internal fixation of femoral neck fractures; provides mechanical support to femur and femoral head during bone regeneration. Implanted by surgeons in clinical/OR settings; compatible with dedicated Corifix or Stratec (Synthes) instrumentation. Designed to withstand in vivo physiological loads. Benefits patient by stabilizing fracture site to facilitate natural healing.
Clinical Evidence
No clinical data provided. Substantial equivalence based on design, material, and dimensional similarity to predicate device with established clinical history.
Technological Characteristics
Materials: 316 LVM stainless steel certified to BS 7252 Part 1. Mechanical fixation system consisting of compression plates and lag screws. Dimensions: equivalent to predicate DHS system (plate thickness, width, and barrel diameter).
Indications for Use
Indicated for patients with femoral neck fractures in pertrochanteric, intertrochanteric, and high subtrochanteric regions of the femur, including combinations of these fractures.
Regulatory Classification
Identification
Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.
Predicate Devices
Stratec (Synthes) DHS System
Related Devices
K152757 — SIGN Hip Construct (SHC) · Sign Fracture Care International · Nov 13, 2015
K971654 — OSTEO COMPRESSION HIP SCREW SYSTEM · Osteonics Corp. · Jul 9, 1997
K023851 — PRECIMED HIP SCREW SYSTEM · Precimed, Inc. · Jan 28, 2003
K962644 — VIGOR COMPRESSION HIP SCREW SYSTEM · Deutschland Surgical, Inc. · Sep 6, 1996
K221787 — Auxein's DHS/DCS Plate System · Auxein Medical Private Limited · Nov 30, 2023
Submission Summary (Full Text)
{0}------------------------------------------------
# 510(k) SUMMARY
| Name of Company: | Corifix Ltd<br>The Corinium Centre<br>Cirencester<br>Gloucestershire<br>GL7 1YJ<br>England |
|------------------|--------------------------------------------------------------------------------------------|
| Name of Device: | Corifix Dynamic Hip Screw System |
The Dynamic Hip Screw System comprises stainless steel Dynamic Device Description: Compression Plates and Compression Lag Screws and optional Hexagonal Screw for use in the fixation of femoral neck fractures.
The devices provide support to the femur and femoral head during the natural process of bone regeneration to repair the femoral neck fracture. The devices are designed to withstand the in vivo loads experienced during normal activity throughout the life of the implant.
A dedicated Corifix Dynamic Hip Screw instrument set is used to implant the devices or Stratetc (Synthes) instruments may be used.
Each of the implantable devices is manufactured form stainless steel which in certified to BS 7252 Part 1.
#### SUBSTANTIAL EQUIVALENCE AND SAFETY AND SUMMARY OF EFFECTIVENESS
The Corifix Dynamic Hip Screw System is substantially equivalent to the Stratec (Synthes) DHS System for the following reasons:
- Both are manufactured from 316 LVM stainless steel. a)
- Both have a similar range of plate angles and holes. b)
- c) Both are designed to be used to treat proximal hip fractures.
- Both of the plates are the same thickness and width. d)
- Both have the same outside barrel diameter. e)
This type of device has been in clinical use for over twenty years and no significant problems have been reported.
The Corifix Hip Screw has been in clinical use outside the USA for three years and no postoperative problems have been reported.
The Corifix product design is generic and equivalent to the Synthes DHS System.
{1}------------------------------------------------
Image /page/1/Picture/1 description: The image shows the seal of the Department of Health & Human Services - USA. The seal is circular, with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. In the center of the seal is an abstract symbol that resembles a stylized eagle or bird, with three human profiles incorporated into its design. The overall design is simple and monochromatic.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
NOV 20 1997
Mr. Craig Corrance President Corin U.S.A. 10500 University Center Drive Suite 130 Tampa, Florida 33612
K973231 Re: Corifix Dynamic Hip Screw Trade Name: Regulatory Class: II Product Code: COKTT Land: Comm Dated: August 12, 1997 Received: August 27, 1997
#### Dear Mr. Corrance:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions The general controls provisions of the Act of the Act. include requirements for annual reqistration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Requlations, Title 21, Parts 800 to 895. ਖੇ substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Failure to Administration (FDA) will verify such assumptions. comply with the GMP requlation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{2}------------------------------------------------
### Page 2 - Mr. Craig Corrance
This letter will allow you to begin marketing your device as described in your 510 (k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
G. Celia M. Whitton, Ph.D., M.
Cella M. Witten, Ph.D., M.D. Director Division of General and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
510(k) Number (if known):
The Corifix Dynamic Hip Screw System Device Name:
# INDICATIONS FOR USE
The Dynamic Hip Screw System devices are used in the treatment of femoral neck fractures in the pertrochanteric, intertrochanteric and high subtrochanteric regions of the femur, or in the treatment of combinations of these fractures.
Pertrochanteric fractures occur from the greater to lesser trochanter. Intertrochanteric High subtrochanteric fractures occur fractures occur within the trochanteric area. immediately below the trochanter.
The devices provide support to the femur and femoral head during the natural process of bone regeneration to repair the femoral fracture.
The devices are designed to withstand the in vivo loads experienced during normal activity throughout the life of the implant.
## (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation
signature
(Division Sign-Off) Division of General Restorative Devices
510(k) Number K973231
Prescription Use (per 21 CFR 801.109)
OR
Over-The-Counter Use
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.