VIGOR COMPRESSION HIP SCREW SYSTEM

K962644 · Deutschland Surgical, Inc. · JDO · Sep 6, 1996 · Orthopedic

Device Facts

Record IDK962644
Device NameVIGOR COMPRESSION HIP SCREW SYSTEM
ApplicantDeutschland Surgical, Inc.
Product CodeJDO · Orthopedic
Decision DateSep 6, 1996
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3030
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Vigor Compression Hip Screw is indicated as a means of rigid internal fixation of stable and unstable fractures of the proximal femur in which a stable medial buttress can be reconstructed and is intended for the management of intertrochanteric, subtrochanteric and basilar neck fractures and selected trochanteric non-unions.

Device Story

System provides rigid internal fixation for proximal femur fractures. Components include bone screws, compression screws, lag screws, and side plates. Lag screws feature flattened square base for rotational stability. Implants available in various lengths and plate configurations. Used by surgeons in clinical settings to stabilize fractures, facilitating bone healing. Instrumentation set included for surgical procedure. Implants are single-use.

Clinical Evidence

Bench testing only; no clinical data provided.

Technological Characteristics

Material: 316L Stainless Steel. Components: Bone screws (22-60mm), lag screws (50-150mm), and side plates (4, 5, 6-hole; 135°, 140°, 145° angles). Non-sterile; requires user sterilization to 10^-6 SAL.

Indications for Use

Indicated for patients with stable or unstable proximal femur fractures, including intertrochanteric, subtrochanteric, basilar neck fractures, and selected trochanteric non-unions, where a stable medial buttress can be reconstructed.

Regulatory Classification

Identification

Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.

Predicate Devices

Submission Summary (Full Text)

{0} SEP - 6 1996 K962644 # SUMMARY OF SAFETY AND EFFECTIVENESS ## VIGOR COMPRESSION HIP SCREW SYSTEM ### I. General Information Classification Name: Device, Fixation, Proximal Femoral Implant Common Name: Compression Hip Screw Device Trade Name: Vigor Compression Hip Screw System Classification Code: 87JDO Submitter's Name & Address: Deutschland Surgical Incorporated 1222 Commerce Street, Suite #415 Dallas, Texas 75202-4306 Establishment Registration No: 1646853 Contact Person: Thomas Wolter Summary Preparation Date: July 5, 1996 ### II. Predicate Device The Vigor Compression Hip Screw is claimed to be substantially equivalent in material, design, and function to the Synthes Dynamic Hip Screw cleared by FDA under 510(k) K791619 on August 28, 1979. ### III. Device Description The Vigor Compression Hip Screw System is intended as a means of rigid internal fixation for fractures of the proximal femur. The Vigor System implants consist of Bone Screws; Compression Screws, Lag Screws and Side Plates, and are offered in a wide, incremental range of sizes to ensure optimum selection for a proper fixation. The Vigor full-length threaded Bone Screws and Compression Bone Screws are available in a range of sizes from 22.0 mm to 60.0 mm in length. Vigor Lag Screws and Super Lag Screws feature a flattened, square base that engages the inner surface of the implant barrel for increased rotational stability. These Lag Screws are available in a range of sizes from 50.0 mm to 150.0 mm in length, in 5.0 mm increments. And, finally, a trio of low-profile Side Plates are included in the system. Vigor offers the standard 4-hole, 5-hole and 6-hole plates, each available in a 135°, 140° or 145° configuration. The Vigor System instrumentation set consists of the standard instrumentation required for each indicated procedure. Premarket Notification: Vigor Compression Hip Screw Deutschland Surgical Incorporated: Submitted: July 5, 1996 Page 83 {1} # IV. Sterilization Vigor Implants and Instrumentation are provided non-sterile. Implants are for single use only. Both implants and instruments must be sterilized prior to use in accordance with the recommended sterilization parameters described in the package insert in order to achieve a sterility assurance level of $10^{-6}$. # V. Indications for Use The Vigor Compression Hip Screw is indicated as a means of rigid internal fixation of stable and unstable fractures of the proximal femur in which a stable medial buttress can be reconstructed and is intended for the management of intertrochanteric, subtrochanteric and basilar neck fractures and selected trochanteric non-unions. # VII. Substantial Equivalence The Vigor Compression Hip Screw System is considered to be substantially equivalent to the Synthes Dynamic Hip Screw[K791619] distributed by Synthes (U.S.A) at 1690 Russell Road, Paoli, PA 19301. | Feature | Vigor Compression Hip Screw | Synthes Dynamic Hip Screw | | --- | --- | --- | | Material | 316L Stainless Steel | 316L Stainless Steel | | Design | Bone, Compression and Lag Screws, low-profile, angled side plates and associated instrumentation. | Bone, Compression and Lag Screws, low-profile, angled side plates and associated instrumentation | | Function | Rigid internal fixation to fractures of the proximal femur. | Rigid internal fixation to fractures of the proximal femur. | # VIII. Conclusion The Vigor Compression Hip Screw System is considered to be substantially equivalent in design, material and function to the Synthes Dynamic Hip Screw and is believed to perform as well as the Synthes System. Premarket Notification: Vigor Compression Hip Screw Deutschland Surgical Incorporated: Submitted: July 5, 1996
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...