The INTEGRA® Freedom Wrist Arthroplasty System is indicated for intractable pain resulting from traumatic arthritis, osteoarthritis, rheumatoid arthritis, trauma-induced osteoarthritis of the radial/carpal joint and is intended to replace functionality of the joint due to deformity or elements stated above. The INTEGRA® Freedom Wrist Arthroplasty System is intended for cemented use.
Device Story
Three-part semi-constrained wrist arthroplasty system; replaces radiocarpal joint (distal radius and proximal carpal row). Components: Cobalt Chromium Molybdenum (CrCoMo) radial implant; Titanium alloy carpal plate with variable angle screws/locking caps; Ultra-High-Molecular-Weight Polyethylene (UHMWPe) carpal bearing. Portions of radial implant and carpal plate feature titanium plasma-sprayed coating. System implanted via cemented fixation by orthopedic surgeons to alleviate pain and restore joint mobility/functionality. Provided with necessary surgical instrumentation.
Clinical Evidence
Bench testing only. Performance verified via fatigue life testing, carpal poly bearing removal, movement to instability, and screw thread verification (axial pull-out, insertion/breaking torque/angle per ASTM F543). Porous coating evaluated for shear fatigue (ASTM F1160), static shear (ASTM F1044), static tensile (ASTM F1147), and abrasion (ASTM F1978). All tests met acceptance criteria.
Indicated for patients with intractable pain and loss of joint function due to traumatic arthritis, osteoarthritis, rheumatoid arthritis, or trauma-induced osteoarthritis of the radial/carpal joint. Intended for cemented fixation.
Regulatory Classification
Identification
A wrist joint metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace a wrist joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that have either a one-part radial component made of alloys, such as cobalt-chromium-molybdenum, with an ultra-high molecular weight polyethylene bearing surface, or a two-part radial component made of alloys and an ultra-high molecular weight polyethylene ball that is mounted on the radial component with a trunnion bearing. The metallic portion of the two-part radial component is inserted into the radius. These devices have a metacarpal component(s) made of alloys, such as cobalt-chromium-molybdenum. This generic type of device is limited to those prostheses intended for use with bone cement (§ 888.3027).
K191525 — KinematX Total Wrist Arthroplasty System · Extremity Medical, LLC · Mar 4, 2020
K233574 — Total Wrist Arthroplasty System (TWA) · Skeletal Dynamics · Jul 3, 2024
K020554 — MODIFICATION TO UNIVERSAL TOTAL WRIST SYSTEM · Kinetikos Medical, Inc. · May 3, 2002
K961051 — UNIVERSAL TOTAL WRIST SYSTEM (MODIFICATION) · Kinetikos Medical, Inc. · Jun 6, 1996
K233674 — Freedom Wrist Arthroplasty System · Ascension Orthopedics, Inc. · May 30, 2024
Submission Summary (Full Text)
{0}------------------------------------------------
INTEGRA® Freedom Wrist Arthroplasty System
MAR 2 4 2014
Section 5: 510(k) Summary
## 510(k) Summary
| 807.92(a)(1) - Submitter Information | |
|---------------------------------------------------------------------------|-------------------------------------------------------------------------------------|
| Name | Integra LifeSciences Corporation |
| Address | 311 Enterprise Drive<br>Plainsboro NJ 08536 |
| Name of Contact Person | Sally K. Wixson, VMD, RAC |
| Phone Number | (609) 936-2454 |
| Fax Number | (609) 750 4277 |
| Establishment<br>Registration Number | 3004608878 |
| Date Prepared | July 8, 2013 |
| 807.92(a)(2) - Name of device | |
| Trade or Propriety Name | INTEGRA® Freedom Wrist Arthroplasty System |
| Common or Usual Name | Prosthesis, wrist, 3 part metal-plastic-metal articulation,<br>semi-constrained |
| Classification Name | Wrist joint metal/polymer semi-constrained cemented<br>prosthesis (21CFR §888.3800) |
| Classification Panel | Orthopedic |
| Regulation | Class II, under 21CFR §888.3800 |
| Product Code | JWJ |
| 807.92(a)(3) - Legally marketed device(s) to which equivalence is claimed | |
Universal Total Wrist System (K030037), cleared February 05, 2003 ●
. Maestro™ Total Wrist (K042032), cleared October 7, 2004
### 807.92(a)(4) - Device description
The INTEGRA® Freedom Wrist Arthroplasty (IFW) System three part semi-constrained implant system designed to replace the radiocarpal joint (distal radius and proximal row of carpal bones) and is intended to alleviate pain while restoring functionality and mobility of the joint. IFW is intended for patients with intractable pain and loss function due to traumatic arthritis, osteoarthritis, rheumatoid arthritis, and trauma-induced osteoarthritis. The system consists of three major components:
- Radial implant made from Cobalt Chromium Molybdenum Alloy (CrCoMo) .
- . Carpal plate - (includes two variable angle screws with locking caps) - made from Titanium Alloy
- . Carpal Poly bearing - made from Ultra-High-Molecular-Weight Polyethylene (UHMWPe)
Portions of the radial implant and carpal plate have a titanium plasma sprayed coating. The implant system is provided with instrumentation necessary to complete the procedures for which the system is indicated. The components are intended to be
{1}------------------------------------------------
Integra LifeSciences Corporation-Traditional 510(k)
INTEGRA® Freedom Wrist Arthroplasty System
Section 5: 510(k) Summary
implanted together as a system, not individually as hemi-arthroplasty components and are intended for cemented fixation.
| 807.92(a)(5) - Intended Use of the device | |
|------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for Use | The INTEGRA® Freedom Wrist Arthroplasty System is indicated<br>for intractable pain resulting from traumatic arthritis, osteoarthritis,<br>rheumatoid arthritis, trauma-induced osteoarthritis of the<br>radial/carpal joint and is intended to replace functionality of the<br>joint due to deformity or elements stated above. The INTEGRA®<br>Freedom Wrist Arthroplasty System is intended for cemented use. |
| 807.92(a)(6) Summary of the technological characteristics of the device compared to<br>the predicate | |
| | The proposed INTEGRA® Freedom Wrist Arthroplasty (IFW) system and the predicate<br>systems (UTW and MTW) are designed to replace the radiocarpal joint and are intended<br>to alleviate pain while restoring functionality and mobility of the joint. The overall design<br>and materials of the proposed device and the predicate devices are similar. The systems |
consist of titanium, cobalt chrome, and Ultra-High-Molecular-Weight Polyethylene (UHMWPe) components which replace the articulation of the distal radius and proximal row of carpal bones of the wrist joint. The devices share three basic design elements in that they employ a radial component, a carpal bearing, and a carpal component. The carpal and radial components of the proposed and predicated devices have porous coated stems that are intended to be fixated into the bone with cement and utilize two screws to secure the carpal component into the carpal bones. The implant devices are provided
The INTEGRA® Freedom Wrist Arthroplasty System has similar indication for use, intended use, and fundamental scientific technology as the predicate devices (K030037 and K042032). The proposed device utilizes similar materials and design features. The differences between the predicates and proposed device do not raise any new issues regarding safety and effectiveness; therefore, The INTEGRA® Freedom Wrist Arthroplasty System is considered substantially equivalent to the predicate devices.
#### 807.92(b)(1-2) - Nonclinical Tests Submitted
Testing to verify the performance for the The The INTEGRA® Freedom Wrist Arthroplasty System included the following:
- Fatigue Life Test .
- . Carpal Poly Bearing Removal
with corresponding instrumentation.
- Movement to Instability .
- . Screw Thread Verification:
- O Axial Screw Pull-out, Insertion Torque, Breaking Torque and Breaking Angle per ASTM F543
- Porous Coating Testing: . Shear Fatigue Strength per ASTM F1160 o
{2}------------------------------------------------
Integra LifeSciences Corporation-Traditional 510(k)
INTEGRA® Freedom Wrist Arthroplasty System
Section 5: 510(k) Summary
- Static Shear Strength per ASTM F1044 o
- Static Tensile per ASTM F1147 o
- Abrasion per ASTM F1978 o
The results of these performance tests met their respective acceptance criteria and demonstrate that the Integra® Freedom Wrist Arthroplasty System is safe for the intended use, and is substantially equivalent to the predicate device identified.
807.92(b)(3) - Conclusions drawn from non-clinical data
The design features, materials, intended use, and overall fundamental scientific technology of the Integra® Freedom Wrist Arthroplasty System are substantially equivalent to the predicate device. The safety and effectiveness of the Integra® Freedom Wrist Arthroplasty System is adequately supported by the substantial equivalence information, materials information, and performance data provided within this Premarket Notification submission.
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized human figure with outstretched arms, representing care and protection. The figure is positioned to the right of the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA", which is arranged in a circular fashion around the figure.
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
March 24, 2014
Integra LifeSciences Corporation Mr. Frederic Testa Director. Regulatory Affairs 311 Enterprise Drive Plainsboro, New Jersey 08536
Re: K132250
Trade/Device Name: INTEGRA® Freedom Wrist Arthroplasty System Regulation Number: 21 CFR 888.3800 Regulation Name: Wrist joint metal/polymer semi-constrained cemented prosthesis Regulatory Class: Class II Product Code: JWJ Dated: February 20, 2014 Received: February 21, 2014
Dear Mr. Testa:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{4}------------------------------------------------
#### Page 2 - Mr. Frederic Testa
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Ronald P. Jean -S for
- Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
Integra LifeSciences Corporation-Traditional 510(k) INTEGRA® Freedom Wrist Arthroplasty System
Section 4: Indications for Use
#### Indications for Use
K132250 510(k) Number (if known):
Device Name:
#### INTEGRA® Freedom Wrist Arthroplasty System
#### Indications For Use:
The INTEGRA® Freedom Wrist Arthroplasty System is indicated for intractable pain resulting from traumatic arthritis, osteoarthritis, rheumatoid arthritis, trauma-induced osteoarthritis of the radial/carpal joint and is intended to replace functionality of the joint due to deformity or elements stated above. The INTEGRA® Freedom Wrist Arthroplasty System is intended for cemented use.
Prescription Use _____________________________________________________________________________________________________________________________________________________________ (Per 21 CFR 801 Subpart D) Over-The Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
AND/OR
Concurrence of CDRH, Office of Device Evaluation (ODE)
Elizabeth₪»§rank -S
Page 1 of 1
Division of Orthopedic Devices
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.