The Skeletal Dynamics Geminus Volar Distal Radius Plate System is intended for fixation of fractures and osteotomies of the distal radius.
Device Story
The Geminus Volar Distal Radius Plate System is a bone fixation system used by orthopedic surgeons for the repair of distal volar radial fractures. The system consists of medical-grade titanium alloy plates, washers, and screws, along with cobalt-chrome cannulated polyaxial locking screws and specialized instrumentation. It includes a Hook Plate Extension to buttress volar marginal fragments. The device is provided non-sterile and must be sterilized at the user facility. Surgeons use the plates and screws to provide mechanical stabilization of bone fragments during the healing process. The system is designed to provide fixed-angle support for distal fragments and cortical fixation to the diaphysis, aiding in the anatomical reduction and fixation of fractures and osteotomies.
Clinical Evidence
Bench testing only. Engineering analysis and performance testing were conducted to demonstrate substantial equivalence to predicate devices.
Technological Characteristics
System components include medical-grade titanium alloy plates, washers, and screws, and cobalt-chrome (CoCr) cannulated polyaxial locking screws. Stainless steel K-wires are provided for provisional fixation only. The system utilizes fixed-angle peg technology and polyaxial locking mechanisms. Provided non-sterile; requires facility-based sterilization.
Indications for Use
Indicated for fixation of fractures and osteotomies of the distal radius in patients requiring bone stabilization.
Regulatory Classification
Identification
Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.
Predicate Devices
Geminus Volar Distal Radius Plate System (K111620)
K111620 — SKELETAL DYNAMICS GEMINUS VOLAR DISTAL RADIUS PLATE MODEL STANDARD 3 HOLE RIGHT, STANDARD 3 HOLE LEFT, SKELETAL DYNAMICS · Skeletal Dynamics, LLC · Aug 23, 2011
K122737 — GEMINUS FOSSA SPECIFIC PLATE SYSTEM · Skeletal Dynamics, LLC · Oct 3, 2012
K252949 — Geminus Volar Distal Radius Plating System · Skeletal Dynamics, Inc. · Nov 13, 2025
K122310 — GEMINUS FOSSA SPECIFIC PLATE SYSTEM · Skeletal Dynamics, LLC · Jan 18, 2013
K030198 — HAND INNOVATIONS, INC. DISTAL RADIUS FRACTURE REPAIR SYSTEM · Hand Innovations, Inc. · Apr 1, 2003
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
Skeletal Dynamics, LLC Ana Escagedo President 7300 North Kendall Drive, Suite 400 Miami, Florida 33156
October 10, 2018
Re: K182492
Trade/Device Name: Geminus Volar Distal Radius Plate System Regulation Number: 21 CFR 888.3030 Regulation Name: Single/Multiple Component Metallic Bone Fixation Appliances And Accessories Regulatory Class: Class II Product Code: HRS Dated: September 7, 2018 Received: September 11, 2018
Dear Ana Escagedo:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Jesse Muir -S 2018.10.10 12:54:36 -04'00'
For: Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K182492
Device Name
Geminus Volar Distal Radius Plate System
Indications for Use (Describe)
The Skeletal Dynamics Geminus Volar Distal Radius Plate System is intended for fixation of fractures and osteotomies of the distal radius
| Type of Use (Select one or both, as applicable) |
|--------------------------------------------------------------------------------------|
| <label><input type="checkbox"/> Research Use (Part 21 CFR 601.21, Subpart B)</label> |
| <label><input type="checkbox"/> Compassionate Use (21 CFR 601.21, Subpart E)</label> |
> Prescription Use (Part 21 CFR 801 Subpart D)
| | Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
## Skeletal Dynamics Geminus Volar Distal Radius Plate System
August 21, 2018
#### Submitter:
Skeletal Dynamics, LLC 7300 North Kendall Drive Suite 400 Miami, FL 33156 Tel: (305) 596-7585 Fax: (305) 596-7591 Contact: Ana M. Escagedo, President Email: aescagedo@skeletaldynamics.com
Establishment Registration Number: 3006742481
#### Name and Classification:
Name Geminus Volar Distal Radius Plate System Common Name Plate, Fixation, Bone 21 CFR §888.3030 Classification Product Code HRS Class Class II
#### Predicate Devices:
Primary: Geminus Volar Distal Radius Plate System (K111620) Synthes LCP Diaphyseal-metaphyseal (dia-meta) Volar Distal Radius Plate (K070946)
#### Description of the Device:
The Skeletal Dynamics Geminus Volar Distal Radius Plate System contains bone plates for the repair of distal volar radial fractures. Included in the set are titanium bone screws, fixation pegs, fragment plates, and specialized instrumentation. Also included are a Hook Plate Extension to buttress a volar marginal fragment, and cannulated cobalt chrome polyaxial locking screws for trajectories different than those of the fixed angled bone plates.
The Geminus Volar Distal Radius Plates are available in various sizes and are made of medical grade titanium alloy. Cortical screws affix the plate to the diaphysis and fixed angle pegs are used for distal bone fragments. The system is provided non-sterile and is sterilized in the user facility.
The Geminus Volar Distal Radius Plate System is comprised of:
- Titanium alloy plates, washers and screws ●
- CoCr Cannulated Polyaxial Locking Screw (PLS) ●
- Stainless steel K-wires (for provisional fixation; not for implantation) ●
- System specific instrumentation ●
{4}------------------------------------------------
## Indications for Use:
The Geminus Volar Distal Radius Plate System is intended for fixation of fractures and osteotomies of the distal radius.
# Technological Characteristics:
The substantial equivalence of the Geminus Volar Distal Radius Plate System to the predicate device is demonstrated by similarities in intended use, indications for use, materials, design (fundamental scientific technology), performance, sterility and packaging, and does not present any new issues of safety or effectiveness.
# Performance Testing:
Engineering analysis and testing demonstrated that the Geminus Volar Distal Radius Plate System is substantially equivalent to predicate devices currently marketed. Therefore, the subject device is as safe and effective as legally marketed predicate devices.
#### Conclusion:
The Skeletal Dynamics Geminus Volar Distal Radius Plate System is substantially equivalent to the predicate device identified in this premarket notification.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.