Activ Ankle

K173641 · Newclip Technics · HRS · Jan 16, 2018 · Orthopedic

Device Facts

Record IDK173641
Device NameActiv Ankle
ApplicantNewclip Technics
Product CodeHRS · Orthopedic
Decision DateJan 16, 2018
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3030
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Activ Ankle range is intended for the fixation of fractures, osteotomies and pseudarthroses of the distal and the diaphyseal fibula, the distal tibia and for the syndesmotic repair in adults.

Device Story

Activ Ankle range consists of metallic bone plates, screws, and washers for orthopedic fixation. Used by surgeons in clinical settings for stabilization of distal/diaphyseal fibula and distal tibia fractures, osteotomies, and pseudarthroses, as well as syndesmotic repair. Implants provided either non-sterile (requiring healthcare professional sterilization) or sterile via gamma radiation. Instruments provided non-sterile for professional sterilization. Single-use kits available. Device functions as mechanical fixation hardware; surgeon selects appropriate plate/screw configuration based on fracture pattern to provide structural support during bone healing. Benefits include restoration of skeletal alignment and stability.

Clinical Evidence

No clinical studies were performed. Evidence is based on bench testing, including comparative static and fatigue bend tests (ASTM F382) and LAL quantitative kinetic chromogenic endotoxin testing.

Technological Characteristics

Materials: Titanium alloy Ti-6Al-4V ELI (ASTM F 136, ISO 5832-3). Components: Plates, screws, washers. Sterilization: Gamma radiation or professional steam sterilization. Mechanical testing: Static and fatigue bend tests per ASTM F382.

Indications for Use

Indicated for fixation of fractures, osteotomies, and pseudarthroses of distal/diaphyseal fibula, distal tibia, and syndesmotic repair in adults.

Regulatory Classification

Identification

Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ January 16, 2018 Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" and "ADMINISTRATION" in blue text. Newclip Technics % J.D. Webb The OrthoMedix Group, Inc. 1001 Oakwood Blvd Round Rock, Texas 78681 # Re: K173641 Trade/Device Name: Activ Ankle Regulation Number: 21 CFR 888.3030 Regulation Name: Single/Multiple Component Metallic Bone Fixation Appliances And Accessories Regulatory Class: Class II Product Code: HRS, HWC Dated: November 17, 2017 Received: November 24, 2017 Dear Mr. Webb: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {1}------------------------------------------------ Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, # Katherine D. Kavlock -S for Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ### Indications for Use 510(k) Number (if known) K173641 Device Name Activ Ankle Indications for Use (Describe) The Activ Ankle range is intended for the fixation of fractures, osteotomies and pseudarthroses of the distal and the diaphyseal fibula, the distal tibia and for the syndesmotic repair in adults. | Type of Use (Select one or both, as applicable) | | |----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------| | <span style="font-size:12px">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) | ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | | CONTINUE ON A SEPARATE PAGE IF NEEDED. | | | This section applies only to requirements of the Paperwork Reduction Act of 1995. | | | <b>*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*</b> | | | The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: | | | Department of Health and Human Services<br>Food and Drug Administration<br>Office of Chief Information Officer<br>Paperwork Reduction Act (PRA) Staff<br>PRAStaff@fda.hhs.gov | | "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ Image /page/3/Picture/1 description: The image shows the logo for NEWCLIP-TECHNICS. The logo features the letters 'nc' stacked on top of each other inside of a circle. The 't' in 'nc' is stylized with an upward-pointing arrow. The words 'NEWCLIP-TECHNICS' are written in all caps below the circle. #### 510 (k) Summary for the ACTIV ANKLE range 4. In accordance with 21 CFR 807.92 of the Federal Code of Regulations, the following 510(k) summary is submitted for the Activ Ankle range. Summary preparation date: October 25, 2017 | 1. Submitter: | Contact Person: | |-------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | NEWCLIP TECHNICS<br>P.A. de la Lande Saint Martin<br>45 rue des Garottières<br>F-44115 Haute-Goulaine - France<br>Telephone: (33) 2 28 21 37 12 | J.D. Webb<br>The OrthoMedix Group, Inc.<br>1001 Oakwood Blvd<br>Round Rock, TX 78681<br>Telephone: 512-388-0199 | | 2. Trade name: | Activ Ankle | | Common Name: | Plate, Fixation, Bone / Screw, Fixation, bone | | Product code: | HRS - Plate, Fixation, Bone<br>HWC - Screw, Fixation, Bone | | Classification Name: | Single/multiple component metallic bone<br>fixation appliances and accessories.<br>(21 CFR part. 888.3030)<br>Smooth or threaded metallic bone<br>fixation fastener (21 CFR part. 888.3040) | {4}------------------------------------------------ Image /page/4/Picture/1 description: The image shows the logo for NEWCLIP-TECHNICS. The logo features the letters "nct" inside of a circle. The "t" in "nct" has an upward-pointing arrow as part of the letter. The words "NEWCLIP-TECHNICS" are written in all caps below the circle. # 3. Primary predicate or legally marketed devices which are substantially equivalent: - Activ Ankle Locking Plating System of Newclip Technics (K143061) . ## Secondary predicate or legally marketed devices which are substantially equivalent: - Ortholoc® 3Di Ankle Fracture System of Wright Medical Technology . (K131093) - Small Fragment Set 3.5 of AAP Implantate AG (K113652) . - . Alians Elbow Locking Plating System (K152289) - Large Screws of Newclip Technics (K172596) . | 4. Description of the device: | The Activ Ankle range consists of plates with screws designed for fixation of fractures, osteotomies and pseudarthroses of the distal and the diaphyseal fibula, the distal tibia and for the syndesmotic repair in adults. Washers are also available for use with screws. The implants of the Activ Ankle range will be provided non sterile for sterilization by health care professionals prior to use or provided sterile by gamma sterilization. The instruments of the Activ Ankle range will be provided non sterile for sterilization by health care professionals prior to use. Single use kits (Initial A) contain implants and instruments, or instruments only | |-------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| |-------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| Materials: Titanium alloy Ti-6Al-4V ELI (conform to ASTM F 136 and ISO 5832-3). provided sterile by gamma sterilization. - Function: The implants of Activ Ankle range are intended for the fixation of fractures, osteotomies and pseudarthroses of the distal and the diaphyseal fibula, the distal tibia and for the syndesmotic repair in adults. {5}------------------------------------------------ Image /page/5/Picture/1 description: The image shows the logo for New Clip-Technics. The logo consists of a circle containing the letters "nct" in lowercase. The "t" in "nct" is stylized with an upward-pointing arrow. Below the circle, the words "NEW CLIP-TECHNICS" are written in a sans-serif font, with a hyphen separating "CLIP" and "TECHNICS". # 5. Substantial equivalence claimed to predicate devices: The Activ Ankle range is substantially equivalent to the predicate devices in terms of intended use, design, materials used, mechanical safety and performance. ### 6. Intended use: The Activ Ankle range is indicated for the fixation of fractures, osteotomies and pseudarthroses of the distal and the diaphyseal fibula, the distal tibia and for the syndesmotic repair in adults. # 7. Non-clinical Test Summary: The following tests were conducted: - Comparative static bend test. (ASTM F382) ● - . Comparative fatigue bend test. (ASTM F382) - Endotoxins testing is performed using LAL quantitative kinetic . chromogenic method. # 8. Clinical Test Summary: No clinical studies were performed. # 9. Conclusions Non-clinical and Clinical: Newclip Technics considers the Activ Ankle range to be equivalent to the predicate devices listed above. This conclusion is based upon the device's similarities in principles of operation, technology, materials, and indications for use.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...