Wright's ORTHOLOC® 3Di Ankle Plating System is intended for fixation of fractures, osteotomies, and non-unions of the distal tibia and fibula such as: Lateral Malleolar Fractures Syndesmosis injuries Medial Malleolar Fractures Bi-Malleolar Fractures Tri-Malleolar Fractures Posterior Malleolar Fractures Distal Anterior Tibia Fractures Vertical Shear Fractures of the Medial Malleolous Pilon Fractures Distal Tibia Shaft Fractures Distal Fibula Shaft Fractures Distal Tibia Periarticular Fractures Medial Malleolar Avulsion Fractures Lateral Malleolar Avusion Fractures. The ORTHOLOC® 3Di locking screws are intended for use with Wright's ORTHOLOC® 3Di Plating Systems of the same base material. ORTHOLOC® Screws are indicated for use in bone reconstruction, osteotomy, arthrodesis, joint fusion, fracture repair, and fracture fixation, appropriate for the size of the device. Wright's washers are intended to prevent a screw head from breaking through the cortex of the bone by distributing the forces/load over a large area when used for fracture fixation of bone fragments.
Device Story
ORTHOLOC® 3Di Ankle Plating System is a metallic bone fixation system for distal tibia and fibula. System comprises titanium alloy plates with polyaxial locking screw holes, compression slots, and syndesmosis slots. Used by orthopedic surgeons in clinical settings for fracture repair and stabilization. Plates accept 2.7mm locking screws, 2.7mm/3.5mm/4.0mm bone screws, and 4.0mm compression screws. Washers distribute load to prevent screw head breakthrough. Device provides rigid fixation to facilitate bone healing and restore structural integrity of the ankle joint.
Clinical Evidence
No clinical data. Substantial equivalence is based on mechanical analysis and design comparison to the predicate device.
Technological Characteristics
Titanium alloy bone plates with polyaxial locking screw holes. Includes compression slots, syndesmosis slots, and k-wire holes. Dimensions vary by anatomical contouring (distal tibia/fibula). System uses metallic bone screws and washers for fixation. Sterilization method not specified.
Indications for Use
Indicated for patients requiring fixation of fractures, osteotomies, and non-unions of the distal tibia and fibula, including malleolar, pilon, and shaft fractures, syndesmosis injuries, and avulsion fractures. Also indicated for bone reconstruction, arthrodesis, and joint fusion.
Regulatory Classification
Identification
Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.
{0}------------------------------------------------
K131093 page 1 of 2
Create Motion.® WRIC
### 510(K) SUMMARY OF SAFETY AND EFFECTIVENESS
In accordance with the Food and Drug Administration Rule to implement provisions of the Safe Medical Devices Act of 1990 and in conformance with 21 CFR 807.92, this information serves as a Summary of Safety and Effectiveness for the use of the ORTHOLOC® 3Di Ankle Plating System.
Wright Medical Technology, Inc. (a)(1). Submitted By: 5677 Airline Road Arlington, TN 38002 April 17, 2013 Date: Ryan Bormann Contact Person: Regulatory Affairs Specialist II Office - (901) 867-4409 Fax - (901) 867-4190 (a)(2). Proprietary Name: ORTHOLOC® 3Di Ankle Plating System Bone Plate System Common Name: 21 CFR 888.3030 - Class II Classification Name and Reference: HRS: Plate, Fixation Bone Device Product Code, Device Panel: (a)(3). Predicate Device: K 102429 ORTHOLOC® 3Di Ankle Plating System
## (a)(4). Device Description .
4
The ORTHOLOC® 3Di .Ankle Plating System contains plates belonging to 1 of 2 general categories-distal tibia and fibula-based on the contouring of each plate: All plates feature polyaxial locking screw holes. Some plates have k-wire holes, compression slots, or syndesmosis slots. The plates are made from titanium alloy and accept 2.7mm ORTHOLOC® 3Di locking screws. 2.7mm. 3.5mm and 4.0mm ORTHOLOC® Bone Screws, and 4.0mm Wright™ Compression Screws (cleared under K082320, now branded DART-FIRE®). Washers are also available for use with the ORTHOLOC® Bone Screws.
The design features of the ORTHOLOC® 3Di Ankle Plating System - Line Addition are substantially equivalent to the design features of the predicate ORTHOLOC® 3Di Ankle Plating System
## (a)(5). INTENDED USE
#### Ankle Plates:
Wright's ORTHOLOC® 3Di Ankle Plating System is intended for fixation of fractures, osteotomies, and non-unions of the distal tibia and fibula such as:
| headquarters | | |
|---------------------------------|-------------------|---------|
| Wright Medical Technology, Inc. | 5677 Airline Road | wmt.com |
で、この上、このことになっていました。 この日には、このため、この時には、このため、この時には、この時には、この時には、いつでも、このと、あり、このことになるので、
A | - |
{1}------------------------------------------------
K131093 page 2 of 2
- Lateral Malleolar Fractures
- Syndesmosis injuries .
- Medial Malleolar Fractures ●
- Bi-Malleolar Fractures .
- Tri-Malleolar Fractures ●
- Posterior Malleolar Fractures �
- Distal Anterior Tibia Fractures ●
- Vertical Shear Fractures of the Medial Malleolous .
- Pilon Fractures .
- Distal Tibia Shaft Fractures �
- Distal Fibula Shaft Fractures .
- Distal Tibia Periarticular Fractures .
- Medial Malleolar Avulsion Fractures
- Lateral Malleolar Avulsion Fractures ●
#### ORTHOLOC® 3Di Locking Screws:
The ORTHOLOC® 3Di locking screws are intended for use with Wright's ORTHOLOC® 3Di Plating Systems of the same base material.
### ORTHOLOC® Bone Screws:
ORTHOLOC® Screws are indicated for use in bone reconstruction, osteotomy, arthrodesis, joint fusion, fracture repair, and fracture fixation, appropriate for the size of the device.
#### Washer
Wright's washers are intended to prevent a screw head from breaking through the cortex of the bone by distributing the forces/load over a large area when used for fracture fixation of bone fragments.
## (a)(6). Technological Characteristics Comparison
The ORTHOLOC® 3Di Ankle Plating System - Line Addition and the legally marketed predicate ORTHOLOC® 3Di Ankle Plating System have identical indications, utilize the same instrumentation, and are identical in material, sterilization methods, and selection.
#### (b)(1). Substantial Equivalence - Non-Clinical Evidence
Through mechanical analysis the new plates do not represent a new worst-case. ●
#### (b)(2). Substantial Equivalence - Clinical Evidence
N/A
## (b)(3). Substantial Equivalence - Conclusions
The design characteristics of the subject system do not raise any new types of questions of safety or effectiveness. From the evidence submitted in this 510(k), the subject line addition can be expected to perform at least as well as the predicate systems.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract symbol that resembles an eagle or a stylized human figure with outstretched arms. The logo is printed in black and white.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
### May 21, 2013
Wright Medical Technology, Incorporated % Mr. Ryan Bormann Regulatory Affairs Specialist II 5677 Airline Road Arlington, Tennessee 38002
Re: K131093
Trade/Device Name: ORTHOLOC® 3Di APS - Straight Tubular Plates Regulation Number: 21 CFR 888.3030 Regulation Name: Single/multiple component metallic bone fixation appliances and accessories Regulatory Class: Class II Product Code: HRS Dated: April 17, 2013 Received: April 23, 2013
Dear Mr. Bormann:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent ffor the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21
{3}------------------------------------------------
## Page 2 - Mr. Ryan Bormann
CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Image /page/3/Picture/8 description: The image shows the name "Erin D. Keith" in a stylized font. The first name, "Erin," is in a simple, bold typeface. The middle initial "D" is enclosed in a decorative, geometric shape. The last name, "Keith," is also in a bold font, with some decorative elements.
For
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# Indications for Use
510(k) Number (if known): K131093
page 1 of 2 ·
A3-1
Device Name: ORTHOLOC® 3Di APS - Straight Tubular Plates
Indications For Use:
Wright's ORTHOLOC® 3Di Ankle Plating System is intended for fixation of fractures, osteotomies, and non-unions of the distal tibia and fibula such as:
- . Lateral Malleolar Fractures
- Syndesmosis injuries .
- Medial Malleolar Fractures �
- Bi-Malleolar Fractures .
- Tri-Malleolar Fractures .
- Posterior Malleolar Fractures .
- Distal Anterior Tibia Fractures ●
- Vertical Shear Fractures of the Medial Malleolous .
- . Pilon Fractures
- Distal Tibia Shaft Fractures .
- Distal Fibula Shaft Fractures .
- Distal Tibia Periarticular Fractures .
- Medial Malleolar Avulsion Fractures .
- . Lateral Malleolar Avusion Fractures
#### ORTHOLOC® 3Di Locking Screws:
The ORTHOLOC® 3Di locking screws are intended for use with Wright's ORTHOLOC® 3Di Plating Systems of the same base material.
#### ORTHOLOC® Bone Screws:
ORTHOLOC® Screws are indicated for use in bone reconstruction, osteotomy, arthrodesis, joint fusion, fracture repair, and fracture fixation, appropriate for the size of the device.
#### Washer
Wright's washers are intended to prevent a screw head from breaking through the cortex of the bone by distributing the forces/load over a large area when used for fracture fixation of bone fragments.
{5}------------------------------------------------
# K131093 page 2 of 2
## Prescription Use _ xxx (Part 21 CFR 801 Subpart D)
# AND/OR .
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) . . · ·
Elizabet标订%Erank -S
Division of Orthopedic Devices
Concurrence of CDRH, Office of Device Evaluation (ODE) 1 of 1
A3-2
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.