VARIAX 2 COMPRESSION PLATING SYSTEM

K130009 · Stryker Trauma AG · HRS · Mar 22, 2013 · Orthopedic

Device Facts

Record IDK130009
Device NameVARIAX 2 COMPRESSION PLATING SYSTEM
ApplicantStryker Trauma AG
Product CodeHRS · Orthopedic
Decision DateMar 22, 2013
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3030
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Stryker VariAx 2 Compression Plating System is intended for internal fixation of bones in adult patients.

Device Story

VariAx 2 Compression Plating System is an internal bone fixation device. System consists of metallic plates and compatible screws used to stabilize fractures in radius, ulna, humerus, clavicle, distal tibia, and fibula. Used by surgeons in clinical settings to treat fractures, osteotomies, mal-unions, and non-unions. Components available in sterile and non-sterile formats; plate lengths range from 30-246mm. Device provides mechanical stabilization of bone segments to facilitate healing. Substantial equivalence established via dynamic fatigue testing compared to predicate systems.

Clinical Evidence

No clinical data. Substantial equivalence supported by non-clinical bench testing, specifically dynamic fatigue testing.

Technological Characteristics

Metallic bone fixation plates and screws. Metallic construction. Dimensions: 30-246mm length. Mechanical fixation principle. Sterile and non-sterile availability.

Indications for Use

Indicated for internal fixation of fractures in the radius, ulna, humerus, clavicle, distal tibia, and fibula in adult patients. Includes osteotomies, mal-unions, non-unions, and single, segmental, or comminuted fractures in patients with normal or osteopenic bone density.

Regulatory Classification

Identification

Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K130009 Page 1/2 · #### 510(k) Summary ### MAR 2 2 2013 Proprietary Name: VariAx 2 Compression Plating System Common Name: Bone plates Bone Screws Classification Name and Reference: Single/multiple component metallic bone fixation appliances and accessories 21 CFR §888.3030 Smooth or threaded metallic bone fixation fastener 21 CFR §888.3040 Regulatory Class: Class II 87 HRS: Plate, Fixation, Bone Product Codes: 87 HWC : Screw, Fixation, Bone Stryker Trauma AG Sponsor: Bohnackerweg 1 CH-2545 Selzach Switzerland Contact Person: Estela Celi Regulatory Affairs Specialist Date Prepared: December 29, 2012 estela.celi@stryker.com 325 Corporate Drive Mahwah, NJ 07430 Phone: (201) 831-6461 Fax: (201) 831-3461 ## Description This Traditional 510(k) submission is being supplied to the U.S. FDA to provide authorization to market the new VariAx 2 Compression Plating System. The VariAx 2 Compression Plating System is an internal fixation device that consists of various plates used with compatible screws to fit different types of fractures in the radius, ulna, humerus, clavicle, distal tibia and fibula. The subject components will be available sterile and non-sterile. The plates will be available in sizes ranging from 30-246mm in length. {1}------------------------------------------------ #### Intended Use The Stryker VariAx 2 Compression Plating System is intended for internal fixation of bones in adult patients. #### Indications The Stryker VariAx 2 Compression Plating System is indicated for internal fixation of fractures in the Radius, Ulna, Humerus, Clavicle, Distal tibia and Fibula for the following indications: - osteotomies, mal-unions, and non-unions t - single, segmental, and comminuted fractures . - . normal bone density and osteopenic bone #### Summary of Technologies Device comparison showed that the proposed device is substantially equivalent in intended use, materials and performance characteristics to the following predicate device: - . K000684 Small Fragment Dynamic Compression Locking (DCL) System - K082807- SYNTHES (USA) 3.5 AND 4.5 MM LOCKING COMPRESSION PLATE SYSTEM WITH EXPANDED INDICATIONS #### Non-Clinical Testing Non-clinical laboratory testing was performed for the VariAx 2 Compression Plating System components to determine substantial equivalence. Testing demonstrated that the VariAx 2 Compression Plating System is substantially equivalent to the predicate device currently cleared for marketing. The following testing was performed Dynamic Fatigue Testing . #### Clinical Testing Clinical testing was not required for this submission. #### Conclusion The VariAx 2 Compression Plating System is substantially equivalent to the predicate device identified in this premarket notification. {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three lines representing its body and wings. The eagle is facing right. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" is arranged in a circular fashion around the eagle. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 March 22, 2013 Stryker Trauma AG % Ms. Estela Celi Regulatory Affairs Specialist 325 Corporate Drive Mahwah, New Jersey 07430 Re: K130009 Trade/Device Name: VariAx 2 Compression Plating System Regulation Number: 21 CFR 888.3030 Regulation Name: Single/multiple component metallic bone fixation appliances and accessories Regulatory Class: Class II Product Code: HRS, HWC Dated: January 18, 2013 Received: January 24, 2013 Dear Ms. Celi: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical {3}------------------------------------------------ #### Page 2 - Ms. Estela Celi device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours. # Erin DKeith Mark N. Melkerson -Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ #### Indications for Use 510(k) Number (if known): _K130009 Device Name: VariAx 2 Compression Plating System Indications for Use: The Stryker VariAx 2 Compression Plating System is indicated for internal fixation of fractures in the Radius, Ulna, Humerus, Clavicle, Distal tibia and Fibula for the following indications: - osteotomies, mal-unions, and non-unions . - single, segmental, and comminuted fractures . - normal bone density and osteopenic bone · ● Prescription Use X Over-The-Counter Use AND/OR (Part 21 CFR 801 Subpart D) (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) # Elizabet 标签Frank -S Division of Orthopedic Devices
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...