Synthes 1.5mm Mini Fragment LCP System is indicated for fixation of small bones and small fragments, osteotomies, arthrodeses, replantations, and reconstructions of small bones and small fragments, particularly in osteopenic bone.
Device Story
The Synthes 1.5mm Mini Fragment LCP System is a metallic bone fixation system consisting of low-profile plates and locking/cortex screws. It is used by surgeons to stabilize small bones and fragments during orthopedic procedures such as osteotomies, arthrodeses, and reconstructions. The system provides mechanical support to bone segments, facilitating healing, particularly in cases of osteopenic bone where traditional fixation may be less effective. The device is used in clinical settings (OR) by orthopedic surgeons.
Clinical Evidence
No clinical data provided; substantial equivalence is based on design and technological characteristics.
Technological Characteristics
Low-profile plates and screws available in stainless steel and titanium. System utilizes locking and cortex screw technology for bone fixation. Class II device under 21 CFR 888.3030.
Indications for Use
Indicated for fixation of small bones and small fragments, including osteotomies, arthrodeses, replantations, and reconstructions, particularly in patients with osteopenic bone.
Regulatory Classification
Identification
Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.
Predicate Devices
Synthes Modular Mini Fragment LCP System
Synthes Stainless Steel Modular Hand System
Related Devices
K090492 — 1.5MM SMALL LOCKING PLATE, 1.5MM NON-LOCKING AND LOCKING SCREWS · DePuy Orthopaedics, Inc. · Mar 31, 2009
K113652 — LOQTEQ SMALL FRAGMENT SET LOQTEQ CORTICAL SCREW3.5, SMALL HEAD, T15, SELF TAPPING LOQTEQ CORTICAL SCREW 3.5, T15, SELF T · Aap Implantate AG · Aug 30, 2012
K202959 — PolyLock Plating System · Fusion Orthopedics, LLC · Nov 24, 2020
K081546 — SMALL BONE LOCKING PLATING SYSTEM · DePuy Orthopaedics, Inc. · Aug 26, 2008
K200062 — Foot and Ankle System · Leith Medical, LLC · Feb 22, 2021
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for Synthes. The logo consists of a stylized symbol on the left and the word "SYNTHES" in bold, sans-serif font on the right. A horizontal line is present under the word "SYNTHES".
.
K09004749 1/1)
:
| 3.0 | 510(k) Summary | Page 1 of 1 |
|-----|-----------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Sponsor: | Synthes (USA)<br>Karl J. Nittinger<br>1301 Goshen Parkway<br>West Chester, PA 19380<br>(610) 719-6941<br>APR -1 200 |
| | Device Name: | Synthes (USA) 1.5mm Mini Fragment LCP System |
| | Classification: | Class 11, §888.3030 - Single / multiple component metallic bone<br>fixation appliance and accessories. |
| | Predicate Device: | Synthes Modular Mini Fragment LCP System<br>Synthes Stainless Steel Modular Hand System |
| | Device Description: | The Synthes 1.5mm Mini Fragment LCP System consists of low-<br>profile plates of various shapes as well as locking and cortex screws<br>which are intended to treat small bones and small fragments. The<br>plates and screws of the 1.5mm Mini Fragment LCP System are<br>available in stainless steel and titanium. |
| | Intended Use: | Synthes 1.5mm Mini Fragment LCP System is indicated for<br>fixation of small bones and small fragments, osteotomies,<br>arthrodeses, replantations, and reconstructions of small bones and<br>small fragments, particularly in osteopenic bone. |
| | Substantial<br>Equivalence: | Information presented supports substantial equivalence. |
・
{1}------------------------------------------------
Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized eagle with three stripes forming its body and wing. The logo is surrounded by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Synthes (USA) % Mr. Karl J. Nittinger 1301 Goshen Pkwy. West Chester, PA 19380
APR - 1 2009
Re: K090047 Trade/Device Name: Synthes (USA) 1.5mm Mini Fragment LCP System Regulation Number: 21 CFR 888.3030 Regulation Name: Single/multiple component metallic bone fixation appliances and ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
-accessories-Regulatory Class: Class II
Product Code: HRS Dated: January 5, 2009 Received: January 7, 2009
Dear Mr. Nittinger:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{2}------------------------------------------------
## Page 2 - Mr. Karl J. Nittinger
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (240) 276-3474. For questions regarding of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at (240).276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at toll-free number (800) 638-2041 or (240) 276-3150 or the Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Barbara Bachus
Mark N. Melk Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
| 2.0 | Indications for Use |
|---------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Number (if known): | K090047 (pg1/1) |
| Device Name: | Synthes (USA) 1.5mm Mini Fragment LCP System |
| Indications for Use: | Synthes 1.5mm Mini Fragment LCP System is indicated for fixation of small bones and small fragments, osteotomies, arthrodeses, replantations, and reconstructions of small bones and small fragments, particularly in osteopenic bone. |
Prescription Use X (Per 21 CFR 801.109)
AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Charles Beecher for
(Division Sign-Off)
Division of General, Restorative, and Neurological Devices
**510(k) Number** K090047
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.