FLUOROPERM 92, 60, 30 AND 151 AND PARAGON HDS, HDS 100 AND THIN RIGID GAS PERMEABLE CONTACT LENSES
Applicant
Paragon Vision Sciences
Product Code
HQD · Ophthalmic
Decision Date
May 1, 2007
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 886.5916
Device Class
Class 2
Indications for Use
FluoroPerm® 92 rigid gas permeable spherical or aspheric contact lenses are indicated for daily wear and extended wear from 1 to 7 days between removals for cleaning and disinfection as recommended by the eye care practitioner. FluoroPerm® 92 rigid gas permeable bifocal, or toric contact lenses are indicated for daily wear only. FluoroPerm® 92 rigid gas permeable spheric and bifocal contact lenses are indicated for the correction of visual acuity in not-aphakic persons with nondiseased eyes that are nearsighted (myopic), farsighted (hyperopic), and may exhibit corneal astigmatism up to 4.00 diopters that does not interfere with visual acuity. FluoroPerm® 92 toric contact lenses are indicated to correct astigmatism of up to 6.00 diopters. FluoroPerm® 92 bifocal lenses are indicated to treat presbyopia up to +4.00 D add power. FluoroPerm® 92 contact lenses are indicated for management of irregular corneal conditions such as keratoconus, pellucid marginal degeneration, or following penetrating keratoplasty, radial keratotomy, or LASIK surgery in otherwise nondiseased eyes. [Note: Similar statements provided for FluoroPerm® 60, Paragon HDS®, FluoroPerm® 30, Paragon Thin™, FluoroPerm® 151, and Paragon HDS® 100, with varying wear schedules.]
Device Story
Rigid gas permeable (RGP) contact lenses; lathe-cut fluorosilicone acrylate polymers; daily or extended wear (1-7 days) depending on specific lens model. Corrects refractive errors and manages irregular corneal conditions. Practitioner-prescribed; patient-worn. Lenses feature spherical, aspheric, bifocal, or toric designs. May include UV-blocking additives (Uvinul D-49) and color tints (D&C dyes) for handling. Provides visual acuity correction and corneal surface regularization.
Clinical Evidence
No clinical data provided; substantial equivalence based on material safety and performance characteristics of previously approved RGP lens materials.
Technological Characteristics
Lathe-cut rigid gas permeable contact lenses; materials: paflufocon A, B, C, D (fluorosilicone acrylate polymers). Features: spherical, aspheric, bifocal, or toric designs; optional UV absorber (Uvinul D-49) blocking up to 97% of light <380 nm; optional color tints (D&C Green No. 6, Yellow No. 9, Violet No. 2, Red No. 17).
Indications for Use
Indicated for non-aphakic patients with nondiseased eyes for correction of myopia, hyperopia, and corneal astigmatism (up to 4.00D, or up to 6.00D for toric). Bifocal lenses indicated for presbyopia (up to +4.00D add). Also indicated for irregular corneal conditions including keratoconus, pellucid marginal degeneration, or post-surgical (penetrating keratoplasty, radial keratotomy, LASIK).
Regulatory Classification
Identification
A rigid gas permeable contact lens is a device intended to be worn directly against the cornea of the eye to correct vision conditions. The device is made of various materials, such as cellulose acetate butyrate, polyacrylate-silicone, or silicone elastomers, whose main polymer molecules generally do not absorb or attract water.
Predicate Devices
Boston® XO (hexafocon A)
Boston EO® (enflufocon B)
Boston ES® (enflufocon A)
Related Devices
K052983 — FLUOROPERM 151 (PAFLUFOCON D) AND PARAGON HDS 100 (PAFLUFOCON D) RIGID GAS PERMEABLE CONTACT LENSES · Paragon Vision Sciences · Feb 10, 2006
K253822 — GPA (hexafocon A) Rigid Gas Permeable Contact Lenses · Zhuhai Fitlens Medical Technology Co., Ltd. · Mar 27, 2026
K202860 — Valley Contax Single Vision (SV) Gas Permeable Contact Lens (roflufocon D, roflufocon E, tisilfocon A, hexafocon A, enflufocon A), Valley Contax Goldeneye Aspheric Front Multifocal (AFM) Gas Permeable Contact Lens (roflufocon D, roflufocon E, tisilfocon A, hexafocon A, enflufocon A), Valley Contax V Kone (VK) Gas Permeable Contact Lens (roflufocon D, roflufocon E, tisilfocon A, hexafocon A, enflufocon A) · Valley Contax, Inc. · May 11, 2021
K241398 — SDJMASTERVISION (fluoroxyfocon A) Rigid Gas Permeable Contact Lenses · Tianjin Mastervision Technology Co., Ltd. · Oct 22, 2024
K240618 — Acuity 200 (fluoroxyfocon A) Rigid Gas Permeable Contact Lens; Acuity 100 (hexafocon A) Rigid Gas Permeable Contact Lens; Acuity 200 with Tangible® Hydra-PEG® (fluoroxyfocon A) Rigid Gas Permeable Contact Lens; Acuity 100 with Tangible® Hydra-PEG® (hexafocon A) Rigid Gas Permeable Contact Lens · Acuity Polymers, Inc. · Apr 3, 2024
Submission Summary (Full Text)
{0}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Image /page/0/Picture/2 description: The image shows a logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three lines representing its wings, and the text "U.S. Department of Health & Human Services" is arranged in a circle around the eagle. The logo is black and white.
MAY - 1 2007
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Paragon Vision Sciences William E. Mevers, Ph.D. Vice President, Science & Technology 947 East Impala Mesa, AZ 85204-6619
Re: K070637
FluoroPerm® 30 and Paragon Thin™ (paflufocon C); Trade/Device Name: FluoroPerm® 60 and Paragon HDS® (paflufocon B): FluoroPerm® 92 (paflufocon A); and, FluoroPerm® 151 and Paragon HDS® 100 (paflufocon D) RGP Contact Lenses Regulation Number: 21 CFR 886.5916 Regulation Name: Rigid gas permeable contact lenses Regulatory Class: Class II Product Code: HQD Dated: March 5. 2007 Received: March 7, 2007
Dear Dr. Meyers:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
Page 2 - Dr. William Meyers
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address (car) supply industr. Support inationali.
Sincerely yours,
M.B. Egclehner SiwD
Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications For Use
K070637
510(k) Number (if known):
Device Name:
FluoroPerm® 92 (paflufocon A) rigid gas permeable contact lenses FluoroPerm® 60 (paflufocon B) rigid gas permeable contact lenses Paragon HDS® (paflufocon B) rigid gas permeable contact lenses FluoroPerm 30 (paflufocon C) rigid gas permeable contact lenses FluoroPerm® 151 (paflufocou D) rigid gas permeable contact lenses Paragon HDS® 100 (paflufocon D) rigid gas permeable contact lenses
#### Indications For Use:
#### FluoroPerm® 92
FluoroPerm® 92 rigid gas permeable spherical or aspheric contact lenses are indicated for daily wear I horor chil. 92 ngla ga permission between removals for cleaning and disinfection as recommended by the eye care practitioner. FluoroPerm® 92 rigid gas permeable bifocal, or toric contact lenses are indicated for daily wear only.
FluoroPerm® 92 rigid gas permeable spheric and bifocal contact lenses are indicated for the correction of visual acuity in not-aphabic persons with nondiseased eyes that are nearsighted (myopic), farsighted (byperopic), and may exhibit corneal astigmatism up to 4.00 diopters that does not interfere with visual acuity. FluoroPerm® 92 toric contact lenses are indicated to correct astigmatism of up to 6.00 diopters. FluoroPerm® 92 bifocal leases are indicated to treat presbyopia up to +4.00 D add power.
FluoroPerm® 92 contact lenses are indicated for management of irregular corneal conditions such as kerator ones, pellucid marginal degeneration, or following penctrating keratoplasty, radial keratotomy, or LASIK surgery in otherwise nondiseased eyes.
# FluoroPerm® 60
FluoroPerm® 60 rigid gas permeable spherical or aspheric contact lenses are indicated for daily wear I tobior and out from 1 to 7 days between removals for cleaning and disinfection as recommonded by the eye care practitioner. FluoroPerm® 60 rigid gas permeable bifocal, or toric contact leases are indicated for daily wear only.
FluoroPerm® 60 rigid gas permeable spheric and bifocal contact leases are indicated for the correction of visual acuity in not-aphakic persons with nondiseased eyes that are nearsighted (myopic), farsighted (byperopic), and may exhibit corneal astigmatism up to 4.00 diopters that does not interfece with visual acuity. FivoroPerm® 60 toric contact lenses are indicated to correct astigmatism of up to 6.00 diopters. FluoroPerm® 60 bifocal lenses are indicated to treat presbyopia up to +4.00 D add power.
FluoroPerm® 60 contact leases are indicated for management of irregular comcal conditions such as keratoconus, pellucid marginal degeneration, or following penetrating keratoriasty, radial keratotomy, or LASIK surgery in otherwise nondiseased eyes.
Page 1 of 3
{3}------------------------------------------------
# Paragon HDS40
Paragon HDS® rigid gas permeable spheric contact lenses are indicated for daily wear and extended wear from 1 to 7 days between removals for cleaning and disinfection as recommended by the cye care practitioner. Paragon HDS® rigid gas permeable bifocal, or toric contact lenses are indicated for daily wear only.
Paragon HDS rigid gas permeable spherical, asphenc and bifocal contact lenses are indicated for the correction of visual aculty in not-aphakic persons with nondiseased eyes that are nearsighted (myopic), farsighted (hyperopic), and may exhibit corneal astigmatism up to 4.00 diopters that does not interfere with visual acuity. Paragon HDS tone contact leases are indicated to correct astigmatism of up to 6.00 diopters. Faragon HDS® bifocal lenses are indicated to treat presbyopia up to +4.00 D add power.
Paragon HDS contact lenses are indicated for management of irregular corneal conditions such as keratoconus, pellucid marginal degeneration, or following penetrating keratoplasty, radial keratotorny, or LASIK surgeryin otherwise nondiseased eyes.
# FluoroPerm® 30
FluoroPerm® 30 rigid gas permeable contact leases are indicated for daily wear as recommended by the eye care practitioner.
FluoroPerm* 30 rigid gas permeable spheric and bifocal contact lenses are indicated for the correction of visual acuity in not-aphakic persons with nondiseased eyes that are nearsighted (myopic), farsighted (hyperopic), and may exhibit corneal astigmatism up 10 4.00 diopters that does not interfere with visual acuity. FluoroPerm® 30 toric contact lenses are indicated to correct astigmatism of up 10 6.00 diopters. FluoroPerm* 30 bifocal lenses are indicated to treat presbyopia up to +4.00 D add power.
FluoroPerm® 30 contact lenses are indicated for management of irregular comeal conditions such as keratoconus, pellucid marginal degeneration, or following penetrating keratoplasty, radial keratotomy, or LASIK surgery in otherwise nondiseased eyes.
## Paragon Thin"
Paragon Thin " rigid gas permeable contact lenses are indicated for daily wear as recommended by the eve care practitioner.
Paragon Thin rigid gas permeable spheric and bifocal contact lenses are indicated for the correction of visual acuity in not-aphakic persons with nondiseased tyes that are nearsighted (myopic), farsighted (hyperopic), and may exhibit comes astigmatism up to 4.00 diopters that does not interfere with visual acuity. Paragon Thin" witc contact lenses are indicated to correct astigmatism of up to 6.00 diopters. Paragon Thin bifocal lenses are indicated to treat presbyopia up 10 +4.00 D add power.
Paragon Thin" contact lenses are indicated for management of iregular corneal conditions such as keratoconus, pellucid marginal degeneration, or following penetrating keratoplasty, radial keratotomy, or LASIK surgery in otherwise nondiseased eyes.
Page 2 of 3
{4}------------------------------------------------
# FluoroPerm® 151
FluoroPerm® 151 rigid gas permeable spheric contact lenses are indicated for daily wear and extended wear from 1 to 7 days between removals for cleaning and disinfection as recommended by the eye care practitioner. FluoroPerme 151 rigid gas permeable bifocal, or toric contact lenses are indicated for daily wear only.
FluoroPerm 151 rigid gas permeable spherical, aspheric and bifocal contact lenses are indicated for the correction of visual acuity in not-aphakic persons with nondiseased cycs that are nearsighted (myopic), farsighted (hyperopic), and may exhibit comeal astigmatism up to 4.00 diopters that does not interfere with visual acuity. FluoroPenn® 151 toric contact lenses are indicated to correct astigmatism of up to 6.00 diopers. FluoroPerm 151 bifocal lenses are indicated to treat presbyopia up to +4.00 D add power.
FluoroPerm 151 contact lenses are indicated for management of irregular corneal conditions such as keratoconus, pellucid marginal degeneration, or following penetrating keratoplasty, radial keratotomy, or LASIK surgery in otherwise nondiseased eyes.
#### Paragon HDS 100
Paragon HDS 100 rigid gas permeable spheric contact lenses are indicated for daily wear and extended wear from 1 to 7 days between removals for cleaning and disinfection as recommended by the eye care practitioner. Paragon HDS 100 rigid gas permeable bifocal, or toric contact lenses are indicated for daily wear only.
Paragon HDS 100 rigid gas permeable spheric and bifocal contact lenses are indicated for the correction of visual scuity in not-aphakic persons with nondiseased eves that are nearsighted (myopic), farsighted (hyperopic), and may exhibit comeal astigmatism up to 4.00 diovers that docs not interfere with visual acuity. Paragon HDS® 100 toric contact lenses are indicated to correct astigmatism of up to 6.00 diopters. Paragon HDS 100 bifocal lenses are indicated to treat presbyopia up to +4.00 D add power.
Paragon HDS 100 contact leases are indicated for management of irregular corneal conditions such as keratoconus, pellucid marginal degeneration, or following peactrating keratoplasty, radial keratotomy, or LASIK surgery in otherwise nondiseased eyes.
Prescription Use X (Part 21 CFR 801 Sub) bart D)
Over -
(2) CFR
Over - the - Counter Use (2) CFR 801 Subpan C).
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IS NEEDED)
AND/OR
Concurrence of CDRH, Office of Device Evaluation (ODE)
Marc Roblez
0
(Division Sign-Off)
Division of Ophthalmic Ear,
Nose and Throat Devises
Page 3 of 3
umbal
{5}------------------------------------------------
# 510(k) Summary of Safetv and Effectiveness
MAY - 1 2007
# Submitter Information
Paragon Vision Sciences 947 E. Impala Ave. Mesa, AZ 85204
Contact Person:
William Meyers PhD Vice President, Science & Technology
Email: wemevers@MSN.com Phone: 480 507 7606
## Device Name
FluoroPerm® 92 (patlufocon A) rigid gas permeable contact lenses FluoroPerm® 60 (paflufocon B) rigid gas permeable contact lenses Paragon HDS (paflufocon B) rigid gas permeable contact lenses FluoroPerm 30 {paflufocon C) rigid gas permeable contact lenses " (paflufocon C) rigid gas permeable contact lenses Paragon Thin FluoroPerm 151 (paflufocon D) rigid gas permeable contact lenses Paragon HDS* 100 (paflufocon D) rigid gas permeable contact lenses
## Reason for Submission
Expanded Indication
## Predicate Devices
Classification Name: rigid gas permeable (hydrophobic) contact lens
Proprietary Names: Boston® XO* (chexafocon A),
Boston EO® (enflufocon EO® (enflufocon B)
Boston ES* (enflufocon A)
### Description of the device:
FluoroPerm® 92, FluoroPerm® 60, Paragon HDS*, FluoroPerm* 30, Paragon Thin"*, and FluoroPerm® 151, and Paragon HDS® 100 lenses are available as lath cut rigid gas permeable contact lenses for daily wear only. The lenses are manufactured from these FDA approved contact lens materials; pastufocon A (FluoroPerm" 02), pallufocon B (FluoroPerm) 60, Paragon HDS®), patlufocon C (Fluorosco (Film The Paragon Time), and Poly San Poly San Parised Bron fluorosilicone acrylate monomers.
The lenses may be tinted to offer a handling and for locating the lens. The tinted lenses contain one or more of the following approved color additives: D&C Green No. 6, Peroxide Yellow No. 9, D&C Violet No. 2 and D&C Red No. 17.
The lenses may be available with an ultraviolet absorber (not in all colors and materials). The ultraviolet absorber, Uvinul D-49, has been integrated as an additive within the polymer matrix, blocking up to 97% of light below 380 nm. The UV absorber is 2.2'-dihydroxy-4,4'-dimethoxybenzophenone.
The lens designs have a posterior surface consisting of a base curve and a series of up to four annular spherical or aspherical curves peripheral to the base curve.
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.