SDJMASTERVISION (fluoroxyfocon A) Rigid Gas Permeable Contact Lenses
Applicant
Tianjin Mastervision Technology Co., Ltd.
Product Code
HQD · Ophthalmic
Decision Date
Oct 22, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 886.5916
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The SDJMASTERVISION (fluoroxyfocon A) Rigid Gas Permeable Contact Lenses are indicated for daily wear for the correction of refractive error (myopia, hyperopia, presbyopia and/or astigmatism) in aphakic and non-aphakic persons with non-diseased eyes. The lenses may be prescribed for daily wear in otherwise non-diseased eyes that require a rigid contact lens for the management of irregular corneal conditions such as keratoconus, pellucid marginal degeneration, or following penetrating keratoplasty or refractive (e.g., LASIK) surgery. The SDJMASTERVISION (fluoroxyfocon A) Rigid Gas Permeable Contact Lenses are also indicated for therapeutic use in eyes with ocular surface disease (e.g. ocular Graft-versus-Host disease, Sjögren's syndrome, dry eye syndrome and Filamentary Keratitis), limbal stem cell deficiency (e.g. Stevens-Johnson syndrome, chemical radiation and thermal burns), disorders of the skin (e.g. atopy, ectodermal dysplasia), neurotrophic keratitis (e.g. Herpes simplex, Herpes zoster, Familial Dysautonomia), and corneal exposure (e.g. anatomic, paralytic) that might benefit from the presence of an expanded tear reservoir and protection against an adverse environment. When prescribed for therapeutic use for a distorted cornea or ocular surface disease, the lens may concurrently provide correction of refractive error. The SDJMASTERVISION (fluoroxyfocon A) Rigid Gas Permeable Contact Lenses may be cleaned and disinfected using a chemical (not heat) lens care system.
Device Story
Rigid gas permeable (RGP) contact lens; manufactured from siloxanyl fluoromethacrylate copolymers (fluoroxyfocon A); lathe-cut production. Available in spherical, toric, multifocal, scleral, and aspheric designs; includes visibility tints and optional UV absorber. Supplied dry/non-sterile; requires chemical cleaning/disinfection. Prescribed by eye care professionals for daily wear to correct refractive errors or manage irregular corneas/ocular surface diseases. Provides expanded tear reservoir and environmental protection for therapeutic applications. Benefits include vision correction and ocular surface management consistent with standard RGP lenses.
Clinical Evidence
No clinical data. Bench testing performed to verify manufacturing repeatability and adherence to ANSI Z80.20 specifications. Biocompatibility testing conducted per ISO 10993-5 (cytotoxicity), ISO 10993-11 (systemic toxicity), and ISO 10993-23 (acute ocular irritation) confirmed safety. Bioburden testing confirmed <100 CFU/lens.
Technological Characteristics
Material: fluoroxyfocon A (siloxanyl fluoromethacrylate copolymer); FDA Group #3 fluoro silicone acrylate. Dk: 200 x 10^-11. Refractive index: 1.430. Modulus: 1194 MPa. Hardness: 78 Shore D. Specific gravity: 1.18. Water content: <1%. Production: Lathe-cut. Connectivity: None. Sterilization: Supplied non-sterile.
Indications for Use
Indicated for daily wear for correction of refractive error (myopia, hyperopia, presbyopia, astigmatism) in aphakic/non-aphakic persons with non-diseased eyes; management of irregular corneal conditions (keratoconus, pellucid marginal degeneration, post-penetrating keratoplasty, post-refractive surgery); and therapeutic use for ocular surface disease, limbal stem cell deficiency, skin disorders, neurotrophic keratitis, and corneal exposure.
Regulatory Classification
Identification
A rigid gas permeable contact lens is a device intended to be worn directly against the cornea of the eye to correct vision conditions. The device is made of various materials, such as cellulose acetate butyrate, polyacrylate-silicone, or silicone elastomers, whose main polymer molecules generally do not absorb or attract water.
Predicate Devices
Acuity 200™ (fluoroxyfocon A), Acuity 100™ (hexafocon A), Acuity 200™ with Tangible® Hydra-PEG® (fluoroxyfocon A), and Acuity 100™ with Tangible® Hydra-PEG® Rigid Gas Permeable Contact Lenses (K240618)
Acuity 200™ (fluoroxyfocon A) Rigid Gas Permeable Contact Lens (K201194)
Acuity 200™ (fluoroxyfocon A) Rigid Gas Permeable Contact Lens (K203571)
Submission Summary (Full Text)
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
October 22, 2024
Tianjin MasterVision Technology Co., Ltd. % Bret Andre Principal Consultant Andre Vision and Device Research 6119 Canter Lane West Linn, OR 97068
Re: K241398
Trade/Device Name: SDJMASTERVISION (fluoroxyfocon A) Rigid Gas Permeable Contact Lenses Regulation Number: 21 CFR 886.5916 Regulation Name: Rigid Gas Permeable Contact Lens Regulatory Class: Class II Product Code: HQD Dated: May 3, 2024 Received: September 20, 2024
Dear Bret Andre:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{2}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents titled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801): medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketingsafety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-devicereporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn).
{3}------------------------------------------------
Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/deviceadvice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
## J Angelo Green -S
J.Angelo Green, Ph.D. Assistant Director DHT1A: Division of Ophthalmic Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT, and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
{4}------------------------------------------------
## Indications for Use
Submission Number (if known)
K241398
Device Name
SDJMASTERVISION (fluoroxyfocon A) Rigid Gas Permeable Contact Lenses
Indications for Use (Describe)
The SDJMASTERVISION (fluoroxyfocon A) Rigid Gas Permeable Contact Lenses are indicated for daily wear for the correction of refractive error (myopia, hyperopia, presbyopia and/or astigmatism) in aphakic and non-aphakic persons with non-diseased eyes. The lenses may be prescribed for daily wear in otherwise non-diseased eyes that require a rigid contact lens for the management of irregular corneal conditions such as keratoconus, pellucid marginal degeneration, or following penetrating keratoplasty or refractive (e.g., LASIK) surgery.
The SDJMASTERVISION (fluoroxyfocon A) Rigid Gas Permeable Contact Lenses are also indicated for therapeutic use in eyes with ocular surface disease (e.g. ocular Graft-versus-Host disease, Sjögren's syndrome, dry eye syndrome and Filamentary Keratitis), limbal stem cell deficiency (e.g. Stevens-Johnson syndrome, chemical radiation and thermal burns), disorders of the skin (e.g. atopy, ectodermal dysplasia), neurotrophic keratitis (e.g. Herpes simplex, Herpes zoster, Familial Dysautonomia), and corneal exposure (e.g. anatomic, paralytic) that might benefit from the presence of an expanded tear reservoir and protection against an adverse environment. When prescribed for therapeutic use for a distorted cornea or ocular surface disease, the lens may concurrently provide correction of refractive error.
The SDJMASTERVISION (fluoroxyfocon A) Rigid Gas Permeable Contact Lenses may be cleaned and disinfected using a chemical (not heat) lens care system.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{5}------------------------------------------------
# 510 (k) SUMMARY
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.
#### K241398 The assigned 510(k) number is:
#### I. SUBMITTER
| Date Prepared: | October 17, 2024 |
|-------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Name:<br>Address: | Tianjin MasterVision Technology Co., Ltd.<br>1st Floor No.15 Building. No. 89 Heyuan Road, Jing-jin Technology<br>Valley, Wuqing District, Tianjin. P.R. China |
| Contact Person: | Delia Ke<br>Managing Director |
| Phone number: | +86 18920270903 |
| Consultant: | Bret Andre<br>Andre Vision and Device Research<br>6119 Canter Ln.<br>West Linn, OR 97068 |
| Phone number: | (503) 372-5226 |
#### II. DEVICE
| Trade Name: | SDJMASTERVISION (fluoroxyfocon A) Rigid Gas Permeable<br>Contact Lenses |
|-------------|-------------------------------------------------------------------------|
|-------------|-------------------------------------------------------------------------|
| Common<br>Name: | Rigid gas permeable contact lens |
|-------------------------|-----------------------------------------------------|
| Classification<br>Name: | Rigid gas permeable contact lens. (21 CFR 886.5916) |
| Regulatory<br>Class: | Class II |
| Product Code: | HQD |
#### ~Reason for Submission~ New Device
{6}------------------------------------------------
#### III. PREDICATE DEVICE
The SDJMASTERVISION (fluoroxyfocon A) Rigid Gas Permeable Contact Lenses for daily wear are substantially equivalent to the following predicate devices:
- "Acuity 200™ (fluoroxyfocon A), Acuity 100™ (hexafocon A), Acuity 200™ with . Tangible® Hydra-PEG® (fluoroxyfocon A), and Acuity 100™ with Tangible® Hydra-PEG® Rigid Gas Permeable Contact Lenses" By Acuity Polymers, Inc. 510(k) number; K240618 Product Code: HQD
- I "Acuity 200™ (fluoroxyfocon A) Rigid Gas Permeable Contact Lens" By Acuity Polymers, Inc. 510(k) number; K201194 Product Code: HQD
- I "Acuity 200™ (fluoroxyfocon A) Rigid Gas Permeable Contact Lens" By Acuity Polymers, Inc. 510(k) number: K203571 Product Code: HQD
#### IV. DEVICE DESCRIPTION
The SDJMASTERVISION (fluoroxyfocon A) Rigid Gas Permeable Contact Lenses are manufactured from machine latheable rigid gas permeable materials composed of siloxanyl fluoromethacrylate copolymers that are tinted for visibility and available with or without an ultraviolet (UV) light absorber.
The SDJMASTERVISION (fluoroxyfocon A) Rigid Gas Permeable Contact Lenses for daily wear are made available in spherical, toric, multifocal, scleral and aspheric designs in the following lens parameters:
| Parameter | Range | Tolerance |
|--------------------|--------------------|------------------------------------------------------------------------------------------------------------------|
| Base Curve | 4.00mm to 11.50mm | ± 0.05 mm |
| Center Thickness | 0.08mm to 0.75mm | ± 0.02 mm |
| Diameter | 7.0mm to 21.0mm | ± 0.10mm |
| Spherical Power | -20.00D to +20.00D | ± 0.12 (0 to = 5D)<br ± 0.18 (5 to = 10.0D)<br ± 0.25 (10 to = 15D)<br ± 0.37 (15 to = 20D)<br ± 0.50 (over 20D) |
| Cylindrical Power | Up to 9.00D | ± 0.25 (0 to = 2D)<br ± 0.37 (2 to = 4D)<br ± 0.50 (over 4D) |
| Multifocal Power | +1.00D to 4.00D | ± 0.25D |
| Surface Appearance | | Lenses should be clear with no surface<br>defect |
{7}------------------------------------------------
The following table depicts the physical properties of the SDJMASTERVISION (fluoroxyfocon A) Rigid Gas Permeable Contact Lenses:
| | Fluoroxyfocon A Rigid Gas Permeable Contact Lens | |
|-----------------------------|--------------------------------------------------------------------------------|--|
| Refractive Index | 1.430 | |
| Modulus (MPa) | 1194 | |
| Hardness (Shore D) | 78 | |
| Specific Gravity | 1.18 | |
| Oxygen Permeability<br>(Dk) | $200 x 10^{-11} (cm^2/sec) (ml O_2/ml x mm Hg @ 35°C)$ | |
| Color Additives | Visibility Tints – D&C Green #6, D&C Violet #2, Solvent Yellow 18, D&C Red #17 | |
The SDJMASTERVISION (fluoroxyfocon A) Rigid Gas Permeable Contact Lenses are supplied dry (without storage solution) and non-sterile in a screw top case and must be cleaned and conditioned prior to use.
#### V. INDICATIONS FOR USE
The SDJMASTERVISION (fluoroxyfocon A) Rigid Gas Permeable Contact Lenses are indicated for daily wear for the correction of refractive error (myopia, hyperopia, presbyopia and/or astigmatism) in aphakic and non-aphakic persons with non-diseased eyes. The lenses may be prescribed for daily wear in otherwise non-diseased eyes that require a rigid contact lens for the management of irregular corneal conditions such as keratoconus, pellucid marginal degeneration, or following penetrating keratoplasty or refractive (e.g., LASIK) surgery.
The SDJMASTERVISION (fluoroxyfocon A) Rigid Gas Permeable Contact Lenses are also indicated for therapeutic use in eves with ocular surface disease (e.g. ocular Graft-versus-Host disease, Sjögren's syndrome, dry eye syndrome and Filamentary Keratitis), limbal stem cell deficiency (e.g. Stevens-Johnson syndrome, chemical radiation and thermal burns), disorders of the skin (e.g. atopy, ectodermal dysplasia), neurotrophic keratitis (e.g. Herpes simplex, Herpes zoster, Familial Dysautonomia), and corneal exposure (e.g. anatomic, paralytic) that might benefit from the presence of an expanded tear reservoir and protection against an adverse environment. When prescribed for therapeutic use for a distorted cornea or ocular surface disease, the lens may concurrently provide correction of refractive error.
The SDJMASTERVISION (fluoroxyfocon A) Rigid Gas Permeable Contact Lenses may be cleaned and disinfected using a chemical (not heat) lens care system.
{8}------------------------------------------------
#### COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH VI. PREDICATE DEVICE
The SDJMASTERVISION (fluoroxyfocon A) Rigid Gas Permeable Contact Lenses are substantially equivalent to the predicate device (cleared under K240618) in terms of the following:
- I Intended use - daily wear contact lenses
- I Indications for use
- I Classification - Lenses, Rigid Gas Permeable, Daily Wear (21 CFR 886.5916)
- 트 FDA material group - group # 3 fluoro silicone acrylate
- 트 USAN material (fluoroxyfocon A)
- I Production method - lathe cut
The SDJMASTERVISION (fluoroxyfocon A) Rigid Gas Permeable Contact Lenses are substantially equivalent to the predicate device (cleared under K201194) in terms of the following:
- 트 Intended use - daily wear contact lenses
- I Classification - Lenses, Rigid Gas Permeable, Daily Wear (21 CFR 886.5916)
- FDA material group group # 3 fluoro silicone acrylate l
- I USAN material (fluoroxyfocon A)
- I Production method - lathe cut
The SDJMASTERVISION (fluoroxyfocon A) Rigid Gas Permeable Contact Lenses are substantially equivalent to the predicate device (cleared under K203571) in terms of the following:
- 트 Intended use - daily wear contact lenses
- I Classification - Lenses, Rigid Gas Permeable, Daily Wear (21 CFR 886.5916)
- I FDA material group - group # 3 fluoro silicone acrylate
- 트 USAN material (fluoroxyfocon A)
- I Production method - lathe cut
{9}------------------------------------------------
The following matrix illustrates the production method, lens function and material characteristics of the SDJMASTERVISION (fluoroxyfocon A) Rigid Gas Permeable Contact Lenses, as well as the predicate devices.
| | SDJMASTERVISION<br>Rigid Gas Permeable<br>Contact Lenses | Acuity 200™ Acuity<br>100™ Rigid Gas<br>Permeable Contact Lens | Acuity 200™ Rigid Gas<br>Permeable Contact Lens | Acuity 200™ Rigid Gas<br>Permeable Contact Lens | | Indications for Use |
|-----------------------|---------------------------------------------------------------------------|---------------------------------------------------------------------------|---------------------------------------------------------------------------|---------------------------------------------------------------------------|--------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Subject Device | Predicate Device<br>(K240618) | Predicate Device<br>(K201194) | Predicate Device<br>(K203571) | SDJMASTERVISION<br>Rigid Gas Permeable<br>Contact Lenses<br>(subject device) | The SDJMASTERVISION (fluoroxyfocon A) Rigid Gas Permeable Contact Lenses are indicated<br>for daily wear for the correction of refractive error (myopia, hyperopia, presbyopia and/or<br>astigmatism) in aphakic and non-aphakic persons with non-diseased eyes. The lenses may be<br>prescribed for daily wear in otherwise non-diseased eyes that require a rigid contact lens for the<br>management of irregular corneal conditions such as keratoconus, pellucid marginal degeneration,<br>or following penetrating keratoplasty or refractive (e.g., LASIK) surgery. |
| Classification | Class II<br>Lenses, Rigid Gas<br>Permeable, Daily Wear<br>21 CFR 886.5916 | Class II<br>Lenses, Rigid Gas<br>Permeable, Daily Wear<br>21 CFR 886.5916 | Class II<br>Lenses, Rigid Gas<br>Permeable, Daily Wear<br>21 CFR 886.5916 | Class II<br>Lenses, Rigid Gas<br>Permeable, Daily Wear<br>21 CFR 886.5916 | | The SDJMASTERVISION (fluoroxyfocon A) Rigid Gas Permeable Contact Lenses are also<br>indicated for therapeutic use in eyes with ocular surface disease (e.g. ocular Graft-versus-Host<br>disease, Sjögren's syndrome, dry eye syndrome and Filamentary Keratitis), limbal stem cell<br>deficiency (e.g. Stevens-Johnson syndrome, chemical radiation and thermal burns), disorders of<br>the skin (e.g. atopy, ectodermal dysplasia), neurotrophic keratitis (e.g. Herpes simplex, Herpes<br>zoster, Familial Dysautonomia), and corneal exposure (e.g. anatomic, paralytic) that might benefit<br>from the presence of an expanded tear reservoir and protection against an adverse environment.<br>When prescribed for therapeutic use for a distorted cornea or ocular surface disease, the lens may<br>concurrently provide correction of refractive error. |
| Product Code | HQD | HQD | HQD | HQD | | The SDJMASTERVISION (fluoroxyfocon A) Rigid Gas Permeable Contact Lenses may be<br>cleaned and disinfected using a chemical (not heat) lens care system. |
| FDA Group # | Group # 3 Fluoro Silicone<br>Acrylate | Group # 3 Fluoro Silicone<br>Acrylate | Group # 3 Fluoro Silicone<br>Acrylate | Group # 3 Fluoro Silicone<br>Acrylate | Acuity 200™ Acuity<br>100™ Rigid Gas<br>Permeable Contact<br>Lens (K240618) | The Acuity 200™ (fluoroxyfocon A), Acuity 100™ (hexafocon A), Acuity 200™ with Tangible®<br>Hydra-PEG® (fluoroxyfocon A), and Acuity 100™ with Tangible® Hydra-PEG® Rigid Gas<br>Permeable Contact Lenses are indicated for daily wear for the correction of refractive error<br>(myopia, hyperopia, presbyopia and/or astigmatism) in aphakic and non-aphakic persons with non-<br>diseased eyes. The lenses may be prescribed for daily wear in otherwise non-diseased eyes that<br>require a rigid contact lens for the management of irregular corneal conditions such as<br>keratoconus, pellucid marginal degeneration, or following penetrating keratoplasty or refractive<br>(e.g., LASIK) surgery.<br>The Acuity 200™ (fluoroxyfocon A), Acuity 100™ (hexafocon A), Acuity 200™ with Tangible®<br>Hydra-PEG® (fluoroxyfocon A), and Acuity 100™ with Tangible® Hydra-PEG® Rigid Gas<br>Permeable Contact Lenses are also indicated for therapeutic use in eyes with ocular surface<br>disease (e.g. ocular Graft-versus-Host disease, Sjögren's syndrome, dry eye syndrome and<br>Filamentary Keratitis), limbal stem cell deficiency (e.g. Stevens-Johnson syndrome, chemical<br>radiation and thermal burns), disorders of the skin (e.g. atopy, ectodermal dysplasia), neurotrophic<br>keratitis (e.g. Herpes simplex, Herpes zoster, Familial Dysautonomia), and corneal exposure (e.g.<br>anatomic, paralytic) that might benefit from the presence of an expanded tear reservoir and<br>protection against an adverse environment. When prescribed for therapeutic use for a distorted<br>cornea or ocular surface disease, the lens may concurrently provide correction of refractive error.<br>The Acuity 200™ (fluoroxyfocon A), Acuity 100™ (hexafocon A), Acuity 200™ with Tangible®<br>Hydra-PEG® (fluoroxyfocon A), and Acuity 100™ with Tangible® Hydra-PEG® Rigid Gas<br>Permeable Contact Lenses may be cleaned and disinfected using a chemical (not heat) lens care<br>system. |
| Material (USAN) | fluoroxyfocon A | hexafocon A,<br>fluoroxyfocon A | fluoroxyfocon A | fluoroxyfocon A | Acuity 200™<br>(fluoroxyfocon A)<br>Rigid Gas Permeable<br>Contact Lens<br>(K201194) | The Acuity 200™ (fluoroxyfocon A) Rigid Gas Permeable Contact Lenses are indicated for daily<br>wear for the correction of refractive error (myopia, hyperopia, presbyopia and/or astigmatism) in<br>aphakic and non-aphakic persons with non-diseased eyes. The lens may be prescribed in spherical<br>and aspheric powers ranging from -20.00 D to +20.00 D for daily wear. |
| Production Method | Lathe-Cut | Lathe-Cut | Lathe-Cut | Lathe-Cut | Acuity 200™<br>(fluoroxyfocon A)<br>Rigid Gas Permeable<br>Contact Lens<br>(K203571) | The Acuity 200™ (fluoroxyfocon A) Rigid Gas Permeable Contact Lens is indicated for daily<br>wear for the correction of refractive error (myopia, hyperopia, presbyopia and/or astigmatism) in<br>aphakic and non-aphakic persons with non-diseased eyes. The lenses may be prescribed for daily<br>wear in otherwise non-diseased eyes that require a rigid contact lens for the management of<br>irregular corneal conditions such as keratoconus, pellucid marginal degeneration, or following<br>penetrating keratoplasty or refractive (e.g., LASIK) surgery. |
| Intended Use | Daily Wear | Daily Wear | Daily Wear | Daily Wear | | The Acuity 200™ (fluoroxyfocon A) Rigid Gas Permeable Contact Lens may be cleaned and<br>disinfected using a chemical (not heat) lens care system. |
| Water Content (%) | <1% | <1% | <1% | <1% | | |
| UV Absorber Available | Yes | Yes | Yes | Yes | | |
{10}------------------------------------------------
{11}------------------------------------------------
#### VII. PERFORMANCE DATA
### ~ Non-Clinical Studies ~
The following testing was performed on finished SDJMASTERVISION (fluoroxyfocon A) Rigid Gas Permeable Contact Lenses:
Bench Testing-manufacturing verification testing was conducted to demonstrate the ability of Tianjin MasterVision Technology Co., Ltd. to manufacture lenses, on a repeatable basis, from fluoroxyfocon A supplied lens blanks to a variety of prescribed parameters. All lenses were manufactured to established finished product specifications within the ANSI Z80.20 tolerance.
Bioburden Testing-bioburden testing conducted on fluoroxyfocon A rigid gas permeable lenses manufactured at Tianjin MasterVision Technology Co., Ltd. demonstrated that the colony forming units (CFU) per lens was within the established acceptance criteria of less than 100 CFU per lens.
Biocompatibility Testing-biocompatibility studies were conducted on finished fluoroxyfocon A gas permeable contact lenses produced at the Tianjin MasterVision Technology Co., Ltd. facility. The following biocompatibility/toxicology tests were conducted in accordance with Good Laboratory Practice (GLP):
- · In-Vitro Cytotoxicity: Cytotoxicity testing was performed in accordance with ISO 10993-5 with results indicating that the finished lenses are not cytotoxic.
- · Systemic Toxicity: The finished lenses meet the requirements of the systemic injection test in accordance with ISO 10993-11, and the extracts from finished lenses did not induce acute systemic toxicity.
- · Acute Ocular Irritation: Acute ocular irritation testing was performed in accordance with ISO 10993-23, and the extracts from finished lenses did not induce ocular irritation.
## ~ Clinical Studies ~
Finished contact lenses manufactured from fluoroxyfocon A, which is a contact lens material adopted by United States Adopted Name (USAN) council, were previously cleared by the United States Food and Drug Administration (FDA). Based on recommendations in the FDA guidance titled Class II Daily Wear Contact Lenses -Premarket Notification [510(k)] Guidance Document, clinical studies are not required for this 510(k) premarket notification.
{12}------------------------------------------------
## VIII. CONCLUSIONS
## Validity of Scientific Data
Laboratories under Good Laboratory Practice regulations conducted biocompatibility studies following scientific protocols. The data were determined to be scientifically valid under 21 CFR 860.7.
## Substantial Equivalence
Information presented in this premarket notification establishes that the SDJMASTERVISION (fluoroxyfocon A) Rigid Gas Permeable Contact Lenses for daily wear are as safe and effective as the predicate devices when used in accordance with the labeled directions for use and for the proposed indications.
### Risks and Benefits
The risks of the SDJMASTERVISION (fluoroxyfocon A) Rigid Gas Permeable Contact Lenses are the same as those normally attributed to the wearing of rigid gas permeable (RGP) daily wear contact lenses. The benefits to the patient are the same as those for other RGP contact lenses.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.