FLUOROPERM 151 (PAFLUFOCON D) AND PARAGON HDS 100 (PAFLUFOCON D) RIGID GAS PERMEABLE CONTACT LENSES

K052983 · Paragon Vision Sciences · HQD · Feb 10, 2006 · Ophthalmic

Device Facts

Record IDK052983
Device NameFLUOROPERM 151 (PAFLUFOCON D) AND PARAGON HDS 100 (PAFLUFOCON D) RIGID GAS PERMEABLE CONTACT LENSES
ApplicantParagon Vision Sciences
Product CodeHQD · Ophthalmic
Decision DateFeb 10, 2006
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 886.5916
Device ClassClass 2

Indications for Use

FluoroPerm® 30 rigid gas permeable contact lenses are indicated for daily wear as recommended by the eye care practitioner. FluoroPerm® 30 rigid gas permeable spherical, aspheric and bifocal contact lenses are indicated for the correction of visual acuity in not-aphakic persons with nondiseased eyes that are nearsighted (myopic), farsighted (hyperopic), and may exhibit corneal astigmatism up to 4.00 diopters that does not interfere with visual acuity. FluoroPerm® 30 contact lenses are indicated for persons requiring keratoconus management with otherwise non-diseased eyes. FluoroPerm® 30 toric contact lenses are indicated to correct astigmatism of up to 6.00 diopters. FluoroPerm® 30 bifocal lenses are indicated to treat presbyopia up to +4.00 D add power. Paragon Thin™ rigid gas permeable contact lenses are indicated for daily wear as recommended by the eye care practitioner. Paragon Thin™ rigid gas permeable spheric and bifocal contact lenses are indicated for the correction of visual acuity in not-aphakic persons with nondiseased eyes that are nearsighted (myopic), farsighted (hyperopic), and may exhibit corneal astigmatism up to 4.00 diopters that does not interfere with visual acuity. Paragon Thin™ contact lenses are indicated for persons requiring keratoconus management with otherwise non-diseased eyes. Paragon Thin™ toric contact lenses are indicated to correct astigmatism of up to 6.00 diopters. Paragon Thin™ bifocal lenses are indicated to treat presbyopia up to +4.00 D add power. FluoroPerm® 60 rigid gas permeable spherical or aspheric contact lenses are indicated for daily wear and extended wear from 1 to 7 days between removals for cleaning and disinfection as recommended by the eye care practitioner. FluoroPerm® 60 rigid gas permeable bifocal, or toric contact lenses are indicated for daily wear only. FluoroPerm® 60 rigid gas permeable spheric and bifocal contact lenses are indicated for the correction of visual acuity in not-aphakic persons with nondiseased eyes that are nearsighted (myopic), farsighted (hyperopic), and may exhibit corneal astigmatism up to 4.00 diopters that does not interfere with visual acuity. FluoroPerm® 60 contact lenses are indicated for persons requiring keratoconus management with otherwise non-diseased eyes, daily wear application only. FluoroPerm® 60 toric contact lenses are indicated to correct astigmatism of up to 6.00 diopters. FluoroPerm® 60 bifocal lenses are indicated to treat presbyopia up to +4.00 D add power. Paragon HDS® rigid gas permeable spherical or aspheric contact lenses are indicated for daily wear and extended wear from 1 to 7 days between removals for cleaning and disinfection as recommended by the eye care practitioner. Paragon HDS® rigid gas permeable bifocal, or toric contact lenses are indicated for daily wear only. Paragon HDS® rigid gas permeable spherical, aspheric and bifocal contact lenses are indicated for the correction of visual acuity in not-aphakic persons with nondiseased eyes that are nearsighted (myopic), farsighted (hyperopic), and may exhibit corneal astigmatism up to 4.00 diopters that does not interfere with visual acuity. Paragon HDS® contact lenses are indicated for persons requiring keratoconus management with otherwise non-diseased eyes, daily wear application only. Paragon HDS® toric contact lenses are indicated to correct astigmatism of up to 6.00 diopters. Paragon HDS® bifocal lenses are indicated to treat presbyopia up to +4.00 D add power. FluoroPerm® 92 rigid gas permeable spherical or aspheric contact lenses are indicated for daily wear and extended wear from 1 to 7 days between removals for cleaning and disinfection as recommended by the eye care practitioner. FluoroPerm® 92 rigid gas permeable bifocal, or toric contact lenses are indicated for daily wear only. FluoroPerm® 92 rigid gas permeable spheric and bifocal contact lenses are indicated for the correction of visual acuity in not-aphakic persons with nondiseased eyes that are nearsighted (myopic), farsighted (hyperopic), and may exhibit corneal astigmatism up to 4.00 diopters that does not interfere with visual acuity. FluoroPerm® 92 contact lenses are indicated for persons requiring keratoconus management with otherwise non-diseased eyes, daily wear application only. FluoroPerm® 92 toric contact lenses are indicated to correct astigmatism of up to 6.00 diopters. FluoroPerm® 92 bifocal lenses are indicated to treat presbyopia up to +4.00 D add power. FluoroPerm® 151 rigid gas permeable spheric contact lenses are indicated for daily wear and extended wear from 1 to 7 days between removals for cleaning and disinfection as recommended by the eye care practitioner. FluoroPerm® 151 rigid gas permeable bifocal, or toric contact lenses are indicated for daily wear only. FluoroPerm® 151 rigid gas permeable spheric and bifocal contact lenses are indicated for the correction of visual acuity in not-aphakic persons with nondiseased eyes that are nearsighted (myopic), farsighted (hyperopic), and may exhibit corneal astigmatism up to 4.00 diopters that does not interfere with visual acuity. FluoroPerm® 151 contact lenses are indicated for persons requiring keratoconus management with otherwise non-diseased eyes, daily wear application only. FluoroPerm® 151 toric contact lenses are indicated to correct astigmatism of up to 6.00 diopters. FluoroPerm® 151 bifocal lenses are indicated to treat presbyopia up to +4.00 D add power. Paragon HDS® 100 rigid gas permeable spheric contact lenses are indicated for daily wear and extended wear from 1 to 7 days between removals for cleaning and disinfection as recommended by the eye care practitioner. Paragon HDS® 100 rigid gas permeable bifocal, or toric contact lenses are indicated for daily wear only. Paragon HDS® 100 rigid gas permeable spherical, aspheric and bifocal contact lenses are indicated for the correction of visual acuity in not-aphakic persons with nondiseased eyes that are nearsighted (myopic), farsighted (hyperopic), and may exhibit corneal astigmatism up to 4.00 diopters that does not interfere with visual acuity. Paragon HDS® 100 contact lenses are indicated for persons requiring keratoconus management with otherwise non-diseased eyes, daily wear application only. Paragon HDS® 100 toric contact lenses are indicated to correct astigmatism of up to 6.00 diopters. Paragon HDS® 100 bifocal lenses are indicated to treat presbyopia up to +4.00 D add power.

Device Story

Lath-cut rigid gas permeable (RGP) contact lenses; manufactured from fluorosilicone acrylate thermoset copolymers (paflufocon A, B, C, D). Designed for daily or extended wear (1-7 days) depending on specific lens model. Posterior surface features base curve plus four progressive spherical/aspherical curves. Optional UV absorber (Uvinul D-49) and color tints. Prescribed by eye care practitioners for visual acuity correction and keratoconus management. Lenses improve patient vision by correcting refractive errors and managing corneal irregularities associated with keratoconus.

Clinical Evidence

Safety and efficacy of the paflufocon A, B, C, and D materials were previously demonstrated in PMA P870024 and its supplements.

Technological Characteristics

Materials: paflufocon A, B, C, D (fluorosilicone acrylate thermoset copolymers). Geometry: base curve with four progressive spherical/aspherical curves. Additives: D&C colorants, Uvinul D-49 UV absorber. Manufacturing: lath-cut, optional plasma treatment. Properties: Dk range 29 to >145, water content <1%.

Indications for Use

Indicated for not-aphakic persons with non-diseased eyes requiring correction of myopia, hyperopia, corneal astigmatism (up to 4.00D, or up to 6.00D for toric), or presbyopia (up to +4.00D add), and for keratoconus management.

Regulatory Classification

Identification

A rigid gas permeable contact lens is a device intended to be worn directly against the cornea of the eye to correct vision conditions. The device is made of various materials, such as cellulose acetate butyrate, polyacrylate-silicone, or silicone elastomers, whose main polymer molecules generally do not absorb or attract water.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ # FEB 1 0 2006 K052983 # 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS FluoroPerm® 30 and Paragon Thin" (paflufocon C); FluoroPerm® 60 and Paragon HDS® (paflufocon B); FluoroPerm® 92 (patufocon A); and, FluoroPerm® 151 and Paragon HDS® 100 (paflufocon D) RGP Contact Lenses for Daily Wear Keratoconus Indication Date of Submission January 31, 2006 Submitter Company Name: Address: Phone: Fax: Paragon Vision Sciences 945 East Impala Ave., Mesa, AZ 85204 480-892-7602 480-892-3226 Manufacturer Information Company Name: Address: Phone: Fax: Paragon Vision Sciences 945 East Impala Ave., Mesa, AZ 85204 480-892-7602 480-892-3226 Official Correspondent Name: Address: Address: Phone: Fax: William E. Meyers, Ph.D. % Paragon Vision Sciences 945 East Impala Ave., Mesa, AZ 85204 480-507-7606 480-892-3226 Device Identification Common Name: Trade Name: Classification: Predicate Device Contact lens FluoroPerm® 30, FluoroPerm® 60, FluoroPerm® 92, FluoroPerm® 151, Paragon Thin", Paragon HDS and Paragon HDS 100 21 CFR 886.5916; rigid gas permeable contact lens, Class II - daily wear contact lens A claim of substantial equivalence to the Boston® II (itafocon A) rigid gas permeable contact lens in keratoconus lens designs approved in PMA P820065. on 11/17/1983 #### Device Description Keratoconus contact lens designs are available as lath cut rigid gas permeable contact lenses for daily wear only. The lenses are manufactured from these FDA approved contact lens materials; paflufocon A (FluoroPerm® 92), paflufocon B (FluoroPerm® 60, Paragon HDS®), paflufocon C (FluoroPerm® 30, Paragon Thin""), and paflufocon D (FluoroPerm® 151. Paragon HDS® 100). These materials are thermoset copolymers derived from fluorosilicone acrylate monomers. {1}------------------------------------------------ The lenses may be tinted to offer a handling aid for locating the lens. The tinted lenses contain one or more of the following approved color additives: D&C Green No. 6, Peroxide Yellow No. 9, D&C Violet No. 2 and D&C Red No. 17. These products may be plasma treated. The lenses may be available with an ultraviolet absorber (not in all colors and materials). The ultraviolet absorber. Uvinul D-49, has been integrated as an additive within the polymer matrix, blocking up to 97% of light below 380 nm. The UV absorber is 2,2'-dihydroxy-4,4'dimethoxybenzophenone. The lens designs have a posterior surface consisting of a base curve and a series of four spherical or aspherical curves progressively flatter than the base curve. ## Properties of the Contact Lenses | | FluoroPerm | | | | | Paragon | | | |------------------------------|------------|-------|-------|-------|-------|---------|---------|--| | Property | 30 | 60 | 92 | 151 | Thin | HDS | HDS 100 | | | Specific Gravity | 1.14 | 1.15 | 1.10 | 1.10 | 1.145 | 1.16 | 1.10 | | | Refractive Index | 1.466 | 1.453 | 1.453 | 1.442 | 1.463 | 1.449 | 1.442 | | | % Light<br>Transmittance* | 94 | 95 | 93 | 93 | 94 | 95 | 93 | | | % Water Content | < 1 | < 1 | < 1 | < 1 | < 1 | < 1 | < 1 | | | Oxygen<br>Permeability(Dk)** | 30 | 60 | 92 | 151 | 29 | 58 | >145 | | *Blue tinted material **Revised Method of I. Fatt #### Safety and Efficacy The safety and efficacy of paflufocon A (FluoroPerm® 92), paflufocon B (FluoroPerm® 60 and Paragon HDS®), paflufocon C (FluoroPerm® 30 and Paragon Thin") and paflufocon D (FluoroPerm® 151 and Paragon HDS® 100) materials as rigid gas permeable contact lenses have been demonstrated in PMA P870024 and various of its supplements. ### Intended Use Device Name: FluoroPerm® 30 (paflufocon C) rigid gas permeable contact lenses FluoroPerm® 30 rigid gas permeable contact lenses are indicated for daily wear as recommended by the eye care practitioner. FluoroPerm® 30 rigid gas permeable spherical, aspheric and bifocal contact lenses are indicated for the correction of visual acuity in not-aphakic persons with nondiseased eyes that are nearsighted (myopic), farsighted (hyperopic), and may exhibit corneal astigmatism up to 4.00 diopters that does not interfere with visual acuity. FluoroPerm® 30 contact lenses are indicated for persons requiring keratoconus management with otherwise non-diseased eves. FluoroPerm® 30 toric contact lenses are indicated to correct astigmatism of up to 6.00 diopters. FluoroPerm® 30 bifocal lenses are indicated to treat presbyopia up to +4.00 D add power. Paragon Thin (paflufocon C) rigid gas permeable contact lenses Device Name: Paragon Thin" rigid gas permeable contact lenses are indicated for daily wear as recommended by the eye care practitioner. Paragon Thin" rigid gas permeable spheric and bifocal contact tenses are indicated for the correction of visual acuity in not-aphakic persons with nondiseased eves that are nearsighted (myopic), {2}------------------------------------------------ farsighted (hyperopic), and may exhibit corneal astigmatism up to 4.00 diopters that does not interfere with visual acuity. Paragon Thin" contact lenses are indicated for persons requiring keratoconus management with otherwise non-diseased eyes. Paragon Thin" toric contact lenses are indicated to correct astigmatism of up to 6.00 diopters. Paragon Thin" bifocal lenses are indicated to treat presbvopia up to +4.00 D add power. FluoroPerm® 60 (paflufocon B) rigid gas permeable contact lenses Device Name: FluoroPerm® 60 rigid gas permeable spherical or aspheric contact lenses are indicated for daily wear and extended wear from 1 to 7 days between removals for cleaning and disinfection as recommended by the eye care practitioner. FluoroPerm® 60 rigid gas permeable bifocal, or toric contact lenses are indicated for daily wear only. FluoroPerm® 60 rigid gas permeable spheric and bifocal contact lenses are indicated for the correction of visual acuity in not-aphakic persons with nondiseased eyes that are nearsighted (myopic), farsighted (hyperopic), and may exhibit corneal astigmatism up to 4.00 diopters that does not interfere with visual acuity. FluoroPerm® 60 contact lenses are indicated for persons requiring keratoconus management with otherwise non-diseased eyes, daily wear application only. FluoroPerm® 60 toric contact lenses are indicated to correct astigmatism of up to 6.00 diopters. FluoroPerm® 60 bifocal lenses are indicated to treat presbyopia up to +4.00 D add power. Paragon HDS (paflufocon B) rigid gas permeable contact lenses Device Name: Paragon HDS® rigid gas permeable spherical or aspheric contact lenses are indicated for daily wear and extended wear from 1 to 7 days between removals for cleaning and disinfection as recommended by the eye care practitioner. Paragon HDS® rigid gas permeable bifocal, or toric contact lenses are indicated for daily wear only. Paragon HDS® rigid gas permeable spherical, aspheric and bifocal contact lenses are indicated for the correction of visual acuity in not-aphakic persons with nondiseased eyes that are nearsighted (myopic), farsighted (hyperopic), and may exhibit corneal astigmatism up to 4.00 diopters that does not interfere with visual acuity. Paragon HDS® contact lenses are indicated for persons requiring keratoconus management with otherwise non-diseased eyes, daily wear application only. Paragon HDS toric contact lenses are indicated to correct astigmatism of up to 6.00 diopters. Paragon HDS® bifocal lenses are indicated to treat presbyopia up to +4.00 D add power. Device Name: FluoroPerm® 92 rigid gas permeable spherical or aspheric contact lenses are indicated for daily wear and extended wear from 1 to 7 days between removals for cleaning and disinfection as recommended by the eye care practitioner. FluoroPerm® 92 rigid gas permeable bifocal, or toric contact lenses are indicated for daily wear only. FluoroPerm® 92 rigid gas permeable spheric and bifocal contact lenses are indicated for the correction of visual acuity in not-aphakic persons with nondiseased eyes that are nearsighted (myopic), farsighted (hyperopic), and may exhibit corneal astigmatism up to 4.00 diopters that does not interfere with visual acuity. FluoroPerm® 92 contact lenses are indicated for persons requiring keratoconus management with otherwise non-diseased eyes, daily wear application only. FluoroPerm® 92 toric contact lenses are indicated to correct astigmatism of up to 6.00 diopters. FluoroPerm® 92 bifocal lenses are indicated to treat presbyopia up to +4.00 D add power. FluoroPerm® 151 (paflufocon D) rigid gas permeable contact lenses Device Name: FluoroPerm® 151 rigid gas permeable spheric contact lenses are indicated for daily wear and extended wear from 1 to 7 days between removals for cleaning and disinfection as recommended {3}------------------------------------------------ by the eye care practitioner. FluoroPerm® 151 rigid gas permeable bifocal, or toric contact lenses are indicated for daily wear only. FluoroPerm® 151 rigid gas permeable spheric and bifocal contact lenses are indicated for the correction of visual acuity in not-aphakic persons with nondiseased eyes that are nearsighted (myopic), farsighted (hyperopic), and may exhibit corneal astigmatism up to 4.00 diopters that does not interfere with visual acuity. FluoroPerm® 151 contact lenses are indicated for persons requiring keratoconus management with otherwise non-diseased eyes, daily wear application only. F FluoroPerm® 151 toric contact lenses are indicated to correct astigmatism of up to 6.00 diopters. FluoroPerm® 151 bifocal lenses are indicated to treat presbyopia up to +4.00 D add power. Paragon HDS® 100 (paflufocon D) rigid gas permeable contact lenses Device Name: Paragon HDS® 100 rigid gas permeable spheric contact lenses are indicated for daily wear and extended wear from 1 to 7 days between removals for cleaning and disinfection as recommended by the eye care practitioner. Paragon HDS® 100 rigid gas permeable bifocal, or toric contact lenses are indicated for daily wear only. Paragon HDS® 100 rigid gas permeable spheric and bifocal contact lenses are indicated for the correction of visual acuity in not-aphakic persons with nondiseased eves that are nearsighted (myopic), farsighted (hyperopic), and may exhibit corneal astigmatism up to 4.00 diopters that does not interfere with visual acuity. Paragon HDS® 100 contact lenses are indicated for persons requiring keratoconus management with otherwise non-diseased eves, daily wear application only. Paragon HDS 100 toric contact lenses are indicated to correct astigmatism of up to 6.00 diopters. Paragon HDS® 100 bifocal lenses are indicated to treat presbyopia up to +4.00 D add power. #### Substantial Equivalence The FluoroPerm® 30 and Paragon Thin" (paflufocon C); FluoroPerm® 60 and Paragon HDS® (paflufocon B); FluoroPerm® 92 (paflufocon A); and, FluoroPerm® 151 and Paragon HDS® 100 (paflufocon D) Rigid Gas Permeable Contact Lenses are similar to the currently marketed Boston® II (itafocon A) Rigid Gas Permeable Contact Lens, in that all fall into the same FDA contact lens classification (Class II) and all are manufactured by the same manufacturing process. The following Substantial Equivalence Summary Table lists the similarities and differences among the devices. | Similarities | | FluoroPerm® | | | | Paragon<br>Thin™ | Paragon<br>HDS® | Paragon<br>HDS® 100 | |----------------------------------|---------------|---------------|---------------|---------------|---------------|------------------|-----------------|---------------------| | | Boston® II | 30 | 60 | 92 | 151 | Daily Wear | Daily Wear | Daily Wear | | Indications<br>for Use | Daily<br>Wear | Daily<br>Wear | Daily<br>Wear | Daily<br>Wear | Daily<br>Wear | | | | | Differences | | paflufocon | | | | | | | | Material | itafocon | C | B | A | D | C | B | D | | Indication<br>for<br>Keratoconus | YES | NO | NO | NO | NO | NO | NO | NO | Substantial Equivalence Summary Table As the comparison table shows there are virtually no differences between the products offered by Paragon Vision Sciences and the Boston® II product, except material and the indication for keratoconus. {4}------------------------------------------------ Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes coiled around it, and the words "DEPARTMENT OF HEALTH & HUMAN SERVICES. USA" arranged in a circular fashion around the symbol. The logo is black and white. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 FEB 1 0 2006 Paragon Vision Sciences c/o William E. Meyers, PhD Vice President, Science and Technology 947 East Impala Ave. Mesa, AZ 85204-6619 # Re: K052983 Trade/Device Name: FluoroPerm® 92 (paflufocon A) Rigid Gas Permeable Contact Lenses FluoroPerm® 60 (paflufocon B) Rigid Gas Permeable Contact Lenses Paragon HDS® (paflufocon B) Rigid Gas Permeable Contact Lenses FluoroPerm® 30 (paflufocon C) Rigid Gas Permeable Contact Lenses Paragon Thin™ (paflufocon C) Rigid Gas Permeable Contact I enses FluoroPerm® 151 (paflufocon D) Rigid Gas Permeable Contact ' enses Paragon HDS® 100 (paflufocon D) Rigid Gas Permeable Contact Lenses Regulation Number: 21 CFR 886.5916 Regulation Name: Rigid Gas Permeable Contact Lens Regulatory Class: Class II Product Code: HOD Dated: January 13, 2006 Received: January 17, 2006 Dear Dr. Meyers: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {5}------------------------------------------------ # Page 2 - William Meyers, PhD Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 827-8910. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours. M B Eycholtz, mD Malvina B. Eydelman, M.D. Acting Division Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {6}------------------------------------------------ # Indications For Use 510(k) Number (if known): K052983 Device Name: FluoroPerm® 92 (paflufocon A) rigid gas permeable contact lenses FluoroPerm® 60 (paflufocon B) rigid gas permeable contact lenses Paragon HDS® (paflufocon B) rigid gas permeable contact lenses FluoroPerm® 30 (paflufocon C) rigid gas permeable contact lenses Paragon Thin" (paflufocon C) rigid gas permeable contact lenses FluoroPerm® 151 (paflufocon D) rigid gas permeable contact lenses Paragon HDS® 100 (paflufocon D) rigid gas permeable contact lenses Indications For Use: FluoroPerm® 92 FluoroPerm® 92 rigid gas permeable spheric contact lenses are indicated for daily wear and extended wear from 1 to 7 days between removals for cleaning and disinfection as recommended by the eye care practitioner. FluoroPerm® 92 rigid gas permicable bifocal, or toric contact lenses are indicated for daily wear only. FluoroPerm® 92 rigid gas permeable spherical, aspheric and bifocal contact lenses are indicated for the correction of visual acuity in not-aphakic persons with nondiseased eyes that are nearsighted (myopic), farsighted (hyperopic), and may exhibit corneal astigmatism up to 4.00 diopters that does not interfere with visual acuity. FluoroPerm® 92 contact lenses are indicated for persons requiring keratoconus management with otherwise non-diseased eyes, daily wear application only. FluoroPerm® 92 toric contact lenses are indicated to correct astigmatism of up to 6.00 diopters. FluoroPerm® 92 bifocal lenses are indicated to treat presbyopia up to +4.00 D ad power. ## FluoroPerm® 60 FluoroPerm® 60 rigid gas permeable spherical or aspheric contact lenses are indicated for daily wear and extended wear from 1 to 7 days between removals for cleaning and disinfection as recommended by the eye care practitioner. FluoroPerm® 60 rigid gas permeable bifocal, or toric contact lenses are indicated for daily wear only. FluoroPerm® 60 rigid gas permeable spherical, aspheric and bifocal contact lenses are indicated for the correction of visual acuity in not-aphakic persons with nondiseased eyes that are nearsighted (myopic), farsighted (hyperopic), and may exhibit corneal astigmatism up to 4.00 diopters that does not interfere with visual acuity. FluoroPerm® 60 contact lenses are indicated for persons requiring. keratoconus management with otherwise non-diseased eyes, daily wear application only. FluoroPerm® 60 toric contact lenses are indicated to correct astigmatism of up to 6.00 diopters. FluoroPerm® 60 bifocal lenses are indicated to treat presbyopia up to +4.00 D add power. J. Savita vision of Ophthalmic Ea Nose and Throat Devises 510(k) Number Page 1 of 3 {7}------------------------------------------------ #### Paragon HDS® Paragon HDS® rigid gas permeable spheric contact lenses are indicated for daily wear and extended wear from 1 to 7 days between removals for cleaning and disinfection as recommended by the eye care practitioner. Paragon HDS® rigid gas permeable bifocal, or toric contact lenses are indicated for daily wear only. Paragon HDS® rigid gas permeable spheric and bifocal contact lenses are indicated for the correction of visual acuity in not-aphakic persons with nondiseased eyes that are nearsighted (myopic), farsighted (hyperopic), and may exhibit corneal astigmatism up to 4.00 diopters that does not interfere with visual acuity. Paragon HDS® contact lenses are indicated for persons requiring keratoconus management with otherwise non-diseased eyes, daily wear application only. Paragon HDS® toric contact lenses are indicated to correct astigmatism of up to 6.00 diopters. Paragon HDS® bifocal lenses are indicated to treat presbyopia up to +4.00 D add power. #### FluoroPerm® 30 FluoroPerm® 30 rigid gas permeable contact lenses are indicated for daily wear as recommended by the eye care practitioner. FluoroPerm® 30 rigid gas permeable spheric and bifocal contact lenses are indicated for the correction of visual acuity in not-aphakic persons with nondiseased eyes that are nearsighted (myopic), farsighted (hyperopic), and may exhibit corneal astigmatism up to 4.00 diopters that does not interfere with visual acuity. FluoroPerm® 30 contact lenses are indicated for persons requiring keratoconus management with otherwise non-diseased eyes. FluoroPerm® 30 toric contact lenses are indicated to correct astigmatism of up to 6.00 diopters. FluoroPerm® 30 bifocal lenses are indicated to treat presbyopia up to +4.00 D add power. Paragon Thin™ Paragon Thin" rigid gas permeable contact lenses are indicated for daily wear as recommended by the eye care practitioner. Paragon Thin" rigid gas permeable spheric and bifocal contact lenses are indicated for the correction of visual acuity in not-aphakic persons with nondiseased eyes that are nearsighted (myopic), farsighted (hyperopic), and may exhibit corneal astigmatism up to 4.00 diopters that does not interfere with visual acuity. Paragon Thin" contact lenses are indicated for persons requiring keratoconus management with otherwise non-diseased eyes. Paragon Thin" toric contact lenses are indicated to correct astigmatism of up to 6.00 diopters. Paragon Thin"" bifocal lenses are indicated to treat presbyopia up to +4.00 D add power. FluoroPerm® 151 FluoroPerm® 151 rigid gas permeable spheric contact lenses are indicated for daily wear and extended wear from 1 to 7 days between removals for cleaning and disinfection as recommended by the eye care practitioner. FluoroPerm® 151 rigid gas permeable bifocal, or toric contact lenses are indicated for daily wear only. Savirata 510(k) Number g Page 2 of 3 {8}------------------------------------------------ # FluoroPerm® 151 (continued) FluoroPerm® 151 rigid gas permeable spheric and bifocal contact lenses are indicated for the correction of visual acuity in not-aphakic persons with nondiseased eyes that are nearsighted (myopic), farsighted (hyperopic), and may exhibit corneal astigmatism up to 4.00 diopters that does not interfere with visual acuity. FluoroPerm® 151 contact lenses are indicated for persons requiring keratoconus management with otherwise non-diseased eyes, daily wear application only. FluoroPerm® 151 toric contact lenses are indicated to correct astigmatism of up to 6.00 diopters. FluoroPerm® 151 bifocal lenses are indicated to treat presbyopia up to +4.00 D add power. # Paragon HDS® 100 Paragon HDS® 100 rigid gas permeable spheric contact lenses are indicated for daily wear and extended wear from 1 to 7 days between removals for cleaning and disinfection as recommended by the eye care practitioner. Paragon HDS® 100 rigid gas permeable bifocal, or toric contact lenses are indicated for daily wear only. Paragon HDS® 100 rigid gas permeable spherical, aspheric and bifocal contact lenses are indicated for the correction of visual acuity in not-aphakic persons with nondiscased eyes that are nearsighted (myopic), farsighted (hyperopic), and may exhibit corneal astigmatism up to 4.00 diopters that does not interfere with visual acuity. Paragon HDS® 100 contact lenses are indicated for persons requiring keratoconus management with otherwise non-diseased eyes, daily wear application only. Paragon HDS® 100 toric contact lenses are indicated to correct astigmatism of up to 6.00 diopters. Paragon HDS® 100 bifocal lenses are indicated to treat presbyopia up to +4.00 D add power. Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR Over – the – Counter Use (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IS NEEDED) | Concurrence of CDRH, Office of Device Evaluation (ODF) | | |--------------------------------------------------------|--| |--------------------------------------------------------|--| Myna Smith (Division Sign-Off) Division of Ophthalmic Ear, Nose and Throat Devices Page 3 of 3 510(k) Number k052453
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...