KEELER APPLANATION TONOMETER; T-TYPE KAT, K-TYPE KAT, (KAT)

K093445 · Keeler, Ltd. · HKY · Jan 19, 2010 · Ophthalmic

Device Facts

Record IDK093445
Device NameKEELER APPLANATION TONOMETER; T-TYPE KAT, K-TYPE KAT, (KAT)
ApplicantKeeler, Ltd.
Product CodeHKY · Ophthalmic
Decision DateJan 19, 2010
DecisionSESE
Submission TypeAbbreviated
Regulation21 CFR 886.1930
Device ClassClass 2

Intended Use

The Keeler Applanation Tonometer is indicated for measuring intraocular pressure to aid in the screening and diagnosis of Glaucoma.

Device Story

The Keeler Applanation Tonometer is a non-active, purely mechanical device used to measure intraocular pressure (IOP) as part of glaucoma screening and diagnosis. It operates on the Goldmann applanation principle, which involves applying a small flat disc to the cornea. The device is designed to be mounted on a slit lamp and is operated by an ophthalmic professional. The clinician uses a manual dial to adjust the pressure, with the measurement read directly from the device's drum. The output provides an IOP value in mmHg, which the clinician uses to assess the patient's risk for glaucoma. The device requires no electrical components and relies on manual mechanical interaction.

Clinical Evidence

Evidence includes bench testing per BSENISO 8612:2001 and a comparative field trial. Bench testing verified the device against nominal requirements of the Tonometer standard and demonstrated performance equivalent to predicate devices. A field trial conducted by ophthalmic professionals compared the device to a predicate, finding the results to be comparable. No clinical study data (e.g., patient outcomes) was required for this mechanical device.

Technological Characteristics

Manual contact tonometer; slit-lamp mounted; mechanical operation (no electrical components). Prism material: Acrylic (PMMA). Measurement range: 0-80 mmHg. Measurement method: Goldmann applanation; direct reading of IOP in mmHg via manual dial (each division equals 0.2gmf/1.96mN).

Indications for Use

Indicated for measuring intraocular pressure to aid in the screening and diagnosis of glaucoma in patients requiring such assessment.

Regulatory Classification

Identification

A tonometer and accessories is a manual device intended to measure intraocular pressure by applying a known force on the globe of the eye and measuring the amount of indentation produced (Schiotz type) or to measure intraocular tension by applanation (applying a small flat disk to the cornea). Accessories for the device may include a tonometer calibrator or a tonograph recording system. The device is intended for use in the diagnosis of glaucoma.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K093445 ## Abbreviated 510(k) Summary #### Manufacturer's information Company Name: Keeler Ltd . Company Address: Clewer Hill Road Windsor SL4 4AA UK Contact Person: Neil Atkins Telephone: +44 (0)1753 857177 +44 (0)1753 827145 · Fax: ## Submitter's information The submitter of this pre-market notification is: Mr. Eugene VanArsdale (Marketing Manager). Keeler Instruments Inc 456. Parkway, Broomall, PA 19008 Company Phone No: (610) 353 4350 Company Fax No: (610) 353 7814 # Date summary prepared: January 04, 2010 #### Proposed Device Identification | Device Trade Name: | Keeler Applanation Tonometer | |-----------------------|------------------------------| | Common Name: | Goldmann Tonometer | | Class: | 2 | | Classification Panel: | 86 | | Product Code: | HKY | | Regulation Number: | 886.1930 | #### Indications for Use The Keeler Applanation Tonometer is indicated for measuring intraccular pressure to aid in the screening and diagnosis of Glaucoma. #### Device Description and intended use The Keeler Applanation Tonometer is a screening device used to measure intraocular pressure which is one of the factors considered in diagnosing glaucoma. The product is a non-active medical device and is completely mechanical, therefore without electrical components, JAN 1 9 2010 {1}------------------------------------------------ The operation principal is based on Goldmann applanation method. The product is installed on a special support on the slit lamp and is used in conjunction with commercially available slit lamps. #### Predicated Devices Keeler Applanation Tonometer is judged to be substantially equivalent in safety, effectiveness and intended use to the following legally marketed devices: Substantial Equivalence 1: | Predicate Device: | F900 and A900, CSO | |-------------------|--------------------| | 510(k) No: | K022054 | | Manufacturer: | CSO S.R.L | | Substantial Equivalence 2: | | |----------------------------|--------------------| | Predicate Device: | Haag Streit AT 900 | | 510(k) No: | K981432 | | Manufacturer: | Haag-Streit | ### Substantial Equivalence Comparison Table below shows the comparison of Keeler Applanation Tonometer to predicate devices. The comparison of intended use and technological features of KAT indicate that the device is substantially equivalent to legally marketed predicate devices. | Comparison of Goldmann Tonometer Technological Characteristics<br>Keeler Applanation Tonometer vs. CSO Tonometer | | | |------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------| | Criteria | Predicate Device<br>F900 and A900,<br>CSO Tonometers<br>(Re:510(k)<br>K022054) | Keeler Applanation Tonometer (KAT) | | Type | Manual contact<br>Tonometer | Manual contact Tonometer | | Indication | Intraocular Pressure<br>(IOP) measurement | Intraocular Pressure (IOP)<br>measurement | | Design | Slit lamp mounted<br>manual dial | Slit lamp mounted manual dial | | Prism Material | Acrylic (PMMA) | Acrylic (PMMA) | | Measurement Range | 0-80mmHg | 0-80mmHg | | Measurement<br>technique | Applanation | Applanation | | Measurement Method | Direct reading of<br>IOP in mmHg (each<br>division on dial of<br>measuring drum is<br>equal to 0.2gmf<br>(1.96mN)) | Direct reading of IOP in mmHg (each<br>division on dial of measuring drum is<br>equal to 0.2gmf (1.96mN)) | | Calibration | Maintenance and | Maintenance and calibration required | {2}------------------------------------------------ | | calibration required | | |--------------|-----------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------| | Intended use | Instrument is<br>intended to measure<br>pressure by<br>applanation<br>(applying a small flat<br>disc to cornea) | Instrument is intended to measure<br>pressure by applanation (applying a<br>small flat disc to cornea) | | Comparison of Goldmann Tonometer Technological Characteristics<br>Keeler Applanation Tonometer vs. Haag-Streit AT 900 | | | |-----------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------| | Criteria | Predicate Device<br>Haag-Streit AT 900<br>(Re:510(k)<br>K0981432) | Keeler Applanation Tonometer (KAT) | | Type | Manual contact<br>Tonometer | Manual contact Tonometer | | Indication | Intraocular Pressure<br>(IOP) measurement | Intraocular Pressure (IOP)<br>measurement | | Design | Slit lamp mounted<br>manual dial | Slit lamp mounted manual dial | | Measurement Range | 0-80mmHg | 0-80mmHg | | Measurement<br>technique | Applanation | Applanation | | Measurement Method | Direct reading of<br>IOP in mmHg (each<br>division on dial of<br>measuring drum is<br>equal to 0.2gmf<br>(1.96mN)) | Direct reading of IOP in mmHg (each<br>division on dial of measuring drum is<br>equal to 0.2gmf (1.96mN)) | | Calibration | Maintenance and<br>Calibration required | Maintenance and calibration required | | Intended use | Instrument is<br>intended to measure<br>pressure by<br>applanation<br>(applying a small flat<br>disc to cornea) | Instrument is intended to measure<br>pressure by applanation (applying a<br>small flat disc to cornea) | Note: There is no difference between Predicate devices and Keeler Applanation Tonometer. . {3}------------------------------------------------ #### Summary of Performance Testing The performance of Keeler Applanation Tonometer has been evaluated in bench testing, comparative study with its predicative devices and field trial. Bench testing was carried out using arrangement as per Tonometer Standard BSENISO 8612: 2001. Keeler Applanation Tonometer units under test are verified against the nominal requirements of Tonometer standard. Bench testing arrangement was also used to carry out comparative study, i.e. comparing Keeler Applanation Tonometer with its predicate devices. Keeler Applanation Tonometer and its predicate devices were subject to the same nominal requirements as set by the Tonometer standard and the results demonstrate that they are substantially equivalent. Field test was conducted by ophthalmic professional to compare Keeler Applanation Tonometer with its predicate device and the findings are comparable. Based on the results of the performance testing, the Keeler Applanation Tonometer is iudged to be as safe, as effective and performs as well as the predicate devices. {4}------------------------------------------------ Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus, a symbol often associated with medicine and healthcare. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the caduceus. Food and Drug Administration 10903 New Hampshire Avenue Document Mail Center - WO66-G609 Silver Spring, MD 20993-0002 Keeler Instruments, Inc. c/o Mr. Eugene R. VanArsdale Marketing Director 456 Parkway Broomall, PA 19008 JAN 1 9 2010 Re: K093445 Trade/Device Name: Keeler Applanation Tonometer, T-Type KAT, K-Type KAT; (KAT) Regulation Number: 21 CFR 886.1930 Regulation Name: Tonometer and Accessories Regulatory Class: II Product Code: HKY Dated: October 19, 2009 Received: October 21, 2009 Dear Mr. VanArsdale: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {5}------------------------------------------------ If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/cdrl/mdr/ for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours, Kesia Alexander for Malvina B. Eydelman, M.D. Director Division of Ophthalmic, Neurological, and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {6}------------------------------------------------ # Indications for Use 510(k) Number (if known): Device Name: Keeler Applanation Tonometer (KAT) Indications for Use: The Keeler Applanation Tonometer is indicated for measuring intraocular pressure to aid in the screening and diagnosis of Glaucoma. Prescription Use V (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Danl Kauk (Division Sign-Off) Division of Ophthalmic, Neurological and Ear, Nose and Throat Devices 510(k) Number K093445
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...