The Applanation Tonometer AT-1 is indicated to measure intraocular pressure for use as an aid in glaucoma screening and diagnosis.
Device Story
Applanation Tonometer AT-1 is a manual, slit-lamp-mounted device used by clinicians in medical offices, hospitals, and optometry clinics to measure intraocular pressure (IOP). It utilizes the Goldmann applanation method, where a clinician manually adjusts a control knob to apply force via a measuring prism (made of PMMA) to flatten a constant area (3.06mm diameter) of the patient's cornea. The device provides a direct reading of IOP in mmHg via a mechanical dial. It is a non-electrical, mechanical instrument available in Arm and Base models for integration with slit-lamp microscopes. The device aids in glaucoma screening and diagnosis by providing quantitative pressure data. Calibration is verified using a check weight. The measuring prism is reusable and requires manual cleaning and disinfection.
Clinical Evidence
Bench testing only. Performance was validated in accordance with ISO 8612:2009 (Ophthalmic Instruments - Tonometers). Biocompatibility of patient-contacting materials was evaluated per ISO 10993-1:2009. Cleaning and disinfection methods for the reusable measuring prism were validated.
Technological Characteristics
Manual contact tonometer; Goldmann applanation principle. Patient-contacting material: PMMA (acrylic resin). Other components: stainless steel, copper alloy, aluminum alloy. Dimensions: Arm model (71.2x135.0x249mm), Base model (71.2x80x184mm). Measurement range: 0-80mmHg. Mechanical operation; no electrical elements. Calibration via check weight. Complies with ISO 8612:2009.
Indications for Use
Indicated for patients requiring intraocular pressure measurement as an aid in glaucoma screening and diagnosis.
Regulatory Classification
Identification
A tonometer and accessories is a manual device intended to measure intraocular pressure by applying a known force on the globe of the eye and measuring the amount of indentation produced (Schiotz type) or to measure intraocular tension by applanation (applying a small flat disk to the cornea). Accessories for the device may include a tonometer calibrator or a tonograph recording system. The device is intended for use in the diagnosis of glaucoma.
{0}------------------------------------------------
Image /page/0/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of a caduceus, which is a symbol often associated with healthcare and medicine. The caduceus is depicted with a staff entwined by two snakes and topped with wings.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
October 17, 2017
Takagi Seiko Co., Ltd. Shinya Kitamura Official Correspondent Legal & Regulatory Affairs 330-2 Iwafune Nakano-shi, Nagano-ken 383-8585 JP
Re: K172728 Trade/Device Name: Applanation tonometer AT-1 Regulation Number: 21 CFR 886.1930 Regulation Name: Tonometer and Accessories Regulatory Class: Class II Product Code: HKY Dated: September 5, 2017 Received: September 11, 2017
Dear Shinya Kitamura:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Denise L. Hampton -S
for Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K 172728
Device Name Applanation Tonometer AT-1
Indications for Use (Describe)
The Applanation Tonometer AT-1 is indicated to measure intraocular pressure for use as an aid in glaucoma screening and diagnosis.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|---------------------------------------------|
| <span></span> | <span></span> |
| Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
B-TQMS4101Letterhead Rev.1
Image /page/3/Picture/1 description: The image shows the logo for TAKAGI SEIKO CO., LTD. The logo consists of a blue abstract symbol on the left, followed by the company name in bold, black letters. A black line is located underneath the company name. The logo appears to be for a manufacturing or engineering company.
330-2 IWAFUNE, NAKANO-SHI, NAGANO-KEN, 383-8585, JAPAN +81-269-22-4512 FAX. +81-269-26-6321 URL: http://www.takagi-j.com E-mail: info@takagi-j.com
# K172728
# 510(k) Summary
## 1. SUBMITTER
| Company Name: | Takagi Seiko Co., Ltd. |
|--------------------------|------------------------------------------------------|
| Company Address: | 330-2 Iwafune, Nakano-Shi, Nagano-Ken 383-8585 JAPAN |
| Telephone No: | +81-269-22-4511 |
| Fax No: | +81-269-26-6321 |
| Contact Person: | SHINYA KITAMURA |
| Title of Contact Person: | None |
| Telephone No: | +81-269-22-4511 |
| Fax No: | +81-269-26-6321 |
| Email Address: | kitamura.shinya@takagi-j.com |
| Date Prepared: | October 12, 2017 |
# 2. DEVICE
| Name of Device: | Applanation Tonometer AT-1 |
|----------------------|---------------------------------------------|
| Models: | Arm Model, Base Model |
| Common Name: | Goldmann Tonometer |
| Classification Name: | Tonometer and Accessories (21 CFR 886.1930) |
| Regulatory Class: | II |
| Product Code: | HKY |
# 3. PREDICATE DEVICE
| Manufacturer: | HAAG - STREIT |
|--------------------|------------------------------|
| Device trade name: | APPLANATION TONOMETER AT 900 |
| Regulatory Class: | II |
| Product Code: | HKY |
| 510(k) number: | K981432 |
This predicate has not been subject to a design-related recall.
- (1) No records were found with Product: applanation tonometer AT 900 Date Classified from: 1980 Date Classified to: 2015
(https/www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfres/res.cfm)
{4}------------------------------------------------
## 4. DEVICE DESCRIPTION
### 4.1 Device Identification
The device consists of a pressurizing force control unit, tonometer tip (hereinafter referred to as measuring prism), section connected to a slit lamp microscope, and accessories. The measuring prism requires disinfection. The device is available in two models according to the difference in the method of connecting to a slit lamp microscope, while the pressurizing force control unit and measuring prism are the same for the two models. Each model includes no electrical elements. The device is set up to a dedicated support on the slit lamp, and used together with a commercially available slit lamp microscope. The pressurizing force control unit of the device incorporates a measurement dial that indicates pressurizing force, including a pressurizing force control knob, and a prism holder to mount the measuring prism. Also, the pressurizing force control unit includes a connector for mounting a check weight holder to check the pressurizing force of the device. The product is a manual medical device.
### 4.2 Related accessories are as follows:
Check weight / check weight holder used to check the pressurizing force of the device
### 4.3 Models are as follows:
### 1) Arm Model
The arm model can be left on a slit lamp permanently as it is fixed to the mounting base of the microscope of the slit lamp, and it is turned in front of the microscope for examination. The applanation surface is observed monocularly through the left eyepiece lens.
### 2) Base Model
The base model is set on the slit lamp shaft by the guide plate. and it can be moved between two positions on the guide plate for observation through the right or left eyepiece lens.
### 4.4 Device Characteristics
- 1) Patients-contacting material: PMMA.
- 2) The measuring prism of the device is reusable.
- 3) Cleaning and disinfection method
The cleaning and disinfection method of the measuring prism of the device is as follows:
- a) Manual washing.
- b) Immersion in disinfectant.
- c) Manual removal of disinfectant with running.
### 4.5 Environment of Use
4.5.1 Areas of application of the device
The Applanation Tonometer AT-1 is intended for use in medical offices, hospitals, and in optometrist and optician offices and clinics.
{5}------------------------------------------------
4.5.2 Ambient conditions Temperature: +10℃ to +35℃ Air Pressure: 800hPa to 1060hPa Relative humidity: 30% to 75%
## 4.6 Method of Operation
In applanation tonometry the intraocular pressure is calculated from the force required to flatten a constant area of the cornea.
The operation principal based on Goldmann applanation method.
## 4.7 Materials of Use
General type of material used
The Applanation Tonometer AT-1 comes into contact with patients' corneal mucosal membrane surface, the material is acrylic resin and the contact time is limited exposure, respectively. Other component parts of the device are made of general materials (example: stainless Steel, copper alloy, and aluminum alloy).
## 4.8 Calibration
Pressure applied to the pressurizing arm is periodically checked by use of the check weight during factory production/by users.
## 5. INDICATIONS FOR USE
The Applanation Tonometer AT-1 is indicated to measure intraocular pressure for use as an aid in glaucoma screening and diagnosis.
## 6. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE
## 6.1 Comparison of AT900 and AT-1
Table 1. Applanation Tonometer AT-1 (Arm Model & Base Model) vs. HAAG-STREIT Applanation Tonometer AT900
| | Comparison of Goldmann Tonometer Technical Characteristics | |
|--------------------|------------------------------------------------------------|-------------------------------------------|
| | Applanation Tonometer AT-1 vs. Haag-Streit AT900 | |
| Criteria | Predicate Device<br>Haag-Streit AT900 | New Device<br>Applanation Tonometer AT-1 |
| 510(k) Number | K981432 | K172728 |
| Type | Manual contact Tonometer | Manual contact Tonometer |
| Indication | Intraocular Pressure<br>(IOP) measurement | Intraocular Pressure<br>(IOP) measurement |
| Design | Slit lamp mounted<br>Manual dial | Slit lamp mounted<br>Manual dial |
| Prism Material | PMMA | PMMA |
| Measurement Range | 0 – 80mmHg | 0 – 80mmHg |
| Measurement Method | Applanation | Applanation |
{6}------------------------------------------------
| Measurement<br>technique | Direct reading of IOP in<br>mmHg (each division on<br>dial of measuring drum is<br>equal to 0.2gf (1.96mN)) | Direct reading of IOP in<br>mmHg (each division on dial<br>of measuring drum is equal to<br>0.2gf (1.96mN)) |
|--------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------|
| Applanation circle<br>diameter | 3.06mm | 3.06mm |
| Calibration | Maintenance and calibration<br>required | Maintenance and calibration<br>required |
| Intended use | The goldmann manual<br>tonometer is an appliance that<br>serves to measure intraocular<br>pressure, according to the<br>goldmann method. The<br>measuring of the pressure<br>requires to maintain a uniform<br>applanation of the surface of the<br>cornea. It is specially indicated<br>in glaucoma disease. | The Applanation Tonometer<br>AT-1 is indicated to measure<br>intraocular pressure for use as<br>an aid in glaucoma screening<br>and diagnosis. |
### 6.2 Technical Specifications
Table 2. Applanation Tonometer AT-1 (Arm Model & Base Model) vs. HAAG-STREIT Applanation Tonometer AT900
| Models<br>Item | Predicate Device<br>Haag-Streit AT900 | New Device<br>Applanation Tonometer AT-1 | |
|-----------------------------|---------------------------------------------------------------|------------------------------------------|--------------------------|
| | | Arm Model | Base Model |
| Measurement range | 0-80mmHg | | |
| One scale unit | 1.96mN | | |
| Measurement accuracy | ±1.5 % of the nominal value or ±0.49 mN, whichever is greater | | |
| Repeatability | ≤0.50mN | | |
| Reverse span | 0.49mN | | |
| Applanation circle diameter | 3.06mm ± 0.02mm | | |
| Applanation area | 7.354 mm² | | |
| Dimensions (millimeter) | 89(W)×115(D)<br>× 250(H) | 71.2 (W)x135.0(D)<br>x249(H) | 71.2(W)x80(D)<br>x184(H) |
| Weight | 0.5kg | 0.44kg | 0.34kg |
## 7. PERFORMANCE AND SAFETY
### 7.1 Performance Testing
Performance of the Applanation Tonometer AT-1 was tested in accordance with ISO 8612:2009-Ophtalmic Instrument-Tonometers, Annex A.2.
### 7.2 Validation of Cleaning and disinfection Methods
Regarding the reusable measuring prism, validation of the cleaning and disinfection method is carried out by the validation test, and it is confirmed that it passed.
### 7.3 Life
Shelf life and service life of Applanation Tonometer AT-1 and the measuring prism are described.
{7}------------------------------------------------
### 7.4 Material Biocompatibility
Concerning the safety of Applanation Tonometer AT-1 material which comes in direct contact with a patient's eyes, comparative evaluations were performed with the Predicate Device by applying ISO 10993-1:2009-Biological evaluation of medical device-Part 1: Evaluation and testing within a risk management process.
#### 7.5 Labeling
For usage methods and cleaning and disinfection methods of the Applanation Tonometer AT-1, an instruction manual is provided.
## 8. Conclusion
According to each evaluation result as mentioned above, the Applanation Tonometer AT-1 is substantially equivalent to the predicate device in safety, performance, specifications and etc.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.