K101182 · Carl Zeiss Meditec, AG · HJO · Nov 3, 2010 · Ophthalmic
Device Facts
Record ID
K101182
Device Name
IOLMASTER 500
Applicant
Carl Zeiss Meditec, AG
Product Code
HJO · Ophthalmic
Decision Date
Nov 3, 2010
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 886.1850
Device Class
Class 2
Attributes
Real-World Evidence
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K101182 · Nov 3, 2010
IOLMASTER 500
Carl Zeiss Meditec, AG
Published clinical literature; Retrospective clinical cohort of patients presenting for cataract evaluation
Clinical data from literature was reviewed to determine agreement between the IOLMaster and other clinical instruments (Grieshaber Biometric System, handheld keratometer, and Lenstar LS900) for axial length, anterior chamber depth, corneal radius, and white-to-white measurements.
Grieshaber Biometric System (GBS), handheld keratometer (Renaissance, Alcon), and Lenstar LS900
Agreement of axial length, anterior chamber depth, corneal radius, and white-to-white measurements
Indications for Use
The IOLMaster is intended for the biometric determination of ocular measurements for axial length, anterior chamber depth, corneal radius, white-to-white (WTW), and for the measurement of pupil size and deviation of the visual axis from the center of the pupil. For patients who are candidates for intraocular lens (IOL) implantation, the device also performs calculations to assist physicians in determining the appropriate IOL power and type for implantation. This device is intended for use by physicians and eye-care professionals and may only be used under the supervision of a physician.
Device Story
IOLMaster 500 is a non-contact ophthalmic biometry instrument used by physicians and eye-care professionals in clinical settings. It utilizes partial coherence interferometry and traditional ophthalmic biometry to measure axial length, corneal radius, anterior chamber depth, white-to-white distance, pupil size, and visual axis deviation. The device processes these inputs via integrated software algorithms to perform preoperative calculations for IOL power and type selection for cataract surgery. Output is displayed via a graphical user interface (GUI) to assist clinicians in surgical planning. The device features updated computer hardware with USB connectivity for data management. Benefits include precise ocular measurements to optimize IOL selection, potentially improving patient outcomes post-cataract surgery.
Clinical Evidence
Clinical evidence includes literature review and bench testing. Literature compared IOLMaster measurements (axial length, ACD, corneal radius) against Grieshaber Biometric System (N=146-154) and WTW against Lenstar LS 900 (N=112). Mean differences were minimal (e.g., -0.01 mm for axial length, +0.06 mm for WTW). A prospective repeatability/reproducibility study (N=30) assessed inter-device and inter-operator variability; results showed high precision with low standard deviations (e.g., axial length repeatability SD 0.0206 mm).
Technological Characteristics
Non-contact optical biometry instrument using partial coherence interferometry. Measures axial length, corneal radius, anterior chamber depth, and white-to-white distance. Features Windows XP-based software, improved GUI, and USB ports for data export. Operates as an AC-powered slit lamp biomicroscope. Electrical and laser safety verified per applicable standards.
Indications for Use
Indicated for patients who are candidates for intraocular lens (IOL) implantation requiring preoperative ocular biometric measurements (axial length, anterior chamber depth, corneal radius, white-to-white distance, pupil size, and visual axis deviation) and IOL power/type calculations.
Regulatory Classification
Identification
An AC-powered slitlamp biomicroscope is an AC-powered device that is a microscope intended for use in eye examination that projects into a patient's eye through a control diaphragm a thin, intense beam of light.
Special Controls
*Classification.* Class II (special controls). The device, when it is intended only for the visual examination of the anterior segment of the eye, is classified as Group 1 per FDA-recognized consensus standard ANSI Z80.36, does not provide any quantitative output, and is not intended for screening or automated diagnostic indications, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 886.9.
{0}------------------------------------------------
K101182
# 4.0 510(K) SUMMARY
NOV - 3 2010 This 510(k) summary of safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR 807.92(a).
- Submitter's Name, address, telephone number, contact person, and date summary 4.1 PREPARED
| a. | Applicant: | Carl Zeiss Meditec AG<br>Goeschwitzer Strasse 51-52<br>07745 Jena<br>Germany |
|----|-----------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| b. | Contact Person: | Judith A. Brimacombe, MA<br>Director, Clinical/Regulatory Affairs<br>Carl Zeiss Meditec, Inc.<br>5160 Hacienda Drive<br>Dublin, CA 94568<br>USA<br>j.brimacombe@meditec.zeiss.com<br>Tel: (925) 557-4616<br>Fax: (925) 557-4259 |
- October 7, 2010 Summary Prepared: C.
### Name of device, including trade name and classification name 4.2
| a. | Trade/Proprietary Name: | IOLMaster® 500 |
|----|-------------------------|------------------------------------|
| b. | Common/Usual Name: | IOLMaster® |
| c. | Classification Name: | AC-powered slit lamp biomicroscope |
| d. | Product Code and Class: | HJO - Class II |
| e. | Classification Number: | 886.1850 |
{1}------------------------------------------------
### PREDICATE DEVICE 4.3
The IOLMaster 500, the subject of this 510(k), is substantially equivalent to the predicate IOLMaster (Carl Zeiss Meditec, AG) cleared for marketing under 510(k) number K993357 and the Lenstar Model LS 900 (Haag-Streit AG) cleared for marketing under 510(k) number K082891.
### DEVICE DESCRIPTION 4.4
The IOLMaster 500 is a non-contact biometry instrument for measurements of the eye required for preoperative computation of intraocular lens (IOL) type and power. As with the IOLMaster predicate device, the IOLMaster 500 provides measurements of axial length, corneal radius (keratometry), and anterior chamber depth. In the new model of the IOLMaster, the optical measurement technology has remained the same and the measurements of axial length, corneal radius and anterior chamber depth are achieved in the same fashion. In addition to these three parameters, using the same optical techniques with advances in software, the modified IOLMaster also has the capability to measure a fourth parameter, i.e., the "white-to-white" distance (WTW). Incorporation of this additional information extends the physician's choice of computational formulas.
### 4.5 STATEMENT OF INTENDED USE
The IOLMaster is intended for the biometric determination of ocular measurements for axial length, anterior chamber depth, corneal radius, white-to-white (WTW), and for the measurement of pupil size and deviation of the visual axis from the center of the pupil. For patients who are candidates for intraocular lens (IOL) implantation, the device also performs calculations to assist physicians in determining the appropriate IOL power and type for implantation.
This device is intended for use by physicians and eye-care professionals and may only be used under the supervision of a physician.
### Comparison of Technological Characteristics 4.6
The IOLMaster 500 has the same indications for use and operating characteristics as the predicate IOLMaster with the exception that the white-to-white measurement has been added to the indications for use statement.
The IOLMaster utilizes partial coherence interferometry and traditional ophthalmic biometry techniques to obtain measurements for axial length, corneal radius, anterior chamber depth and white-to white distance. Integrated algorithms in the software allow for the use of clinically recognized formulas for the calculation of IOL type and power prior to
{2}------------------------------------------------
cataract surgery. These functions are the same for the IOLMaster 500 and the predicate device.
Software enhancements using the Windows XP platform have improved the Graphic User Interface (GUI) and the usability of the IOLMaster. Additionally, modifications have been made to the IOLMaster computer hardware to incorporate USB ports for importing and exporting data to and from the IOLMaster 500.
Since the measurements are achieved in the same fashion using the identical optical technology, the IOLMaster 500 is therefore substantially equivalent to the predicate IOLMaster (K993357).
### 4.7 CLINICAL EVALUATION
Clinical data reported in the literature was reviewed to determine agreement between the IOL Master, Grieshaber Biometric System, and Lenstar Model LS900. The results are provided in Table 1 on the following page.
## Axial length, anterior chamber depth and corneal radius
IOLMaster axial length (AL) and anterior chamber depth (ACD) measurements were compared to measurements obtained with a high precision immersion ultrasound system (Grieshaber Biometric System (GBS)). A total of 189 consecutive eyes presented for cataract evaluation were measured with the IOLMaster. Excluded from the study were 32 eyes not measured with GBS. Mean difference for axial length was calculated based on 146 phakic eyes comparative axial length measurements between IOLMaster and GBS (10 pseudophakic eyes and 1 aphakic eye were excluded). Mean difference for ACD was calculated based on measurements on 151 eyes comparing IOLMaster and GBS (in 6 pseudophakic eves. ACD could not be measured). Mean difference for the average corneal radius was calculated based on keratometry measurements on 154 eyes comparing IOLMaster and a handheld keratometer ((Renaissance, Alcon). Two eyes with no available keratometry results and one eye with unreliable readings due to dry eye were excluded).2
### White-to-white measurement
IOLMaster white-to-white measurements were compared to measurements obtained with a Lenstar LS 900 (Haag-Streit) in 112 cataract surgery eyes.2
1 Haigis W, Lege B: Akustische und optische Biometrie im klinischen Einsatz; in Wenzel M, Kohnen B. (eds): 14. Kongress der Deutschsprachigen Gesellschaft für Intraokulariinsen-Implantation und refractive Chirurgie, Jahresband der DGII, Luzern, February 2000. Koln, Biermann, 2000, pp 73-78. [ Haigis W, Lege B: Acoustical and Optical Biometry in Clinical Application; in Wenzel M, Kohnen T, Blumer B. (eds): 14. Congress of the german-speaking society for implantation of intraocular lenses and refractive surgery (DGII), Annual book of the DGII, Luzern, February 2000. Koln, Biermann, 2000, pp 73-78.] 4 Buckhurst PJ, Wolffson JS, Shah S, Naroo SA, Davies LN, Berrow EJ. A new optical low coherence reflectometry device for ocular biometry in cataract patients. Br J Ophthalmol 2009; 93 (7): 949-953.
{3}------------------------------------------------
### Table 1. Agreement with Other Instruments
| | N | Mean Difference | Standard Deviation |
|-------------------------|-----|-----------------|--------------------|
| Axial length* | 146 | -0.01 mm | ±0.19 mm |
| Corneal curvature** | 154 | -0.01 mm | ±0.05 mm |
| Anterior chamber depth* | 151 | + 0.03 mm | ±0.18 mm |
| White-to-white*** | 112 | + 0.06 mm | ±0.33 mm |
All referenced studies compared measurements from eyes evaluated for cataract surgery.
- In comparison to high precision immersion ultrasound instrument (Grieshaber Biometric Systems, GBS) *
- ** In comparison to manual keratometer (Renaissance, Alcon)*
- *** In comparison to optical low coherence reflectometry (Lenstar LS900, Haig-Streit)*
## Repeatability and Reproducibility
A study was conducted to determine repeatability and reproducibility of the IOLMaster 500 measurements of axial length (AL), corneal radius (corneal curvature), anterior chamber depth (ACD) and white-to-white (WTW). Phase I of the study enrolled 30 subjects and was designed to determine inter-device variability, wherein each subject was imaged 3 times during a single visit on each of three IOLMaster 500 instruments by one operator. Phase II enrolled the same 30 subjects and was designed to determine inter-operator variability, wherein each subject was imaged three times during a single visit by each of three operators. Subjects in the study were 18 years or older with healthy eyes and no ocular opacities. The precision of axial length measurements and anterior chamber depth measurements may be different in eyes with cataract or corneal abnormalities.
The repeatability and reproducibility results are shown in Table 2 on the following page.
3 Carl Zeiss Meditec, Inc., Clinical Study: Repeatability of IOLMaster 500; 2010.
{4}------------------------------------------------
| | Repeatability | | Reproducibility | |
|------------------------|-----------------------------|--------------------------------|-------------------------------|----------------------------------|
| | Repeatability<br>SD<br>(mm) | Repeatability<br>Limit<br>(mm) | Reproducibility<br>SD<br>(mm) | Reproducibility<br>Limit<br>(mm) |
| Axial length | 0.0206 | 0.0577 | 0.0222 | 0.0623 |
| Corneal curvature | 0.0162 | 0.0455 | 0.0167 | 0.0468 |
| Anterior chamber depth | 0.0347 | 0.0972 | 0.0494 | 0.1383 |
| White-to-white | 0.0558 | 0.1562 | 0.0647 | 0.1811 |
# Table 2. Repeatability and Reproducibility of the IOLMaster 500 Measurements
Repeatability is defined as the standard deviation of measurements taken by a single operator/single instrument. The repeatability standard deviation was estimated by the squared error. Reproducibility is the standard deviation of measurements that include between operator and between instrument variability. The reproducibility standard deviation was estimated by the square root of the sum of random measurement variability, inter-device variability, inter-operator variability, subject and device interaction variability, and subject and operator interaction variability.
The repeatability limit is the upper 95% limit for the differences between repeated results. Per ISO 5725-6, the repeatability limit = 2.8 x repeatability SD.
The reproducibility limit is the upper 95% limit calculated for the difference between individual measurements using different operators and different instruments. Per ISO 5725-6, the reproducibility limit = 2.8 x the reproducibility SD.
#### BRIEF SUMMARY OF PERFORMANCE TESTS AND RESULTS 4.8
The IOLMaster has been designed and tested to the applicable safety standards. The performance data supporting safety and substantial equivalence of the IOLMaster 500 to the predicate device demonstrates that:
- the axial length, corneal radius, anterior chamber depth measurements of the . modified IOLMaster are substantially equivalent to those of the predicate device;
- white-to-white measurements using the IOLMaster 500 are substantially equivalent . to those of the predicate device;
- the IOLMaster meets the requirements for electrical and laser safety; .
- usability of the IOLMaster has been verified in accordance with appropriate . standards.
#### 4.9 CONCLUSION
The IOLMaster 500 is substantially equivalent to the predicate devices.
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the seal of the U.S. Department of Health & Human Services. The seal features a stylized eagle with its wings spread, and three parallel lines extending from the eagle's body. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Carl Zeiss Meditec Incorporated c/o Judith A. Brimacombe, M.A. Director, Clinical and Regulatory Affairs 5160 Hacienda Drive Dublin, CA 94568
Re: K101182
Trade Name: IOLMaster 500 Regulation Number: 21 CFR 886.1850 Regulation Name: AC-powered slit lamp biomicroscope Regulatory Class: Class II Product Code: HJO Dated: September 13, 2010 Received: September 14, 2010
Dear Ms. Brimacombe:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing
NOV - 3 2010
{6}------------------------------------------------
Page 2 - Ms. Judith A. Brimacombe, M.A.
(21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Debra Falls
Malvina B. Eydelman, M. Director Division of Ophthalmic, Neurological, and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
510(k) Number (if known): K101182
Device Name(s): IOLMaster 500 NOV - 3 2010
Indications for Use:
The IOLMaster is intended for the biometric determination of ocular measurements for axial length, anterior chamber depth, corneal radius, white-to-white (WTW), and for the measurement of pupil size and deviation of the visual axis from the center of the pupil. For patients who are candidates for intraocular lens (IOL) implantation, the device also performs calculations to assist physicians in determining the appropriate IOL power and type for implantation.
This device is intended for use by physicians and eye-care professionals and may only be used under the supervision of a physician.
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
DemR
Page 1 1 of of 1 1
(Division Sign-Off)
Division of Ophthalmic, Neurological and Ear,
Nose and Throat Devices
510(k) Number K101182
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.