LipiView II Ocular Surface Interferometer

K152869 · Tearscience, Inc. · HKI · Jan 14, 2016 · Ophthalmic

Device Facts

Record IDK152869
Device NameLipiView II Ocular Surface Interferometer
ApplicantTearscience, Inc.
Product CodeHKI · Ophthalmic
Decision DateJan 14, 2016
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 886.1120
Device ClassClass 2

Indications for Use

The LipiView II Ocular Surface Interferometer is an ophthalmic imaging device intended for use by a physician in adult patients to capture, archive, manipulate and store digital images of: · Specular (interferometric) observations of the tear film. Using these images, LipiView® II measures the absolute thickness of the tear film lipid layer. • Meibomian glands under near-infrared (NIR) illumination · The ocular surface and eyelids under white illumination All of these image types can be photographically documented and visually monitored.

Device Story

Bench-top ophthalmic imaging device; captures digital images/video of tear film, meibomian glands, and ocular surface. Inputs: specular reflection (white light interferometry) for tear film; NIR illumination for meibomian glands; white light for ocular surface/eyelids. Operates via physician-controlled touchscreen interface; includes motion stage, camera, and handheld NIR lid everter. Outputs: visual images/video on display, printed reports, or exported files. Used in-office by physicians to monitor ocular surface health; does not provide diagnosis; no treatment/life-sustaining function. Benefits: enables visualization and documentation of ocular structures to assist clinical assessment.

Clinical Evidence

Bench testing only. Performance testing included usability testing, software validation, and verification of compliance with ISO 15004-2 (light hazard), ANSI/IEC 60601 (electrical safety/EMC), and UL 94V-1 (flammability). No clinical data required.

Technological Characteristics

Bench-top ophthalmic camera; Windows-based OS; touchscreen GUI. Materials: flame retardant per UL 94V-1. Illumination: Class I white LED, NIR LED. Connectivity: USB, wireless network. Sterilization: alcohol disinfection for patient contact parts (chin/forehead rest). Standards: ISO 15004-2, ANSI/IEC 60601-1, IEC 60601-1-2.

Indications for Use

Indicated for adult patients; physician use for capturing, archiving, manipulating, and storing digital images of tear film lipid layer thickness, meibomian glands (via NIR), and ocular surface/eyelids (via white light).

Regulatory Classification

Identification

An ophthalmic camera is an AC-powered device intended to take photographs of the eye and the surrounding area.

Special Controls

*Classification.* Class II (special controls). The device, when it is a photorefractor or a general-use ophthalmic camera, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 886.9.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles a stylized caduceus, with three human profiles forming the staff and a wing-like shape above them. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 January 14, 2016 Tearscience, Inc. % Dr. Christy Coleman Vice President, Clinical, Quality and Regulatory Affairs 5151 McCrimmon Pkwy., Suite 250 Morrisville, North Carolina 27650 Re: K152869 Trade/Device Name: Lipiview II Ocular Surface Interferometer Regulation Number: 21 CFR 886.1120 Regulation Name: Opthalmic Camera Regulatory Class: Class II Product Code: HKI, HJO Dated: December 8, 2015 Received: December 9, 2015 Dear Dr. Coleman: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices. good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {1}------------------------------------------------ If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, Kesia Y. Alexander -A for Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K152869 Device Name LipiView® II Ocular Surface Interferometer Indications for Use (Describe) The LipiView II Ocular Surface Interferometer is an ophthalmic imaging device intended for use by a physician in adult patients to capture, archive, manipulate and store digital images of: · Specular (interferometric) observations of the tear film. Using these images, LipiView® II measures the absolute thickness of the tear film lipid layer. • Meibomian glands under near-infrared (NIR) illumination · The ocular surface and eyelids under white illumination All of these image types can be photographically documented and visually monitored. | Type of Use (Select one or both, as applicable) | | |-------------------------------------------------|--| |-------------------------------------------------|--| X Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ Image /page/3/Picture/1 description: The image shows the logo for TearScience. The logo features a stylized blue eye on the left, followed by the word "TearScience" in a blue, slightly italicized font. A registered trademark symbol is present to the right of the word. # 510(k) SUMMARY | PREPARATION DATE: | September 29, 2015 | |------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | APPLICANT: | TearScience, Inc.<br>5151 McCrimmon Parkway, Suite 250<br>Morrisville, NC 27560 | | | Tel: (919) 467-4007<br>Fax: (919) 467-3300 | | CONTACT PERSON: | Christy Coleman, OD, MPH<br>Vice President, Clinical, Quality and Regulatory Affairs | | DEVICE TRADE NAME: | LipiView® II Ocular Surface Interferometer | | COMMON NAME: | Ophthalmic Imaging Device | | CLASSIFICATION NAME: | Ophthalmic Camera | | DEVICE CLASSIFICATION: | Class II; 21 CFR 886.1120 and 21 CFR 886.1850 | | PRODUCT CODE: | HKI, HJO | | PREDICATE DEVICES: | PRIMARY: LipiView® Ocular Surface Interferometer<br>Class II under 21 CFR 886.1120 and 21 CFR 886.1850;<br>Product Code HKI, HJO; Applicant: TearScience, Inc.;<br>Originally cleared under K091935 on October 23, 2009;<br>Expanded Indications for Use cleared under K122481 on<br>December 31, 2012<br><br>SECONDARY: Optomed Smartscope M5 Digital Camera<br>Class II under 21 CFR 886.1120; Product Code HKI;<br>Applicant: Optomed Oy.;<br>Cleared under K132186 on December 18. 2013 | #### DEVICE DESCRIPTION: The LipiView® II Ocular Surface Interferometer is a bench-top device used as an ophthalmic camera for imaging the lipid layer of the tear film, meibomian glands, ocular surface and eyelids. There are three imaging modes: Lipid, Gland and Ocular. LipiView II contains a computer system and electronics, two position chin and forehead rest, camera and lens, motion stage, illuminator and touch screen display. LipiView II includes pushbutton controls and handheld near-infrared lid everter for gland imaging. {4}------------------------------------------------ In the Lipid Imaging mode, LipiView II operates on the principle of white light interferometry and provides an interferometry color assessment of the tear film by specular reflection. The computer system captures a video image file that is recorded over time since the interference pattern changes as the tear film is distributed across the cornea during blinking. The video image of the ocular surface may be viewed on the computer screen display, in a printed report, or captured on video and exported to USB-attached storage or a mapped network drive. In the Gland Imaging Mode, Lipi View II uses near infrared (NIR) illumination and an NIR-sensitive camera mounted on a motion control system to image the meibomian glands. The tissue between the meibomian glands and the surface of the eyelid are transparent to NIR light. The glands reflect NIR wavelengths, allowing them to be imaged. In the Ocular Imaging mode, LipiView II captures high-resolution still images or video to record relevant findings on the ocular surface and/or eyelids. The images can be taken using room light only or the device's built-in white light. The LipiView II Interferometer does not provide a diagnosis. The device has no treatment or no life-sustaining functions, and presents no potential for serious risk to the health, welfare or safety of patients. ## INTENDED USE: The LipiView® II Ocular Surface Interferometer is intended to image the eye and surrounding area. The LipiView II Interferometer is a prescription device for use by a physician during an in-office exam. Indications for Use: The LipiView II Ocular Surface Interferometer is an ophthalmic imaging device intended for use by a physician in adult patients to capture, archive, manipulate and store digital images of: - . Specular (interferometric) observations of the tear film. Using these images, Lipi View II measures the absolute thickness of the tear film lipid layer. - Meibomian glands under near-infrared (NIR) illumination - . The ocular surface and eyelids under white illumination All of these image types can be photographically documented and visually monitored. As an ophthalmic camera, LipiView II has the same intended use to image the eye and surrounding area as the predicate devices, LipiView Interferometer (K091935, K122481) and Optomed Smartscope M5 Digital Camera (K132186). The Indications for Use for LipiView II are similar to the predicate LipiView device, except that the parts of the eye and surrounding area being imaged have been expanded to include the meibomian glands, ocular surface and eyelids. The expanded Indications for Use do not alter the intended use of the LipiView II Interferometer. LipiView II does not provide a diagnosis, and imaging other parts of the eye does not affect the safety and effectiveness of the device. {5}------------------------------------------------ ### TECHNOLOGICAL CHARACTERISTICS: The LipiView II Interferometer has the same fundamental scientific technology as the predicate devices. As summarized below, most of the technological characteristics of the LipiView II Interferometer remain unchanged from the predicate LipiView device. In addition, LipiView II and Optomed Smartscope M5 with optics module ES2 use similar visible and infrared LED illumination to image the eye. A comparison of the technology between LipiView II and each of the predicate devices is summarized below. Comparison to LipiView Interferometer: The predicate LipiView device and LipiView II have the same principle of operation as an ophthalmic camera to image the eye and for specular observations of the tear film using white light interferometry. Both bench-top devices have similar overall dimensions and components, including computer system and electronics, chin and forehead rest, camera and lens, motion stage, illuminator and touch screen display. Both devices may be used with off-the-shelf USB accessories and operate wirelessly with networks. Similar design features include: adjustable forehead and chin rest with the same patient contact material and disinfection method; digital video camera adjusted using a motion stage; white light Class I LED illuminator; and a visible LED fixation light. Both devices comply with ISO15004-2 standard for light hazard protection. Both devices have an AC power source in compliance with ANSI/IEC 60601 standards for electrical safety and electromagnetic compatibility (EMC). Both devices have flame retardant materials per UL 94V-1 standard. Both devices have ambient temperature of parts of device held by operator or accessible to patient in compliance with ISO 15004-1 standard. The software on both devices runs on a Microsoft Windows-based operating system with a touchscreen display graphical user interface. Similar software design features include: password-protected user login: patient entry into a database: real-time video display to acquire and view images; touchscreen user controls for camera and video playback; image acquisition process with storage of lossless AVI format video images; printed or saved reports; and export of images to USB external media or electronic medical record system via wireless network. In addition, both devices use the same tear film lipid layer image processing for analysis and measurement of the lipid layer thickness. Different design features include new hardware components in LipiView II: pushbutton controls, a two-position adjustable chin rest, a high definition camera, surface NIR illumination, and a handheld NIR lid everter. These design changes are to support imaging other parts of the eye and surrounding area, such as the meibomian glands. The primary software changes in LipiView II are an update to the Windows-based operating system, new graphical user interface appearance; and new meibomian gland imaging and ocular imaging functionality. Other minor hardware and software changes relative to the predicate LipiView device were made for user convenience or product enhancement. Comparison to Optomed M5 Digital Camera: The Optomed M5 Digital Camera with ES2 optics module and LipiView II have the same principle of operation as an ophthalmic camera for digital imaging of the eye. Both devices have a camera, lens, visible and infrared LED illumination and visible LED fixation light. Both devices {6}------------------------------------------------ comply with ISO15004-2:2007 standard for light hazard protection. Both devices have a power source in compliance with ANSI/IEC 60601-1-1 and 60601-1-2 standards for electrical safety and EMC. Also, both devices have display systems to view images and both provide image storage capability on external media. LipiView II is a bench-top device with a computer system, software and a touchscreen display graphical user interface. Conversely, Optomed Smartscope M5 with optic module ES2 consists of a camera handset, attachable ophthalmic lens, cradle for charging and image transfer to a PC by USB connection. LipiView II has an integrated chin rest to support the patient's head, whereas. Optomed Smartscope utilizes an adapter to a slit lamp biomicroscope, which has support for the patient's head. #### PERFORMANCE TESTING: Performance testing demonstrates that the design changes in LipiView II do not raise new questions of safety and effectiveness, and LipiView II is substantially equivalent to the predicate devices. - Usability testing and software validation demonstrate that LipiView II is as safe O and as effective as the predicate devices in the intended use to image the eye and surrounding area, including meibomian glands, ocular surface and eyelids. - As for the predicate devices, LipiView II meets the standards for light hazard o protection, electrical safety and electromagnetic compatibility and material flammability. Therefore, the use of near infrared illumination does not raise new safety or effectiveness questions. - The new hardware features in LipiView II, including two-position chin rest, high o definition camera, NIR illumination and handheld NIR lid everter, meet all design requirements for safety and performance, and therefore, do not adversely affect safety and effectiveness as compared to the predicate device. ## CONCLUSIONS: The LipiView II Ocular Surface Interferometer has the same intended use and the same fundamental scientific technology as the predicate devices. The expanded Indications for Use for LipiView II relative to the predicate devices do not alter the intended use and do not affect safety and effectiveness of the device. Performance testing demonstrates the LipiView II Interferometer is substantially equivalent in intended use and technological characteristics to the predicate devices. The technology differences do not raise new questions of safety and effectiveness, and do not adversely affect safety and effectiveness of the device. Based on overall design verification and validation and bench performance testing, LipiView II is at least as safe and effective as the predicate devices. {7}------------------------------------------------ | TABLE 5-1. PREDICATE DEVICE COMPARISON TABLE | | | | |-------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------| | Comparison Feature | LipiView II Interferometer | Predicate LipiView Interferometer<br>(K091935 and K122481) | Optomed Smartscope M5<br>Digital Camera (K132186) | | Intended Use | Intended to take photographs of the eye and the<br>surrounding area | Intended to take photographs of the eye and the<br>surrounding area | Intended to take photographs of<br>the eye and the surrounding area | | Indications for Use | The LipiView II Ocular Surface Interferometer is<br>an ophthalmic imaging device intended for use by<br>a physician in adult patients to capture, archive,<br>manipulate and store digital images of:<br>• Specular (interferometric) observations of the tear<br>film. Using these images, LipiView II measures<br>the absolute thickness of the tear film lipid layer.<br>• Meibomian glands under near-infrared (NIR)<br>illumination.<br>• The ocular surface and eyelids under white<br>illumination.<br>All of these image types can be photographically<br>documented and visually monitored. | The LipiView Ocular Surface Interferometer is an<br>ophthalmic imaging device that is intended for use<br>by a physician in adult patients to capture, archive,<br>manipulate, and store digital images of specular<br>(interferometric) observations of the tear film,<br>which can be visually monitored and<br>photographically documented. Using these<br>images, the LipiView® Interferometer measures<br>the absolute thickness of the tear film lipid layer. | Optomed Smartscope M5 with<br>optics module ES2 is intended to<br>capture images and video of the<br>surface of the human eye and<br>surrounding areas. | | Rx only/OTC | Prescription Device | Prescription Device | Prescription Device | | Method of Operation<br>Ophthalmic Imaging | Ophthalmic camera with digital imaging | Ophthalmic camera with digital imaging | Ophthalmic camera with digital<br>imaging | | Method of Operation<br>Specular Observation<br>of Tear Film | Tear film images and lipid layer thickness<br>measurement based on specular reflection using<br>white light interferometry | Tear film images and lipid layer thickness<br>measurement based on specular reflection using<br>white light interferometry | None | | Illumination Source | • Visible White Class I LED illumination<br>• Visible LED fixation light<br>• Near-Infrared LED illumination and<br>transillumination of the eyelid | • Visible White Class I LED illumination<br>• Visible LED fixation light | • Visible and Infrared LED<br>illumination<br>• Red LED Fixation Light<br>• White and Cobalt Blue Light | | Illumination Exposure | Complies with ISO 15004-2:2007 Group 2 | Complies with ISO 15004-2:2007 Group 1 | Complies with ISO 15004-2:2007<br>Group 2 instrument for safety | | Parameters | instrument for safety | instrument for safety | | | Brightness Control | No brightness control adjustment by user | No brightness control adjustment by user | Unknown | | Material Flammability | Materials near light source comply with UL 94V-1 | Materials near light source comply with UL 94V-1 | Unknown | | Maximum Temperature of<br>Held or Accessible Parts | Ambient temperature of parts of device held by<br>operator or accessible to patient | Ambient temperature of parts of device held by<br>operator or accessible to patient | Unknown | | TABLE 5-1. PREDICATE DEVICE COMPARISON TABLE | | | | | Comparison Feature | LipiView II Interferometer | Predicate LipiView Interferometer<br>(K091935 and K122481) | Optomed Smartscope M5<br>Digital Camera (K132186) | | Data Collection and/or<br>Display Systems | • Touchscreen display to view images<br>• Stores images on computer and external media<br>• Support for off-the-shelf USB printer and media<br>that are compatible with Windows 7 and USB2.0 | • Touchscreen display to view images<br>• Stores images on computer and external media<br>• Support for off-the-shelf USB printer and media<br>compatible with Windows XP and USB1.0 | • Active matrix color TFT LCD<br>• Stores images on external<br>media (flash memory card)<br>• USB1.1 terminal compatible<br>with Windows XP/Vista/7 | | Electrical Safety and<br>Electromagnetic<br>Compatibility | • AC-powered<br>• Complies with ANSI 60601-1:2005 (3rd edition)<br>and IEC 60601-1-2: 2007 (3rd edition) | • AC-powered<br>• Complies with IEC 60601-1:1995 (2nd edition)<br>and IEC 60601-1-2: 2001 (2nd edition) | • Rechargeable NiMH Battery<br>with charging unit<br>• Complies with IEC 60601-<br>1:2005 (3rd edition) and IEC<br>60601-1-2: 2007 (3rd edition) | | Chin/Forehead Rest<br>Support | Integrated chin rest and forehead support is<br>disinfected with alcohol prior to use and storage | Integrated chin rest and forehead support is<br>disinfected with alcohol prior to use and storage | May be used with slit lamp<br>adapter for chin rest support | {8}------------------------------------------------
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...