K122481 · Tearscience, Inc. · HKI · Dec 31, 2012 · Ophthalmic
Device Facts
Record ID
K122481
Device Name
LIPIVIEW OCULAR SURFACE INTERFEROMETER
Applicant
Tearscience, Inc.
Product Code
HKI · Ophthalmic
Decision Date
Dec 31, 2012
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 886.1120
Device Class
Class 2
Indications for Use
The LipiView® Ocular Surface Interferometer is an ophthalmic imaging device that is intended for use by a physician in adult patients to capture, archive, manipulate and store digital images of specular (interferometric) observations of the tear film, which can be visually monitored and photographically documented. Using these images, the LipiView® Interferometer measures the absolute thickness of the tear film lipid layer. The LipiView® Interferometer is a prescription device for use by a physician during an in-office exam.
Device Story
Bench-top ophthalmic imaging device; uses white light interferometry to capture specular reflections of tear film. Input: video images of ocular surface during blinking. Processing: software version 2.0 performs interferometric color matching and blink detection to calculate absolute tear film lipid layer thickness. Output: digital images, video files, and thickness measurements displayed on touchscreen, printed, or exported via USB/network. Used in-office by physicians. Benefits: provides objective, quantitative assessment of tear film lipid layer, aiding clinical evaluation of ocular surface health.
Clinical Evidence
Bench testing only. Validation performed using physical phantoms representative of pre-corneal tear film with thicknesses verified by ellipsometry. Results demonstrated absolute measurement capability with in vivo variability of 0.31 interferometric color units (ICU) (reporting precision 1 ICU). Performance equal to or better than predicate; higher percentage of correctly identified valid blink frames.
Technological Characteristics
Bench-top device; Class I white light LED illuminator (ISO 15004-2); digital video camera; PC with Windows OS; touchscreen GUI. Materials: standard chin/forehead rests. Connectivity: USB/network export. Electrical: AC power (IEC 60601). Software: version 2.0 with refined interferometric color matching and blink detection algorithms.
Indications for Use
Indicated for use by physicians in adult patients to capture, archive, manipulate, and store digital images of specular (interferometric) observations of the tear film for visual monitoring and documentation, and to measure absolute tear film lipid layer thickness.
Regulatory Classification
Identification
An ophthalmic camera is an AC-powered device intended to take photographs of the eye and the surrounding area.
Special Controls
*Classification.* Class II (special controls). The device, when it is a photorefractor or a general-use ophthalmic camera, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 886.9.
{0}------------------------------------------------
510 (k): K122481
DEC 3 1 2012
# 510(k) SUMMARY
| PREPARATION DATE: | December 12, 2012 |
|------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| APPLICANT: | TearScience, Inc.<br>5151 McCrimmon Parkway, Suite 250<br>Morrisville, NC 27560<br>Tel: (919) 459-4815<br>Fax: (877) 468-5335 |
| CONTACT PERSON: | Christy Stevens, OD, MPH<br>Vice President, Clinical & Regulatory |
| DEVICE TRADE NAME: | LipiView® Ocular Surface Interferometer |
| COMMON NAME: | Ophthalmic Imaging Device |
| CLASSIFICATION NAME: | Ophthalmic Camera |
| DEVICE CLASSIFICATION: | Class II, 21 CFR 886.1120 and 886.1850 |
| PRODUCT CODE: | HKI, HJO |
| PREDICATE DEVICE: | LipiView® Ocular Surface Interferometer<br>Class II under 21 CFR 886.1120 and 21 CFR 886.1850<br>Product Code HKI, HJO; Applicant: TearScience, Inc.;<br>Cleared under K091935 on October 23, 2009 |
## DEVICE DESCRIPTION:
The LipiView Ocular Surface Interferometer is a bench-top imaging device containing a computer system and electronics, chin rest and forehead rest, camera and zoom lens, illuminator and a touch screen display. The LipiView® Interferometer operates on the principle of white light interferometry and provides an interferometry color assessment of the tear film by specular reflection. The computer system captures a video image file that is recorded over time since the interference pattern changes as the tear film is distributed across the cornea during blinking. The video image of the ocular surface may be viewed on the computer screen display, in a printed report, or captured on video and exported to USBattached storage or a mapped network drive.
The LipiView® Interferometer has been modified to software version 2.0, which includes changes to refine the interferometric color matching and blink detection methods used on interferometric images and to provide minor usability enhancements.
## INTENDED USE:
The LipiView® Interferometer is intended to image the tear film. The LipiView® Interferometer is a prescription device for use by a physician during an in-office exam.
Indications for Use: The LipiView® Ocular Surface Interferometer is an ophthalmic imaging device that is intended for use by a physician in adult patients to capture, archive, manipulate and store digital images of specular (interferometric) observations of the tear film, which can be visually monitored and photographically documented. Using these images, the LipiView Interferometer measures the absolute thickness of the tear film lipid layer.
{1}------------------------------------------------
The LipiView® Interferometer with software version 2.0 has the same intended use as the predicate LipiView® Interferometer (K091935). In addition, the Indications for Use are similar to the predicate device, except that the Indications for Use have been expanded for the LipiView® Interferometer with software version 2.0 to include measurement of the tear film lipid laver thickness, as supported by performance testing. The expanded Indications for Use do not alter the intended use of the LipiView® Interferometer.
## TECHNOLOGICAL CHARACTERISTICS:
The LipiView® Interferometer with software version 2.0 has the same fundamental scientific technology as the predicate device. As summarized below, most of the technological characteristics of the LipiView® Interferometer with the modification to software version 2.0 remain unchanged from the predicate device cleared under K091935. Minor differences in technology between the predicate device and the LipiView® Interferometer with software version 2.0 are described below.
Similarities: The LipiView® Interferometer with software version 2.0 and the predicate device share many of the same design features. Both devices have the same operating principle of real-time imaging of tear film dynamics based on the interference pattern from specular reflections. Both devices have an AC power source in compliance with IEC 60601 standards for electrical safety and electromagnetic compatibility. Both devices use the same Class I white light LED illuminator with exposure and level of illumination in compliance with ISO 15004-2 (Group 1 instrument) for safety. The patient contact materials for the chin and forehead rest and the method of disinfection are the same for both devices. Also, both devices have a digital video camera, personal computer with Microsoft Windows-based operating system, touchscreen display graphical user interface and computer accessory support for printing and data storage.
Furthermore, analogous software features on both devices include: password-protected user login; patient database; real-time video display to acquire tear film images; touchscreen user controls for camera and video playback; image acquisition process with storage of lossless A VI format video images; and tear film video playback and analysis.
Differences: Compared to the predicate device, the LipiView® Interferometer with software version 2.0 has refined interferometric color matching and blink detection methods used on interferometric images. In the predicate device, the interferometric color palette was theoretically derived: whereas in software version 2.0. the palette was developed and validated to a known standard for measurement of the tear film lipid layer thickness. Software version 2.0 also has enhancements for user convenience including: playback blink visualization; automated documentation of blinks; example tear film videos; adjustable video capture length; and ability to export video to an external USB storage device or mapped network drive.
## PERFORMANCE TESTING:
The LipiView® Interferometer with software version 2.0 was developed and tested in compliance with design controls and the FDA Guidance documents for software validation in medical devices. To support the expanded Indications for Use, the LipiView® Interferometer was validated to physical phantoms representative of the pre-corneal tear film of a known thickness, as measured independently by ellipsometry. This testing demonstrated the
{2}------------------------------------------------
LipiView® Interferometer can make absolute measurements of the tear film lipid layer thickness by imaging interferometric colors. Test results showed the in vivo device measurement variability is 0.31 interferometric color unit (ICU) on average, or slightly less than one-third of the reporting precision of the device (1 ICU). Verification and validation test results demonstrated that the interferometric color matching performance of the LipiView® Interferometer with software version 2.0 is equal to or better than the predicate device. Tests also showed that version 2.0 had a higher overall percentage of correctly identified valid blink frames as compared to the predicate device. Enhancements for user convenience in software version 2.0 performed as intended and did not introduce any new risks to the device.
## CONCLUSIONS:
The LipiView Ocular Surface Interferometer with software version 2.0 has the same intended use and the same fundamental scientific technology as the predicate device. Performance testing demonstrates the LipiView® Interferometer with software version 2.0 is substantially equivalent in technological characteristics to the predicate device.
{3}------------------------------------------------
# DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized graphic of three human profiles facing to the right, stacked on top of each other.
#### Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
DEC 3 1 2012
Tearscience, Inc. c/o Christy Stevens, OD VP. Clinical & Regulatory Affairs 5151 McCrimmon Parkway, Suite 250 Morrisville, NC 27560
Re: K122481
Trade/Device Name: LipiView Ocular Surface Interferometer Regulation Number: 21 CFR 886.1120 Regulation Name: Ophthalmic Camera Regulatory Class: II Product Code: HKI, HJO Dated: December 18, 2012 Received: December 19, 2012
Dear Dr. Stevens:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21
{4}------------------------------------------------
## Page 2 - Dr. Christy Stevens
CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
# Deborah L. Falls
Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# INDICATIONS FOR USE
510(k) Number:
KIZL 48 | 8 | To Be Assigned By FDA)
Device Name: LipiView® Ocular Surface Interferometer with Software Version 2.0
## Indications for Use:
The LipiView® Ocular Surface Interferometer is an ophthalmic imaging device that is intended for use by a physician in adult patients to capture, archive, manipulate and store digital images of specular (interferometric) observations of the tear film, which can be visually monitored and photographically documented. Using these images, the LipiView® Interferometer measures the absolute thickness of the tear film lipid layer.
Prescription Use X · (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
# (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) .
Concurrence of CDRH, Office of Device Evaluation (ODE)
R
(Division Sign-Off) Division of Ophthalmic and Ear, Nose and Throat Devices 510(k) Number
Page of
Page 40 of 1022
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.