ZEISS CATARACT SUITE MARKERLESS

K141068 · Carl Zeiss Meditec, AG · HJO · Sep 9, 2014 · Ophthalmic

Device Facts

Record IDK141068
Device NameZEISS CATARACT SUITE MARKERLESS
ApplicantCarl Zeiss Meditec, AG
Product CodeHJO · Ophthalmic
Decision DateSep 9, 2014
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 886.1850
Device ClassClass 2

Indications for Use

The ZEISS Cataract Suite markerless workflow uses a preoperative image capture tool from the IOLMaster 500 that permits visualization and guidance during cataract surgery using CALLISTO eye. The ZEISS Cataract Suite markerless utilizes an ophthalmic surgical microscope and the following medical devices: IOLMaster 500: The IOLMaster is intended for the biometric determination of ocular measurements of axial length, anterior chamber depth, corneal radius, white-to-white (WTW), and for the measurement of pupil size and deviation of the visual axis from the center of the pupil. For patients who are candidates for intraocular lens (IOL) implantation, the device also performs calculations to assist physicians in determining the appropriate IOL power and type for implantation. This device is intended for use by physicians and eye-care professionals and may only be used under supervision of a physician. CALLISTO eye: CALLISTO eye is an assistance system that provides non-diagnostic video documentation and image capture for ophthalmic surgeries. The system allows the remote control of the surgical microscope. The graphical guidance tools, as displayed on the CALLISTO eye Panel PC or microscope eye piece, aid the surgeon to insert, align, position, and register an artificial lens. These tools are intended for anterior segment ophthalmic surgical procedures, including positioning and angular alignment of toric intraocular lenses, limbal relaxing incisions and capsulorhexis. The system utilizes surgeon information for position of graphical guidance tools.

Device Story

Workflow integrates IOLMaster 500 and CALLISTO eye for cataract surgery. IOLMaster 500 captures preoperative ocular biometric data (axial length, depth, radius, WTW, pupil size/axis) to calculate IOL power. Data transfers to CALLISTO eye, an assistance system for ophthalmic surgical microscopes. CALLISTO eye provides non-diagnostic video documentation and remote microscope control. System projects graphical guidance templates/overlays into the surgeon's microscope eyepiece or panel PC. Surgeon uses these overlays to align, position, and register toric IOLs, limbal relaxing incisions, and perform capsulorhexis. System aids surgical precision by providing visual reference for angular alignment. Used in operating rooms by surgeons; benefits patient through improved accuracy of IOL placement.

Clinical Evidence

No clinical data. Evidence consists of software verification and validation per IEC 62304:2006 and FDA guidance, electromagnetic compatibility testing, safety testing, and bench testing.

Technological Characteristics

Workflow integrates optical biometer (IOLMaster 500) and surgical assistance system (CALLISTO eye). Uses green LED image capture. Graphical guidance tools projected via microscope eyepiece or panel PC. Software-based workflow; connectivity between biometric device and surgical microscope. Complies with IEC 62304:2006.

Indications for Use

Indicated for patients undergoing cataract surgery requiring biometric ocular measurements and surgical guidance for IOL implantation, toric IOL alignment, limbal relaxing incisions, and capsulorhexis.

Regulatory Classification

Identification

An AC-powered slitlamp biomicroscope is an AC-powered device that is a microscope intended for use in eye examination that projects into a patient's eye through a control diaphragm a thin, intense beam of light.

Special Controls

*Classification.* Class II (special controls). The device, when it is intended only for the visual examination of the anterior segment of the eye, is classified as Group 1 per FDA-recognized consensus standard ANSI Z80.36, does not provide any quantitative output, and is not intended for screening or automated diagnostic indications, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 886.9.

Predicate Devices

Reference Devices

Submission Summary (Full Text)

{1}------------------------------------------------ Image /page/0/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized image of an eagle with three human profiles incorporated into its design. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 December 19, 2014 Carl Zeiss Meditec, Inc. Ms. Sarah Harrington, M.S., MBA Staff Regulatory Specialist 5160 Hacienda Drive Dublin, CA 94568 Re: K141068 Trade/Device Name: Zeiss Cataract Suite markerless Regulation Number: 21 CFR 886.1850 Regulation Name: Biomicroscope, Slit-Lamp, AC-powered Regulatory Class: Class II Product Code: HJO, NFJ, HMR, EPT, HRM Dated: July 29, 2014 Received: July 30, 2014 Dear Ms. Harrington: This letter corrects our substantially equivalent letter of September 9, 2014. We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act {2}------------------------------------------------ or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerelv vours. Kesia Y. Alexander -S for Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ ### 1.0 INDICATIONS FOR USE - ZEISS CATARACT SUITE MARKERLESS 510(k) Number (if known): ZEISS Cataract Suite markerless Device Name(s): The ZEISS Cataract Suite markerless workflow uses a preoperative image capture tool from the IOLMaster 500 that permits visualization and guidance during cataract surgery using CALLISTO eye. The ZEISS Cataract Suite markerless utilizes an ophthalmic surgical microscope and the following medical devices: Indications for Use: #### IOLMaster 500 The IOLMaster is intended for the biometric determination of ocular measurements of axial length, anterior chamber depth, corneal radius, white-to-white (WTW), and for the measurement of pupil size and deviation of the visual axis from the center of the pupil. For patients who are candidates for intraocular lens (IOL) implantation, the device also performs calculations to assist physicians in determining the appropriate IOL power and type for implantation. This device is intended for use by physicians and eye-care professionals and may only be used under supervision of a physician. #### CALLISTO eve CALLISTO eye is an assistance system that provides non-diagnostic video documentation and image capture for ophthalmic surgeries. The system allows the remote control of the surgical microscope. The graphical guidance tools, as displayed on the CALLISTO eye Panel PC or microscope eye piece, aid the surgeon to insert, align, position, and register an artificial lens. These tools are intended for anterior segment ophthalmic surgical procedures, including positioning and angular alignment of toric intraocular lenses, limbal relaxing incisions and capsulorhexis. The system utilizes surgeon information for position of graphical guidance tools. | Prescription Use | X | |-----------------------------|--------| | (Part 21 CFR 801 Subpart D) | | | | AND/OR | | Over-The-Counter Use | | | (21 CFR 807 Subpart C) | | (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Page {4}------------------------------------------------ # 2.0 510(K) SUMMARY This 510(k) summary of safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR 807.92(a). #### Submitter's name, address, telephone number, contact person, and date summary prepared: | 1. Applicant: | Carl Zeiss Meditec AG<br>Goeschwitzer Strasse 51-52<br>07745 Jena<br>Germany | |--------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | 2. Contact Person: | Sarah Harrington, MS, MBA<br>Staff Regulatory Specialist<br>Carl Zeiss Meditec, Inc.<br>5160 Hacienda Drive<br>Dublin, CA 94568<br>sarah.harrington@zeiss.com<br>Tel: (925) 560-5134<br>Fax: (925) 557-4259 | ### Name of device, including trade name and classification name | 1. | Trade/Proprietary Name: | a. IOLMaster / IOLMaster 500<br>b. CALLISTO eye<br>c. Opthalmic Surgical Microscopes | |----|-------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | 2. | Common/Usual Name: | a. Biometer<br>b. Picture Archiving and Communication System<br>c. Ophthalmic Surgical Microscope and accessories | | 3. | Classification Name: | a. AC-powered slit lamp biomicroscope<br>b. System, Image management, Ophthalmic;<br>Ocular Marker; Surgical Microscope<br>c. Surgical Microscope, operating & accessories,<br>AC-powered, ophthalmic | | 4. | Product Code and Class: | a. HJO - Class II<br>b. NFJ, HMR, EPT– Class II<br>c. HRM - Class I | | 5. | Classification Number: | a. 21 CFR 886.1850<br>b. 21 CFR 892.2050<br>c. 21 CFR 878.4700 | {5}------------------------------------------------ # Predicate Device The ZEISS Cataract Suite markerless is similar in function and application in cataract surgery to the predicate device, the TrueVision 3D Visualization and Guidance System (K101861). The visualization system displays real-time images during ophthalmic surgery on a flat-panel digital display device positioned for live video image viewing by the surgeon in the operating room. # Device Description The markerless toric IOL alignment workflow provided with the ZEISS Cataract Suite markerless is a series of medical devices integrated into one workflow for use by a cataract surgeon. The workflow comprises an optical biometric device with green LED image capture from the IOLMaster 500 (K122418) with an optional accessory (Option Reference Image). The captured image is transferred to a cataract surgery assistance system, the CALLISTO eye (K123464), which operates in conjunction with ophthalmic surgical microscopes. Integrated or external data injection systems (IDIS or EDIS) are capable of displaying the graphical templates of the CALLISTO eye in the right ocular of the ophthalmic surgical microscopes. # Indications for Use The ZEISS Cataract Suite markerless workflow uses a preoperative image capture tool from the IOLMaster 500 that permits visualization and guidance during cataract surgery using CALLISTO eye. The ZEISS Cataract Suite markerless utilizes an ophthalmic surgical microscope and the following medical devices: ## IOLMaster 500 The IOLMaster is intended for the biometric determination of ocular measurements of axial length, anterior chamber depth, corneal radius, white-to-white (WTW), and for the measurement of pupil size and deviation of the visual axis from the center of the pupil. For patients who are candidates for intraocular lens (IOL) implantation, the device also performs calculations to assist physicians in determining the appropriate IOL power and type for implantation. This device is intended for use by physicians and eye-care professionals and may only be used under supervision of a physician. {6}------------------------------------------------ ### Indications for Use (cont.) ## CALLISTO eye CALLISTO eye is an assistance system that provides non-diagnostic video documentation and image capture for ophthalmic surgeries. The system allows the remote control of the surgical microscope. The graphical guidance tools, as displayed on the CALLISTO eye Panel PC or microscope eye piece, aid the surgeon to insert, align, position, and register an artificial lens. These tools are intended for anterior segment ophthalmic surgical procedures, including positioning and angular alignment of toric intraocular lenses, limbal relaxing incisions and capsulorhexis. The svstem utilizes surgeon information for position of graphical guidance tools. ### Comparison of Technological Characteristics The ZEISS Cataract Suite markerless and the predicate device, the TrueVision 3D Visualization and Guidance System (K101861) share similar functional features and operating characteristics. The TrueVision system captures an image of the eye that is displayed by the 3D Visualization and Guidance System. As with the CALLISTO eye, the TrueVision 3D Visualization and Guidance System provides graphical templates and overlays that are viewed via an ophthalmic surgical microscope to aid the surgeon during cataract surgery. ### Performance Data The IOLMaster 500 and CALLISTO eye software were tested according to the FDA Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices, May 11, 2005. The software testing also followed the Carl Zeiss Meditec internal software development procedure that is in compliance with the IEC 62304:2006 - Medical device software - Software life cycle processes. Electromagnetic compatibility and safety testing was conducted along with appropriate bench testing to verify changes to the devices. ### Conclusion Based on the software verification and validation, safety testing and bench testing as well as the comparison to the predicate devices, the ZEISS Cataract Suite markerless as well as the contributing devices are safe and effective with respect to their Indications for Use.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...