IOLMASTER 500

K122418 · Carl Zeiss Meditec, AG · HJO · Apr 12, 2013 · Ophthalmic

Device Facts

Record IDK122418
Device NameIOLMASTER 500
ApplicantCarl Zeiss Meditec, AG
Product CodeHJO · Ophthalmic
Decision DateApr 12, 2013
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 886.1850
Device ClassClass 2
AttributesReal-World Evidence

Real-World Evidence

SubmissionDeviceSponsorRWD SourcesRWE Use SummaryKey Tags
K122418 · Apr 12, 2013IOLMASTER 500Carl Zeiss Meditec, AGRetrospective analysis of a previously conducted clinical study datasetA retrospective analysis of 116 astigmatic eyes was used to evaluate the agreement in keratometry function (corneal power, cylindrical power, and axis) between the IOLMaster 500 and the Javal manual keratometer.Retrospective analysis; Keratometry agreement; Clinical study dataset

Clinical Evidence

Study DesignPopulationComparatorKey Endpoints
Retrospective analysis of a previously conducted prospective clinical study116 astigmatic eyes with at least 0.75 D of astigmatism; Sample Size: 116; Number of Sites: 1Javal (Haag-Streit) manual keratometerAgreement in keratometry function (corneal power, cylindrical power, and axis)

Indications for Use

The IOLMaster is intended for the biometric determination of ocular measurements of axial length, anterior chamber depth, corneal radius, white-to-white (WTW), and for the measurement of pupil size and deviation of the visual axis from the center of the pupil. For patients who are candidates for intraocular lens (IOL) implantation, the device also performs calculations to assist physicians in determining the appropriate IOL power and type for implantation. This device is intended for use by physicians and eye-care professionals and may only be used under the supervision of a physician.

Device Story

Non-contact biometry instrument; uses partial coherence interferometry and traditional ophthalmic biometry techniques to measure axial length, corneal radius, anterior chamber depth, WTW, pupil size, and visual axis deviation. Operated by physicians or eye-care professionals in clinical settings. Integrated software algorithms perform calculations using clinically recognized formulas to assist in determining IOL power and type for cataract surgery. Includes remote connection capability (Netviewer one2meet) for service. Output assists physicians in preoperative surgical planning; benefits patient by facilitating accurate IOL selection.

Clinical Evidence

Two prospective clinical studies evaluated keratometry performance. Study 1 (n=61 astigmatic eyes) compared IOLMaster 500 to Marco manual keratometer; Study 2 (n=116 astigmatic eyes) compared IOLMaster 500 to Javal manual keratometer. Endpoints included corneal power, cylindrical power, and axis. Results showed high agreement between devices. Repeatability and reproducibility metrics (SD, %COV) for the IOLMaster 500 were generally superior to manual keratometers. Differences in measurements were deemed inconsequential for toric IOL calculations.

Technological Characteristics

Non-contact biometry instrument; utilizes partial coherence interferometry and traditional ophthalmic biometry. Features integrated software for IOL power calculations (including Holladay 2 formula). Connectivity includes remote service access via Netviewer one2meet. Standalone device form factor.

Indications for Use

Indicated for patients who are candidates for intraocular lens (IOL) implantation requiring ocular biometric measurements (axial length, anterior chamber depth, corneal radius, WTW, pupil size, visual axis deviation) and IOL power/type calculations.

Regulatory Classification

Identification

An AC-powered slitlamp biomicroscope is an AC-powered device that is a microscope intended for use in eye examination that projects into a patient's eye through a control diaphragm a thin, intense beam of light.

Special Controls

*Classification.* Class II (special controls). The device, when it is intended only for the visual examination of the anterior segment of the eye, is classified as Group 1 per FDA-recognized consensus standard ANSI Z80.36, does not provide any quantitative output, and is not intended for screening or automated diagnostic indications, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 886.9.

Predicate Devices

Reference Devices

Submission Summary (Full Text)

{0}------------------------------------------------ # K122418 510(K) SUMMARY # 510(K) SUMMARY # APR 1 2 2013 This 510(k) summary of safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR 807.92(a). ## SUBMITTER'S NAME, ADDRESS, TELEPHONE NUMBER, CONTACT PERSON, AND DATE SUMMARY PREPARED | a. | Applicant: | Carl Zeiss Meditec AG<br>Goeschwitzer Strasse 51-52<br>07745 Jena<br>Germany | |----|-----------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | b. | Contact Person: | Judith A. Brimacombe, MA<br>Director, Clinical/Regulatory Affairs<br>Carl Zeiss Meditec, Inc.<br>5160 Hacienda Drive<br>Dublin, CA 94568<br>USA<br>j.brimacombe@meditec.zeiss.com<br>Tel: (925) 557-4616<br>Fax: (925) 557-4259 | August 6, 2012 c. Summary Prepared: ### NAME OF DEVICE, INCLUDING TRADE NAME AND CLASSIFICATION NAME | a. | Trade/Proprietary Name: | IOLMaster® 500 | |----|-------------------------|------------------------------------| | b. | Common/Usual Name: | Biometer | | c. | Classification Name: | AC-powered slit lamp biomicroscope | | d. | Product Code and Class: | HJO - Class II | | e. | Classification Number: | 886.1850. | {1}------------------------------------------------ ## PREDICATE DEVICE The IOLMaster 500, the subject of this 510(k), is substantially equivalent to the predicate IOL.Master 500 (Carl Zeiss Meditec, AG) cleared for marketing under 510(k) number K101182 in October 2010. #### DEVICE DESCRIPTION The IOLMaster 500 is a non-contact biometry instrument for measurements of the eve required for preoperative computation of intraocular lens (IOL) type and power. As with the IOLMaster 500 predicate device, the IOLMaster 500 provides measurements of axial length, corneal radius (keratometry), anterior chamber depth and the "white" distance (WTW). #### STATEMENT OF INTENDED USE The IOLMaster is intended for the biometric determination of ocular measurements of axial length, anterior chamber depth, corneal radius, white (WTW), and for the measurement of pupil size and deviation of the visual axis from the center of the pupil. For patients who are candidates for intraocular lens (IOL) implantation, the device also performs calculations to assist physicians in determining the appropriate IOL power and type for implantation. This device is intended for use by physicians and eye-care professionals and may only be used under the supervision of a physician. #### COMPARISON OF TECHNOLOGICAL CHARACTERISTICS The IOLMaster 500 has the same indications for use and operating characteristics as the predicate IOLMaster 500. The IOLMaster 500 utilizes partial coherence interferometry and traditional ophthalmic biometry techniques to obtain measurements for axial length, corneal radius, anterior chamber depth and white-to white distance. Integrated algorithms in the software allow for the use of clinically recognized formulas for the calculation of IOL type and power prior to cataract surgery. These functions are identical for the IOLMaster 500 and the predicate device. New software features have been added to the updated IOLMaster 500. These include implementation of the Holladay 2 IOL calculation formula, improvements in displayed parameters, and the use of Netviewer one2meet for remote connection to Carl Zeiss Meditec Service. As the measurements are achieved in the same manner using the identical optical technology; the IOLMaster 500 is therefore substantially equivalent to the predicate IOLMaster 500 (K101182). {2}------------------------------------------------ #### BRIEF SUMMARY OF PERFORMANCE TESTS AND RESULTS A prospective, single site clinical study comparing the performance of the IOLMaster 500 with the Marco (Bausch & Lomb, Inc.) manual keratometer was conducted in 61 astigmatic eyes with at least 0.75 D of astigmatism. This study evaluated the agreement in the keratometry function for corneal power, cylindrical power (i.e., astigmatic power) and axis. The agreement between the instruments is summarized in Table 1. TABLE 1 AGREEMENT BETWEEN THE MARCO KERATOMETER AND THE IOLMASTER IN 61 EYES (ALL VALUES ARE IN D AND ARE MEAN ± SD UNLESS INDICATED) | | Marco<br>Keratometer | IOLMaster | Difference | 95% LoA | |-------------------------|----------------------|----------------|--------------|----------------| | Power in Flat Meridian | 42.30 ± 1.42 | 42.54 ± 1.40 | +0.24 ± 0.13 | -0.07 to +0.55 | | Power in Steep Meridian | 43.68 ± 1.46 | 44.11 ± 1.47 | +0.43 ± 0.21 | +0.02 to +0.84 | | Mean Power (P1+P2)/2 | 42.99 ± 1.39 | 43.33 ± 1.37 | +0.33 ±0.13 | +0.05 to +0.63 | | Astigmatic Power | -1.38 ± 0.80 | -1.56 ± 0.86 | -0.18 ± 0.23 | -0.64 to +0.26 | | Axis [°]* | 110.89 ± 71.59 | 113.97 ± 72.33 | 4.00 ± 3.30 | +10.47 | * The upper limit was derived based on the method of Brand and Altman, DG. Statistical methods for assessing agreement between two methods of clinical measurement. Lancet i:307-310, 1986). The lower limit was not provided since the difference in axis is always a positive number. The results of the study demonstrate a high level of agreement between the instruments for measurements of power in the flat and steep meridians, mean power, astigmatic power and axis. Typically, the manual keratometer measured significantly flatter radii and hence lower corneal power in both meridians. The mean power (±SD) as measured by the IOLMaster was 43.33 ± 1.37 D compared to 42.99 ± 1.39 D for the manual keratometer. Overall, the manual keratometer measured less astigmatic power (-1.38 ± 0.80 D) compared to the IOLMaster (-1.56 ± 0.86 D). The differences between the measurements would be inconsequential in toric IOL calculations. To evaluate the repeatability and reproducibility of the IOLMaster compared to the manual keratometer, the study was conducted in two phases designed to evaluate repeatability and reproducibility of the measurements. - . In Phase 1, inter-instrument variability for both the IOLMaster and the Marco keratometer was evaluated, as was the agreement between the two instruments. Three IOLMaster units and three manual keratometer units were used, and 5 measurements were taken with each unit (30 total measurements per subject). - . In Phase 2, inter-operator variability for both the IOLMaster and the Marco keratometer was evaluated, as was the agreement between the IQLMaster and the Marco keratometer. One IOLMaster unit and one Marco manual keratometer unit were used by three operators, and 5 measurements were taken by each operator (30 total measurements per subject). The results of the analysis for repeatability and reproducibility for the Marco manual Keratometer and the IOLMaster 500 are summarized in Table 2. {3}------------------------------------------------ | | Overall<br>Mean | Repeatability | | | Reproducibility | | | |-----------------------------------------|-----------------|---------------|---------|--------|-----------------|---------|--------| | | | SD | Limit | %COV | SD | Limit | %COV | | MARCO | | | | | | | | | R1, Radius in Flattest<br>Meridian [mm] | 7.88 | 0.0381 | 0.1068 | 0.48% | 0.0492 | 0.1376 | 0.62% | | R2, Radius in Steepest<br>Meridian [mm] | 7.63 | 0.0654 | 0.1832 | 0.86% | 0.0799 | 0.2237 | 1.05% | | P1, Power in Flattest<br>Meridian [D] | 42.30 | 0.2201 | 0.6162 | 0.52% | 0.2800 | 0.7840 | 0.66% | | P2, Power in Steepest<br>Meridian [D] | 43.68 | 0.3953 | 1.1067 | 0.90% | 0.4774 | 1.3367 | 1.09% | | Mean Power,<br>(P1 + P2)/2 [D] | 42.99 | 0.2388 | 0.6686 | 0.56% | 0.3162 | 0.8854 | 0.74% | | Astigmatic Power [D] | -1.38 | 0.4249 | 1.1896 | 30.76% | 0.4604 | 1.2890 | 33.34% | | Axis [°] | 110.89 | 3.1692 | 8.8738 | 2.86% | 4.6560 | 13.0369 | 4.20% | | IOLMaster | | | | | | | | | R1, Radius in Flattest<br>Meridian [mm] | 7.84 | 0.0154 | 0.0431 | 0.20% | 0.0165 | 0.0462 | 0.21% | | R2, Radius in Steepest<br>Meridian [mm] | 7.56 | 0.0179 | 0.0501 | 0.24% | 0.0192 | 0.0539 | 0.25% | | P1, Power in Flattest<br>Meridian [D] | 42.54 | 0.0686 | 0.1921 | 0.16% | 0.0748 | 0.2094 | 0.18% | | P2, Power in Steepest<br>Meridian [D] | 44.11 | 0.0875 | 0.2449 | 0.20% | 0.1010 | 0.2827 | 0.23% | | Mean Power,<br>(P1 + P2)/2 [D] | 43.32 | 0.0563 | 0.1577 | 0.13% | 0.0663 | 0.1855 | 0.15% | | Astigmatic Power [D] | -1.56 | 0.1369 | 0.3833 | 8.75% | 0.1403 | 0.3927 | 8.96% | | Axis [°] | 113.97 | 3.9249 | 10.9897 | 3.44% | 4.1737 | 11.6863 | 3.66% | TABLE 2 REPEATABILITY AND REPRODUCIBILITY MARCO MANUAL KERATOMETER AND IOLMASTER 500 Repeatability includes variation due to measurement error. Reproducibility includes variations due to device, operator, interaction between device and subject, interaction between operator and subject, and measurement error. Repeatability %COV = (Repeatability SD)/|Overall Mean|*100% Reproducibility %COV = (Reproducibility SD)/|Overall Mean|* 100% The prospective clinical study demonstrated excellent agreement between the IOLMaster and Marco manual keratometer for corneal power, astigmatic power and axis. As demonstrated by these results, for all measurements, with the exception of axis, the repeatability and reproducibility limits, and the standard deviations were smaller for the IOLMaster than for the manual keratometer. For axis, the instruments have similar repeatability and reproducibility. {4}------------------------------------------------ Additionally, a retrospective analysis of a previously conducted prospective, single site clinical study comparing the performance of the IOLMaster 500 with the Javal (Haag-Streit) manual keratometer was conducted in 116 astigmatic eyes with at least 0.75 D of astigmatism. This study evaluated the agreement in the keratometry function for corneal power, cylindrical power (i.e., astigmatic power) and axis. In this study, five measurements were taken on each eye with one IOLMaster 500 instrument and one Javal keratometer by a single operator. The agreement between the instruments is summarized in Table 3 with the repeatability of measures shown in Table 4. #### TABLE 3 #### AGREEMENT BETWEEN THE JAVAL KERATOMETER AND THE IOLMASTER IN 116 EYES (ALL VALUES ARE IN D AND ARE MEAN ± SD UNLESS INDICATED) | | Javal<br>Keratometer | IOLMaster | Difference | 95% LoA | |-------------------------|----------------------|----------------|--------------|----------------| | Power in Flat Meridian | 42.02 ± 1.30 | 42.27 ± 1.28 | +0.24 ± 0.16 | -0.00 to +0.48 | | Power in Steep Meridian | 43.31 ± 1.47 | 43.56 ± 1.45 | +0.25 ± 0.17 | -0.08 to +0.58 | | Mean Power (P1+P2)/2 | 42.67 ± 1.36 | 42.91 ± 1.34 | +0.24 ± 0.11 | +0.02 to +0.46 | | Astigmatic Power | -1.29 ± 0.54 | -1.29 ± 0.54 | -0.01 ± 0.19 | -0.38 to +0.36 | | Axis [°]* | 102.12 ± 82.19 | 100.63 ± 82.31 | 3.82 ± 3.51 | +10.70 | *The upper limit was derived based on the method of Brand and Altman (Bland, JM and Altman, DG. Statistical methods for assessing agreement between two methods of clinical measurement, Lancet, i:307-310, 1986). The lower limit was not provided since the difference in axis is always a positive number. | | Overall<br>Mean | SD | Repeatability<br>Limit | %COV | |-----------------------------------------|-----------------|--------|------------------------|--------| | Javal | | | | | | R1, Radius in Flattest<br>Meridian [mm] | 7.91 | 0.0207 | 0.0581 | 0.26% | | R2, Radius in Steepest<br>Meridian [mm] | 7.67 | 0.0252 | 0.0706 | 0.33% | | P1, Power in Flattest<br>Meridian [D] | 42.02 | 0.1104 | 0.3091 | 0.26% | | P2, Power in Steepest<br>Meridian [D] | 43.31 | 0.1387 | 0.3885 | 0.32% | | Mean Power,<br>(P1 + P2)/2 [D] | 42.67 | 0.0919 | 0.2574 | 0.22% | | Astigmatic Power [D] | -1.29 | 0.1709 | 0.4787 | 13.27% | | Axis [°] | 103.67 | 2.8377 | 7.9456 | 2.74% | | IOLMaster | | | | | | R1, Radius in Flattest<br>Meridian [mm] | 7.86 | 0.0109 | 0.0304 | 0.14% | | R2, Radius in Steepest<br>Meridian [mm] | 7.63 | 0.0180 | 0.0503 | 0.24% | | P1, Power in Flattest<br>Meridian [D] | 42.27 | 0.0587 | 0.1644 | 0.14% | | P2, Power in Steepest<br>Meridian [D] | 43.56 | 0.1023 | 0.2864 | 0.23% | | Mean Power,<br>(P1 + P2)/2 [D] | 42.91 | 0.0583 | 0.1632 | 0.14% | | Astigmatic Power [D] | -1.29 | 0.1192 | 0.3337 | 9.22% | | Axis [°] | 103.18 | 2.4242 | 6.7877 | 2.37% | #### TABLE 4 REPEATABILITY OF JAVAL MANUAL KERATOMETER AND IOLMASTER 500 Repeatability includes variation due to measurement error. Repeatability %COV = (Repeatability SD)/|Overall Mean|* 100% {5}------------------------------------------------ ## CONCLUSION · The IOLMaster 500 is substantially equivalent to the predicate device. Additionally, the keratometry function of the IOLMaster 500 provides cylindrical power and axis measurements comparable to manual keratometry, suitable for use in toric lens power calculations. {6}------------------------------------------------ DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/6/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES-USA" around the perimeter. Inside the circle is a stylized image of an eagle with its wings spread. April 12, 2013 Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 Carl Zeiss Meditec AG % Judith A. Brimacombe, M.A. Director, Clinical/ Regulatory Affairs 5160 Hacienda Drive Dublin, CA 94568 Re: K122418 Trade/Device Name: IOLMaster 500 Regulation Number: 21 CFR 886.1850 Regulation Name: AC-powered slit lamp biomicroscope Regulatory Class: Class II Product Code: HJO Dated: March 12, 2013 Received: March 13, 2013 Dear Ms. Brimacombe: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical {7}------------------------------------------------ Page 2 - Ms. Judith A. Brimacombe, M.A. device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerely yours, # Deborah Likalls -S for Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose, and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {8}------------------------------------------------ # Indications for Use 510(k) Number (if known): K122418 Device Name: IOLMaster 500 Indications For Use: The IOLMaster is intended for the biometric determination of ocular measurements of axial length, anterior chamber depth, corneal radius, white-towhite (WTW), and for the measurement of pupil size and deviation of the visual axis from the center of the pupil. For patients who are candidates for intraocular lens (IOL) implantation, the device also performs calculations to assist physicians in determining the appropriate IOL power and type for implantation. This device is intended for use by physicians and eye-care- professionals and may only be use dunder the supervision of a physician. Prescription Use X (part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use_ (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE- CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Ka N. To 2013.04.09 18:57:39 +04'00' (Division Sign-Off) Division of Ophthalmic and Ear, Nose, and Throat Devices 510(k) Number:_ K122418
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...