K123464 · Carl Zeiss Meditec, AG · NFJ · Mar 5, 2013 · Radiology
Device Facts
Record ID
K123464
Device Name
CALLISTO EYE
Applicant
Carl Zeiss Meditec, AG
Product Code
NFJ · Radiology
Decision Date
Mar 5, 2013
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.2050
Device Class
Class 2
Indications for Use
CALLISTO eye is an assistance system that provides non-diagnostic video documentation and image capture for ophthalmic surgeries. The system allows the remote control of the surgical microscope. The graphical guidance tools, as displayed on the CALLISTO eye Panel PC or microscope eye piece, aid the surgeon to insert, align, position, and register an artificial lens. These tools are intended for anterior segment ophthalmic surgical procedures, including positioning and angular alignment of toric intraocular lenses, limbal relaxing incisions, and capsulorhexis. The system utilizes surgeon information for positioning of graphical guidance tools.
Device Story
CALLISTO eye is an ophthalmic surgical assistance system integrated with OPMI LUMERA 700 surgical microscope. System captures real-time video; processes preoperative patient data to generate graphical guidance templates. Surgeon views templates on panel PC or through microscope eyepiece. Assistance functions include Reference, Rhexis, Incision/LRI, Z-Align, and K-track; these aid in capsulorhexis, incision placement, and toric IOL alignment. System allows remote microscope control and data injection via IDIS. Used in operating room by surgeons and staff. Benefits include improved surgical precision and alignment accuracy for anterior segment procedures. Documentation stored in integrated OR database.
Clinical Evidence
Bench testing only. Verification and validation testing confirmed compliance with performance specifications and customer requirements. No clinical study data provided.
Technological Characteristics
System consists of a camera, computer, and 1920x1080 resolution 2D HD flat panel display. Integrates with OPMI LUMERA 700 surgical microscope. Features include OR database for patient data/video storage, touch-screen interface, and Integrated Data Injection System (IDIS) for eyepiece display. Software version 3.0.
Indications for Use
Indicated for surgeons performing anterior segment ophthalmic surgical procedures, including toric intraocular lens (IOL) positioning, limbal relaxing incisions (LRIs), and capsulorhexis. No specific patient population contraindications stated.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
Predicate Devices
TrueVision® 3D Visualization and Guidance System (K101861)
Submission Summary (Full Text)
{0}------------------------------------------------
K123464
510(k) Premarket Notification CALLISTO eye™
## SECTION 5.
MAR 0 5 2013
510(K) SUMMARY
5. 510(K) SUMMARY
510(k) Summary
(per 21 CFR §807.92)
# CALLISTO eye™
GENERAL INFORMATION
| Manufacturer: | Carl Zeiss Meditec AG<br>Carl Zeiss Strasse 22<br>D-73447 Oberkochen, Germany<br>+49 3641220-667 (phone)<br>+49 3641220-756 (fax)<br>Establishment Registration Number: 9615010 |
|-------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Judith Brimacombe, M.A.<br>Director, Clinical/Regulatory Affairs<br>Carl Zeiss Meditec, Inc.<br>5160 Hacienda Drive<br>Dublin, California 94568<br>(925) 557-4616 (phone)<br>(925) 557-4259 (fax)<br>E-mail: j.brimacombe@meditec.zeiss.com |
| Date prepared: | February 28, 2013 |
| Device | System, Image Management, Ophthalmic<br>Ocular Marker<br>Microscope, Surgical |
| Classification: | 21 CFR 892.2050 |
| Device Class: | II |
| Product Code: | NFJ HMR EPT |
| Common Name: | Picture Archiving and Communications System<br>Ophthalmic Surgical Marker<br>Surgical Microscope and Accessories |
| Trade/Proprietary Name: | CALLISTO eyeTM |
{1}------------------------------------------------
#### 510(k) Premarket Notification CALLISTO eye™
SECTION 5.
### PREDICATE DEVICE:
TrueVision Systems, Inc. Company: TrueVision® 3D Visualization and Guidance System Device: (K101861)
### INDICATIONS FOR USE
CALLISTO eye is an assistance system that provides non-diagnostic video documentation and image capture for ophthalmic surgeries. The system allows the remote control of the surgical microscope.
The graphical guidance tools, as displayed on the CALLISTO eye Panel PC or microscope eye piece, aid the surgeon to insert, align, position, and register an artificial lens. These tools are intended for anterior segment ophthalmic surgical procedures, including positioning and angular alignment of toric intraocular lenses, limbal relaxing incisions, and capsulorhexis. The system utilizes surgeon information for positioning of graphical guidance tools.
### DEVICE DESCRIPTION
CALLISTO eye is an assistance system that processes real-time video images that can be displayed on the CALLISTO eye panel PC for viewing by the surgeon and the surgical staff in the operating room. The same video images can be viewed by the surgeon through the evepiece of the OPMI LUMERA 700 surgical microscope. CALLISTO eve provides enhanced visualization and guidance tools to assist the surgeon during procedures such as limbal relaxing incisions, capsulorhexis, and alignment of toric intraocular lenses (IOL). All treatment templates are based on preoperative clinical data of a particular patient and defined by the surgeon prior to the surgery. These templates can be displayed on the CALLISTO eve panel PC and through the eyepiece of the OPMI LUMERA 700 surgical microscope.
CALLISTO eye consists of two product variants: CALLISTO eye BASIC and CALLISTO eye ASSISTANCE.
CALLISTO eve BASIC displays video images and patient data, stores patient data and video recordings, imports and exports patient data and exports videos and images after the surgery. This product variant also allows the remote control of the OPMI LUMERA 700 surgical microscope as well as data injection through the eyepiece via the Integrated Data Injection System (IDIS).
CALLISTO eye ASSISTANCE provides the graphical assistance tools to aid the surgeon during various ophthalmic procedures including limbal relaxing incisions (LRIs), capsulorhexis and toric IOL positioning. The assistance functions include Reference, Rhexis, Incision/LRI, Z-Align and K-track. These functions aid the surgeon in the opening of the capsulorhexis, making incisions and LRIs, aligning the toric intraocular lenses, and
{2}------------------------------------------------
### SECTION 5.
estimating a local corneal curvature during surgery.
- For CALLISTO eye BASIC and ASSISTANCE, a database called OR database contains the patient data. The OR database is installed on the integrated PC. Documentation of videos and images can then be stored in this database as standard (SD) or high definition (HD) format.
## SUBSTANTIAL EQUIVALENCE
The fundamental technological characteristics of CALLISTO eye with software version 3.0 are similar to the predicate device, TrueVision® 3D Visualization and Guidance System (K101861). Both CALLISTO eye and TrueVision 3D Visualization and Guidance System are devices that work in conjunction with a surgical microscope to assist the surgeon during ophthalmic procedures. Both are used in ophthalmic imaging applications and provide the means for capture, storage, or manipulation of image data and processing of patient information. Both provide the surgical guidance templates and documentation of treatments to aid the surgeon during anterior segment ophthalmic surgical procedures. All templates are set up prior to surgery and can be adjusted during the procedure.
CALLISTO eye and the predicate device consist of:
- . a camera connected to a surgical microscope
- . a computer providing the imaging application and the means to capture, store or manipulate images and video data as well as the methods for handling patient information
- a flat panel display unit with a resolution of 1920 x 1080 pixels to display HD/SD . video content.
All systems provide two dimensional graphical templates, also known as assistance functions, to aid the surgeon during anterior segment ophthalmic surgical procedures, to make an incision, to perform a capsulorhexis, or to position a toric intraocular lens.
One minor difference between CALLISTO eye and predicate device is that the predicate device uses the application of a 3D HD flat panel screen and circular polarized glasses to display live or recorded videos images, whereas CALLISTO eye utilizes a standard 2D HD flat panel to display live or recorded videos and images. Furthermore, CALLISTO eye uses the touch screen as an user interface to omit additional keyboards and pointer devices in the OR setting.
CALLISTO eye and the predicate device serve the same principal purpose of clinical usage and provide a wide range of comparable functionality for their users.
It is the opinion of Carl Zeiss Meditec AG that CALLISTO eye with software version 3.0 is substantially equivalent to the predicate device, TrueVision 3D Visualization and Guidance System (K101861). The indications for use for CALLISTO eye are similar to the indications for use of the predicate device. The technological comparison demonstrates that CALLISTO
{3}------------------------------------------------
#### 510(k) Premarket Notification CALLISTO eye™
### SECTION 5.
eve is functionally equivalent to the predicate device. The minor differences between CALLISTO eve and the predicate device do not raise any new questions of safety or effectiveness in comparison to the predicate device.
### PERFORMANCE DATA
Verification and validation testing were completed to demonstrate that the device performance complies with specifications and requirements identified for the CALLISTO eye. A portion of software verification may be considered "bench testing". Each function and/or feature was tested by means of the appropriate test case or test specification. The system verification test report provides the test cases, expected results for each test case and the actual results obtained. All criteria for this testing were met and the results demonstrate that CALLISTO eye meets all performance specifications and requirements. Validation was also conducted for CALLISTO eye according to the Validation Plan to ensure that the device meets the customer's requirements with respect to performance. The objectives defined in the validation plan were achieved according to the validation results.
### SUMMARY
As described in this 510(k) Summary, all testing deemed necessary was conducted on CALLISTO eye to ensure that the device is as safe and effective as the predicate device.
{4}------------------------------------------------
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Image /page/4/Picture/2 description: The image shows a partial view of a seal or logo, featuring the words "DEPARTMENT OF HEALTH & HUM" arranged vertically along the left side. To the right of the text, there are abstract shapes that appear to be stylized figures or symbols, possibly representing people. The image is in black and white, with the text and shapes rendered in a bold, clear font.
March 5, 2013
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
Ms. Judith Brimacombe, M.A. Director, Clinical/Regulatory Affairs Carl Zeiss Meditec, Inc. 5160 Hacienda Drive Dublin, CA 94568
Re: K123464
Trade/Device Name: CALLISTO eye™ Regulation Number: 21 CFR 892.2050 Regulation Name: Picture Archiving and Communications System Regulatory Class: Class II Product Code: NFJ, HMR, EPT Dated: January 31, 2013 Received: February 1, 2013
Dear Ms. Brimacombe:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21
{5}------------------------------------------------
Page 2 - Ms. Judith Brimacombe, M.A.
CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours.
Kesia Y Alexander -S.
for Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose, and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
# Indications for Use
510(k) Number (if known): K123464
Device Name: CALLISTO eye™
Indications For Use:
CALLISTO eye is an assistance system that provides non-diagnostic video documentation and image capture for ophthalmic surgeries. The system allows the remote control of the surgical microscope.
The graphical guidance tools, as displayed on the CALLISTO eye Panel PC or microscope eye piece, aid the surgeon to insert, align, position, and register an artificial lens. These tools are intended for anterior segment ophthalmic surgical procedures, including positioning and angular alignment of toric intraocular lenses, limbal relaxing incisions, and capsulorhexis. The system utilizes surgeon information for positioning of graphical guidance tools.
Prescription Use______________________________________________________________________________________________________________________________________________________________ (part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE- CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Image /page/6/Picture/11 description: The image shows the text "Ka N. To" on the top line. The second line of text reads "2013.02.28 13:22:09 -05'00'". There is a logo in the background of the text.
(Division Sign-Off) Division of Ophthalmic and Ear, Nose, and Throat Devices 510(k) Number: K123464
Page 1 of 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.