Denudation Pipettes

K250838 · Guangzhou Pinzhi Medical Device Co., Ltd. · MQH · Jul 3, 2025 · Obstetrics/Gynecology

Device Facts

Record IDK250838
Device NameDenudation Pipettes
ApplicantGuangzhou Pinzhi Medical Device Co., Ltd.
Product CodeMQH · Obstetrics/Gynecology
Decision DateJul 3, 2025
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 884.6130
Device ClassClass 2

Indications for Use

Denudation Pipettes are used for the removal of cumulus oocyte complexes (COC) from an oocyte prior to Intracytoplasmic Sperm Injection (ICSI) and In Vitro Fertilization (IVF) as well as for the handling of gametes, embryos and biopsied cells (polar bodies, blastomeres and trophectoderm) during assisted reproductive techniques (ART). Denudation Pipettes are not intended for biopsy of cells from oocytes or embryos.

Device Story

Denudation Pipettes are single-use, sterile, polycarbonate capillary tubes used in assisted reproductive technology (ART) laboratories. The device features a tapered tip with a proximal outer diameter of 0.9mm and a length of 90mm, designed to fit into an actuating device. The distal end is available in various inner diameters (80-290μm). The device is operated by embryologists or trained laboratory technicians to manually manipulate oocytes, embryos, and biopsied cells. By providing a controlled suction/transfer mechanism, the pipette facilitates the removal of cumulus cells and the safe handling of delicate reproductive samples, supporting successful fertilization and embryo culture processes.

Clinical Evidence

Bench testing only. Testing included sterilization validation (ISO 11137), package integrity (ASTM F1886, F1929, F88), shelf-life/accelerated aging (ASTM F1980), simulated transportation (ASTM D4169), one-cell mouse embryo assay (MEA) with ≥80% development to expanded blastocyst at 96 hours, and endotoxin testing (USP <85>) showing <0.05 EU/device.

Technological Characteristics

Polycarbonate capillary tubes; 90mm length; 0.9mm proximal OD; distal ID range 80-290μm. Gamma radiation sterilized (10^-6 SAL). Single-use, disposable. No electronic components or software.

Indications for Use

Indicated for removal of cumulus oocyte complexes from oocytes prior to ICSI and IVF, and for handling gametes, embryos, and biopsied cells (polar bodies, blastomeres, trophectoderm) during ART procedures. Not for biopsy of cells from oocytes or embryos.

Regulatory Classification

Identification

Assisted reproduction microtools are pipettes or other devices used in the laboratory to denude, micromanipulate, hold, or transfer human gametes or embryos for assisted hatching, intracytoplasmic sperm injection (ICSI), or other assisted reproduction methods.

Special Controls

*Classification.* Class II (special controls) (mouse embryo assay information, endotoxin testing, sterilization validation, design specifications, labeling requirements, and clinical testing). The device, when the assisted reproduction microtools (pipettes) are manufactured from glass, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 884.9.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0} FDA U.S. FOOD &amp; DRUG ADMINISTRATION July 3, 2025 Guangzhou PINZHI Medical Device Co., Ltd. % Jie Yang Consultant Chonconn Consulting Co., Ltd. Room 504, Block C No. 1029 Nanhai Avenue, Nanshan District Shenzhen, Guangdong 518067 CHINA Re: K250838 Trade/Device Name: Denudation Pipettes Regulation Number: 21 CFR 884.6130 Regulation Name: Assisted Reproduction Microtools Regulatory Class: II Product Code: MQH Received: June 3, 2025 Dear Jie Yang: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" U.S. Food &amp; Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993 www.fda.gov {1} K250838 - Jie Yang Page 2 (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the {2} K250838 - Jie Yang Page 3 Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Monica D. Garcia -S Monica D. Garcia, Ph.D. Assistant Director DHT3B: Division of Reproductive, Gynecology, and Urology Devices OHT3: Office of Gastrorenal, ObGyn, General Hospital, and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {3} DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below. 510(k) Number (if known) K250838 Device Name Denudation Pipettes Indications for Use (Describe) Denudation Pipettes are used for the removal of cumulus oocyte complexes (COC) from an oocyte prior to Intracytoplasmic Sperm Injection (ICSI) and In Vitro Fertilization (IVF) as well as for the handling of gametes, embryos and biopsied cells (polar bodies, blastomeres and trophectoderm) during assisted reproductive techniques (ART). Denudation Pipettes are not intended for biopsy of cells from oocytes or embryos. Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. "DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW." The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." FORM FDA 3881 (6/20) Page 1 of 1 PSC Publishing Services (301) 443-6740 {4} K250838 Page 1 of 4 510(k) Summary K250838 510(k) Owner Guangzhou PINZHI Medical Device Co., Room 405, 4th Floor, No.600-2 Guangshan 2nd Road, Fenghuang Street, Tianhe District, Guangzhou, P.R.China Phone: (8620)85162357 Email: ivfonly@ivfonly.com Contact person: Hestia Xu Title: Technical Regulation Manager Correspondent Yang Jie, Consultant, Chonconn Consulting Co., Ltd. Room 504, Block C No. 1029 Nanhai Avenue, Nanshan District, Shenzhen, Guangdong 518067 CHN Phone: +86-755 33941160 Email: yangjie@chonconn.com Date Prepared July 2, 2025 Product Information Trade Name Denudation Pipettes Common Name Assisted reproduction micropipette Regulation Name Assisted Reproduction Microtools Regulation Number 884.6130 Regulatory Class Class II Product Code MQH - Microtools, Assisted Reproduction (Pipettes) Predicate Device Research Instruments Ltd. EZ-Tip Vial of 20 K161275 The predicate device has not been subject to a design-related recall. Device Description The Denudation Pipettes are extruded polycarbonate capillary tubes that are pulled at one end to form a tapered tip. They have an outer diameter (OD) of 0.9mm at the proximal end and fit into an actuating device. All tips are approximately 90mm in length and depending on the size of the pipette have a volumetric capacity of 18.6-21.9μl. The Denudation Pipettes are supplied in a range of inner diameter (ID) sizes at the distal end as shown below: - 80μm, 130μm, 140μm, 150μm, 160μm, 170μm, 180μm, 190μm, 210μm, 230μm, 250μm, 270μm, 290μm. {5} K250838 Page 2 of 4 The Denudation Pipettes are radiation sterilized and provided in a sterile pouch, supplied in two packaging options. Each pouch contains 5 or 10 pipettes. All pipettes are intended for single use and disposable. ## Indications for Use Denudation Pipettes are used for the removal of cumulus oocyte complexes (COC) from an oocyte prior to Intracytoplasmic Sperm Injection (ICSI) and In Vitro Fertilization (IVF) as well as for the handling of gametes, embryos and biopsied cells (polar bodies, blastomeres and trophectoderm) during assisted reproductive techniques (ART). Denudation Pipettes are not intended for biopsy of cells from oocytes or embryos. ## COMPARISON OF INTENDED USE AND TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE The table below provides a comparison of the intended use and technological characteristics of the subject device and predicate device: | Characteristic | Subject device Denudation Pipettes K250838 | Predicate device EZ-Tip Vial of 20 K161275 | Comments | | --- | --- | --- | --- | | Indications for use | Denudation Pipettes are used for the removal of cumulus oocyte complexes (COC) from an oocyte prior to Intracytoplasmic Sperm Injection (ICSI) and In Vitro Fertilization (IVF) as well as for the handling of gametes, embryos and biopsied cells (polar bodies, blastomeres and trophectoderm) during assisted reproductive techniques (ART). Denudation Pipettes are not intended for biopsy of cells from oocytes or embryos. | EZ-Tip pipettes are for denudation, i.e. removing the cumulus from an oocyte prior to Intracytoplasmic Sperm Injection (ICSI) and In Vitro Fertilization (IVF) and for handling gametes, embryos and biopsied cells (polar bodies, blastomeres and trophectoderm) during assisted reproductive techniques (ART). EZ Tips are not intended for biopsy of cells from oocytes or embryos. | Same | | Materials | Polycarbonate | Polycarbonate | Same | | Sterility | Sterile (10^{-6}) Gamma Irradiated | Sterile (10^{-6}) Gamma Irradiated | Same | | Proximal Outer Dimension | 900 μm | 900 μm | Same | {6} K250838 Page 3 of 4 | Distal Inner Diameter | Size range 80 μm to 290 μm | Size range 75 μm to 600 μm | Different The subject device is within the cleared range of the predicate device. This difference does not raise different questions of safety or effectiveness. | | --- | --- | --- | --- | | Length | 90mm | 90mm | Same | | Mouse Embryo Assay(MEA) | 1-Cell MEA ≥80% Expanded blastocysts at 96 h | 1-Cell MEA ≥80% expanded blastocysts at 120 h | Different The Mouse Embryo Assay (MEA) assessment time is shorter for the subject device, but it is the standard assessment time for this assay. Therefore, this difference does not raise different questions of safety or effectiveness. | | Endotoxin(LAL) | <0.5 EU/device | <20 EU/device | Different The endotoxin specification for the subject device is lower than the predicate device. This difference does not raise different questions of safety or effectiveness. | | Usage | Single Use | Single Use | Same | As shown in the table above, the subject and predicate devices have same indications for use statements and the intended uses of the predicate and subject devices are the same (i.e., denudation of oocytes and handling of oocytes and embryos during assisted reproduction technology procedures). In regards to technological characteristics, the subject and predicate devices have similarities (e.g., material, length, diameter at the proximal end, sterilization method, etc.); however, as noted in the table above, technological differences were identified between the subject and predicate devices (e.g., distal end tip diameter and specifications for MEA and endotoxin testing). The technological differences between the subject and predicate device do not raise different questions of safety and effectiveness. {7} K250838 Page 4 of 4 # Non-Clinical Performance Testing The following performance tests were performed in support of substantial equivalence: 1. Sterilization Validation per ISO 11137-1:2006 and ISO 11137-2:2013 2. Package Integrity Testing after Simulated transportation (ASTM D4169-23) &amp; Accelerated aging (ASTM F 1980-21): a. Visual test of package integrity (ASTM F1886/F1886M-16), b. Dye penetration (ASTM F1929-23), and c. Seal strength testing (ASTM F88/F88M-23) 3. Shelf Life testing assessing device function at Time 0 and at 36 months: a. Simulated transportation (ASTM D4169-22) b. Accelerated aging (ASTM F 1980-21) c. One-cell mouse embryo assay (MEA): ≥ 80% embryos developed to expanded blastocyst at 96 hours per the 2021 FDA guidance Mouse Embryo Assay for Assisted Reproduction Technology Devices. d. Endotoxin (LAL, USP&lt;85&gt;): &lt;0.05 EU/device e. Device dimension validation f. Appearance &amp; Physical property validation # Conclusion The results of the testing described above demonstrate that the Denudation Pipettes is as safe and effective as the predicate device and supports a determination of substantial equivalence.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...