F6/F9 Fetal & Maternal Monitor is intended for non-invasive and invasive monitoring of fetus during antepartum examination, labor and delivery. It is intended to be used only by trained and qualified personnel in antepartum examination rooms, labor and delivery rooms. F6/F9 Fetal & Maternal Monitor provides Non-Stress testing for pregnant women from the 28th week of gestation. It can externally monitor the FHRs using ultrasound and uterine activity via a TOCO transducer. Alternatively, it can internally monitor one of the FHRs with DECG and uterine activity with an IUPC. F6 Express/F9 Express Fetal & Maternal Monitor is intended for monitoring physiological parameters of pregnant women during antepartum examination, labor and delivery. It is intended to be used only by trained and qualified personnel in antepartum examination rooms, labor and delivery rooms. F6 Express/F9 Express Fetal & Maternal Monitor is intended for providing Non-Stress testing or fetal monitoring for pregnant women from the 28th week of gestation. In addition, it provides a solution for maternal vital signs monitoring.
Device Story
Fetal & Maternal Monitor (F6, F6 Express, F9, F9 Express) performs non-invasive/invasive monitoring of fetal heart rate (FHR) and uterine activity; maternal vital signs (ECG, SpO2, NIBP, temperature) included in Express models. Inputs: ultrasound (FHR), TOCO transducer (uterine activity), DECG/IUPC (internal monitoring), and maternal sensors. Device uses pulse Doppler with autocorrelation for FHR and peak-peak detection for ECG. Output: real-time waveforms and numerical data displayed on LCD; quick printing for stored waveforms. Used in antepartum/labor/delivery rooms by trained clinicians. Data transmission to Central Nursing System supported. Benefits: continuous monitoring of fetal/maternal status during labor and delivery to support clinical decision-making.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing, biocompatibility evaluation (ISO 10993-1), electrical safety (IEC 60601-1), and electromagnetic compatibility (IEC 60601-1-2). Acoustic output measured per NEMA UD-2 and IEC 62359.
Technological Characteristics
LCD display; rechargeable Lithium-ion battery; AC/DC power. Sensing: Pulse Doppler (1.0MHz) for FHR, TOCO for uterine activity, ECG/SpO2/NIBP/Temp sensors for maternal monitoring. Connectivity: Wired/wireless (FTS-3) to Central Nursing System. Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-2-27, IEC 60601-2-37, ISO 80601-2-61, ISO 80601-2-56. Software: Moderate level of concern.
Indications for Use
Indicated for pregnant women from the 28th week of gestation for non-invasive and invasive fetal monitoring (FHR, uterine activity) and maternal vital signs monitoring (ECG, SpO2, NIBP, temperature) during antepartum examination, labor, and delivery. Intended for use by trained personnel in clinical settings.
Regulatory Classification
Identification
A perinatal monitoring system is a device used to show graphically the relationship between maternal labor and the fetal heart rate by means of combining and coordinating uterine contraction and fetal heart monitors with appropriate displays of the well-being of the fetus during pregnancy, labor, and delivery. This generic type of device may include any of the devices subject to §§ 884.2600, 884.2640, 884.2660, 884.2675, 884.2700, and 884.2720. This generic type of device may include the following accessories: Central monitoring system and remote repeaters, signal analysis and display equipment, patient and equipment supports, and component parts.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
July 1, 2015
Edan Instruments, Inc. % Doug Worth Sr. Dir. Us RA/QA Edan Medical 1200 Crossman Ave, Suite 200 Sunnyvale, California 94089
Re: K150901
Trade/Device Name: Fetal & Maternal Monitor, models F6, F6 Express Regulation Number: 21 CFR 884.2740 Regulation Name: Perinatal monitoring system and accessories Regulatory Class: Class II Product Code: HGM Dated: March 27, 2015 Received: April 3, 2015
Dear Mr. Worth:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device
{1}------------------------------------------------
related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Benjamin R. Fisher -S
Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K150901
Device Name
Fetal & Maternal Monitor, models F6, F6 Express, F9, F9 Express
Indications for Use (Describe)
F6/F9 Fetal & Maternal Monitor (hereinafter called F6/F9):
F6/F9 Fetal & Maternal Monitor is intended for non-invasive monitoring of fetus during antepartum examination, labor and delivery. It is intended to be used only by trained and qualified personnel in antepartum examination rooms, labor and delivery rooms.
F6/F9 Fetal & Maternal Monitor provides Non-Stress testing for pregnant women from the 28th week of gestation. It can externally monitor the FHRs using ultrasound and uterine activity via a TOCO transducer. Alternally monitor one of the FHRs with DECG and uterine activity with an IUPC.
F6 Express/F9 Express Fetal & Maternal Monitor (hereinafter called F6 Express/F9 Express):
F6 Express/F9 Express Fetal & Maternal Monitor is intended for monitoring physiological parameters of pregnant women during antepartum examination, labor and delivery. It is intended to be used only by trained and qualified personnel in antepartum examination rooms, labor and delivery rooms.
F6 Express/F9 Express Fetal & Maternal Monitor is intended for providing Non-Stress testing or fetal monitoring for pregnant women from the 28th week of gestation. In addition, it provides a solution for maternal vital signs monitoring.
Type of Use (Select one or both, as applicable)
| Prescription Use (Part 21 CER 801 Subpart D) | Over-The-Counter Use (21 CER 801 Subpart C) |
|----------------------------------------------|---------------------------------------------|
|----------------------------------------------|---------------------------------------------|
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(K) Summary
# Prepared in accordance with the content and format regulatory requirements of 21 CFR Part 807.92
| 1. Submitter: | Edan Instruments, Inc<br>3/F - B, Nanshan Medical<br>Equipments Park, Nanhai Rd 1019#,<br>Shekou, Nanshan Shenzhen,<br>518067 P.R. China<br>Tel: +86(0755) 26858736<br>Fax: +1 (408) 418-4059 |
|-----------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact person:<br>Preparing date: | Queena Chen<br>June 29, 2015 |
| 2. Device name and<br>classification: | Device Name: Fetal & Maternal Monitor, models F6, F6 Express,<br>F9, F9 Express (hereinafter called F series)<br>Classification Name/ Product code:<br>884.2740 Perinatal monitoring system and accessories/ HGM<br>Regulatory Class: Class II |
| 3.Premarket<br>Notification Class III<br>Certification and<br>Summary | Not applicable, the subject device is Class II. |
| 4. Predicate Device(s): | 1) Edan Instruments, Inc., Fetal & Maternal Monitor/F9 Express<br>/K100797.<br>2) Philips Medical Systems, Inc./Avalon CTS cordless Fetal<br>Transducer System/K023931 |
| 5. Reason for<br>Submission | Modification for previous cleared product Fetal & Maternal<br>Monitor, models F6, F6 Express, F9, F9 Express. |
| 6. Pre-Submission,<br>IDE | Not applicable, there is no prior submission. |
| 7. Device Description: | F series Fetal & Maternal Monitor, including F6, F6 Express,<br>F9, F9 Express which provides the following primary features<br>that can be available for the multiple configurations:<br>• Basic parameters: FHR, TOCO, Event Mark, AFM<br>• Dual FHR monitoring<br>• Internal parameters: IUP/DECG<br>• FHR limit alarm |
{4}------------------------------------------------
• Software for data transmission to PC, which means the subject device can be connected to Central Nursing System provided by EDAN, which has been cleared by K100358.
- Quick printing for stored waveform
- Lithium battery for continuous working
- · Maternal ECG monitoring
- · Maternal SpO2 monitoring
- · Maternal NIBP
- · Maternal temperature monitoring
- Used with FTS-3 Fetal Telemetry System (hereinafter called FTS-3)
The above is the maximum configuration for F series Fetal & Maternal Monitor, the user may select different measurement parameters in according with their requirements, which is also show in the following table.
| Model | F6 | F9 | F6<br>Express | F9<br>Express |
|-------------------------------------------------------|-----|-----|---------------|---------------|
| Measurement | | | | |
| Single-FHR | √ | √ | √ | √ |
| Dual-FHR | √ | √ | √ | √ |
| TOCO | √ | √ | √ | √ |
| FM | √ | √ | √ | √ |
| AFM | √ | √ | √ | √ |
| DECG/IUP | Opt | Opt | × | Opt |
| MECG | × | × | √ | √ |
| NIBP | × | × | √ | √ |
| MSpO2 | × | × | √ | √ |
| TEMP | × | × | √ | √ |
| NOTE: √ = Standard; Opt = Optional; × = Not Available | | | | |
#### 8. Indications for Use:
F6/F9 Fetal & Maternal Monitor (hereinafter called F6/F9): F6/F9 Fetal & Maternal Monitor is intended for non-invasive and invasive monitoring of fetus during antepartum examination, labor and delivery. It is intended to be used only by trained and qualified personnel in antepartum examination rooms, labor and delivery rooms.
F6/F9 Fetal & Maternal Monitor provides Non-Stress testing for pregnant women from the 28th week of gestation. It can externally monitor the FHRs using ultrasound and uterine activity via a TOCO transducer. Alternatively, it can internally monitor one of the FHRs with DECG and uterine activity with an IUPC.
{5}------------------------------------------------
F6 Express/F9 Express Fetal & Maternal Monitor (hereinafter called F6 Express/F9 Express):
F6 Express/F9 Express Fetal & Maternal Monitor is intended for monitoring physiological parameters of pregnant women during antepartum examination, labor and delivery. It is intended to be used only by trained and qualified personnel in antepartum examination rooms, labor and delivery rooms.
F6 Express/F9 Express Fetal & Maternal Monitor is intended for providing Non-Stress testing or fetal monitoring for pregnant women from the 28th week of gestation. In addition, it provides a solution for maternal vital signs monitoring.
#### 9. Predicate Device Comparison
The subject devices share the same characteristics in all items with the predicate device, concluding from using the same technology and principle. All the technological differences existed between the subject and predicate devices are only some performance parameters items, detailed comparison information can be found in the following tables.
| Item | F6 / F6 Express/ F9 / F9 Express | F9 Express | Substantial Equivalence (SE) |
|----------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------|
| Manufacturer | Edan Instruments, Inc. | Edan Instruments, Inc. | / |
| K# | N/A | K100797 | / |
| Intended use | The F9 Express fetal & maternal monitor is intended for monitoring physiological parameters of pregnant women during antepartum examination, labor and delivery. It is intended to be used only by trained and qualified personnel in antepartum examination rooms, labor and delivery rooms. It is not intended for use in intensive care units, operating rooms or for home use. F9 Express fetal & maternal monitor is intended for providing | The F9 Express fetal & maternal monitor is intended for monitoring physiological parameters of pregnant women during antepartum examination, labor and delivery. It is intended to be used only by trained and qualified personnel in antepartum examination rooms, labor and delivery rooms. It is not intended for use in intensive care units, operating rooms or for home use. F9 Express fetal & maternal monitor is intended for providing | Same |
| | Non-Stress testing or fetal<br>monitoring for pregnant<br>women .In addition, it<br>provides a solution for<br>maternal vital signs<br>monitoring. | Non-Stress testing or fetal<br>monitoring for pregnant<br>women. In addition, it<br>provides a solution for<br>maternal vital signs<br>monitoring. | |
| Electrical Safety | | | |
| Anti-electric-shock<br>degree: | FHR1, FHR2, TOCO,<br>FM, IUP: BF;<br>SpO2, NIBP: BF<br>(Defibrillator-proof);<br>DECG:CF;<br>ECG, TEMP: CF<br>(Defibrillator-proof);<br>FTS-3: FHR1, FHR2,<br>TOCO: BF<br>(Defibrillator-proof). | FHR1, FHR2, TOCO, FM,<br>IUP: BF;<br>SpO2, NIBP: BF<br>(Defibrillator-proof);<br>DECG: CF;<br>ECG, TEMP: CF<br>(Defibrillator-proof). | Different |
| Degree of safety of<br>application in the<br>presence of<br>flammable gas: | Equipment not suitable<br>for use in presence of<br>flammable gases | Equipment not suitable for<br>use in presence of<br>flammable gases | Same |
| Working mode: | Continuous running<br>equipment | Continuous running<br>equipment | Same |
| FHR | | | |
| Technique: | Pulse Doppler with<br>autocorrelation processing | Pulse Doppler with<br>autocorrelation processing | Same |
| Pulse Repetition<br>Frequency: | 2 KHz | 2 KHz | Same |
| Pulse Duration: | 92 μs | 92 μs | Same |
| Ultrasound<br>Frequency: | 1.0MHz±10% | 1.0MHz±10% | Same |
| Spatial-Peak<br>Temporal Average<br>Intensity: | Ispta < 100 mW/cm2 | Ispta < 100 mW/cm2 | Same |
| FHR Range: | 50 bpm-240 bpm | 50 bpm-240 bpm | Same |
| Resolution: | 1 bpm | 1 bpm | Same |
| Accuracy: | ±1 bpm(F9)<br>±2 bpm(F6,FTS-3) | ±1 bpm(F9)<br>±2 bpm(F6) | Same |
| TOCO | | | |
| TOCO Range: | 0%-100% | 0%-100% | Same |
| Resolution: | 1%/ | 1%/ | Same |
| ECG | | | |
| Technique: | Peak-peak detection<br>technique | Peak-peak detection<br>technique | Same |
| Heart Rate<br>Counting Range: | 30 bpm ~ 240 bpm | 30 bpm ~ 240 bpm | Same |
| Heart Rate<br>Resolution: | ±1 BPM | ±1 BPM | Same |
| Input Impedance: | >10ΜΩ (Differential,)<br>>20ΜΩ (Common Mode) | >10ΜΩ (Differential,)<br>>20ΜΩ (Common Mode) | Same |
| Countable Input<br>Signal Range: | 20uV-3mV peak-to-peak | 20uV-3mV peak-to-peak | Same |
| Pressure Range: | 0~100mmHg | 0~100mmHg | Same |
| Sensitivity: | 5uV/V/mmHg | 5uV/V/mmHg | Same |
| Resolution: | 1% | 1% | Same |
| Zero Mode: | Automatic / Manual | Automatic / Manual | Same |
| Heart Rate Range | 30-240BPM | 30-240BPM | Same |
| Heart Rate<br>Resolution | ±1 BPM | ±1 BPM | Same |
| Input Signal Range | ±8 mV PP | ±8 mV PP | Same |
| Defibrillator<br>Protection | YES | YES | Same |
| Tall T-wave<br>Rejection | 1.2mV(12mm) | 1.2mV(12mm) | Same |
| HR averaging<br>method | Heart rate is computed by<br>averaging the 12 most<br>recent RR intervals. | Heart rate is computed by<br>averaging the 12 most<br>recent RR intervals. | Same |
| Accuracy and<br>Response to<br>Irregular Rhythm | Ventricular bigeminy:<br>80bpm±1bpm<br>Slow alternating<br>ventricular bigeminy:<br>60bpm±1bpm<br>Rapid alternating<br>ventricular bigeminy:<br>120bpm±1bpm<br>Bidirectional systoles:<br>91bpm±1bpm | Ventricular bigeminy:<br>80bpm±1bpm<br>Slow alternating<br>ventricular bigeminy:<br>60bpm±1bpm<br>Rapid alternating<br>ventricular bigeminy:<br>120bpm±1bpm<br>Bidirectional systoles:<br>91bpm±1bpm | Same |
| Response time to<br>Change in HR | HR range: 80bpm ~<br>120bpm<br>Range: 7s ~ 8s (average:<br>7.5s)<br>HR range: 80bpm ~<br>40bpm<br>Range : 7s ~ 8s (average:<br>7.5s) | HR range: 80bpm ~<br>120bpm<br>Range: 7s ~ 8s (average:<br>7.5s)<br>HR range: 80bpm ~ 40bpm<br>Range : 7s ~ 8s (average:<br>7.5s) | Same |
| SpO2 | | | |
| Measuring Range: | 50% ~ 100% | 50%~100% | Same |
| Resolution: | 1% | 1% | Same |
| | | | |
| Measuring<br>Accuracy<br>(EDAN): | $90\% \sim 100\% \pm 2\%$<br>$70\% \sim 90\% \pm 4\%$<br>$< 70\%$<br>unspecified | $90\% \sim 100\% \pm 2\%$<br>$70\% \sim 90\% \pm 4\%$<br>$< 70\%$<br>unspecified | Same |
| Measuring<br>Accuracy<br>(Nellcor): | $70\% \sim 100\% \pm 2\%$<br>$< 70\%$<br>unspecified | $70\%~100\% \pm 2\%$<br>$< 70\%$<br>unspecified | Same |
| Pulse Rate<br>Measurement<br>Range: | 30-240BPM | 30-240BPM | Same |
| Pulse Rate<br>Measuring<br>Accuracy: | $\pm$ ulse | $\pm$ ulse | Same |
| Emitted light<br>energy | < 15 mW | < 15 mW | Same |
| NIBP | | | |
| Blood Pressure<br>Range | Systolic<br>pressure:40mmHg ~<br>270mmHg Diastolic<br>pressure:10mmHg ~<br>215mmHg | Systolic pressure:40mmHg<br>~ 270mmHg Diastolic<br>pressure:10mmHg ~<br>215mmHg | Same |
| Measuring<br>Accuracy: | $\le 8mmHg$ | $\le 8mmHg$ | Same |
| Cuff Pressure<br>Protection | 300mmHg | 300mmHg | Same |
| TEMP | | | |
| Channel: | 1 | 1 | Same |
| Measurement<br>Range: | 0 °C ~ 50 °C | 0 °C ~ 50 °C | Same |
| Accuracy: | 0°C ~ +25°C (+32°F ~<br>+77°F): $\pm$ 0.2°C<br>( $\pm$ 0.36°F)<br>+25°C ~ +45°C (+77°F ~<br>+113°F): $\pm$ 0.1°C<br>( $\pm$ 0.18°F)<br>+45°C ~ +50°C (+113°F ~<br>+122°F): $\pm$ 0.2°C<br>( $\pm$ 0.36°F) | 0°C ~ +25°C (+32°F ~<br>+77°F): $\pm$ 0.2°C<br>( $\pm$ 0.36°F)<br>+25°C ~ +45°C (+77°F ~<br>+113°F): $\pm$ 0.1°C<br>( $\pm$ 0.18°F)<br>+45°C ~ +50°C (+113°F ~<br>+122°F): $\pm$ 0.2°C<br>( $\pm$ 0.36°F) | Same |
| Printer | | | |
| Paper width: | 152mm (GE), 150mm<br>(PHILIPS) | 152mm (GE), 150mm<br>(PHILIPS) | Same |
| Effective printing<br>width: | 110mm (American<br>Standard)<br>120mm (International | 110mm (American<br>Standard)<br>120mm (International | Same |
#### Comparison between the subject F9 series and the previous cleared F9 Express
{6}------------------------------------------------
{7}------------------------------------------------
{8}------------------------------------------------
{9}------------------------------------------------
| | Standard) | Standard) | |
|-------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------|------|
| FHR printout<br>width: | 70mm (American<br>Standard)<br>80mm (International<br>Standard) | 70mm (American<br>Standard)<br>80mm (International<br>Standard) | Same |
| FHR scaling: | 30bpm/cm (American<br>Standard)<br>20bpm/cm (International<br>Standard) | 30bpm/cm (American<br>Standard)<br>20bpm/cm (International<br>Standard) | Same |
| TOCO printout<br>width: | 40mm | 40mm | Same |
| TOCO scaling: | 25%/cm | 25%/cm | Same |
| Standard Speed<br>(Real-Time<br>Traces ): | 1 cm/min, 2 cm/min, 3<br>cm/min | 1 cm/min, 2 cm/min, 3<br>cm/min | Same |
| Paper: | Z-fold, thermosensitive<br>(compatible with GE and<br>PHILIPS record papers) | Z-fold, thermosensitive<br>(compatible with GE and<br>PHILIPS record papers) | Same |
| <b>Physical Specification</b> | | | |
| Screen | LCD | LCD | Same |
| Screen Diagonal: | F9 Express / F9 :<br>12.1"<br>F6 Express / F6 :<br>10.1" | F9 Express / F9 :<br>12.1"<br>F6 Express / F6 :<br>10.1" | Same |
| Power Supply: | AC or battery | AC or battery | Same |
| Operating Voltage: | a.c.100 V-240 V | a.c.100 V-240 V | Same |
| Line Frequency: | 50/60 Hz | 50/60 Hz | Same |
| Pmax : | 1.0A-0.5A | 1.0A-0.5A | Same |
| Battery : | Rechargeable Lithium-ion<br>Battery | Rechargeable Lithium-ion<br>Battery | Same |
| Dimensions: | 347mm × 330mm ×<br>126mm | 347mm × 330mm ×<br>126mm | Same |
| Weight: | F6: Approx. 5.3 kg<br>F6 Express: Approx. 6.1<br>kg<br>F9: Approx. 5.5 kg<br>F9 Express: Approx. 6.3<br>kg<br>FTS-3:<br>1.8 kg | F6: Approx. 5.3 kg<br>F6 Express: Approx. 6.1<br>kg<br>F9: Approx. 5.5 kg<br>F9 Express: Approx. 6.3<br>kg | Same |
| Operating<br>Temperature: | 5 °C ~ 40 °C | 5 °C ~ 40 °C | Same |
{10}------------------------------------------------
| Transport/ Storage<br>Temperature: | -20 °C ~ 55 °C | -20 °C ~ 55 °C | |
|-----------------------------------------------|-------------------------------------------------…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.