FETAL MONITORS, MODELS F2 AND F3

K102140 · Edan Instruments, Inc. · HGM · Jan 13, 2011 · Obstetrics/Gynecology

Device Facts

Record IDK102140
Device NameFETAL MONITORS, MODELS F2 AND F3
ApplicantEdan Instruments, Inc.
Product CodeHGM · Obstetrics/Gynecology
Decision DateJan 13, 2011
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 884.2740
Device ClassClass 2

Intended Use

The F2 and F3 Fetal Monitor are intended for non-invasive and invasive monitoring of fetus during antepartum examination, labor and delivery. They are intended to be used only by trained and qualified personnel in antepartum examination rooms, labor and delivery rooms. They both provide Non-stress testing for pregnant women from the 28th week of gestation. They can externally monitor the FHRs using ultrasound and uterine activity via a TOCO transducer. Alternatively, they can internally monitor one of the FHRs with DECG and uterine activity with an IUPC.

Device Story

Fetal monitor provides non-invasive and invasive monitoring of fetal heart rate (FHR) and uterine activity. Inputs: ultrasound Doppler signals (via external transducer), TOCO transducer signals (external uterine activity), direct fetal ECG (DECG) signals, and intrauterine pressure catheter (IUPC) signals. Device processes these inputs to display real-time fetal and maternal data. Used in antepartum examination rooms, labor, and delivery rooms by trained clinical personnel. Connectivity: RJ45 interface for central monitoring system; DB9 interface for wireless network module. Output: visual monitoring data for clinical assessment of fetal well-being during labor and delivery. Benefits: enables continuous fetal surveillance to support clinical decision-making.

Clinical Evidence

No clinical data provided. Substantial equivalence supported by bench testing, including software, hardware, safety, and environmental testing, alongside risk analysis and final validation.

Technological Characteristics

Non-invasive ultrasound Doppler, external TOCO, and direct fetal ECG (DECG) sensing. Connectivity via RJ45 (wired) and DB9 (wireless module). Form factor: Fetal monitor console. Software-based signal processing. Sterilization: Not specified.

Indications for Use

Indicated for non-invasive and invasive fetal monitoring during antepartum examination, labor, and delivery in pregnant women from the 28th week of gestation. Contraindicated for use in intensive care units, operating rooms, or home use.

Regulatory Classification

Identification

A perinatal monitoring system is a device used to show graphically the relationship between maternal labor and the fetal heart rate by means of combining and coordinating uterine contraction and fetal heart monitors with appropriate displays of the well-being of the fetus during pregnancy, labor, and delivery. This generic type of device may include any of the devices subject to §§ 884.2600, 884.2640, 884.2660, 884.2675, 884.2700, and 884.2720. This generic type of device may include the following accessories: Central monitoring system and remote repeaters, signal analysis and display equipment, patient and equipment supports, and component parts.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ F2, F3 Fetal Monitors TRADITIONAL 510(K) SUBMISSION Section 1 # 510(k) Summary of Safety and Effectiveness JAN 1 3 201 This summary of 510k safety and effectiveness is being submitted in according with 21CFR part 807.92 Submitter: Edan Instruments, Inc 3/F - B, Nanshan Medical Equipments Park, Nanhai Rd 1019#, shekou, Nanshan Shenzhen, 518067 P.R. China Tel: 86-755-26882220. . . . . . . . . . . . . Fax:86-755-26882223 Contact person: Yue Qiuhong Official correspondent: William Stern Multigon Industries, Inc. 1 Odell Plaza Yonkers, N.Y. 10701 Phone: 914 376 5200 X27 Fax: 914 376 6111 Date of Preparation: 2010-7-22 Proprietary Name: F2, F3 Fetal Monitor (Models: F2, F3) Classification Name: System, Monitoring, Perinatal Product code: HGM, HGL Predicate Devices: | Predicate devices | CADENCE, CADENCE DUAL,<br>CADENCE II FETAL MONITOR | |-------------------|----------------------------------------------------| | Manufacturer | Edan Instruments, Inc | | K # | K082369 | Device Description: With non-invasive ultrasound Doppler, external TOCO and direct fetal ECG technique, Fetal Monitor provides Non-Stress testing for pregnant women from the 28th week of gestation. You can externally monitor the fetal heart rate using ultrasound and uterine activity via a TOCO transducer. Alternatively, you can internally monitor one of the fetal heart rate with direct fetal ECG technique and uterine activity with an Intrauterine Pressure Catheter. {1}------------------------------------------------ ## F2, F3 Fetal Monitors TRADITIONAL 510(K) SUBMISSION · Section 1 The Fetal Monitor has six models: Cadence, Cadence DUAL, Cadence PRO, Cadence II, F2 and F3. The former four models Cadence, Cadence DUAL, Cadence PRO, Cadence II have already been cleared under K082369, while the other two models F2, F3 are new models of Fetal Monitor in this Traditional 510(K) application, which have the same intended use and constructions to the existing models. The Fetal monitor can be connected with Central Monitoring System via RJ45 interface. Also it can be connected to wireless network module via a DB9 interface, and the wireless network module will complete the data switch of the monitor and the Obstetrical Central Monitoring System. ### Comparison with predicate device . The F2, F3 fetal monitors have the same device characteristics as the predicate approved device cleared under K082369. Both models use the same technology and circuitry as the already approved device models Cadence, Cadence DUAL, Cadence PRO, Cadence II cleared under K082369. Hence the models F2, F3 above are substantially equivalent to the predicate devices cited. #### Intended Use The F2, F3 Fetal Monitor are intended for non-invasive and invasive monitoring of fetus during antepartum examination, labor and delivery. They are intended to be used only by trained and qualified personnel in antepartum examination rooms, labor and delivery rooms. They both provide Non-stress testing for pregnant women from the 28th week of gestation. They can externally monitor the FHRs using ultrasound and uterine activity via a TOCO transducer. Alternatively, they can internally monitor one of the FHRs with DECG and uterine activity with an IUPC. Contraindications: It is not intended for use in intensive care units, operating rooms or for home use. Test Summary: The following quality assurance measures were applied to the development of the F2, F3 Fetal Monitors. - · Software testing - · Hardware testing - · Safety testing - · Environment test - · Risk analysis · · .. - · Final validation #### Conclusion: Verification and validation testing was done on the Fetal Monitor models F2, F3. This premarket notification submission demonstrates 5/22 {2}------------------------------------------------ , KIDZL 40 , કર્ભ રે page ## F2, F3 Fetal Monitors TRADITIONAL 510(K) SUBMISSION Section 1 that Fetal Monitor models F2, F3 is substantially equivalent to the predicate device. {3}------------------------------------------------ # DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/3/Picture/1 description: The image shows the logo for the Department of Health & Human Services USA. The logo features a stylized image of an eagle with its wings spread, enclosed within a circular border. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged around the border of the circle. Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002 Mr. Yue Qiuhong Certification Engineer Edan Instruments, Inc. 3/F-B, Nanshan Medical Equipment Park Nanhai Road 1019# Shenzhen, Guangdong, 518067 CHINA APR 2 6 2012 K102140 Re: Trade/Device Name: Fetal Monitor model F2 and F3 Regulation Number: 21 CFR § 884.2740 Regulation Name: Perinatal monitoring system and accessories Regulatory Class: II Product Code: HGM, HGL Dated: December 20, 2010 Received: December 27, 2010 Dear Mr. Oiuhong: This letter corrects our substantially equivalent letter of January 13, 2011. We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. This determination of substantial equivalence applies to the following transducers intended for use with the Fetal Monitor model F2 and F3, as described in your premarket notification: ## MS3-109301 If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found {4}------------------------------------------------ Page 2 - Mr. Yue Oiuhong in the Code of Federal Regulations, Title 21, Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus permits your device to proceed to market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. If you have any questions regarding the content of this letter, please contact Ms. Kathryn Daws- Kopp at (301) 796-6535. は Sincerely yours, t Herbert Hoover Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health ## Enclosures {5}------------------------------------------------ F2, F3 Fetal Monitors TRADITIONAL 510(K) SUBMISSION Section 1 ## Indication for Use 510(k) Number (if.known): K102140 Device Name: F2, F3 Fetal Monitors (Models: F2, F3) The F2 and F3 Fetal Monitor are intended for non-invasive and invasive monitoring of fetus during antepartum examination, labor and delivery. They are intended to be used only by trained and qualified personnel in antepartum examination rooms, labor and delivery rooms. They both provide Non-stress testing for pregnant women from the 28th week of gestation. They can externally monitor the FHRs using ultrasound and uterine activity via a TOCO transducer. Alternatively, they can internally monitor one of the FHRs with DECG and uterine activity with an IUPC. × Prescription Use (21 CFR Part 801 Subpart D) And/Or Over the Counter Use _ (21 CFR Part 801 Subpart C) Concurrence of CDRH, Office of Device Evaluation (ODE) ivision Sign of Reproductive. Gastro-R {6}------------------------------------------------ ## Diagnostic Ultrasound indications for Use Form Fill out one form for each ultrasound system and each transducer. 1MHz PW fetal probe Intended use : Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows | Clinical Application | Mode Operation | | | | | | | | | | | |----------------------------------|----------------|---|---|-----|-----|------------------|----------------------|------------------------------|-----------------------|--------------------|--| | | A | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(specify) | Other<br>(specify) | | | Ophthalmic | | | | | | | | | | | | | Fetal | | | | P | | | | | | | | | Abdominal | | | | | | | | | | | | | Intraoperative(specify) | | | | | | | | | | | | | Intraoperative Neurological | | | | | | | | | | | | | Pediatric | | | | | | | | | | | | | Small Organ(specify) | | | | | | | | | | | | | Neonatal Cephalic | | | | | | | | | | | | | Cardiac | | | | | | | | | | | | | Transesophageal | | | | | | | | | | | | | Transrectal | | | | | | | | | | | | | Transvaginal | | | | | | | | | | | | | Transurethral | | | | | | | | | | | | | Intravascular | | | | | | | | | | | | | Peripheraln Vascular | | | | | | | | | | | | | Laparoscopic | | | | | | | | | | | | | Musculo-skeletal<br>Conventional | | | | | | | | | | | | | Musculo-skeletal Superficial | | | | | | | | | | | | | Other | | | | | | | | | | | | ndication; P=previously cleared by FDA; e=ADDED UNDER appendix E Additional Comments: _________________________________________________________________________________________________________________________________________________________ The above is a 1MHz PW transducer for the fetal heart rate detection PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE I PNEEDED CONCURRENCE OF cdrh, Office of Device Evaluation (ODE) Tracy r. 13, 2012 Section 1
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...