K082602 · Edan Instruments, Inc. · HGM · Nov 26, 2008 · Obstetrics/Gynecology
Device Facts
Record ID
K082602
Device Name
FETAL & MATERNAL MONITOR, MODEL F6, F9
Applicant
Edan Instruments, Inc.
Product Code
HGM · Obstetrics/Gynecology
Decision Date
Nov 26, 2008
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 884.2740
Device Class
Class 2
Indications for Use
The Fetal & Maternal Monitor is intended for non-invasive and invasive monitoring of fetus during antepartum examination, labor and delivery. It is intended to be used only by trained and qualified personnel in antepartum examination rooms, labor and delivery rooms. Fetal & Maternal Monitor provides Non-Stress testing for pregnant women from the 28th week of gestation. It can externally monitor the FHRs using ultrasound and uterine activity via a TOCO transducer. Alternatively, it can internally monitor one of the FHRs with DECG and uterine activity with an IUPC.
Device Story
Fetal & Maternal Monitor (Models F6/F9) performs non-invasive/invasive fetal monitoring; inputs include ultrasound signals (FHR), TOCO transducer data (uterine activity), DECG, and IUPC. Device processes signals to provide real-time waveform display, FHR limit alarms, and data storage (12-hour playback). Used in antepartum/labor/delivery rooms by trained clinicians. Features include optional wireless module, fetal stimulator, and PC data transmission. Output assists clinicians in assessing fetal status during labor and delivery; supports Non-Stress testing. Benefits include continuous fetal heart rate and uterine activity monitoring to guide clinical management.
Clinical Evidence
No clinical data provided. Substantial equivalence established via bench testing, including software, hardware, safety, and environmental testing, alongside risk analysis and final validation.
Technological Characteristics
Perinatal monitoring system; 1 MHz PW fetal ultrasound transducers; TOCO, DECG, and IUPC sensors. Features lithium battery, wireless module, and PC connectivity. Software-based waveform processing and alarm management. Class II device.
Indications for Use
Indicated for pregnant women from the 28th week of gestation for non-invasive and invasive fetal monitoring during antepartum examination, labor, and delivery. Contraindicated for use in intensive care units, operating rooms, or home use.
Regulatory Classification
Identification
A perinatal monitoring system is a device used to show graphically the relationship between maternal labor and the fetal heart rate by means of combining and coordinating uterine contraction and fetal heart monitors with appropriate displays of the well-being of the fetus during pregnancy, labor, and delivery. This generic type of device may include any of the devices subject to §§ 884.2600, 884.2640, 884.2660, 884.2675, 884.2700, and 884.2720. This generic type of device may include the following accessories: Central monitoring system and remote repeaters, signal analysis and display equipment, patient and equipment supports, and component parts.
Corometrics 120 series maternal/fetal monitor (K991739)
Submission Summary (Full Text)
{0}------------------------------------------------
#### NOV 2 6 2008 510(K)Summary of Safety and Effectiveness
This summary of 510k safety and effectiveness is being submitted in according with 21CFR part 807.92
| Submitter: | Edan Instruments, Inc<br>3/F - B, Nanshan Medical<br>Equipments Park, Nanhai Rd 1019#,<br>shekou, Nanshan Shenzhen,<br>518067 P.R. China<br>Tel: 86-755-26882220<br>Fax:86-755-26882223<br>Responsible person: Jiang Yucai |
|-------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Official correspondent: | William Stern<br>Multigon Industries, Inc.<br>1 Odell Plaza<br>Yonkers, N.Y. 10701<br>Phone: 914 376 5200 X27<br>Fax: 914 376 6111 |
| Date of Preparation: | Jul 17, 2008 |
| Proprietary Name: | Fetal & Maternal Monitor (Models F6 and F9) |
| Classification Name: | 21 CFR 884.2740 Perinatal monitoring system and accessories |
| Product code: | HGM |
| Predicate Devices: | |
| Predicate devices | CADENCE II | Corometrics 120 series<br>maternal/ fetal monitor |
|-------------------|------------------------|---------------------------------------------------|
| Manufacturer | Edan Instruments, Inc. | GE Healthcare |
| K # | K073221 | K991739 |
Device Description:
The Fetal & Maternal Monitor provides the following primary features that can be available for the multiple configurations:
- Basic parameters: FHR, TOCO, Event Mark, AFM
- · Optional Dual FHR monitoring
- · Optional Internal parameters: IUP/DECG
{1}------------------------------------------------
#### Fetal & maternal monitor 510K Submission
- 12 hours waveforms playback
- · Optional Built-in wireless module
- · Software for data transmission to PC
- · Quick printing for stored waveform
- · Lithium battery for 4 hours continuous working
- Probe rack and wall mounting rack
- Optional Fetal Stimulator
- FHR limit alarm
- Pulse wave Doppler waterproof transducer for FHR detection
#### Comparison with predicate device
| Monitoring Mode | Fetal & Maternal<br>Monitor | CADENCE II | Corometrics 120 series<br>maternal/ fetal monitor |
|--------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------|---------------------------------------------------|
| FHR/Dual FHR | yes | yes | yes |
| TOCO | yes | yes | yes |
| Fetal ECG, IUP | yes | yes | yes |
| Intended Use: | The Fetal & Maternal Monitor is intended for non-invasive and invasive<br>monitoring of fetus during antepartum examination, labor and delivery. It is<br>intended to be used only by trained and qualified personnel in antepartum<br>examination rooms, labor and delivery rooms. | | |
| | Fetal & Maternal Monitor provides Non-Stress testing for pregnant women<br>from the 28th week of gestation. It can externally monitor the FHRs using<br>ultrasound and uterine activity via a TOCO transducer. Alternatively, it can<br>internally monitor one of the FHRs with DECG and uterine activity with an<br>IUPC. | | |
| Contraindications: | It is not intended for use in intensive care units, operating rooms or for home<br>use. | | |
| Test Summary: | The following quality assurance measures were applied to the development<br>of the Fetal & Maternal Monitor | | |
| | • Software testing | | |
| | • Hardware testing | | |
| | • Safety testing | | |
| | • Environment test | | |
| | • Risk analysis | | |
| | • Final validation | | |
| Conclusion: | Verification and validation testing was done on the Fetal & Maternal Monitor.<br>This premarket notification submission demonstrates that Fetal & Maternal<br>Monitor is substantially equivalent to the predicate device. | | |
{2}------------------------------------------------
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image shows the seal of the U.S. Department of Health & Human Services. The seal features a stylized caduceus, a symbol often associated with medicine and healthcare, with three parallel lines forming the snake and staff. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the caduceus. The seal is presented in black and white.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Edan Instruments, Inc. c/o Mr. William Stern Official Correspondent Multigon Industries, Inc. 1 Odell Plaza YONKERS NY 10701
NOV 2 6 2008
Rc: K082602
Trade/Device Name: Fetal and Maternal Monitor (Models F6 and F9) Regulation Number: 21 CFR §884.2740 Regulation Name: Perinatal Monitoring System and Accessories Regulatory Class: Class II Product Code: HGM Dated: November 4, 2008 Received: November 5, 2008
Dear Mr. Stern
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
This determination of substantial equivalence applies to the following transducers intended for use with the Fetal and Maternal Monitor (Models F6 and F9), as described in your premarket notification:
> 1 MHz PW fetal probe - model F6 1 MHz PW fetal probe - model F9
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
Page 2 - Mr. William Stern
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus permits your device to proceed to market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html
If you have any questions regarding the content of this letter, please contact Glenn Bell, Ph.D., at (240) 276-3666.
Sincerely yours,
loquítm Whang
Joyce M. Whang, Ph.D. Acting Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosures
{4}------------------------------------------------
Fetal & maternal monitor 510K Submission
### Indication for Use
K0826002
510(k) Number (if known):
Device Name:
Fetal & Maternal Monitor (Models F6 and F9)
The Fetal & Maternal Monitor is intended for non-invasive and invasive monitoring of fetus during antepartum examination, labor and delivery. It is intended to be used only by trained and qualified personnel in antepartum examination rooms, labor and delivery rooms.
Fetal & Maternal Monitor provides Non-Stress testing for pregnant women from the 28th week of gestation. It can externally monitor the FHRs using ultrasound and uterine activity via a TOCO transducer. Alternatively, it can internally monitor one of the FHRs with DECG and uterine activity with an IUPC.
Prescription Use (21 CFR Part 801 Subpart D)
And/Or
Over the Counter Use _ (21 CFR Part 801 Subpart C)
Concurrence of CDRH, Office of In Vitro Diagnostic Device Evaluation and Safety (OIVD)
---
(Division Sign-Off)
Division of Reproductive, Abdominal, and Radiological Devices 510(k) Number
{5}------------------------------------------------
# Diagnostic Ultrasound indications for Use Form
## Fill out one form for each ultrasound system and each transducer. 1 MHz PW fetal probe- model F9
Intended use : Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows
| Clinical Application | Mode Operation | | | | | | | | | |
|-------------------------------------------------------------------------|----------------|---|---|-----|-----|------------------|----------------------|------------------------------|-----------------------|--------------------|
| | A | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(specify) | Other<br>(specify) |
| Ophthalmic | | | | | | | | | | |
| Fetal | | | | N | | | | | | |
| Abdominal | | | | | | | | | | |
| Intraoperative(specify) | | | | | | | | | | |
| Intraoperative Neurological | | | | | | | | | | |
| Pediatric | | | | | | | | | | |
| Small Organ(specify) | | | | | | | | | | |
| Neonatal Cephalic | | | | | | | | | | |
| Cardiac | | | | | | | | | | |
| Transesophageal | | | | | | | | | | |
| Transrectal | | | | | | | | | | |
| Transvaginal | | | | | | | | | | |
| Transurethral | | | | | | | | | | |
| Intravascular | | | | | | | | | | |
| Peripheral Vascular | | | | | | | | | | |
| Laparoscopic | | | | | | | | | | |
| Musculo-skeletal | | | | | | | | | | |
| Conventional | | | | | | | | | | |
| Musculo-skeletal Superficial | | | | | | | | | | |
| Other | | | | | | | | | | |
| N=new indication; P=previously cleared by FDA; e=ADDED UNDER appendix E | | | | | | | | | | |
Additional Comments: ___ The above is a 1 mHz PW transducer for the fetal heart rate detection.
> PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE I FNEEDED CONCURRENCE OF cdrh, Office of Device Evaluation (ODE)
Imri In Whan
(Division Sign-Off) Divisior Regraductive, Abdominal, and Radiological Devic
510(k) Number
{6}------------------------------------------------
# Diagnostic Ultrasound indications for Use Form
## Fill out one form for each ultrasound system and each transducer. 1 MHz PW fetal probe- model F6
Intended use : Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows
| Clinical Application | Mode Operation | | | | | | | | | |
|-------------------------------------------------------------------------|----------------|---|---|-----|-----|------------------|----------------------|------------------------------|-----------------------|--------------------|
| | A | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(specify) | Other<br>(specify) |
| Ophthalmic | | | | | | | | | | |
| Fetal | | | | N | | | | | | |
| Abdominal | | | | | | | | | | |
| Intraoperative(specify) | | | | | | | | | | |
| Intraoperative Neurological | | | | | | | | | | |
| Pediatric | | | | | | | | | | |
| Small Organ(specify) | | | | | | | | | | |
| Neonatal Cephalic | | | | | | | | | | |
| Cardiac | | | | | | | | | | |
| Transesophageal | | | | | | | | | | |
| Transrectal | | | | | | | | | | |
| Transvaginal | | | | | | | | | | |
| Transurethral | | | | | | | | | | |
| Intravascular | | | | | | | | | | |
| Peripheraln Vascular | | | | | | | | | | |
| Laparoscopic | | | | | | | | | | |
| Musculo-skeletal | | | | | | | | | | |
| Conventional | | | | | | | | | | |
| Musculo-skeletal Superficial | | | | | | | | | | |
| Other | | | | | | | | | | |
| N=new indication; P=previously cleared by FDA: e=ADDED UNDER appendix E | | | | | | | | | | |
ਦ
Additional Comments: _ The above is a 1 mHz PW transducer for the fetal heart rate detection.
PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE I FNEEDED
CONCURRENCE OF cdrh, Office of Device Evaluation (ODE)
Irmu In Whay
(Division Sign-Off)
Division of Reproductive, Abdominal, and Radiological Devices
(k) Number K082662
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.