To be used for temporary relief of pain associated with sore and aching muscles in the shoulder, waist, back, neck, upper extremities (arm), and lower extremities (leg) due to strain from exercise or normal household work activities.
Device Story
Prospera OTC TENS Electronic Pulse Massager (models PL009, PL009A, PL029) is a battery-powered transcutaneous electrical nerve stimulator. Device delivers small electrical pulses to user's skin via adhesive electrode pads to activate underlying nerves for temporary pain relief. User selects massage pulse styles and operational modes via device interface; PL009/PL009A use mechanical potentiometer for intensity control, while PL029 uses electronic switches. Device features LCD display showing massage style, intensity, and time. Intended for home use by lay users. Output affects clinical decision-making by providing non-pharmacological pain management for muscle soreness. Benefits include temporary relief of pain associated with muscle strain.
Clinical Evidence
Bench testing only. No clinical data provided. Device validated through electrical safety, electromagnetic compatibility (EMC) testing, and software verification against voluntary standards including IEC 60601-1, IEC 60601-1-2, and IEC 60601-2-10.
Technological Characteristics
Battery-powered (3V or 6V) TENS device. Enclosure material: ABS. Output: Monophasic rectangular pulses. Intensity control: Mechanical potentiometer (PL009/PL009A) or electronic switches (PL029). Connectivity: Standalone. Sterilization: Not applicable (non-sterile). Software: Microprocessor-controlled with automatic shut-off. Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-2-10, EN 60335-1/2.
Indications for Use
Indicated for temporary relief of pain in muscles of shoulder, waist, back, neck, arms, and legs caused by exercise or household/work strain. For over-the-counter use.
Regulatory Classification
Identification
A transcutaneous electrical nerve stimulator for pain relief is a device used to apply an electrical current to electrodes on a patient's skin to treat pain.
K233040 — EMS/TENS Massager Device · Shenzhen Imdk Medical Technology Co., Ltd. · Feb 12, 2024
Submission Summary (Full Text)
{0}------------------------------------------------
510(k) K122744
# 510(k) Summary of Safety and Effectiveness
This 510(k) summary of safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.
## I. Submitter's Information - 21 CFR §807.92(a)(1):
Prospera Corporation 405 Boulder Ct., Ste 500 Pleasanton, CA 94566 (925) 225-0888 Tel. (925) 225-0660 Fax Establishment Registration No. 3009049803
Contact Person: Jane Ding Date Prepared: September 6, 2012
# II. Trade Name, Common Name & Classification - 21 CFR §807.92(a)(2):
| Trade Name: | Prospera OTC TENS Electronic Pulse Massager |
|--------------------|------------------------------------------------------------|
| Models: | PL009; PL009A; PL029 |
| Common Name: | Transcutaneous electrical nerve stimulator, OTC |
| Regulation Number: | 21 CFR 882.5890 |
| Regulation Name: | Transcutaneous Electrical Nerve Stimulator for pain relief |
| Regulatory Class: | II |
| Product Code: | NUJH, NGX |
| Use: | Over-The-Counter Use |
## III. Predicate Devices - 21 CFR §807.92(a)(3):
Prospera OTC TENS electronic pulse massager (herein after as "the subject matter device") is an electrically powered device intended for over the counter use and used to apply an electrical current to electrodes on a user's skin to relieve pain. The subject matter device is substantially equivalent to the following legally marketed devices in safety and effectiveness:
1
{1}------------------------------------------------
| 510(k) No. | Classification | FDA Dated | Model | Manufacturer |
|------------|-----------------------------------------------------------|------------|-------|----------------------------|
| K102598 | NUH<br>21 CFR §882.5890<br>NGX<br>21 CFR §882.5850<br>OTC | 05/13/2011 | JQ-5C | Hi-Dow International, Inc. |
| K060846 | NUH<br>21 CFR §882.5890<br>NGX<br>21 CFR §882.5850<br>OTC | 12/03/2007 | T1040 | Endurance Therapeutics |
| K112392 | NUH<br>21 CFR §882.5890<br>OTC | 03/19/2012 | PP907 | Hivox Biotek Inc. |
The OTC predicate devices' description and indications for use:
.
| 510(k) No. | Description | Indications for Use |
|------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| K102598 | Portable, battery powered (3.7VDC)<br>multi function device offering both<br>Transcutaneous Electrical Nerve<br>Stimulator (TENS) and Powered<br>Muscle Stimulator (PMS) qualities in<br>one device. Double channels that<br>effectively transfers user's desired<br>choice of programmed electrical pulses<br>directly through electrode adhesive<br>pads to the suggested area of the body<br>where the electrodes are placed,<br>causing minimal muscle contractions.<br>There are 6 modes of operation. | Temporary relief of pain<br>associated with sore and aching<br>muscles in the shoulder, waist,<br>back, neck, upper extremities<br>(arm), and lower extremities (leg)<br>due to strain from exercise or<br>normal household work activities. |
| K060846 | Pre-programmed electrical pulses are<br>transferred directly through electrode<br>adhesive pads to the suggested area<br>of the body where the electrodes are<br>placed, causing minimal muscle<br>contractions. | Temporary relief of pain<br>associated with sore and aching<br>muscles in the lower back due to<br>strain from exercise or normal<br>household and work activities. |
| K112392 | The stimulator can generate small<br>pulses of electrical current. Delivered<br>these pulses pass through the skin and<br>activated underlying nerves. | Temporary relief of pain<br>associated with sore and aching<br>muscles in the lower back as well<br>as upper and lower extremities<br>(arm and/or leg) due to strain from<br>exercise or normal household and<br>work activities. |
2 ・
. .
.
.
{2}------------------------------------------------
# IV. Device Description - 21 CFR 21 CFR 21 CFR §807.92(a)(4):
PL009 and PL009A are electrically powered devices intended for over the counter use and used to apply an electrical current to electrodes on a user's skin to relieve pain. It provides three selectable massage pulse styles with five auto operational modes through two channels with four attachment pads and shows graphic information about massage style, intensity and time remaining on a LCD based display which is incorporated with the device body. The device uses DC 6V power supplied by 4 AAA batteries. It is approximately 0.5lbs with dimensions 8" x 2" x 1" plus attachment pads. The only difference between PL009 and PL009A is the former has a button of "Repeat" representative of the function of repeat.
PL029 is an electrically powered device intended for over the counter use and used to apply an electrical current to electrodes on a user's skin to relieve pain. It provides eight selectable massage pulse styles with five auto operational modes through two channels with four attachment pads and shows textual message about massage style, intensity and time remaining on a LCD based display which is incorporated with the device body. The device uses DC 3V power supplied by 2 AAA batteries. It is approximately 0.5lb with dimensions 8" x 2" x 1" plus attachment pads.
The differences between PL009 and PL029 include:
(1) PL009's work voltage is 6V supplied by 4 AAA batteries, while PL029's work voltage is 3V supplied by 2 AAA batteries.
(2) PL009 provides three operational modes, i.e., massage, beat and knead, which can be combined into six stimulation programs for WAIST, SHOULDER, JOINT, HAND-FOOT, REPEAT and SOLE, while PL029 provides eight operational, i.e., massage, beat, knead, slim, acupuncture, scraping, cupping and immune.
(3) PL009 uses mechanical potentiometer to control power and intensity, while PL029 uses electronic switches to control power and intensity.
(4) PL009's LCD display shows body parts in graphics when the device is in operation, while PL029's LCD display shows working modes in text only when the device is in operation.
{3}------------------------------------------------
## V. Indication for Use - 21 CFR §807.92(a)(5):
To be used for temporary relief of pain associated with sore and aching muscles in the shoulder, waist, back, neck, upper extremities (arm), and lower extremities (leg) due to strain from exercise or normal household work activities.
#### VI. Technical Characteristics - 21 CFR §807.92(a)(6):
The operational principle of the above listed OTC predicate devices is to generate small pulses of electrical current and deliver the pulses to an ordinary user's skin through electrode adhesive pads such that the underlying nerves are activated and the pain associated with sore and aching muscles is temporarily relieved.
Similarly, the operational principle of the subject matter device manufactured by the submitter is to generate small pulses of electrical current and deliver the pulses to an ordinary user's skin through electrode adhesive pads such that the underlying nerves are activated and the pain associated with sore and aching muscles is temporarily relieved.
Table 1, attached hereto, illustrates PLO09's and PL009A's comparison with the OTC predicate devices.
Table 2, attached hereto, illustrates PL029's comparison with the OTC predicate devices.
The data in Table 1 and Table 2 indicates that the technical characteristics, features, specifications and intended use of the subject matter device are substantially equivalent to those of the OTC predicate devices.
The design differences between the subject matter device and the OTC predicate devices are insignificant and do neither affect the intended use, nor alter the operational principle of the subject matter device.
#### Non-Clinical Tests Performed - 21 CFR §807.92(b)(1): VII.
Compliance to applicable voluntary standards includes IEC 60601-1 and IEC 60601-1-2.
{4}------------------------------------------------
Non-clinical tests were performed on the subject matter device in order to validate the design and to assure conformance with the following voluntary design standards in connection with medical device electrical safety, electromagnetic compatibility, and particular requirements for the safety of nerve and muscle stimulators:
- a. EN55014-1: Electromagnetic compatibility-Requirements for household appliances, electric tocls, and similar apparatus-Part 1: Emission-Product family standard:
- b. EN55014-2: Electromagnetic compatibility—Requirements for household appliances, electric tocks, and similar apparatus-Part 2: Immunity-Product family standard;
- c. EN60335-1: Household and similar electrical appliances Safety Part 1: General Requirements;
- d. EN60335-2: Safety of household and similar electrical appliances-Part 2: Particular requirements for massage appliances;
- e. EN62233: Measurement methods for electromagnetic fields of household appliances and similar apparatus with regard to human exposure;
- ﺳﺘ IEC 60601-1 "Medical electrical equipment - Part 1: General requirements for safety;
- g. IEC 60601-1-2 "Medical electrical equipment Part 1-2: General requirements for safety - Collateral Standard"; and
- h. IEC 60601-2-10 "Medical electrical equipment Part 2: Particular requirements for the safety of nerve and muscle stimulators".
In addition to the compliance of voluntary standards, the software verification has been carries out according to the FDA Guidance for the Content of Premarket Submission for Software Contained in Medical Devices.
Electrodes, as accessories of the subject matter device, are also meet safety requirement.
#### Conclusion - 21 CFR §807.92(b)(3): VIII.
The product description, the non-clinical tests performed and the comparison of technical characteristics demonstrate that the subject matter device is as safe, as effective, and performs as well as or better than the foregoing identified OTC predicate devices which are legally marketed in the United States.
5
{5}------------------------------------------------
| Parameter | Your Device | Predicate | | | | |
|--------------------------------------------------------------------------------------------|----------------------------------|----------------------------------|----------------------------------|--------------------------------------|------------------------------|------------------------------------|
| 510(k) Number | K122744 | K122744 | K122744 | K102598 | K112392 | K060846 |
| Device Name and Model<br>Manufacturer | PL009<br>Prospera | PL009<br>Prospera | PL029<br>Prospera | JQ-5C<br>Hi-Dow<br>international Inc | PP907<br>HIVOX<br>BIOTEK INC | T1040<br>Endurance<br>Therapeutics |
| Power Source(s) | 6V Battery<br>Battery Supply N/A | 6V Battery<br>Battery Supply N/A | 3V Battery<br>Battery Supply N/A | 3.7V Lithium<br>Battery Supply N/A | 3V Battery<br>Battery Supply | 4.5V Battery<br>Battery Supply |
| Patient Leakage Current | | | | | | |
| - Normal Condition (μA) | 1.8μΑ | 1.8μΑ | 2.0μΑ | 2.0μΑ | NA | NA |
| - Single Fault Condition (μΑ) | 3.3μΑ | 3.3μΑ | 3.0μΑ | 4.2μΑ | NA | NA |
| Average DC current through electrodes when<br>device is on but no pulses are being applied | ΟμΑ | ΟμΑ | ΟμΑ | ΟμΑ | ΟμΑ | ΟμΑ |
| Number of Output Modes | 3 | 3 | 8 | 6 | 1 | 10 |
| Number of Output<br>Channels: | Synchronous | Synchronous | Alternating | Synchronous | Synchronous | Synchronous |
| Method of Channel Isolation | ON/OFF switch | ON/OFF switch | By software | ON/OFF switch | ON/OFF switch | ON/OFF switch |
| Regulated Current or Regulated Voltage? | Voltage control | Voltage control | Voltage control | Voltage control | Voltage control | Voltage control |
| Software/Firmware/Microprocessor Control? | Yes | Yes | Yes | Yes | Yes | Yes |
| Automatic Overload Trip? | No | No | No | No | No | No |
| Automatic No-Load Trip? | No | No | No | No | No | No |
| Automatic Shut Off? | Yes | Yes | Yes | Yes | Yes | No |
| User Override Control? | Yes | Yes | Yes | Yes | Yes | Yes |
| Indicator Display: | | | | | | |
| On/Off Status? | Yes | Yes | Yes | Yes | No | Yes |
| Low Battery? | No | No | No | No | No | No |
| Voltage/Current | No | No | No | No | No | No |
| Timer Range (minutes) | 15 min | 15 min | 5 min, 10min, | | 20min | |
| Compliance with Voluntary Standards? | Yes | yes | Yes. | | Yes | Yes |
| Compliance with 21 CFR 898? | Yes | yes | Yes | Yes | Yes | Yes |
| Weight (lbs., oz.) | 0 lb., 8.21 oz. | 0 lb., 8.21 oz. | 0 lb., 6.40 oz | | 0 lb.,0.53 oz. | 0lbs, 14.4 oz |
| Dimensions (in.) [W x H x D] | 2.24 x 7.80 x 0.91 | 2.24 x 7.80 x 0.91 | 2.32 x 7.87 x 0.83 | 4.45 x 2.76 x 0..38 | 3.55 x | 6 x 1 x 2.8 |
| Housing Materials and Construction | Enclosure: ABS | Enclosure: ABS | Enclosure: ABS | ABS | silicone | ABS |
K122744, 510(K) SUMMARY, TABLE A: Comparisons of PL009/PL009A/PL029 Basic Unit Characteristics with Predicate Devices:
-
:
{6}------------------------------------------------
K122744, 510(K) SUMMARY, TABLE B: PL009 & PL 009A'S COMPARISON WITH PREDICATE DEVICES
.
1 、
:
| Parameter | PL009<br>K122744 | PL009A<br>K122744 | JQ-5C<br>K102598 | PP909<br>K112392 | PP907<br>K112392 | PP904<br>K112392 | T1040<br>K060846 | |
|---------------------------------------------------|-----------------------------|-------------------------|------------------|-------------------------|-------------------------|-------------------------|------------------|-----|
| Waveform | Monophasic | Monophasic | Monophasic | Symmetrical<br>biphasic | Symmetrical<br>biphasic | Symmetrical<br>biphasic | Biphasic | |
| Shape | Rectangular | Rectangular | Rectangular | Rectangular | Rectangular | Rectangular | Rectangular | |
| Maximum output voltage<br>(Volts +/- 20%) at 500Ω | 12.8 | 12.8 | 62.4 | 40.0 | 68.8 | 57.6 | 40.7 | |
| Maximum output voltage<br>(Volts +/- 20%) at 2KΩ | 51 | 51 | 79.2 | 84.0 | 88.0 | 89.6 | 105.1 | |
| Maximum output voltage<br>(Volts +/-20%) at 10kΩ | 368 | 368 | 84 | 92.0 | 95.2 | 96.0 | 154.1 | |
| Maximum output current (mA<br>+/- 20%) at 500Ω | 15 mA | 15 mA | 124.8 | 80.0 | 137.6mA | 115.2 | 81.4 | |
| Maximum output current (mA<br>+/- 20%) at 2KΩ | 3.2 mA | 3.2 mA | 39.6 | 42.0 | 44.0mA | 44.8 | 47.8 | |
| Maximum output current (mA<br>+/- 20%) at 10KΩ | 0.6 mA | 0.6 mA | 8.4 | 9.2 | 9.52mA | 9.6 | 15.4 | |
| Duration of primary<br>(depolarizing phase (µSec) | 40 nsec | 40 nsec | N/A | N/A | N/A | N/A | N/A | |
| Pulse Duration (µSec) | 120-6800 µSec | 120-6800 µSec | 16.3-781mS | 200 µSec<br>(fixed) | 200 µSec (fixed) | 200 µSec (fixed) | 4.1~500mS | |
| Frequency (Hz) | 1-100 | 1-100 | 61.3 | 35 | 2 and 40 | 2, 5, and 40 | 245Hz | |
| Net Charge per pulse (µC) at<br>500Ω | 18000 µC | 18000 µC | N/A | 0.3200 | 1.1008 | 0.2304 | 4.07 | |
| Maximum charge (µC) at<br>500Ω | 23 µC | 23 µC | 17.92 | 16.0 | 27.52 | 23.04 | 16.9 | |
| Maximum current density<br>(mA/cm²) at 500Ω | 1.4 mA/cm² | 1.4 mA/cm² | 9.92 | 1.964 | 3.378 | 2.828 | 2.71 | |
| Maximum average power<br>density (W/cm²) at 500Ω | 0.23 W/cm² | 0.23 W/cm² | 2.72 | 0.078 | 0.232 | 0.163 | 5.35 (mW/cm²) | |
| Burst<br>Mode | (a) Pulse per burst | 1 | 1 | N/A | 1 | N/A | N/A | N/A |
| | (b) Burst per second | 0-25 | 0-25 | N/A | 4 | N/A | N/A | N/A |
| | (c) Burst-duration<br>(sec) | 1 | 1 | N/A | 2 | N/A | N/A | N/A |
| | (d) Duty Cycle | 5 | 5 | N/A | 8 | N/A | N/A | N/A |
| ON time (sec) | Potentiometer<br>switch | Potentiometer<br>switch | N/A | 120 | 120 | 120 | N/A | |
| OFF time (sec) | Potentiometer<br>switch | Potentiometer<br>switch | N/A | 0 | 0 | 0 | N/A | |
| Additional features | N/A | N/A | N/A | N/A | N/A | N/A | N/A | |
.
: :
{7}------------------------------------------------
K122744, 510(K) SUMMARY, TABLE C: PL029'S COMPARISON WITH PREDICATE DEVICES
| Parameters | PL029<br>K122744<br>Monophasic | JQ-5C<br>K102598<br>Monophasic | PP909<br>K112392<br>Symmetrical biphasic | PP907<br>K112392<br>Symmetrical biphasic | PP904<br>K112392<br>Symmetrical biphasic | T1040<br>K060846<br>Biphasic | |
|------------------------------------------------|--------------------------------|--------------------------------|------------------------------------------|------------------------------------------|------------------------------------------|------------------------------|-----|
| Waveform<br>Shape | Rectangular | Rectangular | Rectangular | Rectangular | Rectangular | Rectangular | |
| Maximum output voltage (Volts +/- 20%) at 500Ω | 49.6 | 62.4 | 40.0 | 68.8 | 57.6 | 40.7 | |
| Maximum output voltage (Volts +/- 20%) at 2KΩ | 99.2 | 79.2 | 84.0 | 88.0 | 89.6 | 105.1 | |
| Maximum output voltage (Volts +/- 20%) at 10kΩ | 114 | 84 | 92.0 | 95.2 | 96.0 | 154.1 | |
| Maximum output current (mA +/- 20%) at 500Ω | 18 mA | 124.8 | 80.0 | 137.6mA | 115.2 | 81.4 | |
| Maximum output current (mA +/- 20%) at 2KΩ | 3.2 mA | 39.6 | 42.0 | 44.0mA | 44.8 | 47.8 | |
| Maximum output current (mA +/- 20%) at 10KΩ | 0.6 mA | 8.4 | 9.2 | 9.52mA | 9.6 | 15.4 | |
| Duration of primary (depolarizing phase (µSec) | 40 msec | N/A | N/A | N/A | N/A | N/A | |
| Pulse Duration (µSec) | 100-200 µSec | 16.3-781mS | 200 µSec (fixed) | 200 µSec (fixed) | 200 µSec (fixed) | 4.1~500mS | |
| Frequency (Hz) | 0.5-86Hz | 61.3 | 35 | 2 and 40 | 2, 5, and 40 | 245Hz | |
| Net Charge per pulse (µC) at 500Ω | 18000 µC | N/A | 0.3200 | 1.1008 | 0.2304 | 4.07 | |
| Maximum charge (µC) at 500Ω | 23 | 17.92 | 16.0 | 27.52 | 23.04 | 16.9 | |
| Maximum current density (mA/cm²) at 500Ω | 1.4 mA/cm² | 9.92 | 1.964 | 3.378 | 2.828 | 2.71 | |
| Maximum average power density (W/cm²) at 500Ω | 0.23 W/cm² | 2.72 | 0.078 | 0.232 | 0.163 | 5.35 (mW/cm²) | |
| Burst<br>Mode | A. Pulse per burst | 1 | N/A | 1 | N/A | N/A | N/A |
| | B.Burst per second | 0-7 | N/A | 4 | N/A | N/A | N/A |
| | C.Burst-duration (sec) | 1 | N/A | 2 | N/A | N/A | N/A |
| | D.Duty Cycle | | N/A | 8 | N/A | N/A | N/A |
| ON time (sec) | 7 | N/A | 120 | 120 | 120 | N/A | |
| OFF time (sec) | 40 ms | N/A | 0 | 0 | 0 | N/A | |
| Additional features | 18 ms | N/A | N/A | N/A | N/A | N/A | |
·
{8}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/8/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three tail feathers, representing the three levels of government: federal, state, and local. The eagle is enclosed in a circle with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" around the perimeter.
March 28, 2013
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-002
Prospera Corporation % Mr. Leon E. Jew Attorney at Law Dayhee Law Group 5776 Stoneridge Mall Rd., Suite 288 Pleasanton, CA 94588
Re: K122744
Trade/Device Name: OTC TENS Electronic Pulse Massager Regulation Number: 21 CFR 882.5890 Regulation Name: Transcutaneous electrical nerve stimulator for pain relief Regulatory Class: II Product Codes: NUH, NGX Dated: March 15, 2013 Received: March 18, 2013
Dear Mr. Jew:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{9}------------------------------------------------
Page 2 - Leon E. Jew
forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to:
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address: http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Victor Krauthamer -S
Victor Krauthamer, Ph.D. Acting Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{10}------------------------------------------------
#### INDICATIONS FOR USE
510(k) Number (if known): K122744
Device Name: PROSPERA OTC TENS Electronic Pulse Massager
Models: PL009: PL009A; PL029
Indications for Use:
To be used for temporary relief of pain associated with sore and aching muscles in the shoulder. waist, back, neck, upper extremities (arm), and lower extremities (leg) due to strain from exercise or normal household work activities.
Prescription Use (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use X (21 CFR 807 Subpart C)
(PLEASE IDO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH. Office of Device Evaluation (ODE)
Victor Kra 2013.03.28 G
Division of Neurological and Physical Medicine Devices 510(k) Number: _K122744
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.