To be used for temporary relief of pain associated with sore and aching muscles in the shoulder, waist, back, arm, and leg, due to strain from exercise or normal household and work activities.
Device Story
Electronic Pulse Stimulator is a compact, portable, microprocessor-controlled Transcutaneous Electrical Nerve Stimulator (TENS). Device delivers gentle electrical pulses to user's skin via adhesive electrode pads. User adjusts pulse intensity via front control panel. Intended for over-the-counter use to provide temporary relief of pain in shoulder, waist, back, arm, and leg muscles caused by exercise or daily activities. Device operates by activating underlying nerves through electrical stimulation. Benefits include non-invasive, temporary pain management for muscle soreness.
Clinical Evidence
No clinical data. Substantial equivalence established via bench testing and compliance with voluntary standards including IEC 60601-1, IEC 60601-1-2, and IEC 60601-2-10.
Indicated for temporary relief of pain associated with sore and aching muscles in the shoulder, waist, back, arm, and leg, due to strain from exercise or normal household and work activities in adult patients.
Regulatory Classification
Identification
A transcutaneous electrical nerve stimulator for pain relief is a device used to apply an electrical current to electrodes on a patient's skin to treat pain.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features the department's name in a circular arrangement around a stylized symbol. The symbol consists of three overlapping profiles of human faces, creating a sense of community and connection. The overall design is simple, clean, and professional.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
September 17, 2014
Shenzhen Jingkehui Electronic Co. LTD c/o Bill Quanqin Dai, Ph.D. Application Correspondent 513 Piazza Drive, Unit B Mountain View, CA, 94043
Re: 510(k) Number: K141260 Trade/Device Name: Electronic Pulse Stimulator Regulation Number: 21 CFR 882.5890 Regulation Name: Neurology Regulatory Class: Class II Product Code: NUH Dated: Aug 16th 2014 Received: Aug 20th 2014
Dear Dr. Dai.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{1}------------------------------------------------
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
# Felipe Aguel -S
Carlos Pena, Ph.D. for
Director- Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K141260
Device Name
Electronic Pulse Stimulator
#### Indications for Use (Describe)
To be used for temporary relief of pain associated with sore and aching muscles in the shoulder, waist, back, arm, and leg, due to strain from exercise or normal household and work activities.
Type of Use (Select one or both, as applicable)
- Prescription Use (Part 21 CFR 801 Subpart D)
> Over-The-Counter Use (21 CFR 801 Subpart C)
#### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
#### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
## Date: 2014.09.17 15:39:53 -04'00'
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
#### 510(k) Summary
This 510(k) Summary of safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR 807.92.
#### 1. Submitter's Information
Submitter: Shenzhen Jingkehui Electronic Co., Ltd. Address: 5F, Building 12, Hengmingzhu Industrial Park, Xiangxing Road, Shajing, Baoan District, Shenzhen, China Contact Person: Pu Jiang Tel: +86-755-29970323 Fax: +86-755-23493443 Email: bill@JKHhealth.com Date of Preparation: 05/12/2014
#### 2. Subject Device
Trade/Device Name: Electronic Pulse Stimulator Common Name: Transcutaneous electrical nerve stimulator (TENS) Regulation Description: Transcutaneous electrical nerve stimulator for pain relief Definition: Temporary relief of pain due to sore/aching muscles Regulation Medical Specialty: Neurology Review Panel: Neurology Product Code: NUH Regulation Number: 21 CFR 882.5890 Device Class: II Use: Over-The-Counter
#### 3. Predicate device
Predicate Device: Electronic Pulse Stimulator 510(k) Number: K131921 Use: Over-The-Counter Submitter: Shenzhen Jingkehui Electronic Co., Ltd.
Predicate Device: IQ Technologies 510(k) Number: K131290 Use: Over-The-Counter Submitter: IO Technologies Inc.
#### 4. Description of Subject Device
The subject device is a Transcutaneous Electrical Nerve Stimulator (TENS), intended for the over-thecounter use to temporarily relieve pain in different body areas. The subject device, which is compact, portable, and microprocessor-controlled, delivers a gentle electrical pulse to the user's skin through the electrode pads for pain relief. According to the need of users, the pulse intensity can be adjustable on the front control panel of the device.
## 5. Intended Use of Subject Device
To be used for temporary relief of pain associated with sore and aching muscles in the shoulder, waist, back, arm, and leg, due to strain from exercise or normal household and work activities.
{4}------------------------------------------------
### 6. Summary of Substantial Equivalence
The following table summarizes the comparison between the subject device and predicate devices, indicating the technical characteristics of the subject device are substantially equivalent to those of the predicate devices.
| Parameter | Subject Device | Subject Device | Predicate Device | Predicate Device |
|------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------|----------------------------------------------------------------------------------------------------------------------------|
| 510(k) Number | K141260 | K141260 | K131921 | K131290 |
| Device Name/Model | Electronic<br>Pulse<br>Stimulator/PL-029BL | Electronic<br>Pulse<br>Stimulator/PL-029K | Electronic<br>Pulse<br>Stimulator/PL-029 | IQ Technologies |
| Maximum output<br>voltage (Volts +/-<br>20%) at 500Ω | Mode 1: 41.6<br>Mode 2: 66.4<br>Mode 3: 60.8<br>Mode 4: 35.2<br>Mode 5: 32.8<br>Mode 6: This mode<br>cycles the above<br>modes<br>Mode 7: 75.2<br>Mode 8: 24.0 | Mode 1: 46.0<br>Mode 2: 59.6<br>Mode 3: 49.2<br>Mode 4: 64.8<br>Mode 5: 37.2<br>Mode 6: This mode<br>cycles the above<br>modes<br>Mode 7: 71.2<br>Mode 8: 64.0 | 49.6 | Mode 1: 42<br>Mode 2: 63.2<br>Mode 3: 64<br>Mode 4: 34.4<br>Mode 5: 32<br>Mode 6: This<br>mode cycles the<br>above modes |
| Maximum output<br>voltage (Volts +/-<br>20%) at 2ΚΩ | Mode 1: 80.0<br>Mode 2: 100<br>Mode 3: 83.2<br>Mode 4: 70.4<br>Mode 5: 67.2<br>Mode 6: This mode<br>cycles the above<br>modes<br>Mode 7: 121<br>Mode 8: 46.4 | Mode 1: 86.4<br>Mode 2: 111<br>Mode 3: 80<br>Mode 4: 90.4<br>Mode 5: 64.8<br>Mode 6: This mode<br>cycles the above<br>modes<br>Mode 7: 122<br>Mode 8: 89.6 | 99.2 | Mode 1: 80.8<br>Mode 2: 94.4<br>Mode 3: 87.2<br>Mode 4: 68<br>Mode 5: 64<br>Mode 6: This<br>mode cycles the<br>above modes |
| Maximum output<br>voltage (Volts +/-<br>20%) at 10kΩ | Mode 1: 125<br>Mode 2: 128<br>Mode 3: 84.8<br>Mode 4: 123<br>Mode 5: 116<br>Mode 6: This mode<br>cycles the above<br>modes<br>Mode 7: 134<br>Mode 8: 85.6 | Mode 1: 116<br>Mode 2: 146<br>Mode 3: 138<br>Mode 4: 106<br>Mode 5: 111<br>Mode 6: This mode<br>cycles the above<br>modes<br>Mode 7: 136<br>Mode 8: 96.8 | 114 | Mode 1: 129<br>Mode 2: 129<br>Mode 3: 96.8<br>Mode 4: 128<br>Mode 5: 119<br>Mode 6: This<br>mode cycles the<br>above modes |
| Maximum output<br>current (mA +/- 20%)<br>at 500Ω | Mode 1: 83.2<br>Mode 2: 132.8<br>Mode 3: 121.6<br>Mode 4: 70.4<br>Mode 5: 65.6<br>Mode 6: This mode<br>cycles the above<br>modes | Mode 1: 92.0<br>Mode 2: 119.2<br>Mode 3: 98.4<br>Mode 4: 129.6<br>Mode 5: 74.4<br>Mode 6: This mode<br>cycles the above<br>modes | 18 | Mode 1: 84<br>Mode 2: 126.4<br>Mode 3: 128<br>Mode 4: 68.8<br>Mode 5: 64<br>Mode 6: This<br>mode cycles the |
| | modes | modes | | above modes |
| | Mode 7: 150.4 | Mode 7: 142.4 | | |
| | Mode 8: 48.0 | Mode 8: 128.0 | | |
| Maximum output<br>current (mA +/- 20%)<br>at 2ΚΩ | Mode 1: 40.0 | Mode 1: 43.2 | 3.2 | Mode 1: 40.4 |
| | Mode 2: 50 | Mode 2: 55.5 | | Mode 2: 47.2 |
| | Mode 3: 41.6 | Mode 3: 40 | | Mode 3: 43.6 |
| | Mode 4: 35.2 | Mode 4: 45.2 | | Mode 4: 34 |
| | Mode 5: 33.6 | Mode 5: 32.4 | | Mode 5: 32 |
| | Mode 6: This mode<br>cycles the above<br>modes | Mode 6: This mode<br>cycles the above<br>modes | | Mode 6: This<br>mode cycles the<br>above modes |
| | Mode 7: 60.5 | Mode 7: 61 | | |
| | Mode 8: 23.2 | Mode 8: 44.8 | | |
| Maximum output<br>current (mA +/- 20%)<br>at 10KΩ | Mode 1: 12.5 | Mode 1: 11.6 | 0.6 | Mode 1: 12.9 |
| | Mode 2: 12.8 | Mode 2: 14.6 | | Mode 2: 12.9 |
| | Mode 3: 8.5 | Mode 3: 13.8 | | Mode 3: 9.7 |
| | Mode 4: 12.3 | Mode 4: 10.6 | | Mode 4: 12.8 |
| | Mode 5: 11.6 | Mode 5: 11.1 | | Mode 5: 11.9 |
| | Mode 6: This mode<br>cycles the above<br>modes | Mode 6: This mode<br>cycles the above<br>modes | | Mode 6: This<br>mode cycles the<br>above modes |
| | Mode 7: 13.4 | Mode 7: 13.6 | | |
| | Mode 8: 8.6 | Mode 8: 9.7 | | |
| Pulse Width (μSec) | 100 | 50~100 | 50~140 | 100 |
| Pulse period (mSec) | 6.1~824 | 12~832 | 40 | 10~840 |
| Frequency (Hz) | Mode 1: 69.4 | Mode 1: 54.3 | 86 | Mode 1: 69.4 |
| | Mode 2: 12.8~53.2 | Mode 2: 35.7 | | Mode 2: 12.3~54.3 |
| | Mode 3: 1.2 | Mode 3: 62.5 | | Mode 3: 1.2 |
| | Mode 4: 96.2 | Mode 4: 6.7 | | Mode 4: 100 |
| | Mode 5: 96.2 | Mode 5: 83.3 | | Mode 5: 100 |
| | Mode 6: This mode<br>cycles the above<br>modes | Mode 6: This mode<br>cycles the above<br>modes | | Mode 6: This<br>mode cycles the<br>above modes |
| | Mode 7: 19.2 | Mode 7: 19.8 | | |
| | Mode 8: 164.4 | Mode 8: 1.2 | | |
| Maximum Phase<br>charge (µC) at 500Ω | Mode 1: 17.3 | Mode 1: 19.9 | 23 | Mode 1: 18.1 |
| | Mode 2: 24.4 | Mode 2: 6.0 | | Mode 2: 27.3 |
| | Mode 3: 24.3 | Mode 3: 4.9 | | Mode 3: 25.6 |
| | Mode 4: 14.6 | Mode 4: 6.5 | | Mode 4: 14.9 |
| | Mode 5: 12.6 | Mode 5: 3.7 | | Mode 5: 12.8 |
| | Mode 6: This mode<br>cycles the above<br>modes | Mode 6: This mode<br>cycles the above<br>modes | | Mode 6: This<br>mode cycles the<br>above modes |
| | Mode 7: 31.3 | Mode 7: 33.0 | | |
| | Mode 8: 9.2 | Mode 8: 25.6 | | |
| Maximum current<br>density (mA/cm2) | Mode 1: 3.33 | Mode 1: 5.75 | 1.4 | Mode 1: 3.36 |
{5}------------------------------------------------
{6}------------------------------------------------
| 500Ω | Mode 3: 4.86<br>Mode 4: 2.82<br>Mode 5: 2.62<br>Mode 6: This mode cycles the above modes<br>Mode 7: 6.02<br>Mode 8: 1.92 | Mode 3: 6.15<br>Mode 4: 8.10<br>Mode 5: 4.65<br>Mode 6: This mode cycles the above modes<br>Mode 7: 8.90<br>Mode 8: 8.00 | 0.23 | Mode 3: 5.12<br>Mode 4: 2.75<br>Mode 5: 2.56<br>Mode 6: This mode cycles the above modes |
|------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------|------|-------------------------------------------------------------------------------------------------------------------------------|
| Maximum average power density (mW/cm²) at 500Ω | Mode 1: 2.00<br>Mode 2: 0.83~3.45<br>Mode 3: 0.07<br>Mode 4: 1.98<br>Mode 5: 1.59<br>Mode 6: This mode cycles the above modes<br>Mode 7: 1.81<br>Mode 8: 1.45 | Mode 1: 3.10<br>Mode 2: 0.79<br>Mode 3: 0.95<br>Mode 4: 0.18<br>Mode 5: 0.72<br>Mode 6: This mode cycles the above modes<br>Mode 7: 2.91<br>Mode 8: 0.12 | | Mode 1: 2.11<br>Mode 2: 0.85~3.75<br>Mode 3: 0.08<br>Mode 4: 2.05<br>Mode 5: 1.64<br>Mode 6: This mode cycles the above modes |
## 7. Substantial Equivalence
The operational principle of the above predicate device is to generate small pulses of electrical current and deliver the pulses to the user's skin through adhesive electrode pads such that the underlying nerves are activated and the pain associated with sore and aching muscles is temporarily relieved.
Identically, the subject device generates small pulses of electrical current and delivers the pulses to the user's skin through adhesive electrode pads such that the underlying nerves are activated and the pain associated with sore and aching muscles is temporarily relieved.
The comparison between the subject device and predicate devices demonstrates the technical characteristics, specifications, and intended use of the subject device are substantially equivalent to those of the predicate devices.
The differences, such the output voltage and current, between the subject device and the predicate devices are insignificant in terms of safety or effectiveness. The verification and validation tests, such as IEC 60601-1, IEC 60601-1-2, and IEC 60601-2-10, further demonstrate these differences maintain the same safety and effectiveness as those of the 510(k) cleared predicate devices. In other words, these differences do not affect the intended use or alter the fundamental technology of the device. There are no new safety or effectiveness issues concerning the subject device, which offers substantially equivalent technical specifications, features, intended use, and effective results as the predicate devices.
Concerns of the safe and proper use of the biocompatible electrode pads have been fully addressed by making the use conscious of the proper placement of the electrode pads and appropriate operations of the device through details in the labeling.
#### 8. Non-Clinical Tests Performed
The subject device does not conduct, nor rely upon, clinical tests to determine substantial equivalence. Non-clinical tests were performed on the subject device in order to validate the design and to assure
{7}------------------------------------------------
conformance with the following voluntary design standards in connection with medical device electrical safety, and electromagnetic compatibility.
- (a) IEC 60601-1 "Medical Electrical Equipment Part 1: General requirements for basic safety and essential performance".
- (b) IEC 60601-1-2 "Medical Electrical Equipment Part 1-2: General Requirements for Basic Safety and Essential Performance - Collateral standard: Electromagnetic Compatibility - Requirements and Tests".
- (c) IEC 60601-2-10 "Medical electrical equipment Part 2-10: Particular requirements for the basic safety and essential performance of nerve and muscle stimulators".
In addition to the compliance of voluntary standards, the verification of software used in the subject device has been carried out according to the FDA Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices. The biocompatible electrodes, as the accessory of the subject device, also meet the requirement of safety.
#### 9. Conclusion
The tests performed and the comparison of technical characteristics, specifications, and intended use demonstrate the subject device is substantially equivalent to the predicate device. Therefore, the subject device is as safe and effective as the foregoing identified OTC predicate devices that have been legally marketed in the United States.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.