K131290 · Iq Technologies, Inc. · NUH · Aug 6, 2014 · Neurology
Device Facts
Record ID
K131290
Device Name
IQ TECHNOLOGIES
Applicant
Iq Technologies, Inc.
Product Code
NUH · Neurology
Decision Date
Aug 6, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.5890
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
TENS: To be used for temporary relief of pain associated with sore and aching muscles in the shoulder, waist, back, upper extremities (arm), and lower extremities (leg) due to strain from exercise or normal household work activities. PMS: It is intended to be used to stimulate healthy muscles in order to improve and facilitate muscle performance.
Device Story
Compact, portable, microprocessor-controlled, double-channel TENS/PMS device; delivers electrical pulses via adhesive electrode pads (gel, carbon film, cloth) to user's skin. User adjusts pulse intensity via front control panel. Used OTC for pain relief and muscle stimulation. Output affects nerves/muscles to provide temporary relief or facilitate performance. Benefits include non-invasive pain management and muscle conditioning.
Clinical Evidence
Bench testing only. No clinical data. Validation performed via electrical safety and electromagnetic compatibility testing (IEC 60601-1, IEC 60601-1-2, IEC 60601-2-10) and software verification per FDA guidance.
Technological Characteristics
DC 3.7V Lithium battery-powered; 2-channel output; microprocessor-controlled. Delivers pulsed rectangular waveforms. Pulse width 100µs; frequency 1.2-100Hz. Biocompatible electrode pads (gel/carbon/cloth). Complies with IEC 60601-1, IEC 60601-1-2, IEC 60601-2-10, and 21 CFR 898.
Indications for Use
Indicated for temporary relief of pain in shoulder, waist, back, upper/lower extremities due to exercise or household strain; also for stimulation of healthy muscles to improve/facilitate muscle performance. OTC use.
Regulatory Classification
Identification
A transcutaneous electrical nerve stimulator for pain relief is a device used to apply an electrical current to electrodes on a patient's skin to treat pain.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, with the profiles overlapping each other.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
August 6, 2014
IQ Technologies, Inc. Elli Josef 1631 E. Sunset Rd., Ste. C 103 Las Vegas, NV 89119
Re: K131290 Trade/Device Name: IO Technologies Regulation Number: 21 CFR 882.5890 Regulation Name: Transcutaneous electric nerve stimulator for pain relief Regulatory Class: Class II Product Code: NUH: NGX Dated: June 30, 2014 Received: July 2, 2014
Dear Mr. Josef:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Isting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical
{1}------------------------------------------------
device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Felipe Aguel -S
Carlos L. Peña, Ph.D., M.S. for Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K131290
Device Name IQ Technologies
Indications for Use (Describe)
To be used for temporary relief of pain associated with sore and aching muscles in the shoulder, waist, back, upper extremities (arm), and lower extremities (leg) due to strain from exercise or normal household work activities.
It is intended to be used to stimulate healthy muscles in order to improve and facilitate muscle performance.
Type of Use (Select one or both, as applicable)
_ Prescription Use (Part 21 CFR 801 Subpart D)
> Over-The-Counter Use (21 CFR 801 Subpart C)
#### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
#### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
# Date: 2014.08.06 22:22:02 -04'00'
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
#### 510(k) Summary
This 510(k) Summary of safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR 807.92.
#### 1. Submitter's Information
Submitter: IQ Technologies Inc. Address: 1631 E. Sunset Road, C103, Las Vegas, NV 89119 Contact Person: Elli Josef Tel: 702-260-8829 Fax: 702-260-8840 Email: elijosef57@gmail.com Date of Preparation: 04/11/2013
#### 2. Correspondent's Information
Correspondent: IQ Technologies Inc. Address: 1631 E. Sunset Road, C103, Las Vegas, NV 89119 Contact Person: Elli Josef Tel: 702-260-8829 Fax: 702-260-8840 Email: elijosef57@gmail.com
#### 3. Subject Device
Trade/Device Name: IQ Technologies Common Name: Transcutaneous electrical nerve stimulator (TENS) and Powered Muscle Stimulator (PMS) Classification Name: Stimulator, Nerve, Transcutaneous, Over-The-Counter (OTC) Regulation Description: Transcutaneous electrical nerve stimulator for pain relief Regulation Medical Specialty: Neurology Review Panel: Neurology Product Code: NUH, NGX Regulation Number: 21 CFR 882.5890 Device Class: II Use: Over-The-Counter
## 4. Predicate device
Predicate Device: Powered Muscle Stimulator. JO-5C 510(k) Number: K102598 Use: Over-The-Counter Submitter: Hi-Dow International, Inc.
Predicate Device: TENS&PMS 510(k) Number: K121757 Use: Over-The-Counter Submitter: Healthmate International, LLC
#### 5. Description of Subject Device
The subject device is a Transcutaneous Electrical Nerve Stimulator (TENS) and Powered Muscle
{4}------------------------------------------------
Stimulator (PMS), intended for the over-the-counter use to temporarily relieve pain and stimulate muscle in different body areas. The double-channel subject device, which is compact, portable, and microprocessor-controlled, delivers a gentle electrical pulse through the connecting wires and electrode pads to the user's skin for pain relief and muscle stimulation. The electrode pad used consists of gel, carbon film, cloth backing, and electrode connector, which is 510(k)-cleared and biocompatible. According to the need of users, the pulse intensity can be adjustable on the front control panel of the device.
# 6. Intended Use of Subject Device
# TENS:
To be used for temporary relief of pain associated with sore and aching muscles in the shoulder, waist, back, upper extremities (arm), and lower extremities (leg) due to strain from exercise or normal household work activities.
# PMS:
It is intended to be used to stimulate healthy muscles in order to improve and facilitate muscle performance.
# 7. Summary of Substantial Equivalence
The following table summarizes the comparison between the subject device and predicate devices, indicating the technical characteristics, specifications, and intended use of the subject device are substantially equivalent to those of the predicate devices.
| | Subject Device | Predicate Device | Predicate Device |
|---------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Number | K131290 | K102598 | K121757 |
| Device Name | IQ Technologies | Hi-Dow | HealthmateForever |
| Intended Use | To be used for temporary<br>relief of pain associated<br>with sore and aching<br>muscles in the shoulder,<br>waist, back, upper<br>extremities (arm), and<br>lower extremities (leg)<br>due to strain from<br>exercise or normal<br>household work<br>activities.<br>It is intended to be used<br>to stimulate healthy<br>muscles in order to<br>improve and facilitate<br>muscle performance. | To be used for<br>temporary relief of pain<br>associated with sore<br>and aching muscles in<br>the shoulder, waist,<br>back, neck, upper<br>extremities (arm), and<br>lower extremities (leg)<br>due to strain from<br>exercise or normal<br>household work<br>activities.<br>It is intended to be used<br>to stimulate healthy<br>muscles in order to<br>improve and facilitate<br>muscle performance. | To be used for<br>temporary relief of pain<br>associated with sore<br>and aching muscles in<br>the shoulder, waist,<br>back, neck, upper<br>extremities (arm), and<br>lower extremities (leg)<br>due to strain from<br>exercise or normal<br>household work<br>activities.<br>It is intended to be used<br>to stimulate healthy<br>muscles in order to<br>improve and facilitate<br>muscle performance. |
| Power Source | DC 3.7V Lithium Battery | DC 3.7V Lithium<br>Battery | DC 3.7V Lithium<br>Battery |
| Number of Output Channels | 2 | 2 | 2 |
{5}------------------------------------------------
| Automatic Overload Trip | No | No | No |
|------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------|--------------------------------------------------------------------------------------------------------------------------|
| Automatic No-Load Trip | No | No | No |
| Automatic Shut Off | Yes | Yes | Yes |
| User Override Control | Yes | Yes | Yes |
| Indicator | Yes | Yes | Yes |
| Waveform | Pulsed | Pulsed | Pulsed |
| Shape | Rectangular | Rectangular | Rectangular |
| Maximum output voltage<br>(Volts +/- 20%) at 500Ω | Mode 1: 42<br>Mode 2: 63.2<br>Mode 3: 64<br>Mode 4: 34.4<br>Mode 5: 32<br>Mode 6: This mode cycles the above five modes | 62.4 | 70 |
| Maximum output voltage<br>(Volts +/- 20%) at 2KΩ | Mode 1: 80.8<br>Mode 2: 94.4<br>Mode 3: 87.2<br>Mode 4: 68<br>Mode 5: 64<br>Mode 6: This mode cycles the above five modes | 79.2 | 90 |
| Maximum output voltage<br>(Volts +/- 20%) at 10kΩ | Mode 1: 129<br>Mode 2: 129<br>Mode 3: 96.8<br>Mode 4: 128<br>Mode 5: 119<br>Mode 6: This mode cycles the above five modes | 84 | 100 |
| Maximum output current<br>(mA +/- 20%) at 500Ω | Mode 1: 84<br>Mode 2: 126.4<br>Mode 3: 128<br>Mode 4: 68.8<br>Mode 5: 64<br>Mode 6: This mode cycles the above five modes | 124.8 | 140 |
| Maximum output current<br>(mA +/- 20%) at 2KΩ | Mode 1: 40.4<br>Mode 2: 47.2<br>Mode 3: 43.6<br>Mode 4: 34<br>Mode 5: 32<br>Mode 6: This mode cycles the above five modes | 39.6 | 45 |
| Maximum output current<br>(mA +/- 20%) at 10KΩ | | 8.4 | 10 |
| Mode 1: 12.9<br>Mode 2: 12.9<br>Mode 3: 9.7<br>Mode 4: 12.8<br>Mode 5: 11.9<br>Mode 6: This mode cycles the above five modes | | | |
| Pulse Width (µSec) | 100 | 100 | 90 |
| Pulse period (mSec) | 10~840 | 16.3~781 | 10~836 |
| Frequency (Hz) | Mode 1: 69.4<br>Mode 2: 12.3~54.3<br>Mode 3: 1.2<br>Mode 4: 100<br>Mode 5: 100<br>Mode 6: This mode cycles the above five modes | 61.3 | 100 |
| Maximum Phase charge<br>(µC) at 500Ω | 16.8 | 17.9 | 15.7 |
| Maximum current density<br>(mA/cm²) at 500Ω | Mode 1: 3.36<br>Mode 2: 5.06<br>Mode 3: 5.12<br>Mode 4: 2.75<br>Mode 5: 2.56<br>Mode 6: This mode cycles the above five modes | 9.92 | Mode 1: 6.48<br>Mode 2: 5.6<br>Mode 3: 7.2<br>Mode 4: 5.6<br>Mode 5: 5.6<br>Mode 6: This mode cycles the above modes |
| Maximum average power<br>density (mW/cm²) at 500Ω | Mode 1: 2.11<br>Mode 2: 0.85~3.75<br>Mode 3: 0.08<br>Mode 4: 2.05<br>Mode 5: 1.64<br>Mode 6: This mode cycles the above five modes | 2.72 | Mode 1: 0.52<br>Mode 2: 0.39<br>Mode 3: 0.65<br>Mode 4: 0.39<br>Mode 5: 0.39<br>Mode 6: This mode cycles the above modes |
| Compliance with Voluntary<br>Standards | IEC 60601-1, IEC 60601-1-2, IEC 60601-2-10 | IEC 60601-1, IEC 60601-1-2, IEC 60601-2-10 | IEC 60601-1, IEC 60601-1-2, IEC 60601-2-10 |
| Compliance with 21 CFR<br>898 | Yes | Yes | Yes |
{6}------------------------------------------------
# 8. Substantial Equivalence
The operational principle of the above predicate devices is to generate small pulses of electrical current and delivers the pulses to the user's skin through adhesive electrode pads such that the underlying nerves and/or muscles are activated.
{7}------------------------------------------------
Identically, the subject device generates small pulses of electrical current and delivers the pulses to the user's skin through adhesive electrode pads such that the underlying nerves and/or muscles are activated.
The comparison between the subject device and predicate devices demonstrates the technical characteristics, specifications, and intended use of the subject device are substantially equivalent to those of the predicate devices.
The differences, such the output voltage and current, between the subject device and the predicate devices are insignificant in terms of safety or effectiveness. The verification and validation tests, such as IEC 60601-1, IEC 60601-1-2, and IEC 60601-2-10, further demonstrate these differences maintain the same safety and effectiveness as those of the 510(k) cleared predicate devices. In other words, these differences do not affect the intended use or alter the fundamental technology of the device. There are no new safety or effectiveness issues concerning the subject device, which offers substantially equivalent technical specifications, features, intended use, and effective results as the predicate devices.
Concerns of the safe and proper use of the electrode pads have been fully addressed by making the use conscious of the proper placement of the electrode pads and appropriate operations of the device through details in the labeling. The electrode pads cleared in K090198 are to be used with the subject device.
# 9. Non-Clinical Tests Performed
The subject device does not conduct, nor rely upon, clinical tests to determine substantial equivalence. Non-clinical tests were performed on the subject device in order to validate the design and to assure conformance with the following voluntary design standards in connection with medical device electrical safety, and electromagnetic compatibility.
- (a) IEC 60601-1 "Medical Electrical Equipment Part 1: General requirements for basic safety and essential performance".
- (b) IEC 60601-1-2 "Medical Electrical Equipment Part 1-2: General Requirements for Safety -Collateral standard: Electromagnetic Compatibility - Requirements and Tests".
- (c) IEC 60601-2-10 "Medical electrical equipment -- Part 2-10: Particular requirements for the basic safety and essential performance of nerve and muscle stimulators".
In addition to the compliance of voluntary standards, the verification of software used in the subject device has been carried out according to the FDA Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices. The biocompatible electrodes, as the accessory of the subject device, also meet the requirement of safety.
## 10. Conclusion
The tests performed and the comparison of technical characteristics, specifications, and intended use demonstrate the subject device is substantially equivalent to the predicate device. Therefore, the subject device is as safe and effective as the foregoing identified OTC predicate devices that have been legally marketed in the United States.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.