K230997 · Evolution Devices, Inc. · GZI · Dec 28, 2023 · Neurology
Device Facts
Record ID
K230997
Device Name
EvoWalk 1.0 System
Applicant
Evolution Devices, Inc.
Product Code
GZI · Neurology
Decision Date
Dec 28, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.5810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The EvoWalk 1.0 System is intended to provide ankle dorsiflexion in adult individuals with foot drop and/or to assist knee flexion or extension in adult individuals with muscle weakness related to upper motor neuron disease/injury (e.g., stoke, damage to pathways to the spinal cord). The EvoWalk 1.0 System electrically stimulates muscles in the affected leg to provide ankle dorsiflexion of the flexion or extension; thus, it also may improve the individual's gait.
Device Story
Wearable, non-invasive functional electrical stimulation (FES) device; provides ankle dorsiflexion and knee flexion/extension assistance. Input: motion data from embedded inertial measurement unit (IMU) sensor. Operation: microprocessor-controlled stimulation triggered by gait cycle detection; delivers biphasic symmetric electrical pulses to leg muscles via hydrogel electrodes. Output: electrical stimulation to muscles. Used in home or healthcare settings; operated by patient or clinician. Clinician uses mobile app to set stimulation energy and temporal parameters; patient uses app or device button to toggle modes (standby/active) and tune intensity. Benefits: improved gait for patients with foot drop or upper motor neuron weakness.
Clinical Evidence
No clinical data. Evidence consists of bench testing, including stimulation output waveform verification, motion analysis to validate the gait event detection algorithm, electrical safety testing (IEC 60601-1, IEC 60601-1-11, IEC 60601-2-10), electromagnetic compatibility (IEC 60601-1-2), and software validation (IEC 62304).
Technological Characteristics
Wearable FES device; 2-channel stimulation; biphasic symmetric waveform. Power: 3.7V 700mAh Li-Ion battery. Electrodes: hydrogel (square, rectangular, round). Connectivity: Bluetooth Low Energy (BLE) for mobile app. Microprocessor-controlled. Standards: IEC 60601-1, IEC 60601-1-11, IEC 60601-2-10, IEC 60601-1-2, IEC 62304.
Indications for Use
Indicated for adult patients with foot drop or muscle weakness related to upper motor neuron disease/injury (e.g., stroke, spinal cord damage) to provide ankle dorsiflexion and/or assist knee flexion/extension to improve gait.
Regulatory Classification
Identification
An external functional neuromuscular stimulator is an electrical stimulator that uses external electrodes for stimulating muscles in the leg and ankle of partially paralyzed patients (e.g., after stroke) to provide flexion of the foot and thus improve the patient's gait.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
December 28, 2023
Evolution Devices, Inc. Pierluigi Mantovani CEO 1111 Broadway FL 3. Oakland, California 94607
Re: K230997
Trade/Device Name: EvoWalk 1.0 System Regulation Number: 21 CFR 882.5810 Regulation Name: External Functional Neuromuscular Stimulator Regulatory Class: Class II Product Code: GZI Dated: December 5, 2023 Received: December 18, 2023
Dear Pierluigi Mantovani:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Lauren E. Woodard -S
for Amber Ballard, PhD Assistant Director DHT5B: Division of Neuromodulation and Rehabilitation Devices
{2}------------------------------------------------
OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## Indications for Use
510(k) Number (if known) K230997
Device Name EvoWalk 1.0 System
#### Indications for Use (Describe)
The EvoWalk 1.0 System is intended to provide ankle dorsiflexion in adult individuals with foot drop and/or to assist knee flexion or extension in adult individuals with muscle weakness related to upper motor neuron disease/injury (e.g., stoke, damage to pathways to the spinal cord). The EvoWalk 1.0 System electrically stimulates muscles in the affected leg to provide ankle dorsiflexion of the flexion or extension; thus, it also may improve the individual's gait.
Type of Use (Select one or both, as applicable)
| Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-size: 20px;">☑</span> |
|----------------------------------------------|-----------------------------------------|
|----------------------------------------------|-----------------------------------------|
| Over-The-Counter Use (21 CFR 801 Subpart C) | <span style="font-size: 20px;">☐</span> |
|---------------------------------------------|-----------------------------------------|
|---------------------------------------------|-----------------------------------------|
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(k) Summary
Evolution Devices, Inc. EvoWalk 1.0 System
510(k) Summary: EvoWalk 1.0 System
Company Name: Evolution Devices, Inc.
Contact Person: Pierluigi Mantovani Chief Executive Officer Evolution Devices, Inc. 1111 Broadway FL 3, Oakland CA 94607 United States Phone Number: 6619930685 Email: pierluigi@evolutiondevices.com
Date Prepared: November 15, 2023
Trade Name: EvoWalk 1.0 System
Classification: Name: External functional neuromuscular stimulator Product Code: GZI Regulation Number: 21 CFR 882.5810 External functional neuromuscular stimulator Class: II Classification Panel: Neurology
Predicate devices: Company: Bioness Inc. Device: L300 Go System (K173682)
Purpose of the traditional 510(k): This Traditional 510(k) is submitted to clear a new device, the EvoWalk 1.0 System.
#### Device description:
The EvoWalk 1.0 system is a wearable non-invasive functional electrical stimulation (FES) device intended to provide ankle dorsiflexion in adult individuals with foot drop and/or to assist knee flexion in adult individuals with muscle weakness related to upper motor neuron diseaselinjury (e.g., stroke, damage to pathways in the spinal cord), thus providing treatment for foot drop. EvoWalk 1.0 can be used in a healthcare facility and home settings.
The EvoWalk 1.0 system consists of the EvoWalk device with an elastic band and built-in rechargeable lithium polymer battery, and the accessories including the EvoWalk mobile application, micro USB type charge cable with charge port, lead cables (one single channel and one dual channel), eight sets of electrodes for each channel, instructions for use (IFU), and a carrying case.
#### Indications for use:
The EvoWalk 1.0 system is intended to provide ankle dorsiflexion in adult individuals with foot drop and/or to assist knee flexion or extension in adult individuals with muscle weakness related to upper motor neuron
{5}------------------------------------------------
disease/injury (e.g., stroke, damage to pathways to the spinal cord). The EvoWalk 1.0 system electrically stimulates muscles in the affected leg to provide ankle dorsiflexion of the floxion or extension; thus, it also may improve the individual's gait.
## Substantial Equivalence:
#### Table 1. Predicate Comparison Chart
| Trade Name | EvoWalk 1.0 System<br>(K230997) | L300 Go System<br>(K173682) |
|-------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer | Evolution Devices Inc. | Bioness Inc. |
| 510(k) Number | K230997 | K173682 |
| Product Code | GZI | GZI & IPF |
| Indication for<br>Use | The EvoWalk 1.0 system is intended to<br>provide ankle dorsiflexion in adult individuals<br>with foot drop and/or to assist knee flexion or<br>extension in adult individuals with muscle<br>weakness related to upper motor neuron<br>disease/injury (e.g., stroke, damage to<br>pathways to the spinal cord). The EvoWalk<br>1.0 system electrically stimulates muscles in<br>the affected leg to provide ankle dorsiflexion<br>of the foot and/or knee flexion or extension;<br>thus, it also may improve the individual's gait. | The L300 Go System is intended to provide ankle<br>dorsiflexion in adult and pediatric individuals with<br>foot drop and/or to assist knee flexion or<br>extension in adult individuals with muscle<br>weakness related to upper motor neuron<br>disease/injury (e.g., stroke, damage to pathways<br>to the spinal cord). The L300 Go System<br>electrically stimulates muscles in the affected leg<br>to provide ankle dorsiflexion of the foot and/or<br>knee flexion or extension; thus, it also may<br>improve the individual's gait.<br>The L300 Go System may also:<br>• Facilitate muscle re-education<br>• Prevent/retard disuse atrophy<br>• Maintain or increase joint range of motion<br>• Increase local blood flow |
| Number of<br>Output Modes | Biphasic Symmetric Mode only | 2 modes: Biphasic Asymmetric and Symmetric |
| Number of<br>Programs | • Gait<br>• Tuning<br>• Standby | • Gait<br>• Training/Exercise<br>• Clinician mode |
| Regulated<br>Current or<br>Regulated<br>Voltage | Current | Current |
| Power Source | Rechargeable, Li-Ion, Prismatic, 3.7 V, 700<br>mAh | Control Unit: Li Coin Cell, CR2032, 3 V, 240 mAh<br>EPG: Rechargeable, Li-Ion, Prismatic, 3.7 V,<br>1000 mAh<br>Foot Sensor: Li Coin Cell, CR2032, 3 V, 240 mAh |
| Microprocessor<br>-Controlled | Yes | Yes |
{6}------------------------------------------------
| Max Output<br>Current<br>(± 10%) | Thigh Channel:<br>102.4 mA @ 500<br>Ohm load | Lower Leg Channel:<br>102.4 mA @ 500 Ohm<br>load | Thigh FSC: 100<br>mA @ 500<br>Ohm load | L300 Lower Leg FSC: 100<br>mA @ 500 Ohm load |
|-------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Max Average<br>Current Density<br>[mARMS/cm²]<br>[Over smallest<br>electrode] | Thigh Channel:<br>rectangular<br>electrodes<br>0.26 mArms/cm²<br>(500 Ω, Irms=13.0<br>mA, electrode area<br>of 50 cm²) | Lower Leg Channel:<br>Square electrodes<br>0.56 mArms/cm²<br>(500 Ω, Irms=13.0 mA,<br>electrode area of 25<br>cm²)<br>round electrodes<br>1.37 mArms/cm²<br>(Area 10.2 cm²) | Thigh EPG:<br>0.18 mArms/cm²<br>(500 Ω, Irms=13.0<br>mA, electrode area of<br>74cm²) | Lower Leg EPG:<br>small cuff, gel electrodes<br>1.27 mArms/cm²<br>(500 Ω, Irms=13.0 mA,<br>electrode area of 10.2cm²) |
| Max Average<br>Power Density,<br>(mW/cm²) | Thigh Channel:<br>rectangular<br>electrodes<br>A = 10.1x5.0 cm =<br>50 cm²<br>1.7 mW/cm² @ 500<br>Ω | Lower Leg Channel:<br>square electrodes<br>A = 5x5 cm = 25 cm²<br>3.4 mW/cm² @ 500 Ω<br>round electrodes<br>A =10.2cm²<br>8.3 mW/cm² @ 500 Ω | Thigh EPG:<br>1.1 mW/cm²<br>(500 Ω, Irms=13.0<br>mA, electrode area of<br>74 cm²) | Lower Leg EPG: small cuff,<br>gel electrodes<br>8.3 mW/cm²<br>(500 Ω, Irms=13.0<br>mA, electrode area of 10.2<br>cm²)<br>regular cuff, gel electrodes<br>5.3 mW/cm²<br>(500 Ω, Irms=13.0<br>mA, electrode area of 15.9<br>cm²) |
| Stimulation<br>Channels | 2 channels | | Thigh EPG: 1 | Lower Leg EPG: 2<br>(functioning as a single<br>channel with separately-<br>adjustable medial / lateral<br>stimulation intensity) |
| Electrodes used<br>in the system | Hydrogel Electrodes:<br>Square Electrodes:<br>Male Snap connector;<br>Surface area 25.0 cm²<br>Rectangular Electrodes:<br>Male Snap connector;<br>Surface area 50.0 cm²<br>Round Electrode:<br>Male Snap connector;<br>Surface area: 10.2cm² | | Lower Leg FSC:<br>• 2 Hydro-Gel electrodes assembled on<br>electrode bases, or<br>• 2 non-woven cloth electrodes assembled<br>on electrode bases, or<br>• 2 non-woven cloth electrodes attached with<br>snaps (also called "QuickFit" electrodes), or<br>• 3 non-woven cloth electrodes attached with<br>snaps (segmented electrodes [also called<br>"steering" electrodes], using common anode<br>to allow separate adjustment of medial and<br>lateral stimulation)<br>Thigh FSC:<br>• 2 single, non-woven cloth electrodes<br>attached with snaps | |
{7}------------------------------------------------
| Clinician<br>Control/<br>Programming | Clinician uses the EvoWalk 1.0 companion<br>application to set stimulation energy and<br>temporal parameters related to the functional<br>stimulation performance for dorsiflexion<br>control and/or knee weakness control | Clinician uses the Clinician Programmer (CAPP)<br>to set stimulation energy and temporal<br>parameters related to the functional stimulation<br>performance for dorsiflexion control and/or knee<br>weakness control |
|----------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Clinician<br>Programmer<br>(CAPP)<br>Platform | Mobile Application (iOS and Android) | Tablet PC |
| User Control | Wearable unit:<br>Users can press the button to power on,<br>power off, enter pairing mode and toggle<br>between standby mode (pause all stimulation)<br>and active mode.<br><br>Mobile application:<br>From the app the user can set stimulation<br>parameters, review usage data, toggle<br>between standby mode (pause all<br>stimulation), active mode, and tune the<br>stimulation intensity before and during<br>ambulation. | Using hand-held Control Unit, the mobile<br>application (MAPP), or the EPG-based interface,<br>the user can:<br>• Turn system On/Off (via EPG only) and<br>Start/Stop stimulation<br>• Select Gait/Training program<br>• Fine-tune stimulation intensity around<br>working point set by the clinician<br>• Test L300 Lower Leg EPG & Thigh EPG |
| Stimulation<br>trigger source<br>(when used for<br>gait) | In gait mode, stimulation is triggered by<br>the inertial measurement unit motion sensor<br>embedded in the stimulator printed circuit<br>board | In gait mode, stimulation is triggered by:<br>(1) the motion sensor embedded in the EPG;<br>or<br>(2) Foot Sensor that detects Heel On & Heel<br>Contact events during gait and transmits<br>them wirelessly to the lower and thigh<br>EPGs. |
| Communication<br>method | EvoWalk 1.0 system - mobile user interface<br>application:<br>wireless Bluetooth (Low Energy)<br>communication protocol | Control Unit – Lower Leg /Thigh EPG: wireless<br>Bluetooth (Low Energy) communication protocol<br><br>Gait Sensor - Lower Leg/Thigh EPG: wireless<br>Bluetooth (Low Energy) communication protocol<br><br>Clinician Programmer - EPG: wireless Bluetooth<br>(Low Energy) communication protocol<br><br>MAPP - Lower Leg /Thigh EPG: wireless<br>Bluetooth (Low Energy) communication<br>protocol |
| Maximum Phase<br>Charge, (µC)<br>@500 Ω | Lower leg:<br>$200μs * 102mA = 20μC$<br>Thigh:<br>$200μs * 102mA = 20μC$ | Not Reported |
The indication for use is the same for EvoWalk 1.0 System and Bioness L300 Go System except EvoWalk 1.0 system is for adults only and Bioness L300 Go System is for pediatrics and adults. Additionally, EvoWalk 1.0 System is not indicated for muscle re-education, prevent/retard disuse atrophy, maintain or increase joint range of motion, or increase local blood flow. Both devices are worn on the user's leg, use hydrogel electrodes, have two channels of stimulation, use an embedded sensor to control stimulation, and both devices have a user interface through a mobile application. The EvoWalk 1.0 system eliminates some optional features of the Bioness L300 Go System including the separate thigh cuff, heel sensor, semi-rigid cuff, cloth electrode option, and dedicated remote control. Both systems control the stimulation in the gait program by reading the embedded motion sensor and applying algorithms and user-specified parameters such that the stimulation activates at a specific time within each gait cycle.
{8}------------------------------------------------
## Summary of Nonclinical Tests Submitted
The EvoWalk 1.0 system has been verified and validated successfully for its intended use through a combination of original bench testing and verification and validation of all software and firmware. Based on the result of the nonclinical testing, Evolution Devices, Inc. concludes that the device is substantially equivalent to the predicate Bioness L300 Go System. The tests listed have been conducted to demonstrate that the EvoWalk 1.0 system performs as intended and is substantially equivalent to the predicate device.
- Technological Characteristics of Powered Muscle Stimulation Stimulation Output Waveforms ●
- Motion analysis to validate the gait event detection algorithm ●
- Electrical Safety according to IEC 60601-1; IEC 60601-1-11 ●
- Muscle and Nerve Stimulators according to IEC 60601-2-10 ●
- Electromagnetic compatibility according to IEC 60601-1-2 ●
- Software validation according to IEC 62304 ●
### Conclusion:
Evolution Devices believes the EvoWalk 1.0 system is substantially equivalent to the Bioness L300 GO system (K173682), and does not raise any new issues or concerns of safety or effectiveness. The evidence presented in the bench testing and non-clinical testing is sufficient to support the safety and effectiveness of the device.
The testing ensures that the device is safe for the intended use and will not cause harm to the user or interfere with other medical equipment or electronic devices. The software verification and validation testing ensure that the software used in the device operates as intended and is free of defects or errors.
In conclusion, the non-clinical evidence and testing considered for the bench testing 510k FDA submission for the EvoWalk 1.0 system supports the safety and effectiveness of the device.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.