K200262 · Bioness, Inc. · GZI · Mar 4, 2020 · Neurology
Device Facts
Record ID
K200262
Device Name
L100 Go System
Applicant
Bioness, Inc.
Product Code
GZI · Neurology
Decision Date
Mar 4, 2020
Decision
SESE
Submission Type
Special
Regulation
21 CFR 882.5810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The L100 Go System is intended to provide ankle dorsiflexion in adults with foot drop or muscle weakness related to upper motor neuron disease/injury (e.g., stroke, damage to pathways to the spinal cord). The L100 Go System electrically stimulates muscles in the affected leg to provide ankle dorsiflexion of the foot; thus, it also may improve the adult's gait. The L100 Go System may also: Facilitate muscle re-education. Prevent/retard disuse atrophy. Maintain or increase joint range of motion. Increase local blood flow.
Device Story
L100 Go System is an external functional neuromuscular stimulator for foot drop. Input: motion sensor data (2D tilt) from External Pulse Generator (EPG) worn on lower leg. Operation: EPG delivers transcutaneous electrical stimulation to lower leg muscles to trigger ankle dorsiflexion during gait. Used in clinic (by clinician for configuration) and home (by patient for ambulation/training). Output: electrical stimulation pulses. Clinician uses tablet-based Clinician Application (CAPP) to configure stimulation parameters. Patient uses mobile app (MAPP) or EPG interface to control stimulation intensity, select modes (Gait, Training), and monitor activity. Benefits: improved gait, muscle re-education, atrophy prevention, increased blood flow, and joint mobility maintenance.
Clinical Evidence
Bench testing only. Verification and validation testing performed on the modified cuff design to ensure mechanical strength and functionality. No clinical data presented.
Technological Characteristics
External functional neuromuscular stimulator; battery-powered (3.7V Li-Ion EPG). Regulated current output (max 100mA @ 500 Ohm). Frequency 10-45 Hz; phase duration 100-900μs. Connectivity: Bluetooth Low Energy (BLE) for MAPP and CAPP. Standards: IEC 60601-1, IEC 60601-2-10, IEC 60950-1, IEC 60601-1-11, IEC 60601-1-2, FCC Part 15. Software-controlled with automatic overload/no-load trip and shut-off. Single-channel stimulation.
Indications for Use
Indicated for adults with foot drop or muscle weakness due to upper motor neuron disease/injury (e.g., stroke, spinal cord damage) to provide ankle dorsiflexion and improve gait; also for muscle re-education, atrophy prevention, joint range of motion maintenance, and increased local blood flow.
Regulatory Classification
Identification
An external functional neuromuscular stimulator is an electrical stimulator that uses external electrodes for stimulating muscles in the leg and ankle of partially paralyzed patients (e.g., after stroke) to provide flexion of the foot and thus improve the patient's gait.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA name on the right. The symbol on the left is a stylized image of a human figure, while the FDA name on the right is written in blue letters. The words "U.S. FOOD & DRUG ADMINISTRATION" are written in a clear, sans-serif font.
March 4, 2020
Bioness Inc. Shanna Hu Regulatory Affairs Manager, FES Technology 25103 Rye Canyon Loop Valencia, California 91355
Re: K200262
Trade/Device Name: L100 Go System Regulation Number: 21 CFR 882.5810 Regulation Name: External Functional Neuromuscular Stimulator Regulatory Class: Class II Product Code: GZI, IPF Dated: January 31, 2020 Received: February 3, 2020
Dear Shanna Hu:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
For Vivek Pinto, Ph.D. Director DHT5B: Division of Neuromodulation and Physical Medicine Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### Indications for Use
510(k) Number (if known)
Device Name L100 Go System
Indications for Use (Describe)
The L100 Go System is intended to provide ankle dorsiflexion in adults with foot drop or muscle weakness related to upper motor neuron disease/injury (e.g., stroke, damage to the spinal cord). The L100 Go System electrically stimulates muscles in the affected leg to provide ankle dorsiflexion of the foot; thus, it also may improve the adult's gait.
The L100 Go System may also:
- · Facilitate muscle re-education
- · Prevent/retard disuse atrophy
- · Maintain or increase joint range of motion
- · Increase local blood flow
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
] Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for Bioness. The logo features an orange sun-like symbol on the left, followed by the word "Bioness" in gray, and the words "LiveOn" in a smaller font below it. The logo is simple and modern, and the colors are bright and cheerful.
| 510(k) Summary: | L100 Go System |
|--------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Applicant Name: | Bioness Inc. |
| Contact Person(s): | Shanna Hu<br>Regulatory Affairs Manager, FES Technology<br>Bioness Inc.<br>25103 Rye Canyon Loop<br>Valencia, CA 91355, U.S.A<br>Office Number: (661) 362-6694<br>Email: shanna.hu@bioness.com<br>Mercedes Bayani<br>Global Vice-President, Clinical & Regulatory Affairs<br>Bioness Inc.<br>Office Number: (661) 902-5324<br>Fax Number: (661) 362-4851<br>Email: mercedes.bayani@bioness.com |
| Date Prepared: | March 4, 2020 |
| Trade Name: | L100 Go System |
| Classification: | Name: External functional neuromuscular stimulator<br>Product Code: GZI and IPF<br>Regulation No: 21 CFR § 882.5810, § 890.5850<br>Class: II<br>Classification Panel: Neurology |
Establishment Registration No.: 3004553866
Reason for Submission: Device Modifications
Type of Submission: Special 510(k)Type of Submission:Special 510(k)
# Predicate Device:
Company: Bioness, Inc. Device: L300 Go System, K190285
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the text "510(k) Summary" in a bold, sans-serif font. The text is centered horizontally and appears to be the title or heading of a document. The background is plain white, providing a clean and simple presentation of the text.
Image /page/4/Picture/1 description: The image shows the logo for Bioness. The logo consists of an orange sun-like symbol on the left, followed by the word "Bioness" in gray, and the words "LiveOn" in a smaller gray font below the word "Bioness". The sun-like symbol has eight orange rays, each made up of two lines that curve inward.
### Purpose of this Special 510(k):
This Special 510(k) is submitted for a system called L100 Go System which is the same device as the Lower Leg configuration of the predicate L300 Go System with modified cuff design. None of the modifications affect the intended use of the device nor do they alfer the fundamental scientific technology of the device.
### Device Description:
The L100 Go System is designed to improve gait in adults suffering from foot drop or muscle weakness. The L100 Go System can also deliver transcutaneous stimulation to the muscles in the lower leg to facilitate muscle re-education, prevent/retard disuse atrophy, maintain or increase joint range of motion, and/or increase local blood flow.
The L100 Go system consists of the following L300 Go components/accessories with the exception of a modified lower leg cuff:
- 1) External Pulse Generator (EPG), which can be plugged into Lower Leg Cuff. EPG contains user interface including control and indications. EPG also contains integrated motion sensors enabling detecting gait events.
- 2) Lower Leg Cuff, including cradle for the EPG. This is the only L300 Go component modified for the L100 Go.
- 3) Clinician Application (CAPP), installed on tablet PC. CAPP is used by a trained clinician during configuration of the system for optimal fitting to the patient.
- 4) Power supply (charger) with two USB ports and a proprietary cable to charge the EPG.
- 5) Tester, which is used for trouble shooting to confirm that stimulation is being delivered.
- 6) Optional Mobile Application (MAPP), based on the iOS and Android SmartPhone platform enabling the patients to wirelessly control the EPG(s) and retrieve and monitor their daily activity.
- 7) Electrode and electrode bases which are attached to the inner side of the lower leg cuff.
The L100 Go System can be operated in one of the following modes:
- Gait Mode
- Training Mode •
- . Clinician Mode
Gait Mode is used for walking, and it can be selected by the clinician and also by the patient. Training Mode is used to train muscles when patients are not walking (for example, sitting or lying down), and it can be selected by either the clinician or the patient. Clinician Mode allows the clinician to apply enhanced training and is only available to the clinicians.
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the logo for Bioness LiveOn. The logo features an orange sun-like symbol on the left, followed by the word "Bioness" in gray, and "LiveOn" in a smaller font below it. The logo is simple and modern, with a focus on the company name.
### Indications for Use:
The L100 Go System is intended to provide ankle dorsiflexion in adults with foot drop or muscle weakness related to upper motor neuron disease/injury (e.g., stroke, damage to pathways to the spinal cord). The L100 Go System electrically stimulates muscles in the affected leg to provide ankle dorsiflexion of the foot; thus, it also may improve the adult's gait.
The L100 Go System may also:
- Facilitate muscle re-education .
- Prevent/retard disuse atrophy .
- Maintain or increase joint range of motion ●
- Increase local blood flow ●
#### Modifications Addressed in this Special 510(k)
This Special 510(k) addresses the new L100 Go Cuff which is a modification of the predicate L300 Go lower leg cuffs. The L100 Go Cuff combines the design elements of both the L300 Go lower leg Reqular and Small cuffs.
#### Summary of Technological Characteristics
The Table below summarizes the technological characteristics of the subject device in comparison to those of the predicate device, specifically, the lower leg configuration of the predicate device.
| | L300 Go System<br>(K190285) | L100 Go System (this<br>submission) | Equivalency<br>Assessment |
|-------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer | Bioness Inc. | Bioness Inc. | Same |
| 510(k) number | K190285 | To be assigned | N/A |
| Product code | GZI & IPF | GZI & IPF | Same |
| | L300 Go System<br>(K190285) | L100 Go System (this<br>submission) | Equivalency<br>Assessment |
| Intended use | The L300 Go System is<br>intended to provide ankle<br>dorsiflexion in adult and<br>pediatric individuals with<br>foot drop and/or to assist<br>knee flexion or extension in<br>adult individuals with<br>muscle weakness related to<br>upper motor neuron<br>disease/injury (e.g., stroke,<br>damage to pathways to the<br>spinal cord). The L300 Go<br>System electrically<br>stimulates muscles in the<br>affected leg to provide<br>ankle dorsiflexion of the<br>foot and/or knee flexion or<br>extension; thus, it also may<br>improve the individual's<br>gait.<br>The L300 Go System may<br>also:<br>• Facilitate muscle re-<br>education<br>• Prevent/retard disuse<br>atrophy<br>• Maintain or increase joint<br>range of motion<br>• Increase local blood flow | The L100 Go System is<br>intended to provide ankle<br>dorsiflexion in adults with<br>foot drop or muscle<br>weakness related to upper<br>motor neuron disease/injury<br>(e.g., stroke, damage to<br>pathways to the spinal cord).<br>The L100 Go System<br>electrically stimulates<br>muscles in the affected leg<br>to provide ankle dorsiflexion<br>of the foot; thus, it also may<br>improve the adult's gait.<br>The L100 Go System may<br>also:<br>• Facilitate muscle re-<br>education<br>• Prevent/retard disuse<br>atrophy<br>• Maintain or increase joint<br>range of motion<br>• Increase local blood flow | Same<br>(The intended use of<br>L100 Go System is a<br>subset of the<br>predicate device.<br>The L100 Go System<br>is a lower leg cuff<br>only system,<br>therefore, it does not<br>assist knee flexion or<br>extension) |
| Number of<br>Output Modes | 2 modes: Biphasic<br>Asymmetric and Symmetric | 2 modes: Biphasic<br>Asymmetric and Symmetric | Same |
| Number of<br>Programs | • Gait<br>• Training/Exercise<br>• Cycle Training Mode<br>• Clinician mode | • Gait<br>• Training/Exercise<br>• Clinician mode | Different<br>(Cycle Training Mode<br>is disabled for the<br>L100 Go System) |
| Regulated<br>Current or<br>Regulated<br>Voltage | Current | Current | Same |
| | L300 Go System<br>(K190285) | L100 Go System (This<br>submission) | Equivalency<br>Assessment |
| Power Source | Control Unit: Li Coin Cell,<br>CR2032, 3 V, 240 mAh<br><br>EPG: Rechargeable, Li-Ion,<br>Prismatic, 3.7 V, 1000 mAh<br><br>Foot Sensor: Li Coin Cell,<br>CR2032, 3 V, 240 mAh | EPG: Rechargeable, Li-Ion,<br>Prismatic, 3.7 V, 1000 mAh | Same<br>(optional Control Unit<br>and Foot Sensor are<br>not offered with the<br>L100 Go System) |
| Method of Line<br>Current<br>Isolation | N/A (Battery operated) | N/A (Battery operated) | Same |
| Patient<br>Leakage<br>Current,<br>Normal<br>condition [μA] | Less than 1.0 (IEC 60601-<br>1) | Less than 1.0 (IEC 60601-1) | Same |
| Patient<br>Leakage<br>Current, Single<br>fault condition<br>[μA] | 3.0 (IEC 60601-1) | 3.0 (IEC 60601-1) | Same |
| Number of<br>Output Modes | Two: Symmetric &<br>Asymmetric | Two: Symmetric &<br>Asymmetric | Same |
| Number of<br>Output<br>Channels | Small lower leg cuff: 1<br>Regular lower leg cuff: 1 or<br>2<br>Thigh cuff: 1 | Lower leg cuff: 1 | Same<br>(The L100 Go lower<br>leg cuff has the same<br>output channel as<br>the L300 Go small<br>lower leg cuff) |
| Synchronous<br>or Alternating? | Alternating (at one time<br>only one channel is<br>activated) | N/A (only one channel will<br>be activated) | Same<br>(L100 Go cuff is<br>single channeled.<br>This is the same as<br>L300 Go small lower<br>leg cuff) |
| Method of<br>Channel<br>Isolation | Isolation between lower leg<br>and thigh cuff stimulators:<br>Thigh and Lower cuffs are<br>stimulated by different<br>battery operated EPG's<br>without galvanic connection<br>between them.<br><br>Isolation between channels<br>within the regular lower cuff | Isolation between channels<br>with the lower leg cuff<br>stimulator: N/A<br>(Only one channel will be<br>activated) | Same<br>(The isolation<br>between channels is<br>not applicable to<br>L100 Go lower leg<br>cuff stimulator, this is<br>the same as L300<br>Go small lower leg<br>cuff stimulator) |
| | L300 Go System<br>(K190285) | L100 Go System (This<br>submission) | Equivalency<br>Assessment |
| | same EPG are switched<br>using high voltage FET<br>switches. | | |
| | Isolation between channels<br>within the small lower cuff<br>stimulator: N/A (only one<br>channel will be activated) | | |
| | Isolation between channels<br>within the Thigh cuff<br>stimulator: N/A (only one<br>channel will be activated) | | |
| Regulated<br>Current or<br>Regulated<br>Voltage? | Current | Current | Same |
| Software/Firmw<br>are/Microproce<br>ssor-Controlled | Yes | Yes | Same |
| Automatic<br>Overload Trip? | Yes | Yes | Same |
| Automatic No-<br>Load Trip? | Yes | Yes | Same |
| Automatic Shut<br>Off? | Yes | Yes | Same |
| Patient<br>Override<br>Control? | Yes | Yes | Same |
| Indicator<br>Display:<br>• On/Off<br>Status?<br>• Low Battery?<br>•Voltage/Current Level? | Yes<br>Yes<br>Yes | Yes<br>Yes<br>Yes | Same |
| Timer Range<br>[minutes] | No | No | Same |
| | L300 Go System<br>(K190285) | L100 Go System (This<br>submission) | Equivalency<br>Assessment |
| Max Output<br>Current<br>(± 10%) | Lower leg cuffs:<br>100 mA @ 500 Ohm<br>Thigh cuff:<br>100 mA @ 500 Ohm | Lower leg cuff:<br>100 mA @ 500 Ohm | Same<br>(L100 Go lower leg<br>cuff has the same<br>maximum output<br>current as the L300<br>Go lower leg cuff.<br>Thigh cuff is not<br>available in the L100<br>Go System) |
| Max Average<br>Current Density<br>[mARMS/cm2]<br>[Over smallest<br>electrode] | Lower leg EPG:<br>small cuff, small round<br>electrodes (36mm)<br>$1.27 \text{ mA}_{\text{RMS}}/\text{cm}^{2}$ (500 Ω,<br>Irms=13.0 mA, electrode<br>area of 10.2 cm2)<br>Thigh EPG: $0.18 \text{mA}_{\text{RMS}}/\text{cm}^{2}$<br>(500 Ω, Irms=13.0 mA,<br>electrode area of 74 cm2) | Lower leg EPG:<br>Lower leg cuff, regular round<br>electrodes (45mm)<br>$0.82 \text{mA}_{\text{RMS}}/\text{cm}^{2}$ (500 Ω,<br>Irms=13.0 mA, electrode area<br>of 15.9 cm2) | Different<br>(since L100 Go cuff<br>only comes in the<br>size same as the<br>predicate regular<br>cuff, the calculation<br>is based on the<br>electrode of the<br>regular cuff. Thigh<br>cuff is not available<br>in the L100 Go<br>System) |
| Max Average<br>Power Density,<br>(mW/cm2) | Lower leg EPG:<br>small cuff, small round<br>electrodes (36mm)<br>$8.3 \text{ mW}/\text{cm}^{2}$ (500 Ω,<br>Irms=13.0 mA, electrode<br>area of 10.2 cm2)<br>regular cuff, regular round<br>electrodes (45mm)<br>$5.3 \text{ mW}/\text{cm}^{2}$ (500 Ω,<br>Irms=13.0 mA, electrode<br>area of 15.9 cm2)<br>Thigh EPG:<br>$1.1 \text{mW}/\text{cm}^{2}$ (500 Ω,<br>Irms=13.0mA, electrode area<br>of 74 cm2) | Lower Leg EPG:<br>Lower leg cuff, regular round<br>electrodes (45mm)<br>$5.3 \text{ mW}/\text{cm}^{2}$ (500 Ω,<br>Irms=13.0 mA, electrode area<br>of 15.9 cm2) | Same<br>(L100 Go cuff only<br>comes in regular<br>size) |
| | L300 Go System<br>(K190285) | L100 Go System (This<br>submission) | Equivalency<br>Assessment |
| Stimulation<br>Channels | Regular Cuff: 1 or 2 (In 2-<br>channel<br>mode, both function as a<br>single channel but with<br>separately<br>adjustable medial/ lateral<br>stimulation<br>intensity)<br>Small Cuff: 1<br>Thigh Cuff: 1 | Lower leg cuff: 1 | Same<br>(L100 Go has one<br>stimulation channel<br>which is identical to<br>channel #1 of L300<br>Go Regular Cuff) |
| Frequency/<br>Phase duration | Frequency: 10, 15, 20, 25,<br>30, 35, 40, 45 Hz<br>Symmetric and Asymmetric<br>positive: 100, 150, 200,<br>250, 300μs;<br>Asymmetric Negative: 300,<br>450, 600, 750, 900μs; | Frequency: 10, 15, 20, 25,<br>30, 35, 40, 45 Hz<br>Symmetric and Asymmetric<br>positive: 100, 150, 200, 250,<br>300μs;<br>Asymmetric Negative: 300,<br>450, 600, 750, 900μs; | Same |
| Electrodes<br>used in the<br>system | Lower Leg Cuffs:<br>• 2 Hydro-Gel electrodes<br>assembled on<br>electrode bases, or<br>• 2 non-woven cloth<br>electrodes assembled<br>on electrode bases, or<br>• 2 non-woven cloth<br>electrodes attached<br>with snaps (also called<br>"QuickFit" electrodes),<br>or<br>• 3 non-woven cloth<br>electrodes attached<br>with snaps (segmented<br>electrodes [also called<br>"steering" electrodes],<br>using common anode<br>to allow separate<br>adjustment of medial<br>and lateral stimulation.<br>This electrode is only<br>used for regular cuff)<br>Thigh Cuff:<br>• 2 single, non-woven<br>cloth electrodes<br>attached with snaps | Lower leg Cuff:<br>• 2 Hydro-Gel electrodes<br>assembled on electrode<br>bases, or<br>• 2 non-woven cloth<br>electrodes assembled<br>on electrode bases, or<br>• 2 non-woven cloth<br>electrodes attached<br>with snaps (also called<br>"QuickFit" electrodes) | Different<br>(L100 Go does not<br>use steering<br>electrodes)…
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.