The MStim Drop LGT-233 is intended to provide ankle dorsiflexion in individuals who have a dropped foot as a consequence of upper motor neuron injury. The device electrically stimulates muscles in the affected leg to provide ankle dorsiflexion of the foot and/or knee flexion or extension; thus, it also may improve the individual's gait. The MStim Drop LGT-233 may also: Prevent/retard disuse atrophy Maintain or increase joint range of motion Increase local blood flow
Device Story
Wearable FES device; treats foot drop via electrical stimulation of leg muscles. Components: stimulator unit, electrode lead wire, electrode pads, leg bandage, battery charger. Operates in two modes: Training (seated/lying, muscle recovery/atrophy prevention) and Walk (gait assistance). Input: gait swing phase detection (Walk mode). Output: biphasic square pulse electrical stimulation. Controlled via manufacturer-provided mobile APP for stimulation/time settings. Used by patients; prescribed by physicians. Benefits: improved gait, muscle atrophy prevention, increased blood flow, joint range of motion maintenance.
Indicated for individuals with foot drop due to upper motor neuron injury. Provides ankle dorsiflexion and/or knee flexion/extension to improve gait. Additional uses: prevent/retard disuse atrophy, maintain/increase joint range of motion, increase local blood flow.
Regulatory Classification
Identification
An external functional neuromuscular stimulator is an electrical stimulator that uses external electrodes for stimulating muscles in the leg and ankle of partially paralyzed patients (e.g., after stroke) to provide flexion of the foot and thus improve the patient's gait.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
November 5, 2021
Guangzhou Longest Science & Technology CO., Ltd. % Jet Li Regulation Manager Guangzhou Keda Biological Tech Co., Ltd. 6F, No. I TianTai road, Science City, LuoGang District Guangzhou, Guangdong 510060 China
Re: K202110
Trade/Device Name: MStim Drop Model: LGT-233 Regulation Number: 21 CFR 882.5810 Regulation Name: External Functional Neuromuscular Stimulator Regulatory Class: Class II Product Code: GZI, IPF Dated: October 5, 2021 Received: October 6, 2021
Dear Jet Li:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Amber Ballard, PhD Assistant Director DHT5B: Division of Neuromodulation and Physical Medicine Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K202110
Device Name MStim Drop Model: LGT-233
Indications for Use (Describe)
The MStim Drop LGT-233 is intended to provide ankle dorsiflexion in individuals who have a dropped foot as a consequence of upper motor neuron injury. The device electrically stimulates muscles in the affected leg to provide ankle dorsiflexion of the foot and/or knee flexion or extension; thus, it also may improve the individual's gait. The MStim Drop LGT-233 may also:
·Prevent/retard disuse atrophy
· Maintain or increase joint range of motion
•Increase local blood flow
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary 510K number: K202110
This summary of 510(K) safety and effectiveness information is being submitted in accordance with the requirement of 21 CFR 807.92.
#### Date of the summary prepared: November 04, 2021 1.
#### Submitter'sInformation 2.
Company Name: Guangzhou Longest Science & Technology CO., Ltd. Address: 5&6F, Building B4, No.11, Kaiyuan Avenue, Science City, Guangzhou Hi-tech Industrial Development Zone, 510530 Guangzhou, Guangdong Province, P.R. China Phone: +86 020 6635 3999 Fax:+86 020 6635 3920 URL: www.longest.cn
E-mail: qzlongest@126.com
Contact Person: Xiaobing Luo (Deputy general manager) E-mail: service@longest.cn
# Application Correspondent:
Company: Guangzhou KEDA Biological Tech Co., Ltd. Address: 6F, No.1 TianTai road, Science City, LuoGang District, GuangZhouCity, China
Contact Person: Mr. Jet Li
Tile: Regulation Manager
Tel: +86-18588874857
Email: med-jl@foxmail.com
#### 3. Subject Device Information
Type of 510(k) submission: Traditional Common Name: External functional neuromuscular stimulator Trade Name: MStim Drop Model: LGT-233 Classification Name: Stimulator, neuromuscular, external functional; Stimulator, muscle, powered Review Panel: Neurology, Physical Medicine Product Code: GZI, IPF
Requlation Number: 882.5810, 890.5850
{4}------------------------------------------------
### Regulation Class: 2
| Sponsor | Bioness Inc | Otto Bock Healthcare<br>Products GmbH |
|----------------------|--------------------------------------|----------------------------------------|
| Device Name | L300 Go System | Otto Bock Healthcare<br>Products GmbH |
| 510(k) Number | K162407 | K141812 |
| Product<br>Code | GZI, IPF | GZI; GZI |
| Regulation<br>Number | 21 CFR §882.5810<br>21 CFR §890.5850 | 21 CFR 882.5850 and<br>21 CFR 882.5810 |
| Regulation<br>Class | 2 | 2 |
#### Predicate Device Information 4.
# 5. Device Description
The MStim Drop LGT 233 is a wearable foot drop device, which is mainly consist s of stimulator main unit, electrode lead wire, electrodes pad, leg bandage and battery charger. The stimulator main unit can be charge by the battery charger.
The device can control the out stimulation and treatment time setting by APP application provided by the manufacturer, and it can trigger and control the pulse stimulation on and off by detecting the swing phase of gait during walking. The device was design to use FES (Functional Electrical Stimulation) to treat and improve the patient s foot drop and help to improve walking ability. It has two modes: Training mode, Walk mode. Training mode is suitable for muscle training when the patient is seated or lying down (patient lacking active training), which can promote muscle recover, prevent muscle atrophy, improve joint range of motion, and increase local blood flow. Walk mode is walking with the electrical stimulation.
#### 6. Intended Use / Indicationsfor Use
The MStim Drop LGT-233 is intended to provide ankle dorsiflexion in individuals who have a dropped foot as a consequence of upper motor neuron injury. The device electrically stimulates muscles in the affected leg to provide ankle dorsiflexion of the flexion or extension; thus, it also may improve the individual's gait.
- The MStim Drop LGT-233 may also:
- Prevent/retard disuse atrophy
- Maintain or increase joint range of motion
{5}------------------------------------------------
· Increase local bloodflow
#### 7. Test Summary
The Device has been evaluated the safety and performance by lab bench testing according to the following standards:
- IEC 60601-1-2: 2014: Medical Electrical Equipment -Part 1-2: General requirements for basic salety 1. and essential performance. Collateral Standard: Electromagnetic Compatibility
- IEC 60601-1:2005: Medical Electrical Equipment Part 1: General requirements for basic safety 2. and essential performance
- 3. IEC 60601-1-11: 2015: Medical Electrical Equipment - Part 1-11: General requirements for basic safety and essential performance. Collateral Standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment
- 4. IEC 60601-2-10: 2012: A1 2016 Medical Electrical Equipment - Part 2: Particular requirements for the basic safety and essential performance of nerve and muscle stimulators.
- 5. ANSI IEEE C63.27-2017: American National Standard for Evaluation of Wireless Coexistence
- 6. IEC 62133-2: 2017: Secondary cells and batteries containing alkaline or other non-acid electrolytes-Safety requirements for portable sealed secondary cells, and for batteries made from them, for use in portable applications- Part 2: Lithium systems
- 7. Biological evaluation ofmedical device Part 10: Tests for irritation and skin sensitization(ISO 10993-10:2010)
- 8. Biological evaluation of medical device Part 5: Cytotoxicity test- In vitro method (ISO 10993-5:2009)
# 8. Comparison to predicate device and conclusion
The technological characteristics, features, specifications, mode of operation, and intended use of the MStim Drop Model: LGT-233 is substantially equivalent to the predicate devices quoted above. The differences between the subject device and predicate device do not raise new issues of safety or effectiveness.
| Elements of<br>Comparison | Subject Device | Predicate Device I | Predicate Device II | Remark | |
|-----------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------|----------------------------|
| Device Name<br>and Model | MStim Drop Model: LGT-<br>233 | L300 Go System | MyGait Stimulation<br>System | -- | |
| Elements of<br>Comparison | Subject Device | Predicate Device I | Predicate Device II | Remark | |
| 510 (K)<br>Number | Applying | K162407 | K141812 | -- | |
| Product Code | GZI, IPF | GZI, IPF | GZI, IPF | -- | |
| Regulation<br>Number | 882.5810, 890.5850 | 21 CFR § 882.5810<br><br>21 CFR § 890.5850 | 21 CFR 882.5810, 21 CFR § 890.5850 | -- | |
| Indication for<br>use | The MStim Drop LGT-233 is<br>intended to provide ankle<br>dorsiflexion in individuals who<br>have a dropped foot as a<br>consequence of upper motor<br>neuron injury. The device<br>electrically stimulates muscles<br>in the affected leg to provide<br>ankle dorsiflexion of the foot<br>and/or knee flexion or<br>extension; thus, it also may<br>improve the individual's gait.<br><br>The MStim Drop LGT-233<br>may also:<br><br>Prevent/retard disuse<br>atrophy<br><br>Maintain or increase joint<br>range of motion<br><br>Increase local blood flow | The L300 Go System is<br>intended to provide ankle<br>dorsiflexion in adult and<br>pediatric individuals with<br>foot drop and/or assist<br>knee flexion or extension<br>in adult individuals with<br>muscle w eakness related<br>to upper motor neuron<br>disease/injury (e.g.,<br>stroke, damage to<br>pathways to the spinal<br>cord). The L300 Go<br>Systemelectrically<br>stimulates muscles in the<br>affected leg to provide<br>ankle dorsiflexion of the<br>foot and/or knee flexion or<br>extension; thus, it also<br>may improve the<br>individual's gait. | The stimulation system is<br>intended to provide ankle<br>dorsiflexion and knee flexion<br>or extension in individuals with<br>foot drop and thigh muscle<br>weakness following an upper<br>motor neuron injury or<br>disease.<br><br>During gait, the stimulation<br>system sends electric stimuli<br>to muscles in the affected leg,<br>initiating dorsiflexion of the foot<br><br>and knee extension or flexion<br>and may thus improve the<br>individual's gait. The<br>stimulation system may also<br>prevent or retard atrophy<br>caused by inactivity, facilitate<br>muscle reeducation, maintain<br>or improve the range of motion<br>in the joints and promote local<br>blood circulation. | SE | |
| Apply parts | Leg | Leg and foot | Leg and foot | SE | |
| Power Sources | Adapter model: HYI11-005<br>(USA) and HYI11-005<br>(Europe)<br><br>Adapter supply voltage: AC<br>100~240V, 50/60Hz<br><br>Adapter output: DC 5V, 2A<br><br>Battery: 3.7V, 1200mAh,<br>lithium battery. | Battery operated | Battery operated | Minor<br>difference<br>Note 1 | |
| Indicator<br>display | LED and APP | -- | -- | Minor<br>difference<br>Note 1 | |
| Elements of<br>Compariso<br>n | Subject Device | Predicate Device I | Predicate Device II | Remark | |
| Method of Line<br>Current<br>Isolation | Battery operated | -- | Medical Class II Power<br>Adapter | SE | |
| Number of<br>Modes for<br>Micro current<br>stimulation | 2<br>(For Evaluation mode: no<br>micro current stimulation) | Two modes: Biphasic<br>Symmetric and Biphasic<br>Asymmetric.<br>Applicable to both lower<br>leg and thigh position of<br>the device. | 2 | SE | |
| Number of<br>Channels for<br>Micro current<br>stimulation | 1 | Lower leg small cuff<br>- 1 channel;<br>Lower leg regular<br>cuff - 1 or 2 channels (in 2<br>channel configuration,<br>both<br>channels function as a<br>single channel with<br>separately adjustable<br>medial / lateral stimulation<br>intensity);<br>Thigh cuff - 1 channel | 2 | SE | |
| Synchronous<br>or Alternating | Not applicable – single<br>channel device | Alternating (at one time<br>only one channel is<br>activated) | Synchronous or alternating | SE | |
| Regulated<br>Current or<br>Regulated<br>Voltage | Regulated Voltage | Current | Regulated Current | SE<br>Note 2 | |
| Elements of<br>Comparison | Subject Device | Predicate Device I | Predicate Device II | Remark | |
| Software/Firmware/<br>Microprocessor<br>control | Yes | Yes | Yes | SE | |
| Automatic Overload<br>Trip | Yes | Yes | Yes | SE | |
| Automatic No-load<br>Trip | Yes. | Yes | Yes | SE | |
| Automatic Shut Off | Yes. | Yes | No | SE | |
| Patient Override<br>Control | Yes | Yes | Yes<br>ON/OFF Button | SE | |
| Indic<br>ator<br>Displ<br>ay | On/Off<br>Status | Yes | Yes | Yes | SE |
| | Low<br>Battery | Yes | Yes | Yes | SE |
| | Voltag<br>e/Curr ent<br>Level | Yes | Yes | No | SE |
| Timer Range | | Treatment time: 1min-<br>60min, stepping 1min;<br>Timer tolerance: ±2%;<br>When finish, the device<br>can stop output and<br>prompt | Max<br>stimulation<br>duration (clinician<br>selectable)<br>Training mode: 5-60<br>minutes | 5 to 120min in steps of<br>5 min | Minor difference<br>Note 2 |
| Elements of<br>Comparison | Subject Device | Predicate Device I | Predicate Device II | Remark | |
| Console weight | 60g (only main unit) | Control Unit: 60g, EPG:<br>60g, Lower leg FSC:150g,<br>Thigh cuff: 300g, Foot<br>Sensor: 25g | 78g | SE | |
| Housing<br>Materials and<br>Construction | ABS | Remote Control:<br>Bay State Polymer<br>PA-2000RX<br><br>EPG: Bay State Polymer<br>PA-2000RX<br>(Polycarbonate + ABS)<br>Lower Leg FSC:<br>Biocompatible fabric over<br>plastic (POM Hi) skeleton<br><br>Thigh FSC: TPU Foot<br>Sensor: ABS (Sensor<br>housing), Bay state<br>Polymer<br>PA-2000RX (Electronics<br>housing) | ABS | SE | |
| Waveform | Biphasic square | Biphasic Symmetrical | Biphasic | SE | |
| Shape | Rectangular | Rectangular | Rectangular | SE | |
| Maximum<br>Output Voltage<br>(+/- 10%) | 50V ±10% @ 500Ω<br>90V ±10% @ 1KΩ<br>120V ±10% @ 2KΩ<br>130V ±10% @ 10KΩ | Lower leg: 50V@500Ω<br>Thigh: 50V@500Ω<br><br>Lower leg: 130V@2K Ω<br>Thigh: 130V@2K Ω<br>lower leg: 130V@10K Ω<br>Thigh: 130V@10K Ω | 45V +-10% @ 500Ω<br>90V +-10% @ 1KΩ<br>120V +-10%@2K Ω<br>n/a @10K Ω<br>(no stimulation possible at<br>10kΩ) | Minor difference<br>Note 3 | |
| Elements of<br>Comparison | Subject Device | Predicate Device I | Predicate Device II | Remark | |
| Maximum<br>output Current | 100mA ±10% @ 500Ω<br>90mA ±10% @ 1KΩ<br>60mA ±10% @ 2KΩ<br>13mA ±10% @ 10KΩ | Lower leg: 100mA@500Ω<br>Thigh: 100mA@500Ω<br>Lower leg: 65mA@2KΩ<br>Thigh: 65mA@2KΩ<br>Lower leg: 13mA@10KΩ<br>Thigh: 13mA@10KΩ | 90mA (+/- 10%) @ 500Ω<br>90mA (+/- 10%) @ 1KΩ<br>60mA (+/- 10%)@ 2K Ω<br>n/a @10K Ω<br>(no stimulation possible at<br>10KΩ) | Minor difference<br>Note 3 | |
| Frequency<br>range | 10~80Hz | 10, 15, 20, 25, 30, 35, 40,<br>45 Hz…
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.