The stimulation system is intended to provide ankle dorsiflexion or extension in individuals with foot drop and thigh muscle weakness following an upper motor neuron injury or disease. During gait, the stimulation system sends electric stimuli to muscles in the affected leg, initiating dorsiflexion of the foot and knee extension or flexion and may thus improve the individual's gait. The stimulation system may also prevent or retard atrophy caused by inactivity, facilitate muscle reeducation, maintain or improve the range of motion in the joints and promote local blood circulation.
Device Story
MyGait Stimulation System is a battery-operated, two-channel external functional neuromuscular stimulator for gait improvement. Input: wireless heel switch (worn in sock) registers heel lift/ground contact to detect swing/stance phases. Processing: stimulator (worn in cuff) receives wireless signals and delivers electrical stimulation to common peroneal nerve (dorsiflexion) and/or thigh muscles (knee flexion/extension). Output: controlled muscle contraction. Used in everyday environments by patients; settings adjusted by clinicians via MyGait Adjustment Tool and MyGait Soft PC software. Benefits: improved gait, muscle reeducation, atrophy prevention, increased local blood circulation, maintained joint range of motion.
Clinical Evidence
Bench testing only. Performance testing included electrical safety (IEC 60601-1, 60601-2-10), electromagnetic compatibility (IEC 60601-1-2), environmental testing (vibration, shock, temperature, humidity), and biocompatibility (ISO 10993-1, 5, 10).
Technological Characteristics
Two-channel neuromuscular stimulator; ABS housing; 3.7V 1100mAh Li-ion battery; wireless heel switch; biphasic symmetrical/asymmetrical waveforms; constant current output; microprocessor-controlled; connectivity via wireless remote and PC software; IP-rated enclosure; medical class II power adapter for charging.
Indications for Use
Indicated for individuals with foot drop and thigh muscle weakness following upper motor neuron injury or disease to provide ankle dorsiflexion and knee flexion/extension during gait.
Regulatory Classification
Identification
An external functional neuromuscular stimulator is an electrical stimulator that uses external electrodes for stimulating muscles in the leg and ankle of partially paralyzed patients (e.g., after stroke) to provide flexion of the foot and thus improve the patient's gait.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized image of three human profiles facing right, with flowing lines representing hair or clothing. The profiles are arranged in a way that suggests unity and forward movement. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the image.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
April 2, 2015
Otto Bock Healthcare Products GmbH Rene Urtz Regulatory Affairs Specialist Brehmstrabe 16 1110 Vienna AUSTRIA
Re: K141812
Trade/Device Name: MyGait Stimulation System Regulation Number: 21 CFR 882.5810 Regulation Name: External Functional Neuromuscular Stimulator Regulatory Class: Class II Product Code: GZI, IPF Dated: February 23, 2015 Received: February 27, 2015
Dear Mr. Urtz:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Isting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21
{1}------------------------------------------------
CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
## Felipe Aquel -S
Carlos L. Peña, PhD, MS for Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known) K141812
Device Name MyGait Stimulation System
#### Indications for Use (Describe)
The stimulation system is intended to provide ankle dorsiflexion or extension in individuals with foot drop and thigh muscle weakness following an upper motor neuron injury or disease.
During gait, the stimulation system sends electric stimuli to muscles in the affected leg, initiating dorsiflexion of the foot and knee extension or flexion and may thus improve the individual's gait. The stimulation system may also prevent or retard atrophy caused by inactivity, facilitate muscle reeducation, maintain or improve the range of motion in the joints and promote local blood circulation.
Type of Use (Select one or both, as applicable)
2 Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
#### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the word "ottobock." in a sans-serif font. The word is written in a dark blue color. The letters are closely spaced together, and the period at the end is small.
# 510(K) SUMMARY
## 1 Submitter Information
| Manufacturer: | Otto Bock Healthcare Products GmbH |
|-----------------------------|----------------------------------------------------------------------|
| Establishment Registration: | 9615892 |
| Adress: | Brehmstraße 16<br>1110 Vienna<br>Austria |
| Contact Person: | Rene Urtz<br>Regulatory Affairs Specialist<br>rene.urtz@ottobock.com |
| Telephone: | +43 1 523 37 86 707 |
| Fax Number: | +43 1 523 22 64 |
| Date Prepared: | March 31, 2015 |
## 2 Device Information
| Trade Name: | MyGait® Stimulation System |
|------------------------------|------------------------------------------------------|
| Common or Usual Name: | Stimulator, Neuromuscular, External |
| Classification Name: | External functional neuromuscular stimulator |
| | (21 CFR 882.5810) |
| Classification Product Code: | GZI (stimulator, neuromuscular, external functional) |
| Subsequent Product Code: | IPF (stimulator, muscle, powered) |
{4}------------------------------------------------
# 3 Identification of Legally Marketed Predicate Devices
| Name: | Walkaide |
|---------------|-------------------------------|
| Manufacturer: | Innovative Neurotronics, Inc. |
| 510k Number: | K123972 |
| Date Cleared: | April 4th, 2013 |
| Name: | Empi 300pv |
|---------------|-----------------|
| Manufacturer: | Empi, Inc |
| 510k Number: | K021100 |
| Date Cleared: | June 18th, 2002 |
| Name: | NESS L300 Plus System* |
|---------------|-----------------------------|
| Manufacturer: | Bioness Neuromodulation Ltd |
| 510k Number: | K103343 |
| Date Cleared: | April 29th, 2011 |
* Primary predicate
{5}------------------------------------------------
## 4 Indications for Use
The stimulation system is intended to provide ankle dorsiflexion or extension in individuals with foot drop and thigh muscle weakness following an upper motor neuron injury or disease. During gait, the stimulation system sends electric stimuli to muscles in the affected leq, initiating dorsiflexion of the foot and knee extension or flexion and may thus improve the individual's gait. The stimulation system may also prevent or retard atrophy caused by inactivity, facilitate muscle reeducation, maintain or improve the range of motion in the joints and promote local blood circulation.
## 5 Device Description
The MyGait® Stimulation System is a battery-operated two-channel stimulator and was developed exclusively for everyday use during walking in an everyday environment.
The patient wears a wireless heel switch in a special sock. If the user is not able to put sufficient weight on the affected leg, he or she can wear the heel switch on the other foot. The heel switch registers heel lift and ground contact and thus the swing and stance phase. It transmits the information to the stimulator wirelessly.
The stimulator is located in a cuff that holds the electrodes. It is easy to put on, even with only one hand. The correct placement of the electrodes is achieved by the cuff. The stimulator delivers the electrical stimulation to the common peroneal nerve. The nerve stimulates for controlled dorsiflexion of the foot during the swing phase.
The second channel can be used independently of the first channel to provide stimulation to the knee flexors or knee extensors in patients with thigh muscle weakness.
Using the wireless remote control, the patient can control and adjust the stimulator settings. When the patient sits for an extended period, the sleep mode helps save energy.
The adjustment tool enables qualified personnel to adjust the stimulation parameters to the patient's individual needs. Stimulation parameters are used to affect movements in the foot. Examples of stimulation parameters are current, pulse width and shape, frequency, stimulation timing,
The "MyGait Soft" PC software is used to manage and analyze gait analysis data and stimulation parameters. The data is loaded from the stimulator to the adjustment tool. The stimulation parameters can be restored to the stimulator from the PC.
{6}------------------------------------------------
## 6 Substantial Equivalence Discussion
### 6.1 Indications for Use
| | New Device | Predicate Device | Predicate Device | Predicate Device |
|-----------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510k Number | K141812 | K123972 | K021100 | K103343 |
| Device<br>Name, Model | MyGait | WalkAide | Empi 300pv | NESS L300 Plus System |
| Manufacturer | Otto Bock Healthcare<br>Products GmbH | Innovative Neurotronics, Inc. | Empi, Inc | Bioness Neuromodulation<br>Ltd. |
| Indications<br>for Use | The stimulation system is<br>intended to provide ankle<br>dorsiflexion and knee flexion or<br>extension in individuals with foot<br>drop and thigh muscle<br>weakness following an upper<br>motor neuron injury or disease.<br>During gait, the stimulation<br>system sends electric stimuli to<br>muscles in the affected leg,<br>initiating dorsiflexion of the foot<br>and knee extension or flexion<br>and may thus improve the<br>individual's gait. The stimulation<br>system may also prevent or<br>retard atrophy caused by<br>inactivity, facilitate muscle<br>reeducation, maintain or<br>improve the range of motion in<br>the joints and promote local<br>blood circulation. | Summary Statement<br>The Innovative Neurotronics<br>WalkAide External Functional<br>Neuromuscular Stimulator<br>(WalkAide System) is intended<br>to address the lack of ankle<br>dorsiflexion in patients who have<br>sustained damage to upper<br>motor neurons or pathways to<br>the spinal cord. During the<br>swing phase of gait, the<br>WalkAide System electrically<br>stimulates the appropriate<br>muscles that cause ankle<br>dorsiflexion and may thus<br>improve the patient's gait.<br>Medical benefits of Functional<br>Electrical Stimulation (FES) may<br>include prevention, retardation<br>of disuse atrophy, increased<br>local blood flow, muscle re-<br>education, and maintained or<br>increased joint range of motion. | Summary Statement<br>The 300 PV is a multifunction<br>electrotherapy device with<br>various treatment modes that<br>allow for neuromuscular<br>electrical stimulation (NMES),<br>transcutaneous electrical nerve<br>stimulation (TENS),<br>interferential current stimulation<br>(IFS) and functional electrical<br>stimulation (FES).<br>As a NMES device, the 300 PV<br>is indicated for the following<br>conditions:<br>Re-educating muscles<br>Relaxation of muscle spasm<br>Increasing local blood circulation<br>Retarding or preventing disuse<br>atrophy<br>Maintaining or increasing range<br>of motion. Prevention of venous<br>thrombosis of the calf muscles<br>immediately after surgery.<br>As a TENS device, the 300 PV<br>is indicated for the following<br>conditions:<br>Symptomatic relief and<br>management of chronic,<br>intractable pain.<br>Adjunctive treatment for post-<br>surgical and post-trauma acute<br>pain. As an IFS device, the 300<br>PV is indicated for the following<br>conditions:<br>Symptomatic relief of acute pain<br>Symptomatic relief and<br>management of chronic pain.<br>As a FES device, the 300 PV is<br>indicated for the following<br>condition:<br>Stimulation of the muscles in<br>the leg and ankle of partially<br>paralyzed<br>patients to provide flexion of the<br>foot and thus improve the<br>patient's gait. | Summary Statement<br>The NESS L300 Plus System is<br>intended to provide ankle<br>dorsiflexion and knee flexion or<br>extension in individuals with foot<br>drop and thigh muscle<br>weakness, following an upper<br>motor neuron injury or disease.<br>During gait, the L300 Plus<br>System electrically stimulates<br>muscles in the affected leg to<br>provide dorsiflexion of the foot<br>and knee flexion or extension,<br>thus it may improve the<br>individual's gait. The L300 Plus<br>System may also:<br>* Facilitate Muscle re-education<br>* Prevent/retard disuse atrophy<br>* Maintain or increase joint<br>range of motion<br>* Increase local blood flow |
| | New Device | Predicate Device | Predicate Device | Predicate Device |
| 510k Number | K141812 | K123972 | K021100 | K103343 |
| Device Name, Model | MyGait | WalkAide | Empi 300pv | NESS L300 Plus System |
| Manufacturer | Otto Bock Healthcare<br>Products GmbH | Innovative Neurotronics,<br>Inc. | Empi, Inc | Bioness Neuromodulation<br>Ltd. |
| Power Sources | | | | |
| Battery | built in | replaceable | replaceable | built in |
| # of batteries | 1 | 1 | 2 | 1 |
| type | Lilon | Alkaline | Alkaline | Lilon |
| size | 3.7V, 1100mAh | AA, LR6 | AA, LR6 | 3.7 V, 750mAh |
| Method of Line Current<br>Isolation | Medical Class II Power<br>Adapter | n/a | n/a | Unknown |
| Patient Leakage<br>Current | | | | |
| normal condition | <1 μA | Unknown | Unknown | Unknown |
| Single Fault condition | <1μA | Unknown | Unknown | Unknown |
| Number of Output<br>Modes | 2 | 1 | 5 | 2 |
| Number of Output<br>Channels | 2…
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.