Receiver Kit, Trial Kit, Spare Lead Kit, Sterile Revision Kit, Wearable Antenna Assembly Kit and Charger Kit
Applicant
Stimwave Technologies Incorporated
Product Code
GZB · Neurology
Decision Date
Dec 16, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.5880
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Freedom Spinal Cord Stimulator (SCS) System is intended as the sole mitigating agent, or as an adjunct to other modes of therapy used in a multidisciplinary approach for chronic, intractable pain of the trunk and/or lower limbs, including unilateral or bilateral pain. The Freedom-8A Trial Lead Kit is only used in conjunction with the Freedom-8A Stimulator Receiver Kit, and the Freedom-4A Trial Lead Kit is used for either the Receiver Kit Freedom-4A Stimulator or the Receiver Kit Freedom-8A Stimulator. The trial devices are solely used for trial stimulation (no longer than 30 days) to determine efficacy before recommendation for a permanent (long term) device.
Device Story
System provides spinal cord stimulation for chronic, intractable pain; utilizes pulsed electrical current to create energy field acting on nerves near spinal column. Components: implantable stimulator (4 or 8 electrodes), receiver, and externally worn Wearable Antenna Assembly (WAA). WAA generates 915 MHz RF power to wirelessly power implant. Used in hospital or home settings; operated by physician (orthopedic, neurosurgeon, anesthesiologist) for implantation and patient for daily use. Patient adjusts stimulation parameters via WAA switch membrane or iPad application (WaveCrest). Output affects nerve signaling to provide pain relief. Benefits include non-invasive power delivery and therapeutic pain management.
Clinical Evidence
No clinical data provided. Substantial equivalence supported by non-clinical bench testing, including biocompatibility (ISO 10993-1), electrical safety (IEC 60601-1), electromagnetic compatibility (IEC 60601-1-2), and mechanical durability testing (AAMI ANSI ISO 14708-3).
Technological Characteristics
Implantable stimulator (Pellethane 55D, Pt-Ir 90:10 electrodes); 4 or 8 electrode array. RF wireless power (915 MHz). WAA includes PCB, silicon rubber switch membrane, lithium-ion battery, and antenna. Connectivity: LE Bluetooth for WAA. Sterilization: Ethylene Oxide. Software: WaveCrest application (iPad).
Indications for Use
Indicated for patients with chronic, intractable pain of the trunk and/or lower limbs, including unilateral or bilateral pain. Used as sole therapy or adjunct in multidisciplinary approach. Trial devices limited to 30 days for efficacy determination.
Regulatory Classification
Identification
An implanted spinal cord stimulator for pain relief is a device that is used to stimulate electrically a patient's spinal cord to relieve severe intractable pain. The stimulator consists of an implanted receiver with electrodes that are placed on the patient's spinal cord and an external transmitter for transmitting the stimulating pulses across the patient's skin to the implanted receiver.
Predicate Devices
Stimwave Freedom SCS System FR8A/FR4A, LBRD-915-2A (K160600)
Stimwave Freedom SCS System FR8A/FR4A, LBRD-915-2A (K150517)
Stimwave Freedom SCS System FRE4-A001, FRT4-A001, WAA-A012 (K141399)
{1}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features the department's name in a circular arrangement around a symbol. The symbol consists of a stylized caduceus, with three human profiles forming the staff and a cloth draped around the base.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
December 16, 2016
Stimwave Technologies Incorporated Elizabeth Greene Vice President of Quality Assurance and Regulatory Affairs 901 East Las Olas Boulevard, Suite 201 Fort Lauderdale, Florida 33301
Re: K162161
Trade/Device Name: Freedom Spinal Cord Stimulation (SCS) System Regulation Number: 21 CFR 882.5880 Regulation Name: Implanted Spinal Cord Stimulator for Pain Relief Regulatory Class: Class II Product Code: GZB Dated: November 16, 2016 Received: November 18, 2016
Dear Elizabeth Greene:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{2}------------------------------------------------
the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
| Sincerely, | |
|------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | William J. Heetderks -A |
| | Digitally signed by William J. Heetderks -A<br>DN: c=US, o=U.S. Government, ou=HHS, ou=NIH,<br>ou=People,<br>0.9.2342.19200300.100.1.1=0010149848,<br>cn=William J. Heetderks -A<br>Date: 2016.12.16 14:42:53 -05'00' |
for
| | Carlos L. Peña, PhD, MS |
|--|-----------------------------------------------------------|
| | Director |
| | Division of Neurological<br>and Physical Medicine Devices |
| | Office of Device Evaluation |
| | Center for Devices and Radiological Health |
Enclosure
{3}------------------------------------------------
# Indications for Use Statement
# 510(k) Number (if known): K162161
Device Name: Freedom Spinal Cord Stimulator (SCS) System
#### Indications For Use:
The Freedom Spinal Cord Stimulator (SCS) System is intended as the sole mitigating agent, or as an adjunct to other modes of therapy used in a multidisciplinary approach for chronic, intractable pain of the trunk and/or lower limbs, including unilateral or bilateral pain. The Freedom-8A Trial Lead Kit is only used in conjunction with the Freedom-8A Stimulator Receiver Kit, and the Freedom-4A Trial Lead Kit is used for either the Receiver Kit Freedom-4A Stimulator or the Receiver Kit Freedom-8A Stimulator. The trial devices are solely used for trial stimulation (no longer than 30 days) to determine efficacy before recommendation for a permanent (long term) device.
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
{4}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for Stimwave Technologies. The word "stim" is in dark blue, and the word "wave" is in green. Below the words "stim wave" is the word "TECHNOLOGIES" in dark blue. To the right of the word "wave" are three curved lines in dark blue that resemble radio waves.
# 510(k) Summary
# for
# Freedom Spinal Cord Stimulator (SCS) System
#### 1. Submission Sponsor
| Stimwave Technologies Incorporated | |
|------------------------------------|------------------------------------------------------------------------------|
| 901 East Las Olas Boulevard | |
| Suite 201 | |
| Fort Lauderdale | |
| Florida 33301 | |
| USA | |
| Phone: | 800.965.5134 |
| Fax: | 800.965.5134 |
| Contact: | Elizabeth Greene, Vice President of Quality Assurance and Regulatory Affairs |
# 2. Date Prepared
September 7, 2016
# 3. Device Identification
| Trade/Proprietary Name: | Freedom Spinal Cord Stimulator (SCS) System |
|----------------------------|--------------------------------------------------|
| Common/Usual Name: | Spinal Cord Stimulator |
| Classification Name: | Stimulator, Spinal-Cord, Implanted (Pain Relief) |
| Classification Regulation: | 882.5880 |
| Product Code: | GZB |
| Device Class: | Class II |
| Classification Panel: | Neurology |
# 4. Legally Marketed Predicate Device(s)
Stimwave Freedom SCS System FR8A/FR4A, LBRD-915-2A (K160600) Stimwave Freedom SCS System FR8A/FR4A, LBRD-915-2A (K150517) Stimwave Freedom SCS System FRE4-A001, FRT4-A001, WAA-A012 (K141399)
# 5. Device Description
This submission is based upon the established Stimwave product family offering additional surgical kitting options (Spare Lead Kit and Sterile Revision Kit), updates to the surgical kits (Receiver Kit and Trial Lead Kit) to include new accessory components (Needle, and RF Stylet), a new Wearable Antenna Assembly (WAA) model utilizing a Low Energy (LE) Bluetooth module, a USB battery charger, and upgrades to the WaveCrest Application.
{5}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for Stimwave Technologies. The word "stim" is in large, blue, lowercase letters. The word "wave" is in large, green, lowercase letters. Below the words is the word "TECHNOLOGIES" in smaller, blue, uppercase letters inside of a blue oval. To the right of the words is a blue graphic of concentric arcs.
The Stimwave Technologies Incorporated (Stimwave) Freedom Spinal Cord SCS System (System) is used for spinal column stimulation to provide therapeutic relief for chronic, intractable pain of the trunk and/or lower limbs including unilateral pain. The therapy utilizes pulsed electrical current to create an energy field that acts on nerves near the spinal column. The System is comprised of an implantable stimulator (Freedom-8A/4A Stimulator), receiver component (Receiver), and an externally worn transmitter (Wearable Antenna Assembly (WAA)) to power the device. The System is implanted only following a successful trial period with the Freedom-8A/4A Trial Lead.
#### A polyurethane (Pellethane 55D) casing with an embedded receiver, Freedom-8A Stimulator. flexible circuit board and electrodes (Platinum Iridium 90:10) that is Freedom-4A placed in the patient's evidural space. The Freedom-8A Stimulator Stimulator has eight (8) electrodes, and the Freedom-4A Stimulator has four (4) electrodes. Identical to K150517 and K160600. Receiver A copper and PEEK cable with dual couplers; placed within the center lumen of the Freedom-8A or Freedom-4A Stimulator with the distal end combination of Receiver and Stimulator being placed under the skin. Two (2) Receivers are provided with each kit. Identical to K160600, but quantity is increased from one to two. A stainless steel wire with a polypropylene handle that is inserted Stylet(s) into the open central lumen of the stimulator to provide rigidity during implantation. Two (2) stylets are provided in the Receiver Kit, one straight and one bent, each with diameter of 0.30 mm. Similar to K150517 and K160600, but with a larger diameter. Needle A 13-gauge stainless steel needle that acts as a conduit for passage of the Stimulator into the epidural space. Similar to K141399, but provided with a larger gauge. Guidewire A stainless steel, rigid, solid core guidewire used to create a hollow pathway in the evidural space for the Stimulator to pass through easily. Identical to K141399, K150517, and K160600. Suture Sleeve Cap A polyurethane (Pellethane 55D) cap that is placed over the proximal end of the Stimulator. The Suture Sleeve Cap is attached to the Freedom-8A/4A Stimulator and can be sutured to tissue to reduce the possibility of device migration. Identical to K141399, K150517, and K160600.
# Freedom-8A and Freedom-4A Stimulator (Receiver Kit)
#### Wearable Antenna Assembly (WAA Kit)
| WAA | The WAA housing includes the following components: |
|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| A. Microwave Field Stimulator (MFS) – A printed circuit board (PCB) that generates 915 MHz RF power with embedded waveform parameter settings and switches for changing parameter settings as needed by the user. Similar to K150517 with design updates for USB charging and PCB layout; | |
| Wearable Antenna Assembly (WAA Kit) | |
| | B. Switch Membrane – A elastomeric silicon rubber pad that |
| | corresponds to switches on the MFS that allows the user to turn |
| | the device on/off or increase or decrease power amplitude as well |
| | as interpret device power status (On, Off, Charging, Transmitting, |
| | and Bluetooth® Connection). Identical to the component used for |
| | K152178 (StimQ PNS System); |
| | C. Battery Assembly – A battery and wire assembly for charging and |
| | the MFS for power delivery. New component for WAA. |
| | Transmitting (Tx) Antenna Assembly – An antenna and coaxial cable |
| | assembly that is attached to the WAA that is used to transmit |
| | microwave energy to the implanted Stimulator. Similar design and |
| | functionality as K152178, with design updates for indication. |
| Charger Kit | |
| Battery Charger | An off-the-shelf battery charger that uses a power adapter and USB to |
| | micro-USB cable to recharge the encased lithium ion battery of the |
| | WAA. New accessory component used to recharge the WAA. |
| Freedom-8A and Freedom-4A Spare Lead (Spare Lead Kit) | |
| Freedom-8A<br>Spare Lead,<br>Freedom-4A<br>Spare Lead | A polyurethane (Pellethane 55D) casing with an embedded receiver,<br>flexible circuit board and electrodes (Platinum Iridium 90:10) that is<br>placed in the patient's epidural space. The Freedom-8A Spare Lead<br>has eight (8) electrodes, and the Freedom-4A Spare Lead has four (4)<br>electrodes. Identical in design and functionality to the Freedom-8A<br>and Freedom-4A Stimulator. |
| RF Stylet | A copper and PEEK cable with dual couplers; placed within the<br>center lumen of the Freedom-8A or Freedom-4A Spare Lead with the<br>distal end combination of RF Stylet and Spare Lead being placed<br>under the skin. Two (2) RF Stylets are provided with each kit.<br>Identical in design and functionality to the Receiver. |
| Stylet(s) | A stainless steel wire with a polypropylene handle that is inserted<br>into the open central lumen of the stimulator to provide rigidity<br>during implantation. Two (2) stylets are provided in the Spare Lead<br>Kit, one straight and one bent, each with diameter of 0.30 mm.<br>Identical in design and functionality to the stylets in the Receiver Kit. |
| Needle | A 13-gauge stainless steel needle that acts as a conduit for passage of<br>the Spare Lead into the epidural space. Identical in design and<br>functionality to the needle in the Receiver Kit. |
| Guidewire | A stainless steel, rigid, solid core guidewire used to create a hollow<br>pathway in the epidural space for the Spare Lead to pass through<br>easily. Identical in design and functionality to the guidewire in the<br>Receiver Kit. |
| | |
| Freedom-8A<br>Trial Lead,<br>Freedom-4A<br>Trial Lead | A polyurethane (Pellethane 55D) casing with an embedded receiver, flexible circuit board and electrodes (Platinum Iridium 90:10) that is placed percutaneously in the patient's epidural space. The Freedom-8A Trial Lead has eight (8) electrodes, and the Freedom-4A Trial Lead has four (4) electrodes. Identical in design and functionality to the Freedom-8A and Freedom-4A Stimulator. |
| RF Stylet | A copper and PEEK cable with dual couplers; placed within the center lumen of the Freedom-8A or Freedom-4A Trial Lead with the distal end combination of RF Stylet and Trial Lead being placed under the skin. Two (2) RF Stylets are provided with each kit. Identical in design and functionality to the Receiver. |
| Stylet(s) | A stainless steel wire with a polypropylene handle that is inserted into the open central lumen of the stimulator to provide rigidity during implantation. Two (2) stylets are provided in the Trial Lead Kit, one straight and one bent with a diameter of 0.30 mm. Identical in design and functionality to the stylets in the Receiver Kit. |
| Needle | A 13-gauge stainless steel needle that acts as a conduit for passage of the Trial Lead into the epidural space. Identical in design and functionality to the needle in the Receiver Kit. |
| Guidewire | A stainless steel, rigid, solid core guidewire used to create a hollow pathway in the epidural space for the Trial Lead to pass through more easily. Identical in design and functionality to the guidewire in the Receiver Kit. |
{6}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for Stimwave Technologies. The word "stim" is in a large, purple, sans-serif font. Below and to the right of "stim" is the word "wave" in a green, sans-serif font. Below "stim wave" is the word "TECHNOLOGIES" in a smaller, blue, sans-serif font inside of a rounded rectangle. To the right of the words is a graphic of three concentric arcs in a blue-purple color.
# Wearable Antenna Assembly (WAA Kit)
{7}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for Stimwave Technologies. The word "stim" is in a blue sans-serif font, and the word "wave" is in a green sans-serif font. Below the words "stim wave" is the word "TECHNOLOGIES" in a smaller blue sans-serif font with a blue line above and below the word. To the right of the words is a graphic of blue concentric semi-circles.
Suture Sleeve Cap A polyurethane (Pellethane 55D) cap that is placed over the proximal end of the Stimulator. The Suture Sleeve Cap is attached to the Freedom-8A/4A Spare Lead and can be sutured to tissue to reduce the possibility of device migration. Identical in design and functionality to the Suture Sleeve Cap in the Receiver Kit.
# Freedom-8A and Freedom-4A Trial Lead (Trial Lead Kit)
# Sterile Revision Kit
| Stylet(s) | A stainless steel wire with a polypropylene handle that is inserted into the open central lumen of the stimulator to provide rigidity during implantation. Two (2) stylets are provided in the Sterile Revision Kit, one straight and one bent with a diameter of 0.30 mm. Identical in design and functionality to the stylets in the Receiver Kit. |
|-----------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Needle | A 13-gauge stainless steel needle that acts as a conduit for passage of the Stimulator/Lead into the epidural space. Identical in design and functionality to the needle in the Receiver Kit. |
| Guidewire | A stainless steel, rigid, solid core guidewire used to create a hollow pathway in the epidural space for the Stimulator/Lead to pass through easily. Identical in design and functionality to the guidewire in the Receiver Kit. |
{8}------------------------------------------------
Image /page/7/Picture/0 description: The image contains the logo for Stimwave Technologies. The word "stim" is in blue, and the word "wave" is in green. Below the words is the word "TECHNOLOGIES" in blue. To the right of the words are three curved lines that look like sound waves in blue.
# 6. Indication for Use Statement
The Freedom Spinal Cord Stimulator (SCS) System is intended as the sole mitigating agent, or as an adjunct to other modes of therapy used in a multidisciplinary approach for chronic, intractable pain of the trunk and/or lower limbs, including unilateral or bilateral pain. The Freedom-8A Trial Lead Kit is only used in conjunction with the Freedom-8A Stimulator Receiver Kit, and the Freedom-4A Trial Lead Kit is used for either the Receiver Kit Freedom-4A Stimulator or the Receiver Kit Freedom-8A Stimulator. The trial devices are solely used for trial stimulation (no longer than 30 days) to determine efficacy before recommendation for a permanent (long term) device.
# 7. Substantial Equivalence Discussion
The following table compares the Stimwave Freedom SCS System to the predicate device with respect to intended use, technological characteristics and principles of operation, providing more detailed information regarding the basis for the determination of substantial equivalence.
| Comparator | Stimwave<br>Freedom SCS<br>System<br>(K162161) | Stimwave<br>Freedom SCS<br>System<br>(K160600) | Stimwave<br>Freedom SCS<br>System<br>(K150517) | Stimwave<br>Freedom SCS<br>System<br>(K141399) |
|-------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------|---------------------------------------------------------------------------------|----------------------------------------------------------|
| Product Code | GZB | Same as K162161 | Same as K162161 | Same as K162161 |
| Regulation No. | 882.5880 | Same as K162161 | Same as K162161 | Same as K162161 |
| Regulation Name | Stimulator, Spinal-Cord,<br>Implanted (Pain Relief) | Same as K162161 | Same as K162161 | Same as K162161 |
| Intended Use | Stimulation of spinal cord for<br>chronic, intractable pain of<br>trunk and lower limbs | Same as K162161 | Same as K162161 | Same as K162161 |
| Mode of Action | RF wireless transmission of<br>energy to produce<br>stimulation at Stimulator<br>electrodes. WAA sends a<br>pulsed RF signal on a carrier<br>frequency of 915MHz to the<br>Stimulator | Same as K162161 | Same as K162161 | Same as K162161 |
| Implant Site | Epidural space, L5 to T5 | Same as K162161 | Same as K162161 | Same as K162161 |
| Environmental<br>Use | Hospital, Home | Same as K162161 | Same as K162161 | Same as K162161 |
| Intended Clinician | Orthopedic, Neurosurgeon,<br>Anesthesiologist | Same as K162161 | Same as K162161 | Same as K162161 |
| Intended User | Layperson | Same as K162161 | Same as K162161 | Same as K162161 |
| Electrode<br>Material | Platinum-iridium 90:10 | Same as K162161 | Same as K162161 | Same as K162161 |
| Stimulator Body<br>Material | Polyurethane 2363-55D | Same as K162161 | Same as K162161 | Same as K162161 |
| Cable Features | Multi-lumen Tube | Same as K162161 | Same as K162161 | Same as K162161 |
| Stimulator Length | 45 centimeters | Same as K162161 | Same as K162161 | Same as K162161 |
| Diameter | 1.35 millimeters | Same as K162161 | Same as K162161 | Same as K162161 |
| Electrode Array<br>Length | 24.0 millimeters<br>52.0 millimeters | Same as K162161 | Same as K162161 | Same as K162161 |
| Comparator | Stimwave<br>Freedom SCS<br>System<br>(K162161) | Stimwave<br>Freedom SCS<br>System<br>(K160600) | Stimwave<br>Freedom SCS<br>System<br>(K150517) | Stimwave<br>Freedom SCS<br>System<br>(K141399) |
| No. of Electrodes | 4 or 8 | Same as K162161 | Same as K162161 | 4 |
| Electrode Length | 3.0 millimeters | Same as K162161 | Same as K162161 | Same as K162161 |
| Electrode Spacing | 4.0 millimeters | Same as Freedom | Same as K162161 | Same as K162161 |
| Electrode Surface Area | 12.72 mm² | Same as K162161 | Same as K162161 | Same as K162161 |
| Method of Introduction | Percutaneous and Anchor<br>Incision | Same as K162161 | Same as K162161 | Same as K162161 |
| Tissue Contact | Yes | Same as K162161 | Same as K162161 | Same as K162161 |
| Sterilization | Ethylene Oxide (EO) | Same as K162161 | Same as K162161 | Same as K162161 |
| Labeling | Labeled as Sterile, Single<br>Use, Prescription Device | Same as K162161 | Same as K162161 | Same as K162161 |
| Package | Backer card and two sterile<br>pouches | Same as K162161 | Same as K162161 | Blister<br>Tray/Tyvek Lid |
| Pulse Frequency | 5 to 1500 Hertz | Same as K162161 | Same as K162161 | 2 to 1500 Hertz |
| Pulse Width | 50 to 500 microseconds | Same as K162161 | Same as K162161 | Same as K162161 |
| Current/Voltage Regulated | Current | Same as K162161 | Same as K162161 | Same as K162161 |
| Output Voltage (300 Ω) | 0 to 4.1 V | Same as K162161 | Same as K162161 | 0 to 6.3 V |
| Output Voltage (500 Ω) | 0 to 6.4 V | Same as K162161 | Same as K162161 | 0 to 7.2 V |
| Output Voltage (800 Ω) | 0 to 7.5 V | Same as K162161 | Same as K162161 | 0 to 8.0 V |
| Output Current (300 Ω) | 0 to 13.5 mA | Same as K162161 | Same as K162161 | 0 to 21 mA |
| Output Current (500 Ω) | 0 to 12.8 mA | Same as K162161 | Same as K162161 | 0 to 15 mA |
| Output Current (800 Ω) | 0 to 9.4 mA | Same as K162161 | Same as K162161 | 0 to 10 mA |
| Waveform | Charge Balanced (delayed)<br>Biphasic asymmetrical | Same as K162161 | Same as K162161 | Same as K162161 |
| Polarity | Programmable<br>(Anode, Cathode, or Off) | Same as K162161 | Same as K162161 | Fixed |
| Pulse Shape | Decaying Exponential | Same as K162161 | Same as K162161 | Same as K162161 |
| Avg. Current Density (300 Ω) | 105.0 mA/cm² | Same as K162161 | Same as K162161 | 111.6 mA/cm² |
| Avg. Current Density (500 Ω) | 95.1 mA/cm² | Same as K162161 | Same as K162161 | 96.7 mA/cm² |
| Avg. Current Density (800 Ω) | 69.0 mA/cm² | Same as K162161 | Same as K162161 | 77.0 mA/cm² |
| Max. Phase Charge* (300 Ω) | 6.8 µC/pulse | Same as K162161 | Same as K162161 | 10.5 µC/pulse |
| Max. Phase Charge* (500 Ω) | 6.4 µC/pulse | Same as K162161 | Same as K162161 | 7.2 µC/pulse |
| Max. Phase Charge* (800 Ω) | 4.7 µC/pulse | Same as K162161 | Same as K162161 | 5.0 µC/pulse |
| Max. Charge Density* (300 Ω) | 53.1 µC/cm² | Same as K162161 | Same as K162161 | 82.5 µC/cm² |
| Max. Charge Density* (500 Ω) | 50.3 µC/cm² | Same as K162161 | Same as K162161 | 56.6 µC/cm² |
| Max. Charge Density* (800 Ω) | 36.9 µC/cm² | Same as K162161 | Same as K162161 | 39.3 µC/cm² |
| Comparator | Stimwave<br>Freedom SCS<br>System<br>(K162161) | Stimwave<br>Freedom SCS<br>System<br>(K160600) | Stimwave<br>Freedom SCS<br>System<br>(K150517) | Stimwave<br>Freedom SCS<br>System<br>(K141399) |
| Max. Current<br>Density* (300 Ω) | 106.1 mA/cm² | Same as K162161 | Same as K162161 | 165.1 mA/cm² |
| Max. Current<br>Density* (500 Ω) | 100.6 mA/cm² | Same as K162161 | Same as K162161 | 113.2 mA/cm² |
| Max. Current<br>Density* (800 Ω) | 73.9 mA/cm² | Same as K162161 | Same as K162161 | 78.6 mA/cm² |
| Net Charge | 0 μC | Same as K162161 | Same as K162161 | Same as K162161 |
| Avg. Phase Power<br>(300 Ω) | 0.053 W/phase | Same as K162161 | Same as K162161 | 0.060 W/phase |
| Avg. Phase Power<br>(500 Ω) | 0.073 W/phase | Same as K162161 | Same as K162161 | 0.076 W/phase |
| Avg. Phase Power<br>(800 Ω) | 0.062 W/phase | Same as K162161 | Same as K162161 | 0.060 W/phase |
| Avg. Phase Power<br>Density (300 Ω) | 0.42 W/cm²/phase | Same as K162161 | Same as K162161 | 0.48 W/cm²/phase |
| Avg. Phase Power<br>Density (500 Ω) | 0.58 W/cm²/phase | Same as K162161 | Same as K162161 | 0.59 W/cm²/phase |
| Avg. Phase Power<br>Density (800 Ω) | 0.48 W/cm²/phase | Same as K162161 | Same as K162161 | 0.60 W/cm²/phase |
| Pulse Delivery<br>Mode | Continuous | Same as K162161 | Same as K162161 | Same as K162161 |
| ON/OFF Times | No Cycling | Same as K162161 | Same as K162161 | Same as K162161 |
| Current Path<br>Options | Bipolar | Same as K162161 | Same as K162161 | Same as K162161 |
| Power Delivery | Embedded receiver and<br>coupled receiver in lumen of<br>Stimulator body | Same as K162161 | Coupled receiver,<br>built into<br>Stimulator body | Coupled receiver,<br>built into<br>Stimulator body |
| Transmit<br>Frequency | 915 MHz | Same as K162161 | Same as K162161 | Same as K162161 |
| Material | Platinum-iridium 90:10,<br>Polyurethane 2363-55D | Same as K162161 | Same as K162161 | Same as K162161 |
| Sterile | Yes - ethylene oxide | Same as K162161 | Same as K162161 | Same as K162161 |
| Contract<br>Sterilizer | Steris Isomedix Services | Life Science<br>Outsourcing | Life Science<br>Outsourcing | Oscor, Inc. |
| Single-Use | Yes | Same as K162161 | Same as K162161 | Same as K162161 |
| Shelf Life | 2 year | Same as K162161 | Same as K162161 | 1 year |
| Complies with<br>ISO 10993-1 | Yes | Same as K162161 | Same as K162161 | Same as K162161 |
| Safety Testing<br>Passed | Yes | Same as K162161 | Same as K162161 | Same as K162161 |
| MR Conditional | No, MR Unsafe | Same as K162161 | Same as K162161 | Yes |
| Accessories | Receiver/RF Stylet,<br>Stylet(s), Guidewire,<br>Needle, Suture Sleeve Cap | Receiver,<br>Stylet(s),<br>Guidewire,<br>Needle, Suture<br>Sleeve Cap | Stylet(s),<br>Guidewire,<br>Needle, Suture<br>Sleeve Cap | Stylet(s),<br>Guidewire,<br>Needle, Suture<br>Sleeve Cap |
| Charger | USB Charger | Wireless Charger | Wireless Charger | Wireless Charger |
| Wearable<br>Antenna<br>Assembly | Aluminum transmitter and<br>separate, connected Antenna | ABS plastic<br>transmitter<br>contained in a<br>wearable athletic<br>pouch belt | ABS plastic<br>transmitter<br>contained in a<br>wearable athletic<br>pouch belt | Thermoplastic<br>polyurethane<br>body worn device |
| Comparator | Stimwave<br>Freedom SCS<br>System<br>(K162161) | Stimwave<br>Freedom SCS<br>System<br>(K160600) | Stimwave<br>Freedom SCS<br>System<br>(K150517) | Stimwave<br>Freedom SCS<br>System<br>(K141399) |
| Software Level of<br>Concern | Moderate | Same as K162161 | Same as K162161 | Same as K162161 |
| iPad Application | WaveCrest™ 2.2 | WaveCrest™ | WaveCrest™ | WaveCrest™ |
| Kits | Receiver Kit, Spare Lead Kit,<br>Trial Lead Kit, Sterile Revision<br>Kit, WAA Kit, Charger Kit | Receiver Kit,<br>Trial Lead Kit,<br>WAA Kit | Receiver Kit,<br>Trial Lead Kit,<br>WAA Kit | Receiver Kit,<br>Trial Lead Kit,<br>WAA Kit |
Table 5A. Comparison of Characteristics
{9}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the logo for Stimwave Technologies. The word "stim" is in dark blue, and the word "wave" is in green. The word "TECHNOLOGIES" is in dark blue and is inside of a dark blue rounded rectangle. To the right of the word "wave" are three dark blue curved lines that resemble radio waves.
{10}------------------------------------------------
Image /page/9/Picture/0 description: The image shows the logo for Stimwave Technologies. The word "stim" is in a dark blue font, and the word "wave" is in a green font. To the right of the word "wave" is a graphic of three curved lines that resemble radio waves. Underneath the word "stimwave" is the word "TECHNOLOGIES" in a smaller, dark blue font.
{11}------------------------------------------------
Image /page/10/Picture/0 description: The image contains the logo for Stimwave Technologies. The word "stim" is in a blue sans-serif font on the top left. Below that, the word "wave" is in a green sans-serif font. Below that, the word "TECHNOLOGIES" is in a blue sans-serif font with a blue line above and below it. To the right of the word "wave" are three concentric semi-circles in a blue color.
(*) asterisk denotes that formulas were used for the calculations.
# 8. Biocompatibility Data
The materials of the Freedom-8A/4A Stimulator (K162161) in direct contact with tissue remain unchanged from the Freedom SCS System (K141399, K150517, and K160600) and thus, the biocompatibility tests conducted on representative subassemblies of the Freedom SCS System (Freedom-4, K141399) directly apply to the Freedom SCS System (K150517, K160600, and K162161). The materials, construction and intended use of the Freedom SCS System is comparable to the predicate device, and have a long history of safety with respect to biocompatibility. The biological safety of the Freedom-8A/4A Stimulator (same as the Freedom-4 Stimulator) was evaluated in accordance to ISO 10993-1:2009 and guidance document Blue Book Memorandum G95-1 Use of International Standard ISO-10993, and Biological Evaluation of Medical Devices: Part 1: Evaluation and Testing. Under these, for the stated indications for use, the device was classified as a (C), implant device in contact with tissue/bone. The results for the biocompatible testing of the Freedom-8A/4A Stimulator (same as the Freedom-4 Stimulator) for cytotoxicity, sensitization, irritation or intracutaneous reactivity, acute systemic toxicity, genotoxicity, implantation (4, 8, and 13 weeks), and subchronic toxicity demonstrated no negative impacts from the materials that are used in the Freedom SCS System. The Freedom-8A/4A Stimulator materials in direct tissue contact include Pellethane 55D (Stimulator) and Pt-Ir (90:10) (Stimulator only), both having an extensive record (previously cleared and approved) of chronic and carcinogenetic safety. The Receiver/RF Stylet is never in direct or indirect contact with tissue. The WAA is intended to be on top of a thin shirt or article of clothing around the midsection of the patient. The User Manual provided to the patient describes that the WAA should always be worn on top of a layer of clothing. The WAA does not come into contact with the patient's skin. The categorization by nature of body contact of the WAA is thus "noncontacting device", and not included in the scope of ISO 10993-1:2009. The Freedom SCS System meets biological safety and compatibility requirements of ISO 10993-1:2009 and Blue Book Memorandum G95-1.
# 9. Non-Clinical Performance Data
The Freedom SCS System was tested to verify that the performance meets the system design requirements as well as all applicable voluntary standards. The Freedom SCS System complies with all design requirements and applicable voluntary standards.
{12}------------------------------------------------
Image /page/11/Picture/0 description: The image shows the logo for Stimwave Technologies. The word "stim" is in blue, and the word "wave" is in green. To the right of the word "wave" are three curved lines that resemble radio waves. Underneath the words "stim wave" is the word "TECHNOLOGIES" in blue.
Stimwave Technologies Incorporated Traditional 510(k) Premarket Submission Freedom Spinal Cord Stimulator System
AAMI ANSI ISO 14708-3:2008 - For protection from temperature change including shipping and storage temperature ranges, the Freedom-8A/4A Stimulator was functional. receiving a safe rating following post visual inspection and passed the change of temperature testing performed as specified by AAMI ANSI ISO 14708-3:2008. For atmospheric pressure change, the Freedom-8A/4A Stimulator were functional following post testing functionality inspection and passed atmospheric pressure change testing as specified by AAMI ANSI ISO 14708-3:2008. This testing presented for K162161 is leveraged from K150517 and is directly applicable for demonstration of device safety and efficacy as the packaging and the Freedom-8A/4A Stimulator remains the same. The Receiver/RF Stylet is a passive component that does not impact the outcome of the leveraged tests for safety and effectiveness.
For testing external defibrillation exposure, the Freedom-8A/4A Stimulator and Receiver were verified as functional after exposure to external defibrillation. Thus, the Freedom SCS System complies with testing as specified by AAMI ANSI ISO 14708-3:2008. The Receiver/RF Stylet testing presented for K162161 is leveraged from K160600 and is directly applicable for demonstration of device safety and efficacy as Receiver/RF Stylet remains the same.
Following the thermal shock testing, the Freedom-8A/4A Stimulator was found to have "no irreversible damage" and fully functional as specified by the manufacturer, and to have no physical anomalies present at the time of inspection. Thus, the Freedom-8A/4A Stimulator comply with the thermal shock design requirements and the applicable standard. This testing presented for K162161 is leveraged from K150517 and is directly applicable for demonstration of device safety and efficacy as the Freedom-8A/4A Stimulator remains the same. The Receiver/RF Stylet is a passive component that does not impact the outcome of the leveraged tests for safety and effectiveness.
For leakage current testing, the Freedom-8A/4A Stimulator was produced zero leakage current on all tested paths for…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.