The StimQ Peripheral Nerve Stimulator (PNS) System is indicated for pain management in adults who have severe intractable chronic pain of peripheral nerve origin, as the sole mitigating agent, or as an adjunct to other modes of therapy used in a multidisciplinary approach. The StimQ PNS System is not intended to treat pain in the craniofacial region. The StimQ Trial Lead Kit is only used in conjunction with the StimQ Stimulator Receiver Kit. The trial devices are solely used for trial stimulation (no longer than 30 days) to determine efficacy before recommendation for a permanent (long term) device.
Device Story
System provides peripheral nerve stimulation for chronic intractable pain; utilizes pulsed electrical current to alter pain signal transmission. Comprised of implantable stimulator (receiver) and externally worn transmitter (SWAG). Implanted percutaneously near peripheral nerves in limbs/torso. Transmitter uses 915 MHz RF power to wirelessly power implant. User operates transmitter via membrane switches to adjust power amplitude/on-off status. Healthcare providers (orthopedic, neurosurgeon, anesthesiologist) perform implantation. Output affects pain signal transmission to brain; provides therapeutic relief. Trial period (max 30 days) determines efficacy before permanent implantation.
Clinical Evidence
No clinical data. Substantial equivalence supported by non-clinical bench testing, including biocompatibility (ISO 10993-1), electrical safety (IEC 60601-1), electromagnetic compatibility (IEC 60601-1-2), and mechanical durability (tensile, flex, torsion).
Technological Characteristics
Implant: Polyurethane (Pellethane 55D) casing, Pt-Ir 90:10 electrodes. 4 or 8 electrode array. 915 MHz RF power transfer. Biphasic, charge-balanced, asymmetrical waveform. Pulse frequency 5-1500 Hz, pulse width 50-500 µs. Sterilization: Ethylene Oxide. Connectivity: Wireless inductive/RF coupling. Standards: AAMI ANSI ISO 14708-3, IEC 60601-1, IEC 60529, ISO 10993-1.
Indications for Use
Indicated for adults with severe intractable chronic pain of peripheral nerve origin in limbs or torso. Contraindicated for craniofacial pain. Used as sole therapy or adjunct in multidisciplinary approach.
Regulatory Classification
Identification
An implanted peripheral nerve stimulator for pain relief is a device that is used to stimulate electrically a peripheral nerve in a patient to relieve severe intractable pain. The stimulator consists of an implanted receiver with electrodes that are placed around a peripheral nerve and an external transmitter for transmitting the stimulating pulses across the patient's skin to the implanted receiver.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the circumference. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other, with flowing lines extending downward from the bottom profile.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
March 11, 2016
StimQ LLC Elizabeth Greene Director of Quality Assurance and Regulatory Affairs 901 East Las Olas Boulevard Suite 201 Fort Lauderdale, Florida 33301
Re: K152178
Trade/Device Name: StimQ Peripheral Nerve Stimulator (PNS) System Regulation Number: 21 CFR 882.5870 Regulation Name: Stimulator, Peripheral Nerve, Implanted (Pain Relief) Regulatory Class: Class II Product Code: GZF Dated: March 8, 2016 Received: March 9, 2016
Dear Ms. Greene,
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{1}------------------------------------------------
the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
| William J. Heetderks -A | Digitally signed by William J. Heetderks -A |
|-------------------------|---------------------------------------------|
| | DN: c=US, o=U.S. Government, ou=HHS, |
| | ou=NIH, ou=People, |
| | 0.9.2342.19200300.100.1.1=0010149848, |
| | cn=William J. Heetderks -A |
| | Date: 2016.03.11 15:31:15 -05'00' |
for
Carlos L. Peña, PhD, MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### Indications for Use Statement
#### 510(k) Number (if known):
Device Name: StimO Peripheral Nerve Stimulator (PNS) System
#### Indications For Use:
The StimQ Peripheral Nerve Stimulator (PNS) System is indicated for pain management in adults who have severe intractable chronic pain of peripheral nerve origin, as the sole mitigating agent, or as an adjunct to other modes of therapy used in a multidisciplinary approach. The StimQ PNS System is not intended to treat pain in the craniofacial region. The StimQ Trial Lead Kit is only used in conjunction with the StimQ Stimulator Receiver Kit. The trial devices are solely used for trial stimulation (no longer than 30 days) to determine efficacy before recommendation for a permanent (long term) device.
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for "stimQ". The word "stim" is written in blue, and the "Q" is written in green and is shaped like a magnifying glass. Behind the word "stimQ" are blue concentric circles that look like sound waves.
# 510(k) Summary
for
# StimQ Peripheral Nerve Stimulator (PNS) System
#### 1. Submission Sponsor
| StimQ LLC | |
|-----------------------------|------------------------------------------------------------------------------|
| 901 East Las Olas Boulevard | |
| Suite 201 | |
| Fort Lauderdale | |
| Florida 33301 | |
| USA | |
| Phone: | 786.475.7228 |
| Fax: | 786.475.7228 |
| Contact: | Elizabeth Greene, Vice President of Quality Assurance and Regulatory Affairs |
### 2. Date Prepared
July 31, 2015
# 3. Device Identification
| Trade/Proprietary Name: | StimQ Peripheral Nerve Stimulator (PNS) System |
|----------------------------|-------------------------------------------------------|
| Common/Usual Name: | Peripheral Nerve Stimulator |
| Classification Name: | Stimulator, Peripheral Nerve, Implanted (Pain Relief) |
| Classification Regulation: | 882.5870 |
| Product Code: | GZF |
| Device Class: | 2 |
| Classification Panel: | Neurology |
### 4. Legally Marketed Predicate Device(s)
Stimwave Freedom SCS System FR8A-A1, FR8A-B1, FR4A-A1, FR4A-B1, LBRD-915-2A (K150517) Stimwave Freedom SCS System FRE4-A001, WAA-A012 (K141399) Medtronic Mattrix 3271/3272 Neuromodulation System (K934065) Medtronic Xtrel, Model Number 3425 Receiver (K883780) ANS Renew Neurostimulation System Transmitter, Model 2508, Receiver Model 3408, Antennae Models 1220 and 1230, Lead Models 3143, 3146, 3153, 3156, 3183 and 3186, Extension Models 3382, 3383, 3341, 3342 and 3343 (K000852)
# 5. Device Description
The StimO LLC (StimO) StimO Peripheral Nerve Stimulator System) is used for peripheral neural stimulation to provide therapeutic relief for chronic, intractable pain of peripheral nerve origin. The therapy utilizes pulsed electrical current to create an
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for StimQ. The logo features the word "stim" in a sans-serif font, followed by a green letter "Q" that is shaped like a magnifying glass. Behind the word "stimQ" are three concentric arcs in blue, resembling sound waves or a signal. The overall design is clean and modern, suggesting a focus on research, analysis, or technology.
electrical energy field that acts on peripheral nerves in the limbs and torso to alter the transmission of pain signals to the brain. The System is comprised of an implantable stimulator (Freedom-8A/4A Stimulator) and an externally worn transmitter (StimQ Wearable Antenna Gear (SWAG)) to power the device. The System is implanted only following a successful trial period with the Freedom-8A/4A Trial Lead.
#### Freedom-8A and Freedom-4A Stimulator (Receiver Kit)
| Freedom-8A<br>Stimulator,<br>Freedom-4A<br>Stimulator | A polyurethane (Pellethane 55D) casing with an embedded receiver,<br>flexible circuit board and electrodes (Platinum Iridium 90:10) that is<br>placed next to peripheral nerves in the extremities and torso. The<br>Freedom-8A Stimulator has eight (8) electrodes, and the Freedom-4A<br>Stimulator has four (4) electrodes. |
|-------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Stylet(s) | A stainless steel wire with a polypropylene handle that is inserted into<br>the open central lumen of the stimulator to provide rigidity during<br>implantation. Two (2) stylets are provided in the Receiver Kit, one<br>straight and one bent, each with diameter of 0.25 mm. |
| Needle and<br>Cannula | A 16-gauge stainless steel needle that is packaged inserted in the<br>fluorinated ethylene propylene (FEP) Cannula that acts as a conduit for<br>passage of the Stimulator next to the peripheral nerve. |
| Guidewire | A stainless steel, rigid, solid core guidewire used to create a hollow<br>pathway in tissue for the Stimulator to pass through easily. The<br>manufacturer of the guidewire is Guidewire Technologies. |
#### StimQ Wearable Antenna Gear (SWAG) Transmitter Kit
| Transmitter | The SWAG Transmitter housing includes the following components: |
|-------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| i. | Transmitting (Tx) Antenna - Flexible Dipole Antenna used to transmit microwave energy to the implanted Stimulator; |
| ii. | Microwave Field Stimulator (MFS) – A printed circuit board (PCB) that generates 915 MHz RF power with embedded waveform parameter settings and switches for changing parameter settings as needed by the user; |
| iii. | Switch Membrane - A elastomeric silicon rubber pad that corresponds to switches on the MFS that allows the user to turn the device on/off or increase or decrease power amplitude as well as interpret device power status (On, Off, Charging, Transmitting, and Bluetooth® Connection); |
| iv. | Battery Assembly – A battery and wire assembly coupled with the Wireless Charging Coil Assembly for charging and the MFS for power delivery; |
| v. | Wireless Charging Coil Assembly – Consists of an inductive charging coil and battery that receives energy from the Wireless Battery Charging Pad. The battery charger components on the MFS are used to transfer charge into the 3.7V lithium ion battery by facilitating power transfer and warns the system when battery power is low |
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for StimQ. The logo features the word "stim" in blue, followed by a green letter "Q" that is stylized to resemble a magnifying glass. Behind the "Q" are concentric blue circles, suggesting sound waves or a signal. The overall design is clean and modern, with a focus on the company name and a visual representation of its purpose.
# StimQ Wearable Antenna Gear (SWAG) Transmitter Kit
| Wireless Battery Charging Pad | An off-the-shelf Qi-compliant charging pad (RP-WCN7, RP-WCN12, or RP-WCN13) that uses inductive charging technology to recharge the encased lithium ion battery of the SWAG Transmitter. |
|-------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|-------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
# StimQ Wearable Antenna Gear (SWAG) Accessory Kit (each packaged separately)
| Upper Arm | A wearable unit, sourced “off-the-shelf,” designed to fit an upper extremity, adjustable from small to extra large sizes constructed of lightweight elastic neoprene fabric that is compatible with the Velcro Hook for attaching the SWAG Transmitter. |
|-----------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Lower Arm | A wearable unit, sourced “off-the-shelf,” designed to fit an upper extremity, adjustable from small to extra large sizes constructed of lightweight elastic neoprene fabric that is compatible with the Velcro Hook for attaching the SWAG Transmitter. |
| Torso | A wearable unit, sourced “off-the-shelf,” designed to fit around the torso, adjustable from small to extra large sizes constructed of lightweight elastic neoprene fabric that is compatible with the Velcro Hook for attaching the SWAG Transmitter. |
| Upper Leg | A wearable unit, sourced “off-the-shelf,” designed to fit a lower extremity, adjustable from small to extra large sizes constructed of lightweight elastic neoprene fabric that is compatible with the Velcro Hook for attaching the SWAG Transmitter. |
| Lower Leg | A wearable unit, sourced “off-the-shelf,” designed to fit a lower extremity, adjustable from small to extra large sizes constructed of lightweight elastic neoprene fabric that is compatible with the Velcro Hook for attaching the SWAG Transmitter. |
Freedom-8A and Freedom-4A Trial Lead (Trial Lead Kit)
| Freedom-8A Trial Lead, Freedom-4A Trial Lead | A polyurethane (Pellethane 55D) casing with an embedded receiver, flexible circuit board and electrodes (Platinum Iridium 90:10) that is placed next to peripheral nerves in the extremities and torso. The Freedom-8A Trial Lead has eight (8) electrodes, and the Freedom-4A Trial Lead has four (4) electrodes. |
|----------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Stylet(s) | A stainless steel wire with a polypropylene handle that is inserted into the open central lumen of the stimulator to provide rigidity during implantation. Two (2) stylets are provided in the Trial Lead Kit, one straight and one bent each with a diameter of 0.25 mm. |
| Needle and Cannula | A 16-gauge stainless steel needle that is packaged inserted in the fluorinated ethylene propylene (FEP) Cannula that acts as a conduit for passage of the Stimulator next to the peripheral nerve. |
| Guidewire | A stainless steel, rigid, solid core guidewire used to create a hollow pathway in tissue for the Stimulator to pass through easily. The manufacturer of the guidewire is Guidewire Technologies. |
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for StimQ. The logo has the word "stim" in a dark blue font, followed by a green "Q" that is stylized to look like a magnifying glass. Behind the word and the "Q" are three curved lines that are also dark blue. The curved lines are meant to represent radio waves.
# 6. Indication for Use Statement
The StimO Peripheral Nerve Stimulator (PNS) System is indicated for pain management in adults who have severe intractable chronic pain of peripheral nerve origin, as the sole mitigating agent, or as an adjunct to other modes of therapy used in a multidisciplinary approach. The StimQ PNS System is not intended to treat pain in the craniofacial region. The StimQ Trial Lead Kit is only used in conjunction with the StimQ Stimulator Receiver Kit. The trial devices are solely used for trial stimulation (no longer than 30 days) to determine efficacy before recommendation for a permanent (long term) device.
### 7. Substantial Equivalence Discussion
The following table compares the StimQ PNS System to the predicate device with respect to intended use, technological characteristics and principles of operation, providing more detailed information regarding the basis for the determination of substantial equivalence. Stimwave physically measured Medtronic and ANS predicate devices to obtain the values listed in Table 5A.
| Comparator | StimQ PNS<br>System | Stimwave<br>Freedom<br>SCS System<br>(K150517) | Stimwave<br>Freedom<br>SCS System<br>(K141399) | Medtronic<br>Mattrix<br>3271/3272<br>(K934065) | Medtronic<br>Xtrel, Model<br>Number<br>3425<br>(K883780) | ANS Renew<br>(K000852) |
|-------------------------------------------|----------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------|
| Product Code | GZF | GZB | GZB | GZB and<br>GZF | GZB | GZB and<br>GZF |
| Regulation No. | 882.5870 | 882.5880 | 882.5880 | 882.5880 &<br>882.5870 | 882.5880 | 882.5880 &<br>882.5870 |
| Regulation Name | Stimulator,<br>Peripheral<br>Nerve,<br>Implanted<br>(Pain Relief) | Stimulator,<br>Spinal-Cord,<br>Implanted<br>(Pain Relief) | Stimulator,<br>Spinal-Cord,<br>Implanted<br>(Pain Relief) | Stimulator,<br>Spinal-Cord,<br>Implanted<br>(Pain Relief)<br>Stimulator,<br>Peripheral<br>Nerve,<br>Implanted<br>(Pain Relief) | Stimulator,<br>Spinal-Cord,<br>Implanted<br>(Pain Relief) | Stimulator,<br>Spinal-Cord,<br>Implanted<br>(Pain Relief)<br>Stimulator,<br>Peripheral<br>Nerve,<br>Implanted<br>(Pain Relief) |
| Intended Use | Stimulation<br>of peripheral<br>nerves for<br>chronic,<br>intractable<br>pain of<br>peripheral<br>nerve origin | Stimulation<br>of spinal cord<br>for chronic,<br>intractable<br>pain of trunk<br>and lower<br>limbs | Stimulation<br>of spinal cord<br>for chronic,<br>intractable<br>pain of trunk<br>and lower<br>limbs | Stimulation<br>of spinal cord<br>for chronic,<br>intractable<br>pain of trunk<br>and lower<br>limbs; and<br>peripheral<br>nerve origin | Stimulation<br>of spinal cord<br>for chronic,<br>intractable<br>pain of trunk<br>and lower<br>limbs | Stimulation<br>of spinal cord<br>for chronic,<br>intractable<br>pain of trunk<br>and lower<br>limbs; and<br>peripheral<br>nerve origin |
| Implant Site | Peripheral<br>nerves,<br>excluding<br>craniofacial<br>region | Epidural<br>space, L5 to<br>T5 | Epidural<br>space, L5 to<br>T5 | Epidural<br>space, L5 to<br>T5;<br>Peripheral<br>nerves,<br>excluding<br>craniofacial<br>region | Epidural<br>space, L5 to<br>T5 | Epidural<br>space, L5 to<br>T5;<br>Peripheral<br>nerves,<br>excluding<br>craniofacial<br>region |
| Comparator | StimQ PNS<br>System | Stimwave<br>Freedom<br>SCS System<br>(K150517) | Stimwave<br>Freedom<br>SCS System<br>(K141399) | Medtronic<br>Mattrix<br>3271/3272<br>(K934065) | Medtronic<br>Xtrel, Model<br>Number<br>3425<br>(K883780) | ANS Renew<br>(K000852) |
| Environmental | Hospital,<br>Home | Same as<br>StimQ | Same as<br>StimQ | Same as<br>StimQ | Same as<br>StimQ | Same as<br>StimQ |
| Intended<br>Clinician | Orthopedic,<br>Neurosurgeon,<br>Anesthesiologist | Same as<br>StimQ | Same as<br>StimQ | Same as<br>StimQ | Same as<br>StimQ | Same as<br>StimQ |
| Intended User | Layperson | Same as<br>StimQ | Same as<br>StimQ | Same as<br>StimQ | Same as<br>StimQ | Same as<br>StimQ |
| Electrode<br>Material | Platinum-<br>iridium 90:10 | Same as<br>StimQ | Same as<br>StimQ | Same as<br>StimQ | Same as<br>StimQ | Same as<br>StimQ |
| Stimulator Body<br>Material | Polyurethane<br>2363-55D | Same as<br>StimQ | Same as<br>StimQ | Same as<br>StimQ | Same as<br>StimQ | Same as<br>StimQ |
| Cable Features | Multi-lumen<br>Tube | Multi-lumen<br>Tube | Multi-lumen<br>Tube | Coiled Wires | Coiled Wires | Braided Wire |
| Stimulator<br>Length | 45 cm | 45 cm | 45 cm | 30 to 110 cm | 30 to 110 cm | 30 and 60 cm |
| Diameter | 1.35 mm | 1.35 mm | 1.35 mm | 1.3 mm | 1.3 mm | 1.37 mm |
| Electrode Array<br>Length | 24 mm<br>52 mm | Same as<br>StimQ | Same as<br>StimQ | Same as<br>StimQ | Same as<br>StimQ | Same as<br>StimQ |
| No. of Electrodes | 4 or 8 | 4 or 8 | 4 | 4 | 4 | 4 or 8 |
| Electrode Length | 3.0 mm | Same as<br>StimQ | Same as<br>StimQ | Same as<br>StimQ | Same as<br>StimQ | Same as<br>StimQ |
| Electrode<br>Spacing | 4.0 mm | Same as<br>StimQ | Same as<br>StimQ | Same as<br>StimQ | Same as<br>StimQ | Same as<br>StimQ |
| Electrode<br>Surface Area | 12.72 mm² | 12.72 mm² | 12.72 mm² | 12.25 mm² | 12.25 mm² | "Approx. 13<br>mm²" |
| Method of<br>Introduction | Percutaneous | Same as<br>StimQ | Same as<br>StimQ | Same as<br>StimQ | Same as<br>StimQ | Same as<br>StimQ |
| Tissue Contact | Yes | Same as<br>StimQ | Same as<br>StimQ | Same as<br>StimQ | Same as<br>StimQ | Same as<br>StimQ |
| Sterilization | Ethylene<br>Oxide (EO) | Same as<br>StimQ | Same as<br>StimQ | Same as<br>StimQ | Same as<br>StimQ | Same as<br>StimQ |
| Labeling | Labeled as<br>Sterile,<br>Single Use,<br>Prescription<br>Device | Same as<br>StimQ | Same as<br>StimQ | Same as<br>StimQ | Same as<br>StimQ | Same as<br>StimQ |
| Package | Backer card<br>and two<br>sterile<br>pouches | Blister<br>Tray/Tyvek<br>Lid | Blister<br>Tray/Tyvek<br>Lid | Blister<br>Tray/Tyvek<br>Lid | Blister<br>Tray/Tyvek<br>Lid | Blister<br>Tray/Tyvek<br>Lid |
| Pulse Frequency | 5 to 1500<br>Hertz | 2 to 1500<br>Hertz | 2 to 1500<br>Hertz | 5 to 240<br>Hertz | 5 to 1400<br>Hertz | 10 to 1500<br>Hertz |
| Pulse Width | 50 to 500<br>microseconds | 50 to 500<br>microseconds | 50 to 500<br>microseconds | 50 to 500<br>microseconds | 50 to 1000<br>microseconds | 50 to 500<br>microseconds |
| Current/Voltage<br>Regulated | Current | Current | Current | Voltage | Voltage | Current |
| Output Voltage<br>(300 Ω) | 0 to 4.1 V | 0 to 4.1 V | 0 to 6.3 V | 0 to 7 V | 0 to 5.4 V | 0 to 5.7 V |
| Output Voltage<br>(500 Ω) | 0 to 6.4 V | 0 to 6.4 V | 0 to 7.2 V | 0 to 10.8 V | 0 to 7.1 V | 0 to 7.6 V |
| Output Voltage<br>(800 Ω) | 0 to 7.5 V | 0 to 7.5 V | 0 to 8.0 V | 0 to 11.6 V | 0 to 8.4 V | 0 to 9.6 V |
| Output Current<br>(300 Ω) | 0 to 13.5 mA | 0 to 13.5 mA | 0 to 21 mA | 0 to 23.3 mA | 0 to 18.0 mA | 0 to 19.0 mA |
| Comparator | StimQ PNS<br>System | Stimwave<br>Freedom<br>SCS System<br>(K150517) | Stimwave<br>Freedom<br>SCS System<br>(K141399) | Medtronic<br>Mattrix<br>3271/3272<br>(K934065) | Medtronic<br>Xtrel, Model<br>Number<br>3425<br>(K883780) | ANS Renew<br>(K000852) |
| Output Current<br>(500 Ω) | 0 to 12.8 mA | 0 to 12.8 mA | 0 to 15 mA | 0 to 21.6 mA | 0 to 14.2 mA | 0 to 15.2 mA |
| Output Current<br>(800 Ω) | 0 to 9.4 mA | 0 to 9.4 mA | 0 to 10 mA | 0 to 14.5 mA | 0 to 10.5 mA | 0 to 12.0 mA |
| Waveform | Charge<br>Balanced<br>(delayed)<br>Biphasic<br>asymmetrical | Charge<br>Balanced<br>(delayed)<br>Biphasic<br>asymmetrical | Charge<br>Balanced<br>(delayed)<br>Biphasic<br>asymmetrical | Charge<br>Balanced<br>Biphasic<br>asymmetrical | Charge<br>Balanced<br>Biphasic<br>asymmetrical | Charge<br>Balanced<br>(delayed)<br>Biphasic<br>asymmetrical |
| Pulse Shape | Decaying<br>Exponential | Decaying<br>Exponential | Decaying<br>Exponential | Decaying<br>Exponential | Decaying<br>Exponential | Decaying<br>Exponential |
| Average Current<br>Density (300 Ω) | 105.0 mA/cm² | 105.0 mA/cm² | 111.6 mA/cm² | 175.0 mA/cm² | 125.8 mA/cm² | 117.7 mA/cm² |
| Average Current<br>Density (500 Ω) | 95.1 mA/cm² | 95.1 mA/cm² | 96.7 mA/cm² | 151.7 mA/cm² | 101.7 mA/cm² | 103.1 mA/cm² |
| Average Current<br>Density (800 Ω) | 69.0 mA/cm² | 69.0 mA/cm² | 77.0 mA/cm² | 106.7 mA/cm² | 75.8 mA/cm² | 86.2 mA/cm² |
| Maximum Phase<br>Charge* (300 Ω) | 6.8 µC/pulse | 6.8 µC/pulse | 10.5 µC/pulse | 11.7 µC/pulse | 18.0 µC/pulse | 9.5 µC/pulse |
| Maximum Phase<br>Charge* (500 Ω) | 6.4 µC/pulse | 6.4 µC/pulse | 7.2 µC/pulse | 10.8 µC/pulse | 14.2 µC/pulse | 7.6 µC/pulse |
| Maximum Phase<br>Charge* (800 Ω) | 4.7 µC/pulse | 4.7 µC/pulse | 5.0 µC/pulse | 7.3 µC/pulse | 10.5 µC/pulse | 6.0 µC/pulse |
| Maximum Charge<br>Density* (300 Ω) | 53.1 µC/cm² | 53.1 µC/cm² | 82.5 µC/cm² | 97.2 µC/cm² | 150.0 µC/cm² | 73.1 µC/cm² |
| Maximum Charge<br>Density* (500 Ω) | 50.3 µC/cm² | 50.3 µC/cm² | 56.6 µC/cm² | 90.0 µC/cm² | 118.3 µC/cm² | 58.5 µC/cm² |
| Maximum Charge<br>Density* (800 Ω) | 36.9 µC/cm² | 36.9 µC/cm² | 39.3 µC/cm² | 60.4 µC/cm² | 87.5 µC/cm² | 46.2 µC/cm² |
| Maximum<br>Current Density*<br>(300 Ω) | 106.1 mA/cm² | 106.1 mA/cm² | 165.1 mA/cm² | 194.4 mA/cm² | 150.0 mA/cm² | 146.2 mA/cm² |
| Maximum<br>Current Density*<br>(500 Ω) | 100.6 mA/cm² | 100.6 mA/cm² | 113.2 mA/cm² | 180.0 mA/cm² | 118.3 mA/cm² | 116.9 mA/cm² |
| Maximum<br>Current Density*<br>(800 Ω) | 73.9 mA/cm² | 73.9 mA/cm² | 78.6 mA/cm² | 120.8 mA/cm² | 87.5 mA/cm² | 92.3 mA/cm² |
| Net Charge | 0 µC | 0 µC | 0 µC | 0 µC | 0 µC | 0 µC |
| Average Phase<br>Power (300 Ω) | 0.053 W/phase | 0.053 W/phase | 0.060 W/phase | 0.132 W/phase | 0.068 W/phase | 0.070 W/phase |
| Average Phase<br>Power (500 Ω) | 0.073 W/phase | 0.073 W/phase | 0.076 W/phase | 0.166 W/phase | 0.074 W/phase | 0.090 W/phase |
| Average Phase<br>Power (800 Ω) | 0.062 W/phase | 0.062 W/phase | 0.060 W/phase | 0.131 W/phase | 0.066 W/phase | 0.100 W/phase |
| Average Phase<br>Power Density<br>(300 Ω) | 0.42 W/cm²/phase | 0.42 W/cm²/phase | 0.48 W/cm²/phase | 1.10 W/cm²/phase | 0.57 W/cm²/phase | 0.54 W/cm²/phase |
| Average Phase<br>Power Density<br>(500 Ω) | 0.58 W/cm²/phase | 0.58 W/cm²/phase | 0.59 W/cm²/phase | 1.38 W/cm²/phase | 0.62 W/cm²/phase | 0.69 W/cm²/phase |
| Average Phase<br>Power Density<br>(800 Ω) | 0.48 W/cm²/phase | 0.48 W/cm²/phase | 0.60 W/cm²/phase | 1.09 W/cm²/phase | 0.55 W/cm²/phase | 0.77 W/cm²/phase |
| Comparator | StimQ PNS<br>System | Stimwave<br>Freedom<br>SCS System<br>(K150517) | Stimwave<br>Freedom<br>SCS System<br>(K141399) | Medtronic<br>Mattrix<br>3271/3272<br>(K934065) | Medtronic<br>Xtrel, Model<br>Number<br>3425<br>(K883780) | ANS Renew<br>(K000852) |
| Pulse Delivery<br>Mode | Continuous | Continuous | Continuous | Continuous | Continuous | Continuous |
| ON/OFF Times | No Cycling | No Cycling | No Cycling | ON/OFF<br>Cycling<br>Option | ON/OFF<br>Cycling<br>Option | No Cycling |
| Current Path<br>Options | Bipolar | Bipolar | Bipolar | Bipolar | Bipolar | Bipolar |
| Power Delivery | Coupled<br>receiver built<br>into Stimulator<br>body | Same as<br>StimQ | Same as<br>StimQ | Coupled<br>receiver,<br>hardwired<br>connector | Coupled<br>receiver,<br>hardwired<br>connector | Coupled<br>receiver,<br>hardwired<br>connector |
| Transmit<br>Frequency | 915 MHz | 915 MHz | 915 MHz | 2 MHz | 1.60 MHz | 2 MHz |
| Material | Platinum-<br>iridium<br>90:10,<br>Polyurethane<br>2363-55D | Same as<br>StimQ | Same as<br>StimQ | Same as<br>StimQ | Same as<br>StimQ | Same as<br>StimQ |
| Sterile | Yes -<br>ethylene<br>oxide | Same as<br>StimQ | Same as<br>StimQ | Same as<br>StimQ | Same as<br>StimQ | Same as<br>StimQ |
| Single-Use | Yes | Yes | Yes | Yes | Yes | Yes |
| Shelf Life | 2 year | 2 year | 1 year | 1 year | 1 year | 2 years |
| Complies with<br>ISO 10993-1 | Yes | Yes | Yes | Yes | Yes | Yes |
| Safety Testing<br>Passed | Yes | Yes | Yes | Yes | Yes | Yes |
Table 5A. Comparison of Characteristics
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo for "stimQ". The word "stim" is written in a dark blue sans-serif font. The "Q" is green and stylized to look like a magnifying glass. Behind the word is a series of curved lines that resemble radio waves, also in dark blue.
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the logo for "stimQ". The word "stim" is in blue, and the "Q" is in green. To the right of the word is a graphic of concentric blue circles, which appear to be emanating from the "Q".
{9}------------------------------------------------
Image /page/9/Picture/0 description: The image shows the logo for StimQ. The logo features the word "stim" in dark blue, followed by a green letter "Q" that is shaped like a magnifying glass. Behind the word and the "Q" are three curved lines in dark blue, resembling sound waves or radio waves emanating from the "Q".
(*) asterisk denotes that formulas were used for the calculations.
### 8. Biocompatibility Data
The StimO PNS System uses the same stimulator as the Stimwave Freedom SCS System. Thus, the materials of the StimQ PNS System remain unchanged from the Freedom SCS System (K150517 and K141399). The biocompatibility tests conducted on representative subassemblies of the Freedom SCS System (Freedom-4, K141399) directly apply to the StimQ PNS System. The materials, construction and intended use of the StimQ PNS System is comparable to the predicate device, and have a long history of safety with respect to biocompatibility. The biological safety of the Stimulator was evaluated in accordance to ISO 10993-1:2009 and guidance document Blue Book Memorandum G95-1 Use of International Standard ISO-10993, and Biological Evaluation of Medical Devices: Part 1: Evaluation and Testing. Under these, for the stated indications for use, the device was classified as a (C), implant device in contact with tissue/bone. The results for the biocompatible testing for cytotoxicity, sensitization, irritation or intracutaneous reactivity, acute systemic toxicity, genotoxicity, implantation (4, 8, and 13 weeks), and subchronic toxicity demonstrated no negative impacts from the materials that are used in the StimQ PNS System. The Freedom-8A/4A Stimulator materials in direct tissue contact include Pellethane 55D and Pt-Ir (90:10), both having an extensive record (previously cleared and approved) of chronic and carcinogenetic safety. The SWAG is intended to be
{10}------------------------------------------------
Image /page/10/Picture/0 description: The image shows the logo for StimQ. The logo features the word "stim" in blue, followed by a green letter "Q" that is stylized to resemble a magnifying glass. Behind the "Q" are three curved blue lines that emanate outward, suggesting sound waves or a signal. The overall design is clean and modern, using a combination of text and graphic elements to represent the brand.
on top of an article of clothing. The User Manual provided to the patient describes that the SWAG should always be worn on top of a layer of clothing. The SWAG does not come into contact with the patient's skin. The categorization by nature of body contact of the SWAG is thus "non-contacting device", and not included in the scope of ISO 10993-1:2009. The StimQ PNS System (similar in form to the Freedom SCS System, K141399 and K150517) meets biological safety and compatibility requirements of ISO 10993-1:2009 and Blue Book Memorandum G95-1.
# 9. Non-Clinical Performance Data
The StimQ PNS System was tested to verify that the performance meets the system design requirements as well as all applicable voluntary standards. The StimQ PNS System complies with all design requirements and applicable voluntary standards.
AAMI ANSI ISO 14708-3:2008 - For protection from temperature change including shipping and storage temperature ranges, the StimQ Stimulator was functional, receiving a safe rating following post visual inspection and passed the change of temperature testing performed as specified by AAMI ANSI ISO 14708-3:2008. For atmospheric pressure change, the StimO Stimulator was functional following post testing functionality inspection and passed atmospheric pressure change testing as specified by AAMI ANSI ISO 14708-3:2008.
For testing external defibrillation exposure, the StimQ Stimulator was verified as functional after exposure to external defibrillation. Thus, the StimQ Stimulator complies with testing as specified by AAMI ANSI ISO 14708-3:2008.
Following the thermal shock testing, the StimQ Stimulator was found to have "no irreversible damage" and fully functional as specified by the manufacturer, and to have no physical anomalies present at the time of inspection. Thus, the StimQ Stimulator complies with the thermal shock design requirements and the applicable standard.
For leakage current testing, the StimQ Stimulator produced zero leakage current on all tested paths for all tested samples. Thus, the StimQ Stimulator complies with the leakage design requirements and the applicable standard.
For testing the insertion and withdrawal of the stimulator within the stylet, the StimQ Stimulator was found to require less than 2.5N of insertion or withdrawal force for all tested stylets in all tested stimulator samples. Visual inspection confirmed no damage was present in any stimulator samples. Thus, the StimO Stimulator complies with design specifications for stylet insertion and withdrawal force.
For mechanical testing, the StimQ Stimulator passed all criteria of the test, showing no visible damage to the stimulator body or functional damage to the components. Mechanical testing included tensile testing, flex testing and torsion testing. Thus, the StimO Stimulator complies with all stimulator mechanical design requirements.
{11}------------------------------------------------
Image /page/11/Picture/0 description: The image shows the logo for StimQ. The logo features the word "stim" in blue, followed by a green letter "Q" that is shaped like a magnifying glass. Behind the word and letter is a series of blue concentric arcs, resembling sound waves or radio waves. The logo is simple and modern, using a combination of text and graphic elements to represent the company.
IEC 60601-1 - For testing the external unit for protection from temperature change, including shipping and storage temperature ranges, the external unit met the passing criteria of both of the visual and functional inspections following the testing. It showed no physical damage and was fully operational. Thus, the external unit of the StimO PNS System satisfies the outlined protection from temperature change design requirements and the applicable standard, IEC 60601-1. For atmospheric pressure change testing, the external unit met the passing criteria of both of the visual and functional inspections following the testing. It showed no physical damage and was fully operational. Thus, the external unit of the StimO PNS System satisfies the outlined atmospheric pressure change design requirements and the applicable standard, IEC 60601-1. For the push, drop, and impact testing of the external unit, it was determined through testing that the external unit is robust to withstand expected damage in accordance with general safety standards. The external unit met the passing criteria of both of the visual and functional inspections following the testing. It showed no physical damage and was fully operational. Thus, the SWAG component of the StimQ PNS System satisfies the outlined push, drop, and impact, design requirements and the applicable standard, IEC 60601-1. For the identification, marking and documents of the external unit it was determined through an analysis of the labeling that the external unit complies with the requirements of the standard. All requirements and markings are clearly identified and viewable either from the external case of the product or from within the accompanying documents. For the means of protection, creepage distances, and air clearances of the external unit it was determined through an analysis of the design that the system satisfies the requirements of the applicable standard, IEC 60601-1.
IEC 60529 - For testing the ingress of water, the external unit met the passing criteria of both of the visual and functional inspections f…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.