The Freedom Spinal Cord Stimulator (SCS) System is intended as the sole mitigating agent, or as an adjunct to other modes of therapy used in a multidisciplinary approach for chronic, intractable pain of the trunk and/or lower limbs, including unilateral or bilateral pain. The Freedom-8A Trial Lead Kit is only used in conjunction with the Freedom-8A Stimulator Receiver Kit, and the Freedom-4A Trial Lead Kit is used for either the Receiver Kit Freedom-4A or the Receiver Kit Freedom-8A Stimulator. The trial devices are solely used for trial stimulation (no longer than 30 days) to determine efficacy before recommendation for a permanent (long term) device.
Device Story
Freedom SCS System provides therapeutic relief for chronic, intractable pain via pulsed electrical current acting on nerves near the spinal column. System components: implantable stimulator (4 or 8 electrodes), receiver, and externally worn Wearable Antenna Assembly (WAA). WAA transmits microwave energy (915 MHz) to the implanted receiver to power the device. Used in hospital or home settings; implanted by orthopedic surgeons, neurosurgeons, or anesthesiologists. WAA worn over clothing; user controls stimulation parameters (repetition rate 5 Hz–10 kHz; pulse width 30–1000 µs) via mobile application. Output affects nerve signaling to modulate pain perception. Benefits include non-invasive power delivery and adjustable stimulation parameters to manage chronic pain.
Clinical Evidence
Multicenter, prospective, randomized controlled study compared 10,000 Hz stimulation to 50-1500 Hz stimulation in patients with chronic back/leg pain (FBSS). Primary endpoint: % of subjects achieving ≥50% pain relief (VAS) at 6 months. Results demonstrated non-inferiority (p=0.0082). Secondary endpoints (VAS back/leg pain, ODI, PGIC) also met non-inferiority. Adverse events included migration (15%), incision pain (7%), and infection (1%).
Technological Characteristics
Implantable stimulator: Pellethane 55D casing, Pt-Ir (90:10) electrodes. WAA: PCB-based microwave field stimulator, silicon rubber switch pad, Li-ion battery. Connectivity: Bluetooth for mobile app control. Power: 915 MHz RF wireless transmission. Sterilization: Ethylene Oxide (EO). Standards: AAMI ANSI ISO 14708-3, IEC 60601-1, IEC 60529, IEC 60601-1-2, ISO 10993-1.
Indications for Use
Indicated for patients with chronic, intractable pain of the trunk and/or lower limbs, including unilateral or bilateral pain, as a sole mitigating agent or adjunct to other therapies. Trial devices are indicated for temporary use (up to 30 days) to assess efficacy prior to permanent implantation.
Regulatory Classification
Identification
An implanted spinal cord stimulator for pain relief is a device that is used to stimulate electrically a patient's spinal cord to relieve severe intractable pain. The stimulator consists of an implanted receiver with electrodes that are placed on the patient's spinal cord and an external transmitter for transmitting the stimulating pulses across the patient's skin to the implanted receiver.
Predicate Devices
Stimwave Freedom SCS System FR8A/FR4A, PDBT-915-2A (K180981)
SandShark Injectable Anchor (SIA) System (K172644)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The FDA logo is composed of two distinct elements: the Department of Health and Human Services logo on the left and the FDA acronym along with the full name of the agency on the right. The Department of Health and Human Services logo features a stylized representation of a human figure, while the FDA portion includes the acronym "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
March 29, 2019
Stimwave Technologies Incorporated Elizabeth Greene Chief Compliance Officer Stimwave Technologies Inc. 1310 Park Central Boulevard South Pompano Beach, Florida 33064
## Re: K182720
Trade/Device Name: Freedom Spinal Cord Stimulator (SCS) System Regulation Number: 21 CFR 882.5880 Regulation Name: Implanted Spinal Cord Stimulator For Pain Relief Regulatory Class: Class II Product Code: GZB Dated: February 28, 2019 Received: March 1, 2019
## Dear Elizabeth Greene:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrl/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
# Pamela D. Scott -S
for Carlos L. Peña, PhD, MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K182720
Device Name Freedom Spinal Cord Stimulator (SCS) System
Indications for Use (Describe)
The Freedom Spinal Cord Stimulator (SCS) System is intended as the sole mitigating agent, or as an adjunct to other modes of therapy used in a multidisciplinary approach for chronic, intractable pain of the trunk and/or lower limbs, including unilateral or bilateral pain.
The Freedom-8A Trial Lead Kit is only used in conjunction with the Freedom-8A Stimulator Receiver Kit, and the Freedom-4A Trial Lead Kit is used for either the Receiver Kit Freedom-4A or the Receiver Kit Freedom-8A Stimulator. The trial devices are solely used for trial stimulation (no longer than 30 days) to determine efficacy before recommendation for a permanent (long term) device.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for Stimwave Technologies. The word "stim" is in blue, "wave" is in green, and "TECHNOLOGIES" is in blue. To the right of the words is a blue graphic of concentric arcs, resembling radio waves.
# 510(k) Summary
## For Freedom Spinal Cord Stimulator (SCS) System
#### K182720
#### 1. Submission Sponsor
Stimwave Technologies Incorporated 1310 Park Central Boulevard Pompano Beach Florida 33064 USA Phone: 800.965.5134 Fax: 800.965.5134 Contact: Elizabeth Greene, Chief Compliance Officer
#### 2. Date Prepared
March 27, 2019
#### 3. Device Identification
| Trade/Proprietary Name: | Freedom Spinal Cord Stimulator (SCS) System |
|----------------------------|--------------------------------------------------|
| Common/Usual Name: | Spinal Cord Stimulator |
| Classification Name: | Stimulator, Spinal-Cord, Implanted (Pain Relief) |
| Classification Regulation: | 882.5880 |
| Product Code: | GZB |
| Device Class: | Class II |
| Classification Panel: | Neurology |
#### 4. Legally Marketed Predicate Device(s)
Stimwave Freedom SCS System FR8A/FR4A, PDBT-915-2A (K180981)
#### 5. Device Description
This premarket notification submission is identical to K180981, but includes updates to widen the available range of stimulation parameters to:
- A. Repetition Rate: 5 Hz to 10 kHz;
- B. Pulse Width: 30 us to 1000 us.
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for Stimwave Technologies. The word "stim" is in blue, and the word "wave" is in green. Below the words "stim wave" is the word "TECHNOLOGIES" in blue. To the right of the words is a blue graphic of concentric circles.
No design modifications to the Freedom SCS System of K180981 were made in support of this premarket notification submission. Updates to the labeled product specifications do not change the intended use, indications for use, or adversely affect device safety or effectiveness as compared to the legally marketed predicate device.
The Stimwave Technologies Incorporated (Stimwave) Freedom Spinal Cord SCS System (System) is used for spinal column stimulation to provide therapeutic relief for chronic. intractable pain of the trunk and/or lower limbs including unilateral or bilateral pain. The therapy utilizes pulsed electrical current to create an energy field that acts on nerves near the spinal column. The System is comprised of an implantable stimulator (Freedom-8A/4A Stimulator), receiver component (Receiver), and an externally worn transmitter (Wearable Antenna Assembly (WAA)) to power the device. The System is implanted only following a successful trial period with the Freedom-8A/4A Trial Lead.
| | Freedom-8A and Freedom-4A Stimulator (Receiver Kit) |
|-------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Freedom-8A<br>Stimulator,<br>Freedom-4A<br>Stimulator | A polyurethane (Pellethane 55D) casing with an embedded receiver, flexible<br>circuit board and electrodes (Platinum Iridium 90:10) that is placed in the<br>patient's epidural space. The Freedom-8A Stimulator has eight (8) electrodes,<br>and the Freedom-4A Stimulator has four (4) electrodes. Matches device<br>specifications of K180981 with the exception of the widened range of<br>stimulation parameters for repetition rate (5 Hz to 10 kHz) and pulse width (30<br>$μ$ s to 1000 $μ$ s). |
| Receiver | A copper and PEEK cable with dual couplers; placed within the center lumen of<br>the Freedom-8A or Freedom-4A Stimulator with the distal end combination of<br>Receiver and Stimulator being placed under the skin. Two (2) Receivers are<br>provided with each kit. Identical to K180981. |
| Stylet(s) | A stainless steel wire with a polypropylene handle that is inserted into the open<br>central lumen of the stimulator to provide rigidity during implantation. Two (2)<br>stylets are provided in the Receiver Kit, one straight and one bent, each with<br>diameter of 0.30 mm. Identical to K180981. |
| Needle | A 13-gauge stainless steel needle that acts as a conduit for passage of the<br>Stimulator into the epidural space. Identical to K180981. |
| Guidewire | A stainless steel, rigid, solid core guidewire used to create a hollow pathway in<br>the epidural space for the Stimulator to pass through easily. Identical to<br>K180981. |
#### Freedom-8A and Freedom-4A Stimulator (Receiver Kit)
#### Wearable Antenna Assembly (WAA Kit)
- WAA The WAA housing includes the following components: A. Microwave Field Stimulator (MFS) - A printed circuit board (PCB) that generates RF power with embedded waveform parameter settings and switches for changing parameter settings as needed by the user. Identical to K180981; B. Switch Membrane - An elastomeric silicon rubber pad that corresponds to switches on the MFS that allows the user to turn the device on/off or
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for Stimwave Technologies. The word "stim" is in blue, and the word "wave" is in green. The word "TECHNOLOGIES" is in blue and is located in a blue oval. To the right of the word "wave" are three curved lines that look like sound waves.
#### Wearable Antenna Assembly (WAA Kit)
| Charger Kit | |
|-----------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | increase or decrease power amplitude as well as interpret device power |
| | status (On, Off, Charging, Transmitting, and Bluetooth® Connection). |
| | Identical to K180981; |
| C. | Battery Assembly – A battery and wire assembly for charging and the MFS<br>for power delivery. Identical to K180981. |
| | Transmitting (Tx) Antenna Assembly – An antenna and coaxial cable assembly<br>that is attached to the WAA that is used to transmit microwave energy to the<br>implanted Stimulator. Identical to K180981. |
| Battery Charger | An off-the-shelf battery charger that uses a power adapter and USB to micro-<br>USB cable to recharge the encased lithium ion battery of the WAA. Identical to<br>K180981. |
## Freedom-8A and Freedom-4A Spare Lead (Spare Lead Kit)
| Freedom-8A<br>Spare Lead,<br>Freedom-4A<br>Spare Lead | A polyurethane (Pellethane 55D) casing with an embedded receiver, flexible<br>circuit board and electrodes (Platinum Iridium 90:10) that is placed in the<br>patient's epidural space. The Freedom-8A Spare Lead has eight (8) electrodes,<br>and the Freedom-4A Spare Lead has four (4) electrodes. Matches device<br>specifications of K180981 with the exception of the widened range of<br>stimulation parameters for repetition rate (5 Hz to 10 kHz) and pulse width (30<br>$\u03bcs$ s to 1000 $\u03bcs$ s). |
|-------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| RF Stylet | A copper and PEEK cable with dual couplers; placed within the center lumen of<br>the Freedom-8A or Freedom-4A Spare Lead with the distal end combination of<br>RF Stylet and Spare Lead being placed under the skin. Two (2) RF Stylets are<br>provided with each kit. Identical to K180981. |
| Stylet(s) | A stainless steel wire with a polypropylene handle that is inserted into the open<br>central lumen of the stimulator to provide rigidity during implantation. Two (2)<br>stylets are provided in the Spare Lead Kit, one straight and one bent, each with<br>diameter of 0.30 mm. Identical to K180981. |
| Needle | A 13-gauge stainless steel needle that acts as a conduit for passage of the Spare<br>Lead into the epidural space. Identical to K180981. |
| Guidewire | A stainless steel, rigid, solid core guidewire used to create a hollow pathway in<br>the epidural space for the Spare Lead to pass through easily. Identical to<br>K180981. |
## Freedom-8A and Freedom-4A Trial Lead (Trial Lead Kit)
| Freedom-8A<br>Trial Lead,<br>Freedom-4A<br>Trial Lead | A polyurethane (Pellethane 55D) casing with an embedded receiver, flexible<br>circuit board and electrodes (Platinum Iridium 90:10) that is placed<br>percutaneously in the patient's epidural space. The Freedom-8A Trial Lead has<br>eight (8) electrodes, and the Freedom-4A Trial Lead has four (4) electrodes.<br>Matches device specifications of K180981 with the exception of the widened<br>range of stimulation parameters for repetition rate (5 Hz to 10 kHz) and pulse<br>width (30 $\mu$ s to 1000 $\mu$ s). |
|-------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|-------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image contains the logo for Stimwave Technologies. The word "stim" is in dark blue, and the word "wave" is in green. The word "TECHNOLOGIES" is in black and is located in a black rounded rectangle. There are blue curved lines to the right of the word "wave".
## Freedom-8A and Freedom-4A Trial Lead (Trial Lead Kit)
| RF Stylet | A copper and PEEK cable with dual couplers; placed within the center lumen of<br>the Freedom-8A or Freedom-4A Trial Lead with the distal end combination of<br>RF Stylet and Trial Lead being placed under the skin. Two (2) RF Stylets are<br>provided with each kit. Identical to K180981. |
|-----------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Stylet(s) | A stainless steel wire with a polypropylene handle that is inserted into the open<br>central lumen of the stimulator to provide rigidity during implantation. Two (2)<br>stylets are provided in the Trial Lead Kit, one straight and one bent with a<br>diameter of 0.30 mm. Identical to K180981. |
| Needle | A 13-gauge stainless steel needle that acts as a conduit for passage of the Trial<br>Lead into the epidural space. Identical to K180981. |
| Guidewire | A stainless steel, rigid, solid core guidewire used to create a hollow pathway in<br>the epidural space for the Trial Lead to pass through more easily. Identical to<br>K180981. |
## Sterile Revision Kit
| Stylet(s) | A stainless steel wire with a polypropylene handle that is inserted into the open<br>central lumen of the stimulator to provide rigidity during implantation. Two (2)<br>stylets are provided in the Sterile Revision Kit, one straight and one bent with a<br>diameter of 0.30 mm. Identical to K180981. |
|-----------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Needle | A 13-gauge stainless steel needle that acts as a conduit for passage of the<br>Stimulator/Lead into the epidural space. Identical to K180981. |
| Guidewire | A stainless steel, rigid, solid core guidewire used to create a hollow pathway in<br>the epidural space for the Stimulator/Lead to pass through easily. Identical to<br>K180981. |
## Sterile Single Pack Accessories
| RF Stylet | A copper and PEEK cable with dual couplers; placed within the center lumen of the Freedom-8A or Freedom-4A Stimulator, Spare Lead, or Trial Lead with the distal end combination of RF Stylet and Stimulator, Spare Lead, or Trial Lead being placed under the skin. One (1) RF Stylet is provided in the single pack. Identical to K180981. |
|---------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Needle (4.5") | A 13-gauge stainless steel needle that acts as a conduit for passage of the Stimulator/Lead into the epidural space. Identical to K810981. |
| Needle (6") | A 13-gauge stainless steel needle that acts as a conduit for passage of the Stimulator/Lead into the epidural space. Identical to K180981. |
| Guidewires | A stainless steel, rigid, solid core guidewire used to create a hollow pathway in the epidural space for the Stimulator/Lead to pass through easily. Three (3) guidewires are provided, each with different diameters (1.05mm, 1.30mm, and 1.60mm), but all the same length (45cm). Identical to K180981. |
| Introducer | A stainless steel stylet and polymer sheath that dilates the path over the guidewire for advancing the Stimulator/Lead into place. Identical to K180981. |
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image contains the logo for Stimwave Technologies. The word "stim" is in blue, "wave" is in green, and "TECHNOLOGIES" is in black. To the right of the words is a blue graphic of concentric circles.
| Non-Sterile Single Pack Accessories | |
|------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Small SeaSilk<br>Antenna (12"<br>Cable Length) | An antenna (conductive ink printed onto laminate) bonded to the outside of a<br>fabric shirt and coaxial cable assembly. Attaches to the WAA and used to<br>transmit microwave energy to the implanted Stimulator/Lead. Identical to<br>K180981. |
| Small SeaSilk<br>Antenna (27"<br>Cable Length) | An antenna (conductive ink printed onto laminate) bonded to the outside of a<br>fabric shirt and coaxial cable assembly. Attaches to the WAA and used to<br>transmit microwave energy to the implanted Stimulator/Lead. Identical to<br>K180981. |
| Large SeaSilk<br>Antenna (30"<br>Cable Length) | An antenna (conductive ink printed onto laminate) bonded to the outside of a<br>fabric shirt and coaxial cable assembly. Attaches to the WAA and used to<br>transmit microwave energy to the implanted Stimulator/Lead. Identical to the<br>K180981. |
| Large SeaSilk<br>Antenna (28"<br>Cable Length) | An antenna (conductive ink printed onto laminate) bonded to the outside of a<br>fabric shirt and coaxial cable assembly. Attaches to the WAA and used to<br>transmit microwave energy to the implanted Stimulator/Lead. Identical to the<br>K180981. |
| Small Antenna<br>(12" Cable<br>Length) | A flexible dipole antenna and coaxial cable assembly that is attached to the<br>WAA that is used to transmit microwave energy to the implanted Stimulator.<br>Identical to K180981. |
| Small Antenna<br>(27" Cable<br>Length) | A flexible dipole antenna and coaxial cable assembly that is attached to the<br>WAA that is used to transmit microwave energy to the implanted Stimulator.<br>Identical to K180981. |
| Large Antenna<br>(30" Cable<br>Length) | A flexible dipole array antenna and coaxial cable assembly that is attached to<br>the WAA that is used to transmit microwave energy to the implanted Stimulator.<br>Identical to K180981. |
#### Non-Sterile Single Pack Accessories
#### 6. Indication for Use Statement
The Freedom Spinal Cord Stimulator (SCS) System is intended as the sole mitigating agent, or as an adjunct to other modes of therapy used in a multidisciplinary approach for chronic, intractable pain of the trunk and/or lower limbs, including unilateral or bilateral pain. The Freedom-8A Trial Lead Kit is only used in conjunction with the Freedom-8A Stimulator Receiver Kit, and the Freedom-4A Trial Lead Kit is used for either the Receiver Kit Freedom-4A Stimulator or the Receiver Kit Freedom-8A Stimulator. The trial devices are solely used for trial stimulation (no longer than 30 days) to determine efficacy before recommendation for a permanent (long term) device.
#### 7. Substantial Equivalence Discussion
The following table compares the Stimwave Freedom SCS System to the primary predicate device with respect to intended use, technological characteristics and principles of operation, providing more detailed information regarding the basis for the determination of substantial equivalence.
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the logo for Stimwave Technologies. The word "stim" is in blue, and the word "wave" is in green. Below the words "stim" and "wave" is the word "TECHNOLOGIES" in blue. To the right of the words is a blue circular wave pattern.
| Comparator | Stimwave Freedom SCS System<br>(Predicate K180981) | Stimwave Freedom SCS System<br>(Subject Device K182720) |
|----------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------|
| Product Code | GZB | Same as K180981 |
| Regulation No. | 882.5880 | Same as K180981 |
| Regulation Name | Stimulator, Spinal-Cord,<br>Implanted (Pain Relief) | Same as K180981 |
| Intended Use | Stimulation of spinal cord for chronic,<br>intractable pain of trunk and lower<br>limbs | Same as K180981 |
| Mode of Action | RF wireless transmission of energy<br>to produce stimulation at Stimulator<br>electrodes. WAA sends a pulsed RF<br>signal on a carrier frequency of<br>915MHz to the Stimulator | Same as K180981 |
| Implant Site | Epidural space, L5 to T5 | Same as K180981 |
| Environmental Use | Hospital, Home | Same as K180981 |
| Intended Clinician | Orthopedic, Neurosurgeon,<br>Anesthesiologist | Same as K180981 |
| Intended User | Layperson | Same as K180981 |
| Electrode Material | Platinum-iridium 90:10 | Same as K180981 |
| Stimulator Body Material | Polyurethane 2363-55D | Same as K180981 |
| Cable Features | Multi-lumen Tube | Same as K180981 |
| Stimulator Length | 45 centimeters | Same as K180981 |
| Diameter | 1.35 millimeters | Same as K180981 |
| Electrode Array Length | 24.0 millimeters<br>52.0 millimeters | Same as K180981 |
| No. of Electrodes | 4 or 8 | Same as K180981 |
| Electrode Length | 3.0 millimeters | Same as K180981 |
| Electrode Spacing | 4.0 millimeters | Same as K180981 |
| Electrode Surface Area | 12.72 mm² | Same as K180981 |
| Method of Introduction | Percutaneous and Anchor Incision | Same as K180981 |
| Tissue Contact | Yes | Same as K180981 |
| Sterilization | Ethylene Oxide (EO) | Same as K180981 |
| Labeling | Labeled as Sterile, Single Use,<br>Prescription Device | Same as K180981 |
| Package | Backer card and two sterile<br>pouches | Same as K180981 |
| Pulse Frequency | 5 to 1500 Hertz | 5 to 10,000 Hertz |
| Pulse Width | 50 to 500 microseconds | 30 to 1000 microseconds |
| Current/Voltage Regulated | Current | Same as K180981 |
| Output Voltage (300 Ω) | 0 to 4.1 V | Same as K180981 |
| Output Voltage (500 Ω) | 0 to 6.4 V | Same as K180981 |
| Output Voltage (800 Ω) | 0 to 7.5 V | Same as K180981 |
| Output Current (300 Ω) | 0 to 13.5 mA | Same as K180981 |
| Output Current (500 Ω) | 0 to 12.8 mA | Same as K180981 |
| Output Current (800 Ω) | 0 to 9.4 mA | Same as K180981 |
| Waveform | Charge Balanced (delayed) | Same as K180981 |
| Comparator | Stimwave Freedom SCS System<br>(Predicate K180981) | Stimwave Freedom SCS<br>System<br>(Subject Device K182720) |
| Polarity | Programmable<br>(Anode, Cathode, or Off) | Same as K180981 |
| Pulse Shape | Decaying Exponential | Same as K180981 |
| Avg. Current Density (300 Ω) | 105.0 mA/cm² | Same as K180981 |
| Avg. Current Density (500 Ω) | 95.1 mA/cm² | Same as K180981 |
| Avg. Current Density (800 Ω) | 69.0 mA/cm² | Same as K180981 |
| Max. Phase Charge* (300 Ω) | 6.8 µC/pulse | Same as K180981 |
| Max. Phase Charge* (500 Ω) | 6.4 µC/pulse | Same as K180981 |
| Max. Phase Charge* (800 Ω) | 4.7 µC/pulse | Same as K180981 |
| Max. Charge Density* (300 Ω) | 53.1 µC/cm² | Same as K180981 |
| Max. Charge Density* (500 Ω) | 50.3 µC/cm² | Same as K180981 |
| Max. Charge Density* (800 Ω) | 36.9 µC/cm² | Same as K180981 |
| Max. Current Density* (300 Ω) | 106.1 mA/cm² | Same as K180981 |
| Max. Current Density* (500 Ω) | 100.6 mA/cm² | Same as K180981 |
| Max. Current Density* (800 Ω) | 73.9 mA/cm² | Same as K180981 |
| Net Charge | 0 µC | Same as K180981 |
| Avg. Phase Power (300 Ω) | 0.053 W/phase | Same as K180981 |
| Avg. Phase Power (500 Ω) | 0.073 W/phase | Same as K180981 |
| Avg. Phase Power (800 Ω) | 0.062 W/phase | Same as K180981 |
| Avg. Phase Power Density (300 Ω) | 0.42 W/cm²/phase | Same as K180981 |
| Avg. Phase Power Density (500 Ω) | 0.58 W/cm²/phase | Same as K180981 |
| Avg. Phase Power Density (800 Ω) | 0.48 W/cm²/phase | Same as K180981 |
| Pulse Delivery Mode | Continuous | Same as K180981 |
| ON/OFF Times | No Cycling | Same as K180981 |
| Current Path Options | Bipolar | Same as K180981 |
| Power Delivery | Embedded receiver and coupled<br>receiver in lumen of Stimulator | Same as K180981 |
| Transmit Frequency | 915 MHz | Same as K180981 |
| Material | Platinum-iridium 90:10,<br>Polyurethane 2363-55D | Same as K180981 |
| Sterile | Yes - ethylene oxide | Same as K180981 |
| Contract Sterilizer | Steris Isomedix Services | Same as K180981 |
| Single-Use | Yes | Same as K180981 |
| Shelf Life | 2 year | Same as K180981 |
| Complies with ISO 10993-1 | Yes | Same as K180981 |
| Safety Testing Passed | Yes | Same as K180981 |
| MR Conditional | Yes | Same as K180981 |
| Accessories | Receiver/RF Stylet, Stylet(s),<br>Guidewire, Needle | Same as K180981 |
| Single Pack Accessories | Sterile: RF Stylet, Needle(s),<br>Guidewires, Introducer<br>Non-Sterile: SeaSilk Antenna(s),<br>Antenna Assembly | Same as K180981 |
| Charger | USB Charger | Same as K180981 |
| Wearable Antenna Assembly | Aluminum transmitter & separate,<br>connected Antenna | Same as K180981 |
| Software Level of Concern | Moderate | Same as K180981 |
| iPad Application | WaveCrest™ | Same as K180981 |
| Patient Programmer Application | WaveCrest Mobile | Same as K180981 |
| Comparator | Stimwave Freedom SCS System<br>(Predicate K180981) | Stimwave Freedom SCS<br>System<br>(Subject Device K182720) |
| Kits | Receiver Kit, Spare Lead Kit, Trial<br>Lead Kit, Sterile Revision Kit, WAA<br>Kit, Charger Kit, Single Pack<br>Accessories | Same as K180981 |
#### Table 5A. Comparison of Characteristics
{9}------------------------------------------------
Image /page/9/Picture/0 description: The image contains the logo for Stimwave Technologies. The word "stim" is in blue, and the word "wave" is in green. The word "TECHNOLOGIES" is in blue and is located in a blue rounded rectangle. To the right of the words is a blue graphic that looks like sound waves.
{10}------------------------------------------------
Image /page/10/Picture/0 description: The image contains the logo for Stimwave Technologies. The word "stim" is in blue, and the word "wave" is in green. The word "TECHNOLOGIES" is in blue and underlined. To the right of the word "wave" are blue curved lines that resemble radio waves.
(*) asterisk denotes that formulas were used for the calculations.
## 8. Biocompatibility Data
Materials of this premarket notification submission are identical to K180981. The materials of the Freedom-8A/4A Stimulator in direct contact with tissue remain unchanged from the Freedom SCS System (K170141) and thus, the biocompatibility tests conducted on representative subassemblies of the Freedom SCS System (Freedom-4, K141399) directly apply to the Freedom SCS System (K170141, and K180981). The materials, construction and intended use of the Freedom SCS System is comparable to the predicate devices, and have a long history of safety with respect to biocompatibility. The biological safety of the Freedom-8A/4A Stimulator (same as the Freedom-4 Stimulator) was evaluated in accordance to ISO 10993-1:2009 and guidance document Blue Book Memorandum G95-1 Use of International Standard ISO-10993, and Biological Evaluation of Medical Devices: Part 1: Evaluation and Testing. Under these, for the stated indications for use, the device was classified as a (C), implant device in contact with tissue/bone. The results for the biocompatible testing of the Freedom-8A/4A Stimulator (same as the Freedom-4 Stimulator) for cytotoxicity, sensitization, irritation or intracutaneous reactivity, acute systemic toxicity, genotoxicity, implantation (4, 8, and 13 weeks), and subchronic toxicity demonstrated no negative impacts from the materials that are used in the Freedom SCS System. The Freedom-8A/4A Stimulator materials in direct tissue contact include Pellethane 55D (Stimulator) and Pt-Ir (90:10) (Stimulator only), both having an extensive record (previously cleared and approved) of chronic and carcinogenetic safety. The Receiver/RF Stylet is never in direct or indirect contact with tissue. The WAA is intended to be on top of a thin shirt or article of clothing around the midsection of the patient. The User Manual provided to the patient describes that the WAA should always be worn on top of a layer of clothing. The WAA does not come into contact with the patient's skin. The categorization by nature of body contact of the WAA is thus "non-contacting device", and not included in the scope of ISO 10993-1:2009. The Freedom SCS System meets biological safety and compatibility requirements of ISO 10993-1:2009 and Blue Book Memorandum G95-1.
## 9. Non-Clinical Performance Data
No modifications to the Freedom SCS System of K180981 were made in support of this premarket notification submission. The Freedom SCS System was always designed with the widened stimulation parameters for repetition rate and pulse width, but in predicate
{11}------------------------------------------------
Image /page/11/Picture/0 description: The image shows the logo for Stimwave Technologies. The word "stim" is in blue, and the word "wave" is in green. The word "TECHNOLOGIES" is in blue and is located below the words "stim wave". To the right of the words is a blue graphic of concentric circles.
devices (K180981 and preceding), the software restricted these settings. Performance testing presented in this submission is identical to K180981, and verifies that there is no change to performance, demonstrating substantial equivalence between the subject device and the predicate device (K180981).
The Freedom SCS System was tested to verify that the performance meets the system design requirements as well as all applicable voluntary standards. The Freedom SCS System complies with all design requirements and applicable voluntary standards.
AAMI ANSI ISO 14708-3:2008 - For protection from temperature change including shipping and storage temperature ranges, the Freedom-8A/4A Stimulator was functional, receiving a safe rating following post visual inspection and passed the change of temperature testing performed as specified by AAMI ANSI ISO 14708-3:2008. For atmospheric pressure change, the Freedom-8A/4A Stimulator were functional following post testing functionality inspection and passed atmospheric pressure change testing as specified by AAMI ANSI ISO 14708-3:2008. This testing presented for K180981 is leveraged from K170141, and is directly applicable for demonstration of device safety and efficacy as the packaging and the Freedom-8A/4A Stimulator remains unchanged by electrical component upgrades. The testing for the Freedom-8A/4A Stimulator with the upgraded ASIC demonstrates no change to device performance and specification, and thus, demonstrates continued safety and efficacy compared to the predicates. The Receiver/RF Stylet is identical to K180981 and a passive component that does not impact the outcome of the leveraged tests for safety and effectiveness. The Freedom-8A/4A Stimulator and Receiver/RF Stylet testing presented for this premarket notification submission is leveraged from K180981, and is directly applicable for demonstration of device safety and efficacy as the Freedom-8A/4A Stimulator and Receiver/RF Stylet remains the same the predicate.
For testing external defibrillation exposure, the Freedom-8A/4A Stimulator and Receiver were verified as functional after exposure to external defibrillation. Thus, the Freedom SCS System complies with testing as specified by AAMI ANSI ISO 14708-3:2008. The Freedom-8A/4A Stimulator and Receiver/RF Stylet testing presented for this premarket notification submission is leveraged from K180981, and is directly applicable for demonstration of device safety and efficacy as the Freedom-8A/4A Stimulator design remains unchanged by the widened range of stimulation parameters for repetition rate and pulse width. The Freedom-8A/4A Stimulator stimulation parameter labeling update does not affect the outcome of external defibrillation exposure testing and as such, demonstrates no change to device performance and specification, and continued safety and efficacy compared to the predicate. The Freedom-8A/4A Stimulator and Receiver/RF Stylet testing presented for this premarket notification submission is leveraged from K180981, and is directly applicable for demonstration of device safety and efficacy as the Freedom-8A/4A Stimulator and Receiver/RF Stylet remains the same as the predicate.
{12}------------------------------------------------
Image /page/12/Picture/0 description: The image shows the logo for Stimwave Technologies. The word "stim" is in blue, and the word "wave" is in green. The word "TECHNOLOGIES" is in blue and is located in a blue rounded rectangle. There are blue curved lines to the right of the word "wave".
Following the thermal shock testing, the Freedom-8A/4A Stimulator was found to have "no irreversible damage" and fully functional as specified by the manufacturer, and to have no physical anomalies present at the time of inspection. Thus, the Freedom-8A/4A Stimulator complies with the thermal shock design requirements and the applicable standard. The testing presented for this premarket notification submission is leveraged from K180981, and is directly applicable for demonstration of device safety and efficacy as the Freedom-8A/4A Stimulator design remains unchanged by the widened range of stimulation parameters for repetition rate and pulse width. The Freedom-8A/4A Stimulator stimulation parameter labeling update does not affect the outcome of thermal shock testing and as such, demonstrates no change to device performance and specification, and continued safety and efficacy compared to the predicate. The Receiver/RF Stylet remains the same the predicate and is a passive component that does not impact the outcome of the leveraged tests for safety and effectiveness.
For leakage current testing, the Freedom-8A/4A Stimulator was produced zero leakage current on all tested paths for all tested samples. Thus, the Freedom-8A/4A Stimulator complies with the leakage design requirements and the applicable standard The Freedom-8A/4A Stimulator and Receiver/RF Stylet testing presented for this premarket notification submission is leveraged from K180981, and is directly applicable for demonstration of device safety and efficacy as the Freedom-8A/4A Stimulator design remains unchanged by the widened range of stimulation parameters for repetition rate and pulse width. The Freedom-8A/4A Stimulator stimulation parameter labeling update does not affect the outcome of leakage current testing and as such, demonstrates no change to device performance and specification, and continued safety and efficacy compared to the predicate. The Receiver/RF Stylet remains the same the predicate and is a passive component that does not impact the outcome of the leveraged tests for safety and effectiveness.
For testing the insertion and withdrawal of the stylet within the Stimulator, the stylet was found to require less than 2.5N of insertion or withdrawal force for all tested stylets in all tested stimulator samples. For testing the insertion and withdrawal of the Receiver within the Stimulator, the Receiver/RF Stylet was found to require less than 2.2N of insertion or withdrawal force for all tested stylets in all tested stimulator samples. Visual inspection confirmed no damage was present in any stimulator samples. Thus, the Freedom-8A/4A Stimulator and Receiver/RF Stylet comply with design specifications for stylet insertion and withdrawal force. The Freedom-8A/4A Stimulator and Receiver/RF Stylet testing presented for this premarket notification submission is leveraged from K180981, and is directly applicable for demonstration of device safety and efficacy as the Freedom-8A/4A Stimulator design remains unchanged by the widened range of stimulation parameters for repetition rate and pulse width. The Freedom-8A/4A Stimulation parameter labeling update does not affect the outcome of insertion and withdrawal of the stylet testing
{13}------------------------------------------------
Image /page/13/Picture/0 description: The image shows the logo for Stimwave Technologies. The word "stim" is in blue, and the word "wave" is in green. The word "TECHNOLOGIES" is in blue and is located in a blue rounded rectangle. To the right of the word "wave" are three blue curved lines that resemble radio waves.
and as such, demonstrates no change to device performance and specification, and continued safety and efficacy compared to the predicate. The Receiver/RF Stylet testing is leveraged from K180981, and is directly applicable for demonstration of device safety and effectiveness, as Receiver/RF Stylet remains the same.
For mechanical testing, the Freedom-8A/4A Stimulator passed all criteria of the test, showing no visible damage to the stimulator body or functional damage to the components. Mechanical testing included tensile testing, flex testing and torsion testing. Thus, the Freedom-8A/4A Stimulator complies with all stimulator mechanical design requirements. The Freedom-8A/4A Stimulator and Receiver/RF Stylet testing presented for this premarket notification submission is leveraged from K180981, and is directly applicable for demonstration of device safety and efficacy as the Freedom-8A/4A Stimulator design remains unchanged by the widened range of stimulation parameters for repetition rate and pulse width. The Freedom-8A/4A Stimulator stimulation parameter labeling update does not affect the outcome of mechanical testing and as such, demonstrates no change to device performance and specification, and continued safety and efficacy compared to the predicate. The Receiver/RF Stylet testing is leveraged from K180981, and is directly applicable for demonstration of device safety and effectiveness, as Receiver/RF Stylet remains the same.
For magnetic resonance imaging (MRI) radio frequency (RF) induced heating as related to specific absorbance rate (SAR), the Freedom-8A Stimulator produced a maximum temperature increase lower than the allowable limit for the 1.5T and 3T MRI procedure and thus passed the 1.5T and 3T testing. The Freedom-8A Stimulator produced a maximum temperature increase lower than the allowable limit for the 1.5T and 3T MRI procedure and thus passed both. The Freedom-8A/4A Stimulator and Receiver/RF Stylet testing presented for this premarket notification submission is leveraged from K180981, and is directly applicable for demonstration of device safety and efficacy as the Freedom-8A/4A Stimulator design remains unchanged by the widened range of stimulation parameters for repetition rate and pulse width. The Freedom-8A/4A Stimulation parameter labeling update does not affect the outcome of MRI RF induced heating testing and as such. demonstrates no change to device performance and specification, and continued safety and efficacy compared to the predicate.
ASTM F2182-11a - In accordance with F2182-11a - American Society for Testing and Materials (ASTM) International Standard Test Method for Measurement of Radio Frequency Induced Heating Near Passive Implants During Magnetic Resonance Imaging. the Freedom-8A/4A Stimulator showed that its presence would not cause injury to the patient with the implant during an MRI procedure. The Freedom-8A Stimulator is a passive implant that is not powered while the external unit is not transmitting to it. The Freedom-8A/4A Stimulator and Receiver/RF Stylet testing presented for this premarket notification submission is leveraged from K180981, and is directly applicable for demonstration of device safety and efficacy as the Freedom-8A/4A Stimulator design remains unchanged by
{14}------------------------------------------------
Image /page/14/Picture/0 description: The image shows the logo for Stimwave Technologies. The word "stim" is written in blue, and the word "wave" is written in green. Below the words "stim wave" is the word "TECHNOLOGIES" written in blue inside of a black rounded rectangle. To the right of the words "stim wave" is a blue graphic that looks like a series of concentric arcs.
the widened range of stimulation parameters for repetition rate and pulse width. The Freedom-8A/4A Stimulator stimulation parameter labeling update does not affect the outcome of MRI testing and as such, demonstrates no change to device performance and specification, and continued safety and efficacy compared to the predicate.
ASTM F2119-07 - In accordance with F2119-07 - American Society for Testing and Materials (ASTM) International Standard Test Method for Evaluation of MR Image Artifacts from Passive Implants, the Freedom-8A Stimulator showed that it does not produce image artifacts in 1.5T or 3T MRI procedures. The Freedom-8A/4A Stimulator and Receiver/RF Stylet testing presented for this premarket notification submission is leveraged from K180981, and is directly applicable for demonstration of device safety and efficacy as the Freedom-8A/4A Stimulator design remains unchanged by the widened range of stimulation parameters for repetition rate and pulse width. The Freedom-8A/4A Stimulator stimulation parameter labeling update does not affect the outcome of MRI testing and as such, demonstrates no change to device performance and specification, and continued safety and efficacy compared to the predicate.
ASTM F2052-06 - In accordance with F2052-06 - The American Society for Testing and Materials (ASTM) International Designation Standard Test Method for Measurement of Magnetically Induced Displacement Force on Medical Devices in the Magnetic Resonance Environment, the Freedom-8A Stimulator showed that it does not harm the patient due to its displacement by forces induced by 1.5T or 3T MRI exposure. The Freedom-8A Stimulator passes the ASTM acceptance criteria for deflection angle in a 1.5T and 3T MRI system. The Freedom-8A Stimulator will not present an additional risk or hazard to a patient when used in their tested MR environment. The Freedom-8A/4A Stimulator and Receiver/RF Stylet testing presented for this premarket notification submission is leveraged from K180981, and is directly applicable for demonstration of device safety and efficacy as the Freedom-8A/4A Stimulator design remains unchanged by the widened range of stimulation parameters for repetition rate and pulse width. The Freedom-8A/4A Stimulator stimulation parameter labeling update does not affect the outcome of MRI testing and as such, demonstrates no change to device performance and specification, and continued safety and efficacy compared to the predicate.
ASTM F2213-06 - In accordance with F2213-06 - American Society for Testing and Materials (ASTM) International Standard Test Method for Measurement of Magnetically Induced Torque on Passive Implants in the Magnetic Resonance Environment, the Freedom-8A/4A Stimulator must show that it does not harm the patient due to its torque by forces induced by MRI exposure. The Freedom-8A Stimulator will not present an additional risk or hazard to a patient in the 1.5T MRI or 3T environments or less with regard to torque.
The stimulation waveforms following post-exposure showed no component damage had occurred after MR exposure. During testing, no component damage was observed in any
{15}------------------------------------------------
Image /page/15/Picture/0 description: The image shows the logo for Stimwave Technologies. The word "stim" is in blue, and the word "wave" is in green. The word "TECHNOLOGIES" is in blue and is located in a black rounded rectangle. To the right of the word "wave" are three curved lines that are purple.
waveform. Based on these test results, the Freedom-8A Stimulator will be fully functional following standard 1.5T or 3T MR procedures and its performance or functionality is not affected by such exposures.
The Freedom-8A/4A Stimulator and Receiver/RF Stylet testing presented for this premarket notification submission is leveraged from K180981, and is directly applicable for demonstration of device safety and efficacy as the Freedom-8A/4A Stimulator design remains unchanged by the widened range of stimulation parameters for repetition rate and pulse width. The Freedom-8A/4A Stimulation parameter labeling update does not affect the outcome of MRI testing and as such, demonstrates no change to device performance and specification, and continued safety and efficacy compared to the predicate.
IEC 60601-1 -The WAA was tested for compliance with IEC 60601-1. For testing the WAA for protection from temperature change, including shipping and storage temperature ranges, the WAA met the passing criteria of both of the visual and functional inspections following the testing. It showed no physical damage and was fully operational. Thus, the WAA of the Freedom SCS System satisfies the outlined protection from temperature change design requirements and the applicable standard. IEC 60601-1. For atmospheric pressure change testing, the WAA met the passing criteria of both of the visual and functional inspections following the testing. It showed no physical damage and was fully operational. Thus, the WAA of the Freedom SCS System satisfies the outlined atmospheric pressure change design requirements and the applicable standard, IEC 60601-1. For the push, drop, impact and mold stress relief testing of the WAA, it was determined through testing that the WAA is robust to withstand expected damage in accordance with general safety standards. The WAA met the passing criteria of both of the visual and functional inspections following the testing. It showed no physical damage and was fully operational. Thus, the WAA component of the Freedom-8A/4A SCS System satisfies the outlined push, drop, impact, and mold stress relief design requirements and the applicable standard, IEC 60601-1. For the identification, marking and documents of the WAA it was determined through an analysis of the labeling that the WAA complies with the requirements of the standard. All requirements and markings are clearly identified and viewable either from the external case of the product or from within the accompanying documents. For the means of protection, creepage distances, and air clearances of the WAA it was determined through an analysis of the design that the system satisfies the requirements of the applicable standard, IEC 60601-1. The WAA testing presented for this premarket notification submission is leveraged from K180981, and is directly applicable for demonstration of device safety and efficacy as the WAA design remains unchanged by the widened range of stimulation parameters for repetition rate and pulse width. The Freedom-8A/4A Stimulator stimulation parameter labeling update does not affect the outcome of compliance testing to IEC 60601-1 and as such demonstrates no change to device performance and specification, and continued safety and efficacy compared to the predicate.
{16}------------------------------------------------
Image /page/16/Picture/0 description: The image contains the logo for Stimwave Technologies. The word "stim" is in blue, and the word "wave" is in green. To the right of the word "wave" are concentric purple circles. Below the words "stim wave" is the word "TECHNOLOGIES" in blue.
IEC 60529 - The WAA was tested for compliance with IEC 60529. For testing the ingress of water, the WAA met the passing criteria of both of the visual and functional inspections following the testing. It showed no physical damage and was fully operational. Thus, the WAA component of the Freedom SCS System satisfies the outlined Ingress of Water design requirements and the applicable standard IEC 60529. For particulate matter testing. the WAA met the passing criteria of both of the visual and functional inspections following the testing. It showed no physical damage and was fully operational. Thus, the WAA component of the Freedom SCS System satisfies the outlined Particulate Matter design requirements and the applicable standard, IEC 60529. The WAA testing presented for this premarket notification submission is leveraged from K180981, and is directly applicable for demonstration of device safety and efficacy as the WAA design remains unchanged by the widened range of stimulation parameters for repetition rate and pulse width. The Freedom-8A/4A Stimulation stimulation parameter labeling update does not affect the outcome of compliance testing to IEC 60529 and as such demonstrates no change to device performance and specification, and continued safety and efficacy compared to the predicate.
IEC 60601-1-2 - The WAA was tested for compliance with IEC 60601-1-2. For testing the WAA for electromagnetic compatibility, the unit met all acceptance criteria for emissions, low-frequency magnetic fields, immunity, electrostatic discharge, radiated RF electromagnetic fields, electrical fast transients and magnetic fields. The WAA operated within all test limits and showed no physical damage and was fully operational. Thus, the WAA for the Freedom SCS System satisfies the IEC 60601-1-2 standard. The WAA testing presented for this premarket notification submission is leveraged from K180981, and is directly applicable for demonstration of device safety and efficacy as the WAA design remains unchanged by the widened range of stimulation parameters for repetition rate and pulse width. The Freedom-8A/4A Stimulation parameter labeling update does not affect the outcome of compliance testing to IEC 60601-1-2 and as such demonstrates no change to device performance and specification, and continued safety and efficacy compared to the predicate.
The Freedom SCS System complies with the applicable standards for neurostimulators, electrical safety, electromagnetic interference and compatibility, biocompatibility, packaging, and sterilization. The software of the Freedom SCS System passed all verification tests outlined and the design requirements for Software Verification have been met. The device passed all the testing in accordance with national and international standards.
Following performance testing, it has been determined that the Freedom SCS System is substantially equivalent to legally marketed predicate devices for the therapy for chronic, intractable pain of the trunk and/or lower limbs, including unilateral pain.
{17}------------------------------------------------
Image /page/17/Picture/0 description: The image shows the logo for Stimwave Technologies. The word "stim" is in dark blue, and the word "wave" is in green. Below the words is the word "TECHNOLOGIES" in dark blue inside of a dark blue rounded rectangle. To the right of the words is a series of curved lines that are also in dark blue.
Due to the similarities between the legally marketed predicate device (K180981), and the Freedom SCS System (this premarket notification submission), Stimwave Technologies Incorporated has leveraged applicable performance testing in addition to completed a number of tests that demonstrates substantial equivalence to the legally marketed predicate device. The Freedom SCS System meets all the requirements for overall design, sterilization, biocompatibility, and electrical safety confirms that the output meets the design inputs and specifications. The Freedom SCS System passed all testing stated above as shown by the acceptable results obtained.
## 10. Clinical Performance Data
Stimwave submitted clinical performance data supporting substantial equivalence claims of the subject device to the predicate device. A multicenter, prospective, randomized controlled study demonstrated the efficacy of the Stimwaye Freedom SCS System for the treatment of chronic back, or back and leg pain associated with FBSS using stimulation waveforms with therapeutic frequencies at 10,000 Hz in comparison to stimulation waveforms from 50 Hz…
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.