TOP NEUROPOLE NEEDLE, MODEL SH, BL, SC, TC

K080771 · Top Corp. · GXI · Nov 10, 2008 · Neurology

Device Facts

Record IDK080771
Device NameTOP NEUROPOLE NEEDLE, MODEL SH, BL, SC, TC
ApplicantTop Corp.
Product CodeGXI · Neurology
Decision DateNov 10, 2008
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 882.4725
Device ClassClass 2
AttributesTherapeutic

Intended Use

The Top Neuropole Needles are injection needles which may be used either for percutaneous nerve blocks with local anesthetic solution or for radiofrequency lesioning. A nerve is localized either by using electrostimulation through the needle or by injecting contrast medium through the needle and using radiography concomitantly. The nerve may then be blocked by injecting local anesthetic solution or a radiofrequency lesion may be made.

Device Story

Top Neuropole Needles are sterile, single-use injection needles for percutaneous nerve blocks and radiofrequency (RF) lesioning. Used by clinicians in clinical settings; device facilitates nerve localization through electrical stimulation or contrast medium injection under radiographic guidance. Once localized, clinician performs nerve block via anesthetic injection or creates RF lesion. Device acts as conduit for therapeutic intervention; does not perform automated analysis or processing. Benefits include precise nerve targeting for pain management or therapeutic block.

Clinical Evidence

No clinical data provided; substantial equivalence based on bench testing and comparison of technological characteristics to predicate devices.

Technological Characteristics

Injection needle for RF lesioning; sterile, single-use. Materials, dimensions, and sensing/actuation principles are identical to predicate devices. No software or electronic processing components.

Indications for Use

Indicated for percutaneous nerve blocks via local anesthetic injection or radiofrequency lesioning in patients requiring nerve localization via electrostimulation or contrast-enhanced radiography.

Regulatory Classification

Identification

A radiofrequency lesion probe is a device connected to a radiofrequency (RF) lesion generator to deliver the RF energy to the site within the nervous system where a lesion is desired.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K08077/ Page 1 of 2 g ### 510(k) SUMMARY This summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of 21 CFR Part 807.92(c) Submitter's Name And Address: Top Corporation 19-10 Senju Nakai-cho Adachi-ku Tokyo, Japan 120-0035 NOV 10 2008 Contact Name: Tel. Submission Date: Toshimitsu Suzuki +81 3-3882-3101 December 12, 2007 Device Trade or (Proprietary) Name: Top Neuropole Needles Probe, radiofrequency lesion Pole Needles Common or usual name: Classification Name: Product Code: GXI Legally Marketed Device To Which Claim Substantial Equivalence: Radionics: K870028. K980430. K010202 Top: K062946 Cosman: K060799 Class II devices. (21 C.F.R. § 882.4725) #### DEVICE DESCRIPTION Top Neuropole Needles are injection needles which are used for temperature controlled radiofrequency lesioning. A nerve is localized either by using electrical stimulation through the needle or by injecting contrast medium through the needle and using radiography concomitantly. After injection of anesthetics the nerve may then be blocked by radiofrequency lesion. Top Neuropole Needles are supplied sterile and are labeled for Single Use Only. ### INTENDED USE The Top Neuropole Needles are injection needles which may be used either for percutaneous nerve blocks with local anesthetic solution or for radiofrequency lesioning. A nerve is localized either by using electrostimulation through the needle or by injecting contrast medium through the needle and using radiography concomitantly. The nerve may then be blocked by injecting local anesthetic solution or a radiofrequency lesion may be made. {1}------------------------------------------------ Kv8077/ Page 2 of 2 1 ( # TECHNOLOGICAL CHARACTERISTICS Top Neuropole Needles have the same device characteristics, materials, dimensions and intended use as the predicate device(s). Top Neuropole Needles have been tested to ensure the devices comply with applicable industry standards and US regulations # CONCLUSIONS The intended use and performance characteristics of the Top Neuropole Needles are the same as the predicate(s) and raise no new questions of safety and effectiveness. The Top Neuropole Needles are substantially equivalent to the predicate device. {2}------------------------------------------------ Public Health Service Image /page/2/Picture/2 description: The image shows the seal of the Department of Health & Human Services (HHS) of the United States. The seal features a stylized eagle with its wings spread, clutching a caduceus in its talons. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES. USA" is arranged in a circular pattern around the eagle. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 Top Corporation % Cambria Regulatory Consulting, Inc. Ms. Cathryn N. Cambria 5536 Trowbridge Drive Dunwoody, Georgia 30338 NOV 1 0 2008 Re: K080771 Trade/Device Name: Top Neuropole Needles, Model SH, BL, SC & TC Regulation Number: 21 CFR 882.4725 Regulation Name: Radiofrequency lesion probe Regulatory Class: II Product Code: GXI Dated: October 16, 2008 Received: October 20, 2008 Dear Ms. Cambria: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {3}------------------------------------------------ Page 2 - Ms. Cathryn N. Cambria This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (240) 276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at (240) 276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours, Mark M. Millman - Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ Ko.8077/ ## Indications for Use: 510(k) Number (if known): K080771 Device Name: Top Neuropole Needles, Model SH, BL, SC & TC Indications for Use: The Top Neuropole Needles are injection needles which may be used either for percutaneous nerve blocks with local anesthetic solution or for radiofrequency lesioning. A nerve is localized either by using electrostimulation through the needle or by injecting contrast medium through the needle and using radiography concomitantly. The nerve may then be blocked by injecting local anesthetic solution or a radiofrequency lesion may be made. Prescription Use X AND/OR (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Neil R Ogden for mxm Page 1 of 1 (Division Sign-Off) Division of General, Restorative, and Neurological Devices **510(k) Number** K080771
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from the tree

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...