RF Cannula

K241367 · Shanghai Samedical & Plastic Instruments Co., Ltd. · GXI · Jan 23, 2025 · Neurology

Device Facts

Record IDK241367
Device NameRF Cannula
ApplicantShanghai Samedical & Plastic Instruments Co., Ltd.
Product CodeGXI · Neurology
Decision DateJan 23, 2025
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 882.4725
Device ClassClass 2
AttributesTherapeutic

Intended Use

The RF cannula is intended for use in RF heat lesion procedures for the relief of pain.

Device Story

RF Cannula is a manual, single-use, sterile surgical instrument used in pain management procedures. It consists of a cannula, stylet, hubs, depth stop, and insulating layer. During clinical use, the physician places the cannula's bare tip into target nerve tissue; the stylet is withdrawn and replaced by an RF electrode. The device connects to an external RF generator, which delivers radiofrequency energy through the cannula tip to create thermal lesions for nerve block or pain relief. It is used in clinical settings by physicians. The device facilitates precise energy delivery to target tissues, aiding in therapeutic pain management.

Clinical Evidence

No clinical data. Bench testing only. Performance verified via biocompatibility (ISO 10993 series), sterilization validation (ISO 11135), package stability (ASTM F 1980), and mechanical/electrical testing (ISO 9626, ISO 80369-7, IEC 60601-1, IEC 60601-2-2).

Technological Characteristics

Materials: SUS 304 stainless steel, K-Resin, polyethylene, polyethylene terephthalate. Design: Manual cannula with stylet, depth stop, and insulating layer. Dimensions: 18G-22G gauges; 50-150mm lengths; 2-10mm active tips. Connectivity: Connects to external RF generator. Sterilization: Ethylene Oxide. Standards: ISO 14971, ISO 10993, ISO 11135, ISO 11607, ISO 9626, ISO 80369-7, IEC 60601-1, IEC 60601-2-2, ISO 7864.

Indications for Use

Indicated for use in RF heat lesion procedures for the relief of pain in patients requiring percutaneous nerve blocks or nerve tissue lesioning.

Regulatory Classification

Identification

A radiofrequency lesion probe is a device connected to a radiofrequency (RF) lesion generator to deliver the RF energy to the site within the nervous system where a lesion is desired.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0} FDA U.S. FOOD & DRUG ADMINISTRATION January 23, 2025 Shanghai SAMedical & Plastic Instruments Co., Ltd. Louis Li Level 1-4 & Room 501,506,510, Building No.12, No.2755 Sanlu Highway, Minhang District Shanghai, 201100 China Re: K241367 Trade/Device Name: RF Cannula Regulation Number: 21 CFR 882.4725 Regulation Name: Radiofrequency Lesion Probe Regulatory Class: Class II Product Code: GXI Dated: December 23, 2024 Received: December 23, 2024 Dear Louis Li: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" {1} K241367 - Louis Li Page 2 (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). {2} K241367 - Louis Li Page 3 Sincerely, Adam D. Pierce -S Digitally signed by Adam D. Pierce Date: 2025.01.23 15:24:17 -05'00' Adam D. Pierce, Ph.D. Assistant Director DHT5A: Division of Neurosurgical, Neurointerventional, and Neurodiagnostic Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {3} DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. Submission Number (if known) K241367 Device Name RF Cannula Indications for Use (Describe) The RF cannula is intended for use in RF heat lesion procedures for the relief of pain. Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) # CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {4} K241367 Page 1 of 5 510(k) Summary Submitter: Shanghai SA Medical & Plastic Instruments Co., Ltd. Level 1-4 & Room 501,506,510, Building No.12, No.2755 Sanlu Highway, Minhang District, Shanghai, China Correspondent: Mr. Louis Li MZT Medical Consulting Co., Ltd No. 21 Zhengqiang Road, Xinbei District, Changzhou, Jiangsu, P.R. China. 213000 Tel: +86(0)519-81001811 Email:louisli@mztmedtech.com Device Trade Name: RF Cannula Device Class: Class II Regulation Number: 21 CFR 882.4725 Regulation name: Radiofrequency Lesion Probe Product code: GXI Predicate Device: K112231 Date Prepared: January 23, 2024 Device Description The RF Cannula is indicated for RF heat lesion procedures for the relief of pain. The RF Cannula is composed of Cannula, Cannula hub, Stylet, Stylet hub, Insulating layer, Depth stop and Protection tube. Materials being used include SUS 304 stainless steel, K-Resin, polyethylene, polyethylene terephthalate. RF Cannula is supplied sterile and is only for single-use. It is used in conjunction with a RF Generator and RF Electrode and other accessories to create Radiofrequency (RF) Lesions of nerve tissue or for use in percutaneous nerve blocks. During the RF heat lesion procedure, the RF cannula bare tip is placed into the target tissue; preloaded stylet is withdrawn and replaced by the RF Electrode. The radio frequency is produced by the RF Generator and released by the cannula bare tip. A) Indications for Use: The RF cannula is intended for use in RF heat lesion procedures for the relief of pain. {5} K241367 Page 2 of 5 B) Comparison of Technological Characteristics with the Predicate Device The RF Cannula is substantially equivalent in indication for use, principles of operation and fundamental technological characteristics to the legally marketed predicate LCCS Insulated Spinal Needle (RF Cannula)(K112231). The following table summarizes the similarities and differences in design, materials and dimensions between the subject and predicate device. Table 5-1 Comparison of RF Cannula with predicate device | Feature | Subject Device RF Cannula (K241367) | Predicate Device LCCS Insulated Spinal Needle (RF Cannula) (K112231) | Discussion of Differences | | --- | --- | --- | --- | | Intended/ Indications for use | The RF cannula is intended for use in RF heat lesion procedures for the relief of pain. | LCCS Insulated Spinal Needle (RF Cannula) is indicated for use in RF heat lesion procedures for the relief of pain. | Similar | | Basic Design | The RF Cannula is supplied sterile and is for single-use only. and mainly consists of a Stylet hub, Stylet, Cannula hub, Cannula, Depth Stop, Protection Tube and Insulating layer. | This RF Cannula is provided as a sterile, single use, and consists of a Protection sleeve, Needle tube, Insulating layer, Needle hub, Stylet, Stylet hub. | Similar | | Performance Characteristics | Puncture in RF lesion of peripheral nerve, treatment of pain | In RF heat lesion procedures for the relief of pain | Identical | {6} K241367 Page 3 of 5 | Material type | Stainless Steel | Stainless Steel | Similar | | --- | --- | --- | --- | | Cannula gauge | 18G, 20G, 22G | 18G, 20G, 21G, 22G | Similar | | Cannula length | 50mm | 50mm | Similar | | | 100mm | 100mm | | | | 150mm | 150mm | | | Active tip length | 2mm, 5mm, 10mm | Not available | | | Structure | Straight bevel tip | Straight sharp tip | Identical | | | Curved bevel tip | Curved sharp tip | | | | Curved blunt tip | Curved blunt tip | | | Mechanics of Operation | Manual instrument | Manual instrument | Identical | | Method of Sterilization | Single use, Ethylene | Single use, Ethylene | Identical | | | Oxide sterilization | Oxide sterilization | | The technological characteristics of the subject device are similar to those of predicate device. The subject device has the same basic design as the predicate device. The comparison between the proposed and predicate devices is based on the following: - Indication for use Performance characteristics Sterilization method Basic design C) Performance Testing | Characteristic | Test Method | Result | | --- | --- | --- | | Risk Management | ISO 14971:2019 Medical devices—Application of risk management to medical devices | Pass | | Biocompatibility | ISO 10993-1:2018 Biological evaluation of medical devices — Part 1: Evaluation and testing within a risk management process | Pass | {7} K241367 Page 4 of 5 | | ISO 10993-5:2009 Biological evaluation of medical devices — Part 5: Tests for in vitro cytotoxicity | Pass | | --- | --- | --- | | | ISO 10993-10:2021 Biological evaluation of medical devices — Part 10: Tests for skin sensitization | Pass | | | ISO 10993-11:2017 Biological evaluation of medical devices — Tests for acute systemic toxicity | Pass | | | ISO 10993-11:2017 Biological evaluation of medical devices—Test for material-mediated pyrogenicity | Pass | | | 10993-23:2021 Biological evaluation of medical devices – Part 23: Tests for irritation | Pass | | Ethylene Oxide Sterilization Residuals | ISO 10993-7:2008 Biological evaluation of medical devices — Part 7: Ethylene oxide sterilization residuals | Pass | | Sterilization | ISO 11135:2014 Sterilization of health-care products — Ethylene oxide — Requirements for the development, validation and routine control of a sterilization process for medical devices | Pass | | Sterilization | ISO 11138-1:2017 Sterilization of health care products—Biological indicators—Part 1: General requirements | Pass | | Sterilization | ISO 11737-1:2018 Sterilization of health care products—Microbiological methods—Part 1: Determination of a population of microorganisms on products | Pass | | Package Stability | ASTM F 1980-16 Standard Guide for Accelerated Aging of Sterile Barrier Systems for Medical Devices¹ | Pass | | Packaging | ISO 11607 Packaging for terminally sterilized medical devices | Pass | —Device Shelf-life The proposed devices in their packaging were subjected to accelerated aging to simulate a 5-year shelf life (Treatment: 60±2°C, 194 days, RH70±5%). The aged proposed devices were tested for seal Strength, Dye Penetration, Vacuum Leak and Packaging resistance Bacterial performance. —Performance testing The performance of the subject device was compared to the performance of the predicate device for the following tests: | Characteristic | Test Method | Result | | --- | --- | --- | | Performance | ISO 9626:2016 Stainless steel needle tubing for the manufacture of medical devices—Requirements and test methods | Pass | | | ISO 80369-7:2021 Small-bore connectors for liquids and gases in healthcare applications — Part 7: Connectors for intravascular or hypodermic applications | Pass | {8} K241367 Page 5 of 5 | Electrical safety | IEC 60601-1:2005+A1:2012 Medical electrical equipment—Part 1: General requirements for basic safety and essential performance | Pass | | --- | --- | --- | | | IEC 60601-2-2:2017 Medical electrical equipment—Part 2-2: Particular requirements for the basic safety and essential performance of high frequency surgical equipment and high frequency surgical accessories | Pass | | Surface Finish and Visual Appearance | ISO 7864:2016 Sterile hypodermic needles for single use — Requirements and test methods | Pass | | Bonding Strength Test | ISO 7864:2016 Sterile hypodermic needles for single use - Requirements and test methods | Pass | | Stiffness Test | ISO 9626:2016 Stainless steel needle tubing for the manufacture of medical devices - Requirements and test methods | Pass | | Resistance to Breakage Test | ISO 9626:2016 Stainless steel needle tubing for the manufacture of medical devices - Requirements and test methods | Pass | | Dimension Test | ISO 9626:2016 Stainless steel needle tubing for the manufacture of medical devices - Requirements and test methods | Pass | | Luer Lock Connector Test | ISO 80369-7:2021 Small-bore connectors for liquids and gases in healthcare applications - Part 7: Connectors for intravascular for hypodermic applications | Pass | D) Conclusion Based on the similarities of the indications for use, device design, principles of operation, technological characteristics and the results of the non-clinical performance testing, the subject device, RF Cannula, is substantially equivalent to the legally marketed predicate device LCCS Insulated Spinal Needle (RF Cannula) (K112231).
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...