The Stryker RF Electrodes and Cannulae, in combination with the Stryker RF Generator, are intended for coagulation of soft tissues in orthopedic, arthroscopic, spinal, and neurosurgical applications. They are also used for selective denervation and tissue destruction procedures which may be performed on the lumbar, thoracic, and cervical regions of the spinal cord, peripheral nerves, and nerve roots for the relief of pain. Examples include, but are not limited to, Facette Denervation, Percutaneous Chordotomy/Dorsal Root Entry Zone (DREZ) Lesion, Trigeminus Neuralgia, Peripheral Neuralgia and Rhizotomy.
Device Story
Stryker RF Electrodes and Cannulae function as accessories to Stryker RF Generator; deliver radiofrequency energy to target tissue. Used in orthopedic, arthroscopic, spinal, and neurosurgical procedures. Operated by physicians in clinical settings. Device facilitates coagulation of soft tissue and selective denervation/destruction of nerves for pain management. Output is thermal energy causing tissue lesioning. Benefits include minimally invasive pain relief for conditions like neuralgia and facet joint pain.
Clinical Evidence
No clinical data provided; substantial equivalence based on technological characteristics and intended use compared to existing predicate devices.
Technological Characteristics
Radiofrequency lesion probe and cannula system. Designed for use with RF generator. Class II device (21 CFR 882.4725).
Indications for Use
Indicated for patients requiring soft tissue coagulation or selective nerve denervation/destruction for pain relief in orthopedic, arthroscopic, spinal, and neurosurgical procedures. Applicable to lumbar, thoracic, and cervical spinal regions, peripheral nerves, and nerve roots. Conditions include Facette Denervation, Percutaneous Chordotomy/DREZ Lesion, Trigeminus Neuralgia, Peripheral Neuralgia, and Rhizotomy.
Regulatory Classification
Identification
A radiofrequency lesion probe is a device connected to a radiofrequency (RF) lesion generator to deliver the RF energy to the site within the nervous system where a lesion is desired.
Submission Summary (Full Text)
{0}------------------------------------------------
APR - 1 2004
4100 Easl Milham Avenue Kalamazoo. Mi 49001 Phone (269) 323-7700 (800) 253-3210
## Device Name:
| Trade Name: | Stryker RF Electrodes and Cannulae |
|----------------------|---------------------------------------------------|
| Common Name: | Electrosurgical electrode and cannulae |
| Classification Name: | Probe, Radiofrequency Lesion: 21 CFR 882.4725, GX |
## Device Sponsor:
| Manufacturer: | Stryker Instruments<br>4100 E. Milham Avenue<br>Kalamazoo, MI 49001<br>Registration No.: 1811755 |
|---------------|--------------------------------------------------------------------------------------------------|
|---------------|--------------------------------------------------------------------------------------------------|
Class II Requiatory Class:
## Summary of Safety and Effectiveness:
The Stryker RF Electrodes and Cannulae, in combination with the Stryker RF Generator, are intended for The Stryker RF Electrodes and Calindiae, in confidited with and neurosurgical applications.
coagulation of soft tissues in orthopedic, arthroscopic, spinal, and neurosurgical
They are also used for selective denervation and tissue destruction procedures which may be performed They are also used for selective denervation and insue a manner, and neves, and neves, and neves and neves and neves and neve on the lumbar, thoracle, and cervical regions of the bond portal of entralianeous
relief of pain. Examples include, but are not limited to, Facette Denervation, Percultanera relief of pain. Examples include, but are not limited to, Pacette Benervance, Preripheral Neuralgia, and Rhizotomy.
The Stryker RF Electrodes and Cannulae are equivalent in intended use, safety, and effectiveness to a The Stryker NF EloonSubleted by Stryker Instruments.
The Stryker RF Electrodes and Cannulae do not raise and efficacy concerns when
The Stryker RF and connect of and and annotated - Therefore, the Stryker RF Flectrodes and The Stryker RF Electrodes and Calinual and riotics any now of the Stryker RF Electrodes and compared to similar devious an outly roget of these existing devices.
By: Nicole Ruddy
Nicole Petty Nicole Folly
Associate Manager, Regulatory Affairs
Dated: 1-20-04
{1}------------------------------------------------
Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized graphic of an eagle or bird-like figure with three curved lines representing its wings or body.
APR = 1 2004
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Ms. Nicole Petty, RAC Associate Manager, Regulatory Affairs Stryker Instruments 4100 E. Milham Avenue Kalamazoo, Michigan 49001
Re: K032406
Trade/Device Name: Stryker RF Electrodes and Cannulae Regulation Number: 21 CFR 882.4725, 882.4400 Regulation Name: Radiofrequency lesion probe; Radiofrequency lesion generator Regulatory Class: II Product Code: GXI, GXD Dated: January 20, 2004 Received: January 21, 2004
Dear Ms. Petty:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{2}------------------------------------------------
Page 2 - Ms. Nicole Petty, RAC
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4659. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Miriam C. Provost
Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
Page 1 of
510(k) Number:
KO32406
Device Name:
Indications For Use:
Stryker RF Electrodes and Cannulae
The Stryker RF Electrodes and Cannulae, in combination with the Stryker RF Generator, are intended for coagulation of soft tissues in orthopedic, arthroscopic, spinal, and neurosurgical applications.
They are also used for selective denervation and tissue destruction procedures which may be performed on the lumbar, thoracic, and cervical regions of the spinal cord, peripheral nerves, and nerve roots for the relief of pain. Examples include, but are not limited to, Facette Denervation, Percutaneous Chordotomy/Dorsal Root Entry Zone (DREZ) Lesion, Trigeminus Neuralgia, Peripheral Neuralgia and Rhizotomy.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use
OR
Over-The- Counter Use
(Per 21 CFR 801.109)
Muriam C. Priest (Optional Format 1-2-96)
(Division Sign-Off) Division of General, Restorative, and Neurological Devices
510(k) Number_ Ko32406
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.